  "name": "FDA classification of cardiopulmonary bypass blood pump devices (21 CFR 870.4360, 870.4370, 870.4380)",
  "description": "Compares the four regulated bypass pump device types by identification, FDA class and regulatory basis, compiled from the text of the Code of Federal Regulations as published by the Legal Information Institute.",
  "page": "https://casrai.org/guides/what-is-a-perfusion-pump",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-18T10:00:00Z",
  "temporalCoverage": "2026",
  "spatialCoverage": "United States",
  "columns": [
    {
      "name": "Device",
      "unit": "",
      "description": ""
    },
    {
      "name": "CFR section",
      "unit": "",
      "description": ""
    },
    {
      "name": "FDA class",
      "unit": "",
      "description": ""
    },
    {
      "name": "Regulatory basis",
      "unit": "",
      "description": ""
    },
    {
      "name": "Stated use limit",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Device": "Roller-type cardiopulmonary bypass blood pump",
      "CFR section": "21 CFR 870.4370",
      "FDA class": "II",
      "Regulatory basis": "Performance standards",
      "Stated use limit": "Bypass surgery circuit; no duration stated"
    },
    {
      "Device": "Nonroller-type cardiopulmonary and circulatory bypass blood pump",
      "CFR section": "21 CFR 870.4360(a)",
      "FDA class": "II",
      "Regulatory basis": "Special controls",
      "Stated use limit": "Under 6 hours"
    },
    {
      "Device": "Nonroller-type temporary ventricular support blood pump",
      "CFR section": "21 CFR 870.4360(b)",
      "FDA class": "III",
      "Regulatory basis": "Premarket approval",
      "Stated use limit": "Under 6 hours"
    },
    {
      "Device": "Cardiopulmonary bypass pump speed control",
      "CFR section": "21 CFR 870.4380",
      "FDA class": "II",
      "Regulatory basis": "Performance standards",
      "Stated use limit": "Controls speed of bypass blood pumps; no duration stated"
    }
  ],
  "sources": [
    {
      "title": "21 CFR 870.4370",
      "url": "https://www.law.cornell.edu/cfr/text/21/870.4370",
      "accessed": "2026-11-17"
    },
    {
      "title": "21 CFR 870.4360",
      "url": "https://www.law.cornell.edu/cfr/text/21/870.4360",
      "accessed": "2026-11-17"
    },
    {
      "title": "21 CFR 870.4380",
      "url": "https://www.law.cornell.edu/cfr/text/21/870.4380",
      "accessed": "2026-11-17"
    }
  ]
}