  "name": "BRAND PD-Tips II (repetitive pipette tips): volume range, accuracy and coefficient of variation at nominal volume and at 1% of nominal",
  "description": "Compiled from the accuracy table in the BRAND HandyStep touch brochure (distilled water, ISO 8655 tolerances not exceeded). Adds a calculated column: the permitted systematic error at 1% of nominal volume divided by the error at 100%.",
  "page": "https://casrai.org/guides/what-is-a-repeat-pipette",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-19T23:21:24Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Tip",
      "unit": "",
      "description": ""
    },
    {
      "name": "Nominal volume",
      "unit": "mL",
      "description": ""
    },
    {
      "name": "Minimum set volume",
      "unit": "uL",
      "description": ""
    },
    {
      "name": "Maximum set volume",
      "unit": "uL",
      "description": ""
    },
    {
      "name": "Systematic error limit at 100% of nominal",
      "unit": "%",
      "description": ""
    },
    {
      "name": "Systematic error limit at 1% of nominal",
      "unit": "%",
      "description": ""
    },
    {
      "name": "Ratio of error limit at 1% to 100%",
      "unit": "x",
      "description": "Calculated"
    },
    {
      "name": "CV limit at 100% of nominal",
      "unit": "%",
      "description": ""
    },
    {
      "name": "CV limit at 1% of nominal",
      "unit": "%",
      "description": ""
    }
  ],
  "rows": [
    {
      "Tip": "PD-Tip II 0.1 mL",
      "Nominal volume": 0.1,
      "Minimum set volume": 1,
      "Maximum set volume": 100,
      "Systematic error limit at 100% of nominal": 1,
      "Systematic error limit at 1% of nominal": 8,
      "Ratio of error limit at 1% to 100%": 8,
      "CV limit at 100% of nominal": 0.5,
      "CV limit at 1% of nominal": 12
    },
    {
      "Tip": "PD-Tip II 0.5 mL",
      "Nominal volume": 0.5,
      "Minimum set volume": 5,
      "Maximum set volume": 500,
      "Systematic error limit at 100% of nominal": 0.9,
      "Systematic error limit at 1% of nominal": 5,
      "Ratio of error limit at 1% to 100%": 5.6,
      "CV limit at 100% of nominal": 0.3,
      "CV limit at 1% of nominal": 5
    },
    {
      "Tip": "PD-Tip II 1 mL",
      "Nominal volume": 1,
      "Minimum set volume": 10,
      "Maximum set volume": 1000,
      "Systematic error limit at 100% of nominal": 0.6,
      "Systematic error limit at 1% of nominal": 5,
      "Ratio of error limit at 1% to 100%": 8.3,
      "CV limit at 100% of nominal": 0.3,
      "CV limit at 1% of nominal": 4
    },
    {
      "Tip": "PD-Tip II 1.25 mL",
      "Nominal volume": 1.25,
      "Minimum set volume": 12,
      "Maximum set volume": 1250,
      "Systematic error limit at 100% of nominal": 0.6,
      "Systematic error limit at 1% of nominal": 5,
      "Ratio of error limit at 1% to 100%": 8.3,
      "CV limit at 100% of nominal": 0.15,
      "CV limit at 1% of nominal": 4
    },
    {
      "Tip": "PD-Tip II 2.5 mL",
      "Nominal volume": 2.5,
      "Minimum set volume": 25,
      "Maximum set volume": 2500,
      "Systematic error limit at 100% of nominal": 0.5,
      "Systematic error limit at 1% of nominal": 3.5,
      "Ratio of error limit at 1% to 100%": 7,
      "CV limit at 100% of nominal": 0.15,
      "CV limit at 1% of nominal": 3
    },
    {
      "Tip": "PD-Tip II 5 mL",
      "Nominal volume": 5,
      "Minimum set volume": 50,
      "Maximum set volume": 5000,
      "Systematic error limit at 100% of nominal": 0.5,
      "Systematic error limit at 1% of nominal": 3.5,
      "Ratio of error limit at 1% to 100%": 7,
      "CV limit at 100% of nominal": 0.15,
      "CV limit at 1% of nominal": 3
    },
    {
      "Tip": "PD-Tip II 10 mL",
      "Nominal volume": 10,
      "Minimum set volume": 100,
      "Maximum set volume": 10000,
      "Systematic error limit at 100% of nominal": 0.4,
      "Systematic error limit at 1% of nominal": 3.5,
      "Ratio of error limit at 1% to 100%": 8.8,
      "CV limit at 100% of nominal": 0.15,
      "CV limit at 1% of nominal": 4
    },
    {
      "Tip": "PD-Tip II 12.5 mL",
      "Nominal volume": 12.5,
      "Minimum set volume": 125,
      "Maximum set volume": 12500,
      "Systematic error limit at 100% of nominal": 0.5,
      "Systematic error limit at 1% of nominal": 3.5,
      "Ratio of error limit at 1% to 100%": 7,
      "CV limit at 100% of nominal": 0.15,
      "CV limit at 1% of nominal": 6.5
    },
    {
      "Tip": "PD-Tip II 25 mL",
      "Nominal volume": 25,
      "Minimum set volume": 250,
      "Maximum set volume": 25000,
      "Systematic error limit at 100% of nominal": 0.5,
      "Systematic error limit at 1% of nominal": 3,
      "Ratio of error limit at 1% to 100%": 6,
      "CV limit at 100% of nominal": 0.15,
      "CV limit at 1% of nominal": 6
    },
    {
      "Tip": "PD-Tip II 50 mL",
      "Nominal volume": 50,
      "Minimum set volume": 500,
      "Maximum set volume": 50000,
      "Systematic error limit at 100% of nominal": 0.5,
      "Systematic error limit at 1% of nominal": 3,
      "Ratio of error limit at 1% to 100%": 6,
      "CV limit at 100% of nominal": 0.15,
      "CV limit at 1% of nominal": 9
    }
  ],
  "sources": [
    {
      "title": "BRAND HandyStep touch brochure (English), accuracy table",
      "url": "https://www.brand.de/brand/contentserv_data/Context/BRAND%20GMBH%20+%20CO%20KG/Dokumente/Produkte/Prospekte/HandyStep%20touch/HandyStep_touch_EN.pdf",
      "accessed": "2026-11-19"
    }
  ]
}