Skip to main content
v2026.11,986 entries · CC-BY 4.0

Serial Dilution Technique: How to Perform and Calculate

How to perform and calculate a serial dilution step by step, including the C1V1=C2V2 formula, a worked tenfold dilution series, and back-calculating original concentration from colony counts.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

A serial dilution is a stepwise reduction in the concentration of a solution, produced by diluting a sample by the same factor at each of several successive stages rather than in a single dilution. It is one of the most common quantitative techniques in a research lab, used to prepare standard curves, count colony-forming units, titer viruses and antibodies, and bring concentrated stock reagents into a usable working range. Getting the math and the pipetting technique right matters: a single mislabeled tube or transposed factor propagates through every subsequent step and can silently invalidate an entire assay.

In short: to calculate a serial dilution, divide the total volume of each tube by the volume transferred into it to get the dilution factor per step, then raise that factor to the number of steps. Ten steps of 1:2 give 1:1,024; six steps of 1:10 give 1:1,000,000. Concentration in any tube equals the starting concentration divided by the cumulative dilution factor.

What a serial dilution is (and why not just dilute once)

A single, one-step dilution works fine when the required dilution factor is small — say, 1:10 or 1:20. But many lab applications need much larger dilution factors (1:1,000,000 or more, as in phage titering or bacterial colony counts) or need a whole series of intermediate concentrations at once (as in a standard curve for an ELISA or a spectrophotometric assay). Trying to measure out, for example, 1 microliter of stock into 999,999 microliters of diluent in a single transfer is not practical or accurate with standard lab pipettes, which are calibrated for a limited volume range and lose accuracy at their extremes.

A serial dilution solves this by breaking one large dilution into a chain of smaller, more pipettable steps. Each tube in the series becomes the source for the next, and the dilution factors multiply together to reach the overall target.

The core formula: C₁V₁ = C₂V₂

Every individual dilution step in a serial dilution — and every single dilution generally — is governed by the conservation-of-mass relationship:

C₁V₁ = C₂V₂

  • C₁ = concentration of the stock (starting) solution
  • V₁ = volume of stock solution you need to use
  • C₂ = concentration you want to end up with
  • V₂ = total final volume you want to make

This works because the total amount of solute (moles, or mass, or particles — whatever unit C is expressed in) doesn’t change when you add diluent; only the volume it’s dissolved in changes, so concentration and volume are inversely related and their product (the total quantity of solute) stays constant. To work out the stock concentration first, see molarity and solution calculations for the lab.

Worked example: preparing one dilution

Suppose you have a 1 M stock solution of NaCl and you need 50 mL of a 0.1 M working solution. Solve for V₁, the volume of stock you need:

C₁V₁ = C₂V₂
(1 M)(V₁) = (0.1 M)(50 mL)
V₁ = (0.1 M × 50 mL) / 1 M
V₁ = 5 mL

You would pipette 5 mL of the 1 M stock into a container and add diluent (typically water or buffer) to bring the total volume to 50 mL — not add 45 mL of diluent to the 5 mL, which would only be correct if the diluent volume itself totaled 50 mL after mixing. Always add stock first, then bring the total volume up to the target with diluent (often using a graduated cylinder or volumetric flask, or by adding diluent up to a calibration mark), rather than assuming stock-plus-diluent volumes sum without correction for solution non-ideality in more precise chemistry work.

How to perform a serial dilution, step by step

  1. Decide your dilution factor (DF) per step. The most common is a twofold (1:2) or tenfold (1:10) serial dilution, but any consistent ratio works. A 1:10 dilution means each tube is 1/10th the concentration of the tube before it.
  2. Label every tube with its dilution (e.g., 10⁻¹, 10⁻², 10⁻³ for a tenfold series, or 1:2, 1:4, 1:8 for a twofold series) before you start pipetting. This is the single most common source of downstream error — unlabeled or ambiguously labeled tubes.
  3. Aliquot diluent into every tube first. For a 1:10 series, put 900 µL of diluent in each tube if you’ll be transferring 100 µL; for a 1:2 series, put an equal volume of diluent in each tube as the transfer volume.
  4. Transfer from the stock into tube 1, mix thoroughly (pipette up and down or vortex — do not just tap), and change your pipette tip.
  5. Transfer the same volume from tube 1 into tube 2, mix, change tip, and repeat down the series. Always use a fresh tip between transfers; reusing a tip carries over concentrated solution and silently skews every downstream dilution, an error that is very difficult to detect after the fact because the results still look plausible.
  6. Stop at the dilution(s) you actually need and use those tubes for your assay, plate, or measurement.

Worked example: a tenfold serial dilution

You have a bacterial culture and want dilutions of 10⁻¹ through 10⁻⁶ for a colony-count assay.

  • Label six tubes 10⁻¹ through 10⁻⁶, each containing 9 mL of sterile diluent.
  • Transfer 1 mL of the original culture into tube 10⁻¹ (1 mL into 9 mL = 1:10 total, i.e. 10⁻¹). Mix well.
  • Transfer 1 mL from tube 10⁻¹ into tube 10⁻² (another 1:10 dilution of an already-1:10 solution = 10⁻² overall, or 1:100). Mix well, fresh tip.
  • Repeat the same 1 mL transfer down the chain through 10⁻⁶.

The overall dilution factor at any tube in the chain is the product of each individual step’s factor: after six consecutive 1:10 transfers, the cumulative dilution is 10⁻¹ × 10⁻¹ × 10⁻¹ × 10⁻¹ × 10⁻¹ × 10⁻¹ = 10⁻⁶, i.e. the culture in tube 6 is one-millionth the concentration of the original.

Calculating back to the original concentration

Serial dilutions are frequently used in reverse: you measure something in a diluted tube (for example, count colonies on an agar plate spread from a diluted sample) and need to back-calculate the concentration in the original, undiluted sample. The general relationship is:

Original concentration = (measured value in diluted sample) ÷ (dilution factor) ÷ (volume plated, if applicable)

Worked example: You plate 0.1 mL of the 10⁻⁴ dilution from the series above and count 85 colonies after incubation.

Colony-forming units per mL (CFU/mL) = colonies counted ÷ (dilution factor × volume plated)
CFU/mL = 85 ÷ (10⁻⁴ × 0.1 mL)
CFU/mL = 85 ÷ 0.00001
CFU/mL = 8.5 × 10⁶

In colony-count work, only plates with a countable range (typically 30–300 colonies) are used for this back-calculation; plates with too many or too few colonies are discarded because both very sparse and overcrowded plates introduce disproportionate counting error.

Serial dilution vs. a single dilution: when to use each

Situation Single dilution Serial dilution
Dilution factor needed Small (roughly up to 1:20–1:100) Large (1:1,000+) or multiple concentrations needed at once
Pipetting accuracy Good, if within pipette’s calibrated range Good at every step, because each transfer stays within a practical pipette range
Error propagation Limited to one step Compounds at every step — an error in an early tube affects every tube after it
Typical use Making a single working reagent from a concentrated stock Standard curves, titers, colony counts, viral/antibody dilution series

Common sources of error

  • Reusing pipette tips between tubes. Carryover of even a small volume of the more concentrated preceding tube skews every subsequent dilution in the series, and the resulting error looks like normal experimental noise rather than an obvious mistake.
  • Inadequate mixing. If a tube isn’t thoroughly mixed before the next transfer is taken, the aliquot won’t reflect the tube’s true average concentration.
  • Pipetting outside a pipette’s rated volume range. Micropipettes are least accurate near the bottom of their rated range; where possible, choose a pipette (or a transfer volume) so the volume being measured falls comfortably within the mid-to-upper portion of the instrument’s range.
  • Confusing “dilution factor” with “dilution ratio.” A “1:10 dilution” (1 part sample to 9 parts diluent, total 10 parts) is not the same statement as a “dilution factor of 10” applied loosely — be explicit about whether a ratio describes parts-of-sample-to-parts-of-diluent or sample-to-total-volume, and keep the convention consistent across a protocol.
  • Losing track of the cumulative dilution. Especially past four or five steps, it’s easy to mis-multiply the per-step factors. Writing the cumulative dilution on each tube’s label, not just the per-step factor, reduces this risk.
  • Not accounting for the diluent’s contribution to final volume precisely. For work that requires precise final concentrations (rather than approximate lab dilutions), use volumetric glassware and bring the total solution up to volume with diluent, rather than simply adding a diluent volume to a stock volume and assuming the sum is exact, since mixing certain solutions is not strictly additive in volume.

Dilution factor vs. dilution ratio

Lab protocols write dilutions several ways, and mixing them up is a common source of wrong concentrations. Use the total volume, not the diluent volume, when you calculate the per-step dilution factor.

Written as Parts sample : parts diluent Dilution factor (DF) Sample fraction Example with 1 mL total
1:2 dilution 1 : 1 2 1/2 0.5 mL sample + 0.5 mL diluent
1:4 dilution 1 : 3 4 1/4 0.25 mL sample + 0.75 mL diluent
1:10 dilution 1 : 9 10 1/10 0.1 mL sample + 0.9 mL diluent
1:100 dilution 1 : 99 100 1/100 10 µL sample + 990 µL diluent

For a series, the dilution factor per step is DF = (sample volume + diluent volume) ÷ sample volume, the concentration in tube n is Cn = C0 ÷ DFn, and the cumulative dilution is DFn. Some texts write “1/10” or “10⁻¹” for the same thing; the number that matters is always the total volume divided by the sample volume. The same arithmetic, applied to cell suspensions, is worked through in the cell dilution calculator guide, and for percent solutions in the percent solution dilution calculator.

Twofold (1:2) serial dilution: volumes and concentrations

A twofold series spaces concentrations closely, which makes it a common choice for microplate dilution rows. In a 96-well plate, one example row layout puts 100 µL of diluent in wells 2 to 8, adds 200 µL of the top-concentration solution to well 1, then moves 100 µL from well to well, mixing at each step. Starting from 1,000 µg/mL:

Well Step dilution Cumulative dilution Concentration (µg/mL)
1 none 1:1 1,000
2 1:2 1:2 500
3 1:2 1:4 250
4 1:2 1:8 125
5 1:2 1:16 62.5
6 1:2 1:32 31.25
7 1:2 1:64 15.63
8 1:2 1:128 7.81

Discard 100 µL from the last well so every well ends with the same 100 µL volume. For a series that must keep every tube at the same final volume, the transfer (“move”) volume is the minimum volume needed per tube divided by (DF − 1), as set out in the Omni Calculator serial dilution tool. Check the transfer volumes against your pipette’s rated range, as covered in pipette calibration and what a micropipette is.

Tenfold, half-log and quarter-log series

The tenfold series (a log dilution) is the usual choice for colony counts. When finer spacing is needed on a logarithmic axis, the Wikipedia entry on serial dilution describes half-log dilutions (a factor of about 3.16, which is 100.5) and quarter-log dilutions (about 1.78, which is 100.25). Two half-log steps equal one tenfold step, so a half-log series gives twice as many points per decade of concentration as a tenfold series. The same formula applies: Cn = C0 ÷ DFn, with DF = 3.16 or 1.78.

Serial dilution practice problems with answers

Problem 1: a 1:4 series

A serial dilution series is made in which each tube is diluted 1:4. Tube 1 holds undiluted sample at 6,142 mg/dL, and tubes 2 to 5 each receive a 1:4 transfer from the tube before. Divide by 4 at each step.

Tube Cumulative dilution Concentration (mg/dL)
1 undiluted 6,142
2 1:4 1,535.5
3 1:16 383.9
4 1:64 96.0
5 1:256 24.0

Problem 2: finding the dilution when diluent and transfer volumes are given

Eight tubes each contain 0.79 mL of diluent. 10 µL (0.010 mL) of serum is added to the first tube, mixed, and 10 µL is transferred through each remaining tube. What is the serum dilution in tube 7?

Total volume in each tube after a transfer is 0.79 mL + 0.01 mL = 0.80 mL, so DF per step = 0.80 ÷ 0.01 = 80. Tube 7 has been through seven steps: 807 = 2.097 × 1013, so the serum dilution is 1:2.097 × 1013.

Problem 3: from a plate count to the original concentration

A 0.1 mL aliquot of the 10-5 tube produces 42 colonies. CFU/mL = 42 ÷ (10-5 × 0.1 mL) = 4.2 × 107 CFU/mL in the original sample. If two plates from the same tube give 38 and 46 colonies, average them first (42) and then calculate.

When a dilution series is used to find an endpoint

Serial dilutions are not only for preparing standards. In susceptibility testing, a twofold series of an antibiotic is incubated with the organism and the lowest concentration that prevents visible growth is read as the endpoint; this is the basis of the broth microdilution method described in minimum inhibitory concentration (MIC). For protein quantification, a dilution series of a known standard is read alongside unknowns to build the curve, as shown in the Bradford protein assay standard curve guide.

Final dilution factor: sample preparation x series x plating

In a real assay the number you need is rarely the dilution of one tube. The dilution factor that matters is the product of every dilution between the original sample and the material that is actually measured: the initial sample preparation (for example, a 1:10 homogenate), each transfer in the series, and the fraction of the final tube that is plated or read. The Bacteriological Analytical Manual (BAM) expresses the last part as the volume of original sample represented on a plate: V = dilution ratio x inoculum.

Worked example. A solid sample is homogenized 1:10 (10-1). The homogenate is then diluted in three tenfold steps, so the tube labeled 10-4 holds the sample at 10-4 of its original concentration. If 0.1 mL of that tube is plated, the plate receives 10-4 x 0.1 mL = 10-5 mL of original sample. A count of 85 colonies therefore gives 85 / 10-5 mL = 8.5 x 106 CFU per mL (or per g, if the first step was weight-based) of the original sample. Plating 1 mL of the same tube instead would put 10-4 mL of sample on the plate, so the same count would mean 8.5 x 105 CFU/mL: the plated volume changes the answer by exactly its own factor of ten, which is why it must be written down with the dilution label.

The most frequent arithmetic slip is to use the dilution label alone and forget the plated volume. The tube label describes the concentration of the tube; the plated volume describes how much of the tube reached the medium.

Plate counts: which dilution to count, and what range is countable

For pour plates, only plates whose colony number falls inside a countable range are used for the back-calculation. Different standards bodies set that range differently. The FDA BAM Chapter 3 (Aerobic Plate Count, January 2026 edition) states the ranges as follows: 30 to 300 colonies per plate by USDA and AOAC, 25 to 250 by APHA, and 15 to 300 by ISO. BAM adopted 15 to 300 because ISO is internationally accepted in regulatory work; its revision history records that the suitable counting range changed from 25–250 to 15–300 in March 2025, and that the January 2026 edition corrected the worked examples. Always check which range your own method, accrediting body or regulator specifies before discarding plates. The data table below the article lists each range and the CFU/mL that a count at each limit represents for a 1:100 dilution with 1 mL plated.

The reason for an upper limit is crowding and overlapping colonies; the reason for a lower limit is that a count of a handful of colonies carries a large relative counting error. BAM assumes colony counts follow a Poisson distribution, and it uses a maximum-likelihood approach when plates that are too numerous to count are present.

Worked plate-count calculations following the BAM formula

BAM calculates the concentration as R = N / V, where N is the colonies counted and V is the volume of original sample on the plate (dilution ratio x inoculum).

  • One plate. 27 colonies on a plate from a 1:100 dilution with a 1 mL inoculum: V = 0.01 mL, so R = 27 / 0.01 = 2,700 CFU/mL.
  • Another plate. 150 colonies, V = 0.01 mL: R = 150 / 0.01 = 15,000 CFU/mL.
  • Two countable dilutions. BAM weights plates by the volume of original sample each one received. As an arithmetic illustration, 150 colonies from the 1:100 plate (V = 0.01 mL) and 16 colonies from the 1:1,000 plate (V = 0.001 mL) combine to (150 + 16) / (0.01 + 0.001) = 166 / 0.011, about 15,091 CFU/mL, reported as 15,000 CFU/mL. The plate that received more sample carries proportionally more weight, which is the same as pooling the counts and the volumes.

Technique controls from a standards method

The BAM aerobic plate count method fixes the physical handling around the dilution series, and those details translate to any plate-based dilution work:

  • Decimal dilutions are made by transferring 10 mL of the previous dilution into a 90 mL dilution blank (90 +/- 1 mL of Butterfield’s phosphate-buffered dilution water; 99 +/- 2 mL for milk), using a separate sterile pipet for each transfer and avoiding foam.
  • Each dilution is shaken 25 times in a 30 cm (1 ft) arc within 7 seconds, and re-shaken if it stands for more than 3 minutes before it is pipetted.
  • 1 mL is pipetted into duplicate dishes, and 12–15 mL of plate count agar at 45 +/- 1 °C is added within 15 minutes of the original dilution being prepared.
  • Plates are incubated for 48 +/- 2 hours at 35 °C.
  • Air in the work area is checked with plates left open for 15 minutes; BAM says it should not exceed 15 colonies per plate.

These numbers apply to the aerobic plate count of foods. A different organism or matrix has its own method, so treat them as an example of how tightly a regulatory method specifies handling rather than as universal settings.

Reporting plate-count results

BAM reports only the first two significant digits and rounds only at the final conversion to the aerobic plate count, not at intermediate steps, so 8,468 calculated from the counts is reported as 8,500. If no plate carries any colonies, the result is reported as less than 1 x the lowest dilution used, marked with an asterisk, rather than as zero. For the 1:100 dilution with 1 mL plated in the examples above, that would be less than 100 CFU/mL.

Frequently asked questions

What is a serial dilution?

A serial dilution is a series of sequential dilutions where each dilution uses the previous dilution as its starting material, applying the same (or a defined) dilution factor at each step. It’s used to reach very high dilution factors accurately, or to generate a range of concentrations for a standard curve or titer, without pipetting impractically small or large single-step volumes.

How do you calculate a serial dilution?

Decide the per-step dilution factor (for example, 1:10), then calculate the volume of the previous tube’s solution and the volume of diluent needed at each step using C₁V₁ = C₂V₂. The overall dilution at any point in the series is the product of all the individual step factors up to that point (for example, three consecutive 1:10 steps give an overall 1:1,000 dilution).

What is a dilution factor?

The dilution factor is the ratio of the final total volume to the volume of the original sample used — equivalently, how many times more dilute the new solution is than the source. A 1 mL sample brought to a 10 mL final volume has a dilution factor of 10 (often written 1:10).

Why is serial dilution used instead of one large dilution?

Because standard lab pipettes and volumetric equipment are only accurate within a defined volume range, extremely large single-step dilution factors (1:1,000,000 or more) would require measuring volumes too small to pipette accurately. Breaking the dilution into several smaller, accurately pipettable steps avoids this problem and also naturally produces the intermediate concentrations needed for a standard curve or titer.

What is the difference between a twofold and a tenfold serial dilution?

The difference is the per-step dilution factor: a twofold (1:2) series halves the concentration at each step (useful for generating a finely spaced titer, such as antibody or virus titers), while a tenfold (1:10) series reduces concentration by an order of magnitude at each step (useful for reaching very large overall dilution factors quickly, as in bacterial colony counts).

What does a 1:10 serial dilution mean?

Each step takes one part of the previous tube and adds nine parts of diluent, so the concentration falls tenfold per tube (for example 1 mL into 9 mL). After three steps the cumulative dilution is 1:1,000.

How do you find the concentration in each tube of a serial dilution?

Divide the starting concentration by the dilution factor once for every transfer made. In a twofold series from 1,000 µg/mL the tubes read 1,000, 500, 250, 125 and so on; in a 1:10 series from 1 M they read 1, 0.1, 0.01 and 0.001 M.

How do you calculate the total dilution factor?

Multiply the dilution factors of all the steps, or raise a constant factor to the power of the number of steps: six steps of 1:10 give 106. For CFU/mL, divide the colony count by the cumulative dilution (as a fraction, such as 10-5) multiplied by the volume plated in mL.

How many colonies should a plate have to be countable?

It depends on the method: 30–300 (USDA and AOAC), 25–250 (APHA) or 15–300 (ISO, adopted by FDA BAM Chapter 3 in March 2025).

Why do different standards give different countable ranges?

Each body sets its own balance between crowding at the high end and counting error at the low end. BAM adopted the ISO range because ISO is internationally accepted in regulatory affairs.

How do you calculate CFU/mL from a plate count?

Divide the colonies counted by the volume of original sample on the plate, which is the dilution ratio times the volume plated. 150 colonies from a 1:100 dilution with 1 mL plated gives 150 / 0.01 = 15,000 CFU/mL.

How long can a dilution stand before plating?

In the BAM method a dilution that has stood for more than 3 minutes is re-shaken, and the agar is added within 15 minutes of the original dilution being prepared.

What do you report if no plate has colonies?

BAM reports the result as less than 1 x the lowest dilution used, marked with an asterisk, not as zero.

Serial dilution results are only as trustworthy as the documentation behind them — recording exact dilution factors, lot numbers, and calculation steps at the bench is part of what makes an experiment reproducible rather than a one-off result nobody can retrace. Labs that log dilution series and calculations in an electronic lab notebook rather than loose paper worksheets make that retracing far easier, particularly in regulated or GxP environments where the calculation chain itself may need to be auditable.

Sources

Sources (accessed 2026-11-14):

Data: Countable colony ranges per pour plate by standards body, and the CFU/mL conversion factor they imply at 1 mL inoculum

Countable colony range per pour plate as attributed by the FDA BAM Chapter 3 (January 2026 edition) to USDA, AOAC, APHA and ISO, with BAM's own adopted range; the CFU/mL values are derived with the BAM formula V = dilution ratio x inoculum, R = N / V, for a 1:100 dilution and 1 mL inoculum.

Period covered: 2025/2026.

Countable colony ranges per pour plate by standards body, and the CFU/mL conversion factor they imply at 1 mL inoculum
Body or methodLower limit (colonies per plate)Upper limit (colonies per plate)CFU/mL at lower limit (1:100 dilution, 1 mL plated) (CFU/mL)CFU/mL at upper limit (1:100 dilution, 1 mL plated) (CFU/mL)
FDA BAM Ch. 3 (adopted, Jan 2026)15300150030000
ISO (as cited in BAM)15300150030000
USDA (as cited in BAM)30300300030000
AOAC (as cited in BAM)30300300030000
APHA (as cited in BAM)25250250025000

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Serial Dilution Technique: How to Perform and Calculate

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →