Answers are AI-generated from CASRAI’s own published pages and can be wrong, so check the linked sources before relying on one; your question is logged without personal data — never sold, never used to train a third-party model — to show us what CASRAI is missing, so please do not type personal or confidential details. How we use this

CASRAI is the reference for research administration — bookmark it for the next question.

Huber CC-902 with Pilot ONE

Huber

Huber CC-902 with Pilot ONE

in Chiller/Heater Combos

$36,300.00 - $36,800.00

In stock

Price varies by Certification.

Two-stage refrigerated bath circulator reaching −90 °C, working −90 to 200 °C at ±0.02 °C under Pilot ONE. R452A/R508B cascade, air cooled.

Certification

Choose Certification to continue.

Buying through procurement?

Generate a formal quotation for 1 × this item, plus our W-9 — everything your finance department needs to raise a purchase order.

Choose Certification to continue.

Description

The Refrigeration Bath Circulators are created to control the temperature of externally connected. Applications are: photometers, refractometers, viscometers, double-jacketed reaction vessels and autoclaves. the units can be used, for determining the freezing point, for deep temperature calibration, for petroleum tests, for temperature control of measuring devices and experimental setups as well as for materials testing and quality control.
Download Data Sheet
Click Here
Please observe the notes on installation and/or the space requirements of our products

General information & Temperature Controller

Temperature range

-90…200 °C

Temperature stability ±0.02 °C

±0.02 °C

Controller Pilot ONE (E-grade “Basic”)

Pilot ONE (E-grade “Basic”)

Safety class

III / FL

Protection class

IP20

Dimensions (W x D x H)

550 x 600 x 911 mm | 21.65 x 23.62 x 35.87 inches

Weight

139 kg | 306.5 pounds

Heating / Cooling Capacity

Heating capacity

1,2 kW

Cooling capacity

100 20 0 -20 -40 -60 -80 °C1,2 1,2 1,2 1,1 0,9 0,6 0,2 kW

Cooling machine

air-cooled

Refrigerant 1. stage

R452A (A1, H280) (0,78 kg)

Refrigerant 2. stage

R508B (A1, H280) (0,46 kg)

Circulation Pump

Pressure pump max.

25 lpm ; 0.7 bar | 6.6 gpm ; 10.15 psi

Suction pump max.

18.5 lpm ; 0.4 bar | 4.89 gpm ; 5.8 psi

Pump Connection

M16x1 AG male

Permissible viscosity

50 mm²/s

Connections

Pt100 connection

Pt100

Digital interface

RS232, Ethernet, USB Device, USB Host

Operating Data

Permissible ambient temp.

5…40 °C

Bath opening W x D / BD

120 x 110 / 150 mm | 4.72 x 4.33 / 5.91 inches

Bath volume

5 Liters | 1.32 gal

Power supply

208V 2~ 60Hz

Current max.

18 A

Fusing

2×20 A

Frequently asked questions

How does the two-stage refrigeration circuit work here?+

The unit uses a cascade: a first stage charged with 0.78 kg of R452A and a second stage charged with 0.46 kg of R508B. Both are classified A1 (non-flammable) under ASHRAE. The cascade is what allows a −90 °C bath temperature from an air-cooled machine.

What cooling capacity is available across the range?+

The data sheet gives 1.2 kW at 100 °C, 20 °C and 0 °C, 1.1 kW at −20 °C, 0.9 kW at −40 °C, 0.6 kW at −60 °C and 0.2 kW at −80 °C. Heating capacity is 1.2 kW.

How large is the bath?+

Bath volume is 5 litres, with an opening of 120 × 110 mm and a usable depth of 150 mm (4.72 × 4.33 in, 5.91 in deep). It is a small bath for the machine size — most of the footprint is the cascade refrigeration plant.

What are the electrical and space requirements?+

208 V 2~ 60 Hz, drawing a maximum of 18 A with 2 × 20 A fusing. The unit measures 550 × 600 × 911 mm (21.65 × 23.62 × 35.87 in) and weighs 139 kg (306.5 lb), so a bench check is worth doing before delivery.

What pump performance does it provide?+

25 lpm at 0.7 bar in pressure mode and 18.5 lpm at 0.4 bar in suction mode, through M16x1 male connections, with a permissible fluid viscosity of 50 mm²/s. Note that thermofluid selection matters at −90 °C: viscosity at the low end of the range governs achievable flow.

What controller does it use?+

Pilot ONE at E-grade "Basic" with a 5.7 in colour touchscreen, Pt100 external probe connection, and RS232, Ethernet, USB Device and USB Host interfaces for remote control or data capture.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →