  "name": "Maximum rated RCF and handling notes for Thermo Fisher centrifuge bottles and a reference tube",
  "description": "Compares the manufacturer-stated maximum relative centrifugal force, material, nominal volume and rotor condition for centrifuge bottles of 250 mL to 1 L and one 16 mL tube, compiled from four Thermo Fisher Scientific product records.",
  "page": "https://casrai.org/guides/what-are-centrifuge-bottles",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-20T17:53:10Z",
  "temporalCoverage": "2026",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Product",
      "unit": "",
      "description": ""
    },
    {
      "name": "Material",
      "unit": "",
      "description": ""
    },
    {
      "name": "Nominal volume",
      "unit": "mL",
      "description": ""
    },
    {
      "name": "Maximum RCF",
      "unit": "x g",
      "description": ""
    },
    {
      "name": "Rotor or condition stated",
      "unit": "",
      "description": ""
    },
    {
      "name": "Autoclave note",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Product": "Nalgene PC centrifuge bottle 3122-0500",
      "Material": "Polycarbonate",
      "Nominal volume": 500,
      "Maximum RCF": 13700,
      "Rotor or condition stated": "Maximum speed stated for the bottle; PP screw cap included",
      "Autoclave note": "Autoclavable"
    },
    {
      "Product": "Nalgene PPCO bottle with sealing closure 3141-0250",
      "Material": "Polypropylene copolymer",
      "Nominal volume": 250,
      "Maximum RCF": 27500,
      "Rotor or condition stated": "Maximum speed stated for the bottle; sealing cap assembly with silicone gasket",
      "Autoclave note": "Autoclavable; not recommended for radiation sterilization"
    },
    {
      "Product": "Sorvall high performance bottle 010-1456 (PPCO) or 010-1459 (PC)",
      "Material": "PPCO or PC",
      "Nominal volume": 1000,
      "Maximum RCF": 20584,
      "Rotor or condition stated": "F10-4x1000 LEX rotor in LYNX 6000 or LYNX 4000",
      "Autoclave note": "Do not autoclave with closures engaged; aluminum cap 121 C 20 min or 134 C 3 min"
    },
    {
      "Product": "Sorvall high performance bottle 010-1456 (PPCO) or 010-1459 (PC)",
      "Material": "PPCO or PC",
      "Nominal volume": 1000,
      "Maximum RCF": 17568,
      "Rotor or condition stated": "F9-6x1000 LEX rotor in LYNX 6000",
      "Autoclave note": "Do not autoclave with closures engaged; aluminum cap 121 C 20 min or 134 C 3 min"
    },
    {
      "Product": "Sorvall high performance bottle 010-1456 (PPCO) or 010-1459 (PC)",
      "Material": "PPCO or PC",
      "Nominal volume": 1000,
      "Maximum RCF": 9333,
      "Rotor or condition stated": "F5-10x1000 LEX rotor in RC BIOS 10",
      "Autoclave note": "Do not autoclave with closures engaged; aluminum cap 121 C 20 min or 134 C 3 min"
    },
    {
      "Product": "Nalgene Oak Ridge tube 3139-0016 (tube, for contrast)",
      "Material": "Polypropylene copolymer",
      "Nominal volume": 16,
      "Maximum RCF": 50000,
      "Rotor or condition stated": "Maximum speed stated for the tube; sealing cap advised above 10,000 x g",
      "Autoclave note": "Not stated in the record read"
    }
  ],
  "sources": [
    {
      "title": "Nalgene polycarbonate centrifuge bottle 500 mL, 3122-0500",
      "url": "https://www.thermofisher.com/order/catalog/product/3122-0500",
      "accessed": "2026-11-20"
    },
    {
      "title": "Nalgene PPCO bottle with sealing closure 250 mL, 3141-0250",
      "url": "https://www.thermofisher.com/order/catalog/product/3141-0250",
      "accessed": "2026-11-20"
    },
    {
      "title": "Sorvall high performance centrifuge bottle 1 L, 010-1456",
      "url": "https://www.thermofisher.com/order/catalog/product/010-1456",
      "accessed": "2026-11-20"
    },
    {
      "title": "Nalgene Oak Ridge PPCO tube 16 mL, 3139-0016",
      "url": "https://www.thermofisher.com/order/catalog/product/3139-0016",
      "accessed": "2026-11-20"
    }
  ]
}