Answers are AI-generated from CASRAI’s own published pages and can be wrong, so check the linked sources before relying on one; your question is logged without personal data — never sold, never used to train a third-party model — to show us what CASRAI is missing, so please do not type personal or confidential details. How we use this

CASRAI is the reference for research administration — bookmark it for the next question.

Solvent Recovery Pump - 75L-100L/HR - C1D1 Certified

SKU USA-SRP

USA Lab

Solvent Recovery Pump - 75L-100L/HR - C1D1 Certified

in Recovery Pumps

$7,299.99

In stock

C1D1-rated magnetically coupled solvent recovery pump, 75–100 L/hr at 15 psi, with a 316 stainless body, PEEK gears and an explosion-proof Baldor motor.

Buying through procurement?

Generate a formal quotation for 1 × this item, plus our W-9 — everything your finance department needs to raise a purchase order.

Ordering several items? Add them to your cart, then create one quote for the whole cart.

Description

*NOT FOR USE WITH BUTANE*
This Solvent Pump is a high-precision, explosion-proof gear pump system engineered for transferring flammable solvents in hazardous environments. With durable construction and magnetic drive technology, this pump ensures leak-free, continuous operation in distillation, extraction, and chemical processing applications.

This system is ideal for applications requiring reliable fluid handling in Class I, Div 1 environments. The magnetic coupling and robust materials provide long-lasting performance with minimal maintenance.

Specifications

Flow Rate

75L-100 L/hr @ 15 PSI

Pump Speed

245 – 327 RPM (estimated)

Displacement

5.1 cc/rev

Connections

0.5″ NPT ports

Pump Materials

316 SS body, PEEK gears, 440 SS shaft

Seals

PTFE static O-rings

Magnetic Coupling

4.0 Nm torque

Motor

Baldor 0.5 HP, 1750 RPM, NEMA 56C

Motor Features

Inverter Duty, Explosion Proof (C1D1, GRP D)

Voltage

230/460V, 3-phase, 60Hz

Compatible Solvents

Ethanol, Heptane, Methanol, 2-Propanol

Frequently asked questions

What is the flow rate?+

75–100 L/hr at 15 psi, from a displacement of 5.1 cc/rev at an estimated 245–327 rpm.

What are the wetted materials?+

A 316 stainless steel body, PEEK gears, a 440 stainless steel shaft and PTFE static O-rings.

What motor is fitted?+

A Baldor 0.5 HP, 1750 rpm, NEMA 56C motor — inverter duty and explosion proof for Class I Division 1, Group D — on 230/460 V, 3-phase, 60 Hz.

Which solvents is it rated for?+

The product data lists ethanol, heptane, methanol and 2-propanol.

What are the port connections?+

0.5 in NPT.

How is the drive coupled?+

Through a magnetic coupling rated at 4.0 Nm torque, which removes the dynamic shaft seal.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →