  "name": "Single-Use Vacuum Filtration Unit Specifications and Operational Parameters",
  "description": "Original compilation comparing single-use sterile vacuum filtration systems and membrane architectures across volume capacity, polymer chemistry, pore size, effective filtration area, and vacuum limits, sourced from FDA guidance, ASTM F838 standards, and manufacturer technical datasheets.",
  "page": "https://casrai.org/guides/stericup",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-22T13:25:00Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "Global",
  "columns": [
    {
      "name": "System Configuration",
      "unit": "",
      "description": "Format and product line of the filtration unit"
    },
    {
      "name": "Membrane Polymer",
      "unit": "",
      "description": "Base polymer material of the microporous membrane"
    },
    {
      "name": "Pore Size",
      "unit": "µm",
      "description": "Nominal filtration rating"
    },
    {
      "name": "Funnel Volume",
      "unit": "mL",
      "description": "Maximum process volume capacity"
    },
    {
      "name": "Filtration Area",
      "unit": "cm²",
      "description": "Effective surface area of the membrane"
    },
    {
      "name": "Max Differential Vacuum",
      "unit": "mmHg",
      "description": "Maximum recommended operational vacuum differential at 25 °C"
    },
    {
      "name": "Primary Laboratory Application",
      "unit": "",
      "description": "Typical research or bioprocess use case"
    }
  ],
  "rows": [
    {
      "System Configuration": "Stericup Quick Release GP",
      "Membrane Polymer": "Polyethersulfone (PES)",
      "Pore Size": 0.22,
      "Funnel Volume": 500,
      "Filtration Area": 40,
      "Max Differential Vacuum": 700,
      "Primary Laboratory Application": "Sterile filtration of cell culture media and buffers"
    },
    {
      "System Configuration": "Stericup Quick Release GV",
      "Membrane Polymer": "Polyvinylidene Fluoride (PVDF)",
      "Pore Size": 0.22,
      "Funnel Volume": 500,
      "Filtration Area": 40,
      "Max Differential Vacuum": 700,
      "Primary Laboratory Application": "Sterile clarification of dilute protein and antibody solutions"
    },
    {
      "System Configuration": "Stericup Quick Release VP",
      "Membrane Polymer": "Polyethersulfone (PES)",
      "Pore Size": 0.1,
      "Funnel Volume": 500,
      "Filtration Area": 40,
      "Max Differential Vacuum": 700,
      "Primary Laboratory Application": "Mycoplasma reduction in serum and critical cell lines"
    },
    {
      "System Configuration": "Stericup Quick Release HV",
      "Membrane Polymer": "Polyvinylidene Fluoride (PVDF)",
      "Pore Size": 0.45,
      "Funnel Volume": 500,
      "Filtration Area": 40,
      "Max Differential Vacuum": 700,
      "Primary Laboratory Application": "Coarse clarification and biological fluid pre-filtration"
    },
    {
      "System Configuration": "Stericup E-GP Funnel-Less",
      "Membrane Polymer": "Polyethersulfone (PES)",
      "Pore Size": 0.22,
      "Funnel Volume": 1000,
      "Filtration Area": 40,
      "Max Differential Vacuum": 700,
      "Primary Laboratory Application": "Eco-friendly high-volume media preparation with direct bottle coupling"
    },
    {
      "System Configuration": "Steritop Quick Release GP",
      "Membrane Polymer": "Polyethersulfone (PES)",
      "Pore Size": 0.22,
      "Funnel Volume": 1000,
      "Filtration Area": 40,
      "Max Differential Vacuum": 700,
      "Primary Laboratory Application": "Bottle-top sterile filtration onto standard GL45 glass storage bottles"
    }
  ],
  "sources": [
    {
      "title": "U.S. Food and Drug Administration: Sterile Drug Products Produced by Aseptic Processing — Current Good Manufacturing Practice",
      "url": "https://www.fda.gov/regulatory-information/search-fda-guidance-documents/sterile-drug-products-produced-aseptic-processing-current-good-manufacturing-practice",
      "accessed": "2026-11-21"
    },
    {
      "title": "ASTM International: F838 Standard Test Method for Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration",
      "url": "https://www.astm.org/f0838-20.html",
      "accessed": "2026-11-21"
    },
    {
      "title": "PubMed Central: Single-use solutions for PUPSIT: requirements, challenges, and solutions",
      "url": "https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13134992/",
      "accessed": "2026-11-21"
    }
  ]
}