  "name": "Laboratory Melting and Boiling Point Determination Methods Benchmark",
  "description": "Comparative analytical specifications, temperature limits, heating mechanisms, sample mass requirements, and pharmacopeial compliance across standard laboratory thermal determination methods, compiled from USP General Chapter 741, ASTM E324-16, manufacturer apparatus manuals, and laboratory standards.",
  "page": "https://casrai.org/guides/what-is-a-thiele-tube",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-19T17:25:00Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Method or Apparatus",
      "unit": "",
      "description": "Standard laboratory instrument or manual glassware configuration"
    },
    {
      "name": "Maximum Operating Temperature",
      "unit": "°C",
      "description": "Maximum standard operational thermal ceiling supported by the system"
    },
    {
      "name": "Standard Ramp Rate",
      "unit": "°C/min",
      "description": "Recommended linear heating rate during active phase-transition observation"
    },
    {
      "name": "Nominal Sample Mass",
      "unit": "mg",
      "description": "Typical dry sample mass packed into capillary tube or testing stage"
    },
    {
      "name": "Heating Mode",
      "unit": "",
      "description": "Primary thermodynamic heat transfer mode utilized to warm the specimen"
    },
    {
      "name": "Pharmacopeial Compliance",
      "unit": "",
      "description": "Compliance status with USP Chapter 741 Apparatus I or equivalent compendial standards"
    }
  ],
  "rows": [
    {
      "Method or Apparatus": "Thiele Tube (Manual Glassware)",
      "Maximum Operating Temperature": 250,
      "Standard Ramp Rate": 1,
      "Nominal Sample Mass": 2,
      "Heating Mode": "Passive fluid convective circulation (silicone or mineral oil)",
      "Pharmacopeial Compliance": "Meets USP Chapter 741 Apparatus I requirements with calibrated thermometer"
    },
    {
      "Method or Apparatus": "Mel-Temp / Capillary Block",
      "Maximum Operating Temperature": 400,
      "Standard Ramp Rate": 1,
      "Nominal Sample Mass": 2,
      "Heating Mode": "Electrically heated aluminum block with integrated optical magnification",
      "Pharmacopeial Compliance": "Meets USP Chapter 741 Apparatus I requirements with calibrated platinum sensor"
    },
    {
      "Method or Apparatus": "Automated Optical Melting Point System",
      "Maximum Operating Temperature": 400,
      "Standard Ramp Rate": 0.5,
      "Nominal Sample Mass": 2,
      "Heating Mode": "Digital micro-furnace heating with automated video transmission edge detection",
      "Pharmacopeial Compliance": "Meets USP Chapter 741 and Ph. Eur. 2.2.60 with 21 CFR Part 11 digital audit trails"
    },
    {
      "Method or Apparatus": "Fisher-Johns Hot Stage",
      "Maximum Operating Temperature": 300,
      "Standard Ramp Rate": 2,
      "Nominal Sample Mass": 5,
      "Heating Mode": "Electrically heated aluminum plate between circular microscope cover glasses",
      "Pharmacopeial Compliance": "Non-compliant with capillary standards; rapid qualitative screening only"
    },
    {
      "Method or Apparatus": "Differential Scanning Calorimetry (DSC)",
      "Maximum Operating Temperature": 700,
      "Standard Ramp Rate": 10,
      "Nominal Sample Mass": 1,
      "Heating Mode": "Controlled purge-gas furnace with differential heat flux transducers",
      "Pharmacopeial Compliance": "Meets USP Chapter 891 and ASTM E928 formal thermal analysis standards"
    }
  ],
  "sources": [
    {
      "title": "USP General Chapter 741 Melting Range or Temperature",
      "url": "https://www.uspnf.com",
      "accessed": "2026-11-14"
    },
    {
      "title": "ASTM E324-16 Standard Test Method for Relative Initial and Final Melting Points",
      "url": "https://www.astm.org/e0324-16.html",
      "accessed": "2026-11-14"
    },
    {
      "title": "Lisa Nichols, Organic Chemistry Lab Techniques, Chapter 6: Melting Points",
      "url": "https://chem.libretexts.org/Bookshelves/Organic_Chemistry/Organic_Chemistry_Lab_Techniques_(Nichols)/06%3A_Melting_Points/6.03%3A_Step-by-Step_Procedures_for_Melting_Point_Determination",
      "accessed": "2026-11-14"
    }
  ]
}