  "name": "Commercial Desalting Column Specifications and Operating Limits",
  "description": "Comparative technical parameters of standard commercial laboratory desalting columns compiled from manufacturer instructions for use and published operating procedures, detailing bed volumes, molecular weight cutoffs, permitted loading ranges, and typical salt elimination performance.",
  "page": "https://casrai.org/guides/what-is-a-desalting-column",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-21T08:55:00Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "Global",
  "columns": [
    {
      "name": "Column Format and Model",
      "unit": "",
      "description": "Manufacturer designation and bed format"
    },
    {
      "name": "Stationary Phase Matrix",
      "unit": "",
      "description": "Chemical composition of the porous polymer beads"
    },
    {
      "name": "Molecular Weight Cut-Off",
      "unit": "Da",
      "description": "Nominal exclusion threshold for globular biomolecules"
    },
    {
      "name": "Hydrated Bed Volume",
      "unit": "mL",
      "description": "Total settled resin volume inside the column barrel"
    },
    {
      "name": "Minimum Sample Load",
      "unit": "µL",
      "description": "Lower recommended sample volume limit"
    },
    {
      "name": "Maximum Sample Load",
      "unit": "µL",
      "description": "Upper recommended sample volume limit without salt leakage"
    },
    {
      "name": "Operating Centrifugal Force",
      "unit": "× g",
      "description": "Centrifugation speed for spin formats or 1 for gravity flow"
    },
    {
      "name": "Typical Salt Removal",
      "unit": "%",
      "description": "Percentage of low-molecular-weight ionic solute eliminated"
    }
  ],
  "rows": [
    {
      "Column Format and Model": "Thermo Scientific Zeba Spin (0.5 mL)",
      "Stationary Phase Matrix": "Crosslinked dextran composite",
      "Molecular Weight Cut-Off": 7000,
      "Hydrated Bed Volume": 0.5,
      "Minimum Sample Load": 30,
      "Maximum Sample Load": 130,
      "Operating Centrifugal Force": 1500,
      "Typical Salt Removal": 95
    },
    {
      "Column Format and Model": "Thermo Scientific Zeba Spin (2.0 mL)",
      "Stationary Phase Matrix": "Crosslinked dextran composite",
      "Molecular Weight Cut-Off": 7000,
      "Hydrated Bed Volume": 2,
      "Minimum Sample Load": 200,
      "Maximum Sample Load": 700,
      "Operating Centrifugal Force": 1000,
      "Typical Salt Removal": 95
    },
    {
      "Column Format and Model": "Thermo Scientific Zeba Spin (5.0 mL)",
      "Stationary Phase Matrix": "Crosslinked dextran composite",
      "Molecular Weight Cut-Off": 7000,
      "Hydrated Bed Volume": 5,
      "Minimum Sample Load": 500,
      "Maximum Sample Load": 2000,
      "Operating Centrifugal Force": 1000,
      "Typical Salt Removal": 95
    },
    {
      "Column Format and Model": "Cytiva illustra NAP-5 (Gravity)",
      "Stationary Phase Matrix": "Sephadex G-25 Medium DNA Grade",
      "Molecular Weight Cut-Off": 5000,
      "Hydrated Bed Volume": 1.4,
      "Minimum Sample Load": 100,
      "Maximum Sample Load": 500,
      "Operating Centrifugal Force": 1,
      "Typical Salt Removal": 90
    },
    {
      "Column Format and Model": "G-Biosciences SpinOUT GT-600",
      "Stationary Phase Matrix": "Polyacrylamide / agarose composite",
      "Molecular Weight Cut-Off": 6000,
      "Hydrated Bed Volume": 0.6,
      "Minimum Sample Load": 30,
      "Maximum Sample Load": 130,
      "Operating Centrifugal Force": 1500,
      "Typical Salt Removal": 95
    }
  ],
  "sources": [
    {
      "title": "Thermo Fisher Scientific: Zeba Spin Desalting Columns User Guide",
      "url": "https://www.thermofisher.com/order/catalog/product/89882",
      "accessed": "2026-11-19"
    },
    {
      "title": "National Cancer Institute: SOP 22141 Desalting and Buffer Exchange Using illustra NAP-5 Columns",
      "url": "https://frederick.cancer.gov/sites/default/files/2022-03/Desalting_and_Buffer_Exchange_of_Products_Using_illustra_NAP-5_Columns.pdf",
      "accessed": "2026-11-19"
    },
    {
      "title": "protocols.io: SpinOUT Protein Desalting Protocol",
      "url": "https://www.protocols.io/view/SpinOUT-Protein-Desalting-ewov1oyklr24/v1/v1",
      "accessed": "2026-11-19"
    }
  ]
}