  "name": "Capillary Electrophoresis Reagent Specifications and Analytical Limits",
  "description": "Comparative analytical parameters for automated capillary electrophoresis nucleic acid kits on the Fragment Analyzer platform, compiled from manufacturer technical specifications, assay quick guides, and validation manuals across DNA, RNA, and genomic sequencing formats.",
  "page": "https://casrai.org/guides/what-is-a-fragment-analyzer",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-15T07:39:25Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Assay Kit and Target",
      "unit": "",
      "description": "Reagent kit designation and primary analytical target molecule"
    },
    {
      "name": "Sizing Range Min",
      "unit": "bp",
      "description": "Lower boundary of the analytical sizing window in base pairs or nucleotides"
    },
    {
      "name": "Sizing Range Max",
      "unit": "bp",
      "description": "Upper boundary of the analytical sizing window in base pairs or nucleotides"
    },
    {
      "name": "Lower Limit of Detection",
      "unit": "pg/μL",
      "description": "Minimum detectable fragment concentration"
    },
    {
      "name": "Upper Dynamic Range",
      "unit": "pg/μL",
      "description": "Upper quantification linearity threshold"
    },
    {
      "name": "Separation Run Time",
      "unit": "min",
      "description": "Typical electrophoretic run time per injection"
    },
    {
      "name": "Primary Quality Metric",
      "unit": "",
      "description": "Standard integrity, concentration, or sizing metric generated by software"
    }
  ],
  "rows": [
    {
      "Assay Kit and Target": "Agilent DNF-474 HS NGS Fragment Kit",
      "Sizing Range Min": 100,
      "Sizing Range Max": 6000,
      "Lower Limit of Detection": 5,
      "Upper Dynamic Range": 5000,
      "Separation Run Time": 45,
      "Primary Quality Metric": "Smear Analysis & Molar Concentration"
    },
    {
      "Assay Kit and Target": "Agilent DNF-473 Standard Sensitivity NGS Fragment Kit",
      "Sizing Range Min": 100,
      "Sizing Range Max": 6000,
      "Lower Limit of Detection": 100,
      "Upper Dynamic Range": 50000,
      "Separation Run Time": 45,
      "Primary Quality Metric": "Smear Analysis & Molar Concentration"
    },
    {
      "Assay Kit and Target": "Agilent DNF-468 HS Genomic DNA 50 kb Kit",
      "Sizing Range Min": 75,
      "Sizing Range Max": 60000,
      "Lower Limit of Detection": 50,
      "Upper Dynamic Range": 5000,
      "Separation Run Time": 60,
      "Primary Quality Metric": "Genomic Quality Number (GQN)"
    },
    {
      "Assay Kit and Target": "Agilent DNF-930 dsDNA Reagent Kit",
      "Sizing Range Min": 75,
      "Sizing Range Max": 20000,
      "Lower Limit of Detection": 100,
      "Upper Dynamic Range": 20000,
      "Separation Run Time": 50,
      "Primary Quality Metric": "Fragment Sizing & Peak Quantification"
    },
    {
      "Assay Kit and Target": "Agilent DNF-471 Standard Sensitivity RNA Kit",
      "Sizing Range Min": 15,
      "Sizing Range Max": 6000,
      "Lower Limit of Detection": 5000,
      "Upper Dynamic Range": 500000,
      "Separation Run Time": 40,
      "Primary Quality Metric": "RNA Quality Number (RQN)"
    },
    {
      "Assay Kit and Target": "Agilent DNF-472 High Sensitivity RNA Kit",
      "Sizing Range Min": 15,
      "Sizing Range Max": 6000,
      "Lower Limit of Detection": 50,
      "Upper Dynamic Range": 5000,
      "Separation Run Time": 40,
      "Primary Quality Metric": "RNA Quality Number (RQN)"
    },
    {
      "Assay Kit and Target": "Agilent DNF-935 Small Fragment Kit",
      "Sizing Range Min": 1,
      "Sizing Range Max": 1500,
      "Lower Limit of Detection": 50,
      "Upper Dynamic Range": 10000,
      "Separation Run Time": 30,
      "Primary Quality Metric": "High-Resolution Amplicon Sizing"
    }
  ],
  "sources": [
    {
      "title": "Agilent DNF-474 High Sensitivity NGS Fragment Kit Quick Guide",
      "url": "https://www.agilent.com/en/product/automated-electrophoresis/fragment-analyzer-systems/fragment-analyzer-reagents/high-sensitivity-ngs-fragment-kit",
      "accessed": "2026-11-15"
    },
    {
      "title": "Agilent DNF-468 HS Genomic DNA 50 kb Kit Protocol and User Guide",
      "url": "https://www.agilent.com/en/product/automated-electrophoresis/fragment-analyzer-systems/fragment-analyzer-reagents/hs-genomic-dna-50-kb-kit",
      "accessed": "2026-11-15"
    },
    {
      "title": "Agilent 5200, 5300, and 5400 Fragment Analyzer Systems Specifications",
      "url": "https://www.agilent.com/en/products/automated-electrophoresis/fragment-analyzer-systems",
      "accessed": "2026-11-15"
    }
  ]
}