  "name": "Turbidity Calibration Standards, Regulatory Tolerances, and Stability Specifications",
  "description": "Original comparative metrological dataset compiling primary and secondary turbidity calibration standards, nominal NTU values, chemical matrix compositions, operational shelf lives, calibration check tolerances, and regulatory standards compliance across US EPA Method 180.1, USGS National Field Manual Chapter 6.7, and manufacturer specifications.",
  "page": "https://casrai.org/guides/turbidity-meter-calibration",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-22T23:30:42Z",
  "temporalCoverage": "1993/2026",
  "spatialCoverage": "United States and International",
  "columns": [
    {
      "name": "Standard or Parameter",
      "unit": "",
      "description": "Calibration material, stock suspension, or operational tolerance parameter"
    },
    {
      "name": "Nominal Value",
      "unit": "NTU",
      "description": "Standard nominal turbidity value or upper range limit in nephelometric turbidity units"
    },
    {
      "name": "Chemical Matrix or Optical Design",
      "unit": "",
      "description": "Formulation matrix, polymer chemistry, or optical illumination configuration"
    },
    {
      "name": "Certified Shelf Life",
      "unit": "days",
      "description": "Standard stability and usable lifespan when stored under specified conditions"
    },
    {
      "name": "Verification Tolerance",
      "unit": "%",
      "description": "Acceptable deviation threshold during calibration check verification"
    },
    {
      "name": "Governing Standard",
      "unit": "",
      "description": "Primary regulatory method or standards authority establishing criteria"
    }
  ],
  "rows": [
    {
      "Standard or Parameter": "Formazin Concentrated Stock (EPA 180.1)",
      "Nominal Value": 400,
      "Chemical Matrix or Optical Design": "Hydrazine sulfate (10 g/L) and hexamethylenetetramine (100 g/L) reaction suspension",
      "Certified Shelf Life": 30,
      "Verification Tolerance": 10,
      "Governing Standard": "US EPA Method 180.1"
    },
    {
      "Standard or Parameter": "Formazin Stock Suspension (USGS 4000)",
      "Nominal Value": 4000,
      "Chemical Matrix or Optical Design": "Hydrazine sulfate (5 g/L) and hexamethylenetetramine (50 g/L) in amber polyethylene",
      "Certified Shelf Life": 365,
      "Verification Tolerance": 5,
      "Governing Standard": "USGS TWRI Book 9 Section 6.7"
    },
    {
      "Standard or Parameter": "Dilute Formazin Working Calibrant",
      "Nominal Value": 40,
      "Chemical Matrix or Optical Design": "1:10 aqueous dilution of 400 NTU stock in turbidity-free water (",
      "Certified Shelf Life": 1,
      "Verification Tolerance": 5,
      "Governing Standard": "US EPA Method 180.1 / USGS 6.7"
    },
    {
      "Standard or Parameter": "StablCal Stabilized Formazin 20 NTU",
      "Nominal Value": 20,
      "Chemical Matrix or Optical Design": "Aqueous stabilized formazin polymer with protective colloid matrix",
      "Certified Shelf Life": 730,
      "Verification Tolerance": 5,
      "Governing Standard": "Hach Company 2100Q Specifications"
    },
    {
      "Standard or Parameter": "StablCal Stabilized Formazin 100 NTU",
      "Nominal Value": 100,
      "Chemical Matrix or Optical Design": "Aqueous stabilized formazin polymer with protective colloid matrix",
      "Certified Shelf Life": 730,
      "Verification Tolerance": 5,
      "Governing Standard": "Hach Company 2100Q Specifications"
    },
    {
      "Standard or Parameter": "StablCal Stabilized Formazin 800 NTU",
      "Nominal Value": 800,
      "Chemical Matrix or Optical Design": "Aqueous stabilized formazin polymer with protective colloid matrix",
      "Certified Shelf Life": 730,
      "Verification Tolerance": 5,
      "Governing Standard": "Hach Company 2100Q Specifications"
    },
    {
      "Standard or Parameter": "AMCO-AEPA-1 Polymer Standard",
      "Nominal Value": 40,
      "Chemical Matrix or Optical Design": "Cross-linked styrene divinylbenzene (SDVB) copolymer uniform microspheres",
      "Certified Shelf Life": 365,
      "Verification Tolerance": 5,
      "Governing Standard": "US EPA Method 180.1 Approved Alternative"
    },
    {
      "Standard or Parameter": "Turbidity-Free Reagent Blank",
      "Nominal Value": 0.05,
      "Chemical Matrix or Optical Design": "Deionized water filtered through 0.2 um or 0.45 um membrane filter",
      "Certified Shelf Life": 1,
      "Verification Tolerance": 10,
      "Governing Standard": "US EPA Method 180.1 Section 7.1"
    },
    {
      "Standard or Parameter": "Secondary Sealed Gelex Standard",
      "Nominal Value": 10,
      "Chemical Matrix or Optical Design": "Metal oxide particulate suspension immobilized in sealed silicone gel",
      "Certified Shelf Life": 365,
      "Verification Tolerance": 5,
      "Governing Standard": "USGS TWRI Book 9 Section 6.7"
    },
    {
      "Standard or Parameter": "Drinking Water Filter Effluent Compliance",
      "Nominal Value": 0.3,
      "Chemical Matrix or Optical Design": "Conventional or direct filtration effluent 95th percentile monthly limit",
      "Certified Shelf Life": 30,
      "Verification Tolerance": 5,
      "Governing Standard": "US EPA 40 CFR 141.73 NPDWR"
    }
  ],
  "sources": [
    {
      "title": "Method 180.1: Determination of Turbidity by Nephelometry",
      "url": "https://www.epa.gov/sites/default/files/2015-08/documents/method_180-1_1993.pdf",
      "accessed": "2026-11-22"
    },
    {
      "title": "Chapter A6. Section 6.7. Turbidity",
      "url": "https://pubs.usgs.gov/publication/twri09A6.7",
      "accessed": "2026-11-22"
    },
    {
      "title": "National Primary Drinking Water Regulations",
      "url": "https://www.epa.gov/ground-water-and-drinking-water/national-primary-drinking-water-regulations",
      "accessed": "2026-11-22"
    },
    {
      "title": "2100Q Portable Turbidimeter",
      "url": "https://www.hach.com/p-2100q-portable-turbidimeter/2100Q01",
      "accessed": "2026-11-22"
    }
  ]
}