Answers are AI-generated from CASRAI’s own published pages and can be wrong, so check the linked sources before relying on one; your question is logged without personal data — never sold, never used to train a third-party model — to show us what CASRAI is missing, so please do not type personal or confidential details. How we use this
CASRAI is the reference for research administration — bookmark it for the next question.

Memmert
Memmert® ICO105 CO2 Incubator, 108L | 3.8 cu. Ft
in Incubators
$7,091.00 - $8,635.00
In stockPrice varies by Option.
Memmert ICO105 CO₂ incubator, 107 L, with Peltier-assisted control from +18 to +50 °C at 0.1 °C accuracy and 0–20 % CO₂ on dual-beam NDIR sensing.
Choose Option to continue.
Buying through procurement?
Generate a formal quotation for 1 × this item, plus our W-9 — everything your finance department needs to raise a purchase order.
Choose Option to continue.
Description
The Memmert ICO 105 CO₂ incubator delivers precise, reliable conditions for cell culture, microbiology, and pharmaceutical workflows. Peltier-assisted temperature control, intelligent safety monitoring, and a hygienic stainless chamber keep pH stable and recovery fast after door openings.
Operate from +18 °C to +50 °C with 0.1 °C setting accuracy, manage CO₂ from 0–20% via dual-beam NDIR sensing, and log data or monitor remotely via Ethernet/USB. Passive humidity up to ~93% RH is available with a water dish; the optional K7 module enables active humidity control.
Spacious yet bench-friendly (107 L), ICO 105 supports tissue engineering, diagnostics, and food safety labs that require reproducible growth conditions and audit-ready documentation.
Use distilled/DI water in the humidity dish and run the integrated sterilisation program for best results. Some configurations may be built-to-order.
107 L (~3.8 cu ft) +18–50 °C CO₂ 0–20% (NDIR) Passive RH ~93% (K7 40–90%) TwinDISPLAY Touch
Precise Environment: Peltier heating/cooling with ±0.1 °C fluctuation at 37 °C and dual-beam NDIR CO₂ for stable pH.
Humidity Options: Passive to ~93% RH (water dish) or K7 active control 40–90% RH (or RH-Off).
Safety & Connectivity: Auto diagnostics, visual/audible alarms, Ethernet/USB logging, heated inner glass door, in-situ sensor sterilisation.
User Manual Spec Sheet
Why this Memmert ICO 105?
Accurate temperature & CO₂ — +18–50 °C with 0.1 °C setting accuracy; CO₂ 0–20% via dual-beam NDIR with barometric compensation.
Configurable humidity — passive to ~93% RH (water dish) or select ICO 105 K7 for 40–90% active RH (or RH-Off).
TwinDISPLAY control — dual TFT touchscreens for setup, logging, remote monitoring, and program management via Ethernet/USB.
Hygienic chamber — deep-drawn stainless interior with heated glass inner door; CO₂ and humidity sensors can be sterilised in-situ.
Designed for uptime — automatic diagnostics with visual/audible alarms help protect valuable cultures and timelines.
Technical Specifications
Model
Memmert ICO 105 CO₂ Incubator
Volume approx.
~3.8 cu ft (107 L)
Temperature range
+5 °C above ambient to +50 °C
Setting temperature range
+18 °C to +50 °C
Setting accuracy
0.1 °C
Temperature fluctuations @ 37 °C
±0.1 °C
Temperature variation in chamber @ 37 °C
±0.3 °C
Humidity control (passive)
With water dish, Peltier limits RH to ~93% RH ±2.5%
Humidity setting range (K7 option)
40%–90% RH and RH-Off (active control)
Humidity setting accuracy
0.50%
CO₂ / O₂ control
Digital CO₂ (dual-beam NDIR) with auto diagnostics & pressure compensation; optional O₂ 1%–20%
CO₂ range
0%–20%
CO₂ variation in time
±0.2%
CO₂ setting accuracy
0.10%
O₂ range / accuracy
1%–20% / 0.10% setting accuracy
Controller
TwinDISPLAY dual TFT touchscreens; save logs; email alarms; monitor up to 32 devices; manage via Ethernet/USB
Interfaces
Ethernet and USB
Housing / interior
Textured stainless housing; rear zinc-plated steel; stainless interior 1.4301 (ASTM 304); heated inner glass door; insulated SS door
Automatic sterilisation
Humidity and CO₂ sensors sterilised inside the incubator
Chamber dimensions (W×H×D)
22×18.9×15.7 in (560×480×400 mm)
Max shelves / included
6 total / 2 included
Max load per shelf / chamber
33.1 lb (15 kg) / 198.4 lb (75 kg)
Exterior dimensions (W×H×D)
28.3×33.5×23.3 in (719×850×591 mm)
Standard works calibration cert.
37 °C, 5% CO₂ (chamber center)
Sterilisation program
60 min at 180 °C
Electrical
115 V, 50/60 Hz, 1,100 W (230 V available)
Net / gross weight
~165.3 lb (55 kg) / ~220.5 lb (74 kg)
Shipping dimensions (W×H×D)
31.5×40.6×31.5 in (800×1030×800 mm)
Cell Culture & Tissue Engineering Microbiology Pharmaceutical Academic & Research
Humidity (K7), O₂ option, shelving kits, and accessories.
FAQ
How is ICO 105 different from ICO 105 K7?
Both share the same chamber and temperature/CO₂ control. The K7 adds active humidity control (40–90% RH or RH-Off) for precise RH setpoints and faster recovery; the standard ICO 105 uses passive humidity (water dish) up to ~93% RH.
How accurate is the CO₂ system?
A dual-beam NDIR sensor regulates 0–20% CO₂ with barometric compensation. Expect ±0.2% variation over time and 0.10% setting increments.
Can I monitor and log data remotely?
Yes. TwinDISPLAY supports logging via USB/Ethernet, email alarms, and multi-device monitoring (up to 32 devices).
Is there an automatic sterilisation routine?
Yes. The cabinet includes an automatic sterilisation program (60 min at 180 °C) with sensors sterilised in place.
Specifications
Volume:
108L | 3.8 cu. Ft
[data-sub-layout="e1bc52bf-47e4-4e1d-8e09-99d98110ac26"] {
border-
Frequently asked questions
What is the temperature range and accuracy?+
+18 °C to +50 °C with 0.1 °C setting accuracy, using Peltier-assisted temperature control.
How is CO₂ controlled?+
0–20 %, via dual-beam NDIR sensing.
What humidity can it reach?+
Passive humidity up to approximately 93 % RH with a water dish. The optional K7 module enables active humidity control.
What is the chamber made from?+
A hygienic stainless steel chamber of 107 L.
How is data recorded?+
Logging and remote monitoring via Ethernet and USB.
Related products

Memmert® ICO240 CO2 Incubator, 241L | 8.4 cu. Ft
Memmert ICO240 CO₂ incubator, 241 L, with Peltier-assisted control from +18 to +50 °C at 0.1 °C resolution and 0–20 % CO₂ on a dual-beam NDIR sensor.