  "name": "Published error limits at a 10 µL test volume",
  "description": "Original compilation of BT Lab Systems P10 and Eppendorf Research plus single- and multichannel specifications. BT inaccuracy and imprecision percentages are converted to microliters at 10 µL; Eppendorf systematic and random errors are transcribed in microliters. Values describe each manufacturer’s own test conditions and are not results of comparative testing.",
  "page": "https://casrai.org/guides/what-is-a-p10-pipette",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-14T23:44:53Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Instrument configuration",
      "unit": "",
      "description": ""
    },
    {
      "name": "Test volume",
      "unit": "µL",
      "description": ""
    },
    {
      "name": "Inaccuracy or systematic error limit",
      "unit": "µL",
      "description": "Absolute magnitude of the stated plus/minus limit; BT calls this inaccuracy and Eppendorf calls it systematic error."
    },
    {
      "name": "Imprecision or random error limit",
      "unit": "µL",
      "description": "BT published percentage imprecision converted at 10 µL; Eppendorf published absolute random error."
    },
    {
      "name": "Source location",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Instrument configuration": "BT Lab Systems P10 BT1501, single-channel",
      "Test volume": 10,
      "Inaccuracy or systematic error limit": 0.1,
      "Imprecision or random error limit": 0.08,
      "Source location": "BT P10 page: Single Channel Pipette Specifications, BT1501, 10 µL test row"
    },
    {
      "Instrument configuration": "Eppendorf Research plus 0.5–10 µL, single-channel",
      "Test volume": 10,
      "Inaccuracy or systematic error limit": 0.1,
      "Imprecision or random error limit": 0.04,
      "Source location": "Research plus Operating Manual, p.55, variable single-channel 0.5–10 µL"
    },
    {
      "Instrument configuration": "Eppendorf Research plus 0.5–10 µL, 8-/12-channel",
      "Test volume": 10,
      "Inaccuracy or systematic error limit": 0.2,
      "Imprecision or random error limit": 0.1,
      "Source location": "Research plus Operating Manual, p.57, fixed cone spacing 0.5–10 µL"
    }
  ],
  "sources": [
    {
      "title": "BT Lab Systems P10 BT1501 specifications",
      "url": "https://www.btlabsystems.com/Pipettes/P10_Pipette",
      "accessed": "2026-11-14"
    },
    {
      "title": "Eppendorf Research plus Operating Manual, pp.55 and 57",
      "url": "https://www.eppendorf.com/product-media/doc/en/186591/Eppendorf_Liquid-Handling_Operating-manual_Research-plus_Eppendorf-Research-plus.pdf",
      "accessed": "2026-11-14"
    }
  ]
}