  "name": "Westgard Sigma Rules: rule set, control measurements and runs by sigma metric and number of control levels",
  "description": "Compiled from Westgard QC's Westgard Sigma Rules article (rule set and N/R by sigma metric for two and three control levels) and the classic Westgard rules page (rule definitions), cross-checked against the CLIA control-material requirement in 42 CFR 493.1256. Each row gives the rule set recommended at a sigma-metric level.",
  "page": "https://casrai.org/guides/westgard-rules",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-15T12:46:09Z",
  "temporalCoverage": "2014/2026",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Sigma metric",
      "unit": "",
      "description": ""
    },
    {
      "name": "Rules, two control levels",
      "unit": "",
      "description": ""
    },
    {
      "name": "Measurements and runs, two levels",
      "unit": "",
      "description": ""
    },
    {
      "name": "Rules, three control levels",
      "unit": "",
      "description": ""
    },
    {
      "name": "Measurements and runs, three levels",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Sigma metric": "6",
      "Rules, two control levels": "1-3s",
      "Measurements and runs, two levels": "N=2, R=1",
      "Rules, three control levels": "1-3s",
      "Measurements and runs, three levels": "1 measurement on each of 3 levels"
    },
    {
      "Sigma metric": "5",
      "Rules, two control levels": "1-3s/2-2s/R-4s",
      "Measurements and runs, two levels": "N=2, R=1",
      "Rules, three control levels": "1-3s/2of3 2s/R-4s",
      "Measurements and runs, three levels": "1 measurement on each of 3 levels"
    },
    {
      "Sigma metric": "4",
      "Rules, two control levels": "1-3s/2-2s/R-4s/4-1s",
      "Measurements and runs, two levels": "N=4, R=1 or N=2, R=2",
      "Rules, three control levels": "1-3s/2of3 2s/R-4s/3-1s",
      "Measurements and runs, three levels": "1 measurement on each of 3 levels"
    },
    {
      "Sigma metric": "below 4",
      "Rules, two control levels": "Multirule including 8x",
      "Measurements and runs, two levels": "N=4, R=2 or N=2 across 4 runs",
      "Rules, three control levels": "Multirule including 6x (or 9x)",
      "Measurements and runs, three levels": "N=6, R=1 or N=3, R=2 (9x: N=3, R=3)"
    }
  ],
  "sources": [
    {
      "title": "Westgard Sigma Rules (Westgard QC)",
      "url": "https://www.westgard.com/westgard-sigma-rules.htm",
      "accessed": "2026-11-15"
    },
    {
      "title": "Westgard Rules and Multirules (Westgard QC)",
      "url": "https://www.westgard.com/westgard-rules-and-multirules.htm",
      "accessed": "2026-11-15"
    },
    {
      "title": "42 CFR 493.1256 Control procedures (eCFR)",
      "url": "https://www.ecfr.gov/current/title-42/part-493/section-493.1256",
      "accessed": "2026-11-15"
    }
  ]
}