Title 21 of the Code of Federal Regulations Part 4 (21 CFR Part 4) establishes the United States Food and Drug Administration (FDA) regulatory requirements for combination products composed of drugs, medical devices, and biological products. Codified across Subpart A for Current Good Manufacturing Practice (CGMP) and Subpart B for Postmarketing Safety Reporting (PMSR), Part 4 provides a unified compliance framework that prevents regulatory duplication while ensuring safety and therapeutic performance.
This page describes regulatory requirements and general use for laboratory and manufacturing operations; it is not medical advice.
Regulatory scope and legal authority
Combination products represent some of the most clinically critical innovations in modern healthcare, ranging from prefilled insulin delivery pens and drug-eluting cardiovascular stents to transdermal delivery patches, surgical kits, and monoclonal antibody auto-injectors. Before the formal codification of 21 CFR Part 4, manufacturers faced substantial regulatory ambiguity because drugs, medical devices, and biological products have historically been governed by independent statutory provisions and administrative centers within the FDA. Pharmaceuticals are governed by human drug current good manufacturing practices under 21 CFR Parts 210 and 211, medical devices are regulated under the quality system rules of 21 CFR Part 820, and biological products are subject to licensing and standards under 21 CFR Parts 600 through 680 and Public Health Service (PHS) Act Section 351.
Congress and the FDA addressed this structural challenge under Section 503(g) of the Federal Food, Drug, and Cosmetic Act (FD&C Act), codified at 21 U.S.C. 353(g), which was substantially modernized by Section 3038 of the 21st Century Cures Act. Under 21 CFR Part 3, the FDA assigns premarket review and primary regulatory oversight of a combination product to a single lead center based on the product’s Primary Mode of Action (PMOA). The lead agency may be the Center for Drug Evaluation and Research (CDER), the Center for Biologics Evaluation and Research (CBER), or the Center for Devices and Radiological Health (CDRH). However, assignment to a lead center does not exempt a product from the safety and quality requirements governing its other constituent parts. Promulgated in final form at 78 FR 4321 (January 22, 2013) for Subpart A and 81 FR 92603 (December 20, 2016) for Subpart B, 21 CFR Part 4 establishes how quality management systems and pharmacovigilance programs must function across these intersecting disciplines.
Under 21 CFR 3.2(e) and 21 CFR 4.2, the FDA classifies combination products into three distinct operational architectures:
- Single-Entity Combination Products: Products whose constituent parts are physically, chemically, or otherwise combined or mixed into a single indivisible entity (21 CFR 3.2(e)(1)). Common examples include prefilled syringes, transdermal patches, auto-injectors containing liquid medication, drug-eluting coronary stents, and antimicrobial wound dressings.
- Co-Packaged Combination Products: Products where two or more separate constituent parts (such as a device and a drug) are packaged together in a single commercial package or unit (21 CFR 3.2(e)(2)). Examples include surgical first aid trays containing scalpel blades alongside antiseptic drug swabs, or an empty specialized transfer syringe co-packaged with a lyophilized drug vial and reconstitution diluent.
- Cross-Labeled Combination Products: Separately packaged constituent parts that are intended for use only with another specified drug, device, or biological product, where the labeling of both entities cross-references each other to achieve the intended health outcome (21 CFR 3.2(e)(3) and (4)). Examples include light-activated photodynamic oncology drugs labeled specifically for use with a dedicated wavelength laser delivery console.
Subpart A: Streamlined CGMP operating systems
The central operational breakthrough of 21 CFR Part 4 Subpart A is the establishment of a streamlined Current Good Manufacturing Practice (CGMP) operating system under 21 CFR 4.4(b). If an enterprise manufactures a single-entity or co-packaged combination product, attempting to maintain two completely separate, parallel quality systems—one meeting every clause of 21 CFR Part 211 and another meeting every clause of 21 CFR Part 820—creates massive administrative friction, redundant documentation, and conflicting operational audits. Section 4.4 resolves this by allowing manufacturers to select one primary quality system foundation and integrate a targeted subset of specific requirements from the other regulatory framework.
Under 21 CFR 4.4(a), manufacturers have two compliant pathways. Under 21 CFR 4.4(a)(1), a facility may choose to demonstrate full compliance with all applicable CGMP regulations across all constituent parts (for example, executing all requirements of 21 CFR 210/211 and all requirements of 21 CFR 820). Alternatively, under 21 CFR 4.4(a)(2) and 4.4(b), the facility may implement a streamlined operating system. The streamlined approach depends on which base operating system the facility maintains:
Drug-centric operating system with device provisions (§ 4.4(b)(1))
When a manufacturer of a drug-device or biologic-device combination product already maintains a mature pharmaceutical CGMP operating system compliant with 21 CFR Parts 210 and 211, the facility can establish compliance by incorporating exactly six specific provisions of the medical device quality system regulation (21 CFR Part 820):
- Management Responsibility (21 CFR 820.20): Executive leadership must establish and maintain an overarching quality policy, define organizational authority, and conduct periodic management reviews. In a combination product setting, management reviews must specifically evaluate device performance metrics alongside traditional drug batch release records.
- Design Controls (21 CFR 820.30): The facility must implement structured design controls across the product lifecycle. This includes formal design and development planning, documented design inputs (addressing patient user needs and intended clinical use), design outputs, design verification (confirming physical device mechanics such as injection time, glide force, and break-loose force), design validation under simulated clinical use conditions, design transfer to manufacturing, design change controls, and maintenance of a comprehensive Design History File (DHF).
- Purchasing Controls (21 CFR 820.50): The facility must evaluate and qualify component suppliers, establish clear purchasing specifications for molded parts, elastomers, and subassemblies, and establish documented incoming inspection protocols.
- Corrective and Preventive Action (CAPA) (21 CFR 820.100): The quality system must define procedures to investigate nonconformances, identify root causes, implement corrective solutions, and verify that changes do not adversely affect product safety or device functionality.
- Installation (21 CFR 820.170): Where a combination product requires dedicated installation at a clinical or laboratory site (such as specialized infusion pump drivers), installation procedures and inspection acceptance criteria must be documented.
- Servicing (21 CFR 820.200): For reusable delivery mechanisms or electromechanical hardware, servicing instructions, calibration records, and service report evaluations must be formally maintained.
Device-centric operating system with drug provisions (§ 4.4(b)(2))
Conversely, medical device facilities assembling combination products (such as automated drug delivery systems or drug-eluting orthopedic implants) that already operate under 21 CFR Part 820 may use their device quality management system as the baseline, provided they add eight specific pharmaceutical CGMP provisions from 21 CFR Part 211:
- Testing and Approval of Components and Closures (21 CFR 211.84): Representative sampling, chemical testing, identity verification, and formal quality unit approval or rejection of active pharmaceutical ingredients (APIs), excipients, primary drug containers, and elastomeric closures prior to manufacturing use.
- Calculation of Yield (21 CFR 211.103): Documented calculation of theoretical and actual yields at each manufacturing step to detect unexpected drug product loss or formulation deviations.
- Tamper-Evident Packaging (21 CFR 211.132): Compliance with tamper-evident barrier standards for over-the-counter (OTC) combination drug products to protect consumers against malicious tampering.
- Expiration Dating (21 CFR 211.137): Establishing expiration dating supported by rigorous stability testing that accounts for both drug chemical potency and device physical performance throughout the labeled shelf life.
- Testing and Release for Distribution (21 CFR 211.165): Laboratory batch testing to confirm that each production lot conforms to predetermined standards of identity, strength, quality, and purity before commercial release.
- Stability Testing (21 CFR 211.166): A documented stability testing program that subjects the finished combination product to real-time and accelerated storage conditions in its final container-closure system.
- Special Testing Requirements (21 CFR 211.167): Mandatory batch testing for sterility, bacterial endotoxins (pyrogens), and particulate matter for ophthalmic or injectable combination products.
- Reserve Samples (21 CFR 211.170): Archiving an adequate quantity of representative reserve samples from each commercial lot, retained in primary packaging for at least one year past product expiration.
Biological products and human cellular tissue requirements
Under 21 CFR 4.3(c) and 4.4(b)(3), if a combination product includes a biological constituent part, the manufacturer must satisfy the core biological product standards codified in 21 CFR Parts 600 through 680. These include biological manufacturing standards for safety, purity, and potency, as well as formal reporting of biological product deviations under 21 CFR 600.14. Similarly, under 21 CFR 4.3(d) and 4.4(b)(4), if the product includes Human Cells, Tissues, and Cellular and Tissue-Based Products (HCT/Ps), the operating system must incorporate Current Good Tissue Practice (CGTP) requirements under 21 CFR Part 1271, including strict donor screening, donor eligibility determinations, and recovery records.
Subpart B: Postmarketing safety reporting (PMSR) requirements
Subpart B of 21 CFR Part 4 (§§ 4.100 through 4.105) establishes the postmarketing pharmacovigilance and device vigilance requirements for combination products. In commercial distribution, adverse events may arise from a pharmacological reaction, a mechanical device malfunction, or an interaction between the drug and delivery mechanism. Prior to Subpart B, commercial sponsors were frequently uncertain whether to submit adverse event reports to CDER via the FDA Adverse Event Reporting System (FAERS) or to CDRH via Electronic Medical Device Reporting (eMDR), creating both underreporting and duplicative submissions.
Under 21 CFR 4.102, reporting obligations depend on whether an entity is a “combination product applicant” (holding the master marketing authorization for the entire product) or a “constituent part applicant” (holding an independent application for a specific constituent part). For combination product applicants, Subpart B establishes a unified submission rule: all postmarketing safety reports must be submitted through the electronic gateway associated with the application type under which the product was approved.
Mandatory individual case and expedited safety reports
To ensure comprehensive vigilance, 21 CFR 4.102(c) mandates that combination product applicants submit specific constituent part reports regardless of which application type was used for primary approval:
- Device Constituent Reports: If the combination product includes a device constituent part, the applicant must submit 5-day reports under 21 CFR 803.53 (for events requiring immediate remedial action to prevent substantial harm) and 30-day device malfunction reports under 21 CFR 803.50, as well as correction or removal reports under 21 CFR Part 806.
- Drug Constituent Reports: If the product includes a drug constituent part, the applicant must submit Field Alert Reports (FARs) under 21 CFR 314.81(b)(1) within 3 working days of identifying contamination, significant chemical breakdown, or bacteriological failure. In addition, 15-day expedited adverse drug experience reports under 21 CFR 314.80 must be submitted, though non-expedited ICSR reports are submitted on a 30-calendar-day cycle under Subpart B rules.
- Biological Constituent Reports: If the product incorporates a biologic, the applicant must submit Biological Product Deviation Reports (BPDRs) under 21 CFR 600.14 within 45 calendar days of acquiring information concerning any manufacturing deviation that may affect safety, purity, or potency.
The 5-calendar-day information sharing rule (§ 4.103)
A critical operational requirement that frequently catches sponsors off guard during FDA audits is the mandatory information sharing rule codified in 21 CFR 4.103. When separate corporate entities hold marketing authorizations for constituent parts of the same combination product (such as a pharmaceutical firm supplying a prefilled cartridge and a partner device firm supplying a reusable electronic injector), each applicant must share critical safety information with the other applicant within 5 calendar days of receipt. This requirement applies to any information involving a death or serious injury, or an adverse event report indicating a device malfunction that could lead to death or serious injury if recurring. The sharing entity must maintain records documenting the exact information provided, the date received, the date transferred, and the recipient’s corporate address.
Recordkeeping standards (§ 4.105)
Under 21 CFR 4.105, safety recordkeeping must follow the longest retention period specified among the individual regulations applicable to the constituent parts. For drug constituent parts, records of adverse drug experiences must be retained for at least 10 years. For medical devices, MDR records must be kept for at least 2 years from the date of the report or the expected life of the device, whichever is longer. Consequently, combination product sponsors must configure their electronic document management systems to maintain all Subpart B safety investigations and information-sharing logs for at least 10 years.
Impact of the 2024 FDA QMSR and ISO 13485 harmonization
On February 2, 2024, the FDA issued its landmark Final Rule titled “Medical Devices; Quality System Regulation Amendments” (89 FR 7496), establishing the Quality Management System Regulation (QMSR) with an effective date of February 2, 2026. The QMSR modernizes 21 CFR Part 820 by explicitly incorporating the international quality standard ISO 13485:2016 (“Medical devices — Quality management systems — Requirements for regulatory purposes”) by reference.
This structural change directly impacts 21 CFR Part 4 compliance for combination products. In the QMSR rulemaking, the FDA enacted conforming amendments to 21 CFR Part 4 Subpart A to align terminology and citations. Manufacturers operating streamlined systems under § 4.4(b)(1) must now map their device controls directly to ISO 13485:2016 clauses:
- Management Responsibility (§ 820.20): Maps directly to ISO 13485:2016 Clause 5 (“Management Responsibility”) and Clause 4.1.
- Design Controls (§ 820.30): Maps directly to ISO 13485:2016 Clause 7.3 (“Design and Development”). The FDA specifically preserved supplemental statutory requirements ensuring that design verification includes risk management principles aligned with ISO 14971.
- Purchasing Controls (§ 820.50): Maps directly to ISO 13485:2016 Clause 7.4 (“Purchasing”).
- CAPA (§ 820.100): Maps directly to ISO 13485:2016 Clause 8.5.2 (“Corrective Action”) and Clause 8.5.3 (“Preventive Action”).
- Installation and Servicing (§§ 820.170 and 820.200): Maps directly to ISO 13485:2016 Clause 7.5.1.2 and Clause 7.5.4.
For pharmaceutical manufacturers utilizing a streamlined drug-centric quality system, the QMSR transition requires reviewing their existing device Standard Operating Procedures (SOPs) to ensure that terminology such as Design History Files (DHF) and Device Master Records (DMR) aligns seamlessly with ISO 13485 Medical Device Files (MDF) while maintaining complete FDA compliance.
Laboratory and quality unit operations
Implementing 21 CFR Part 4 in laboratory and manufacturing environments requires active cross-training between analytical chemists, microbiologists, and mechanical quality engineers. In traditional pharmaceutical manufacturing, quality control laboratories focus predominantly on chemical identity, assay purity, dissolution rates, and microbiological sterility. However, when evaluating a combination product, the analytical release testing required under 21 CFR 211.165 must be integrated with mechanical and electromechanical physical testing.
Laboratories evaluating delivery devices must establish validated test methods for physical actuation parameters. For example, testing a prefilled autoinjector requires quantifying cap removal force, needle shield dislocation resistance, injection duration, dispensed dose accuracy, and needle safety shield lockout reliability upon completion of delivery. These mechanical attributes must be verified across lots under the same rigorous data integrity standards governed by electronic records and signatures (21 CFR Part 11).
Stability protocols under 21 CFR 211.166 also present unique challenges. Physical components can interact chemically with drug formulations over time. Elastomeric stoppers can leach organic extractables into aqueous biologics, silicone oil lubricants applied to glass syringe barrels can cause protein aggregation, and mechanical spring mechanisms under constant compression can suffer material fatigue, altering injection speed after 18 months of storage. Analytical protocols must therefore evaluate physical device functionality in parallel with chemical chromatographic purity at every stability time point.
Furthermore, quality units must coordinate premarket documentation files. Device-led combination products often rely on cleared premarket notifications documented in a 510(k) summary or high-risk premarket approval evaluations summarized in FDA SSED reports. Transitioning an early-stage clinical combination asset from initial investigational protocols to commercial manufacturing requires distinct documentation strategies when balancing investigational protocols against commercial licensing, as detailed in our guide on IND vs NDA pathways.
Common compliance mistakes and FDA inspection pitfalls
FDA Form 483 inspection observations and Warning Letters issued to combination product manufacturers frequently highlight recurring systemic vulnerabilities. Understanding these pitfalls allows quality teams to audit and reinforce their operating systems before regulatory inspections:
- Operating Siloed, Disconnected Quality Systems: Some manufacturers maintain a drug quality manual and a device quality manual that never cross-reference each other. During an inspection, investigators look for an integrated Combination Product Quality Manual that explicitly establishes the § 4.4 streamlined operating rationale.
- Treating Co-Packaged Standard Devices as “Exempt”: A widespread error occurs when a pharmaceutical firm co-packages a commercially available, 510(k)-cleared disposable syringe with an injectable drug vial. The manufacturer assumes that because the syringe is already cleared for commercial sale, no design controls are required. Under 21 CFR 4.4, co-packaging creates a combination product; the manufacturer must maintain design controls demonstrating that the syringe is chemically and physically compatible with the specific drug formulation, viscosity, and dosing volume.
- Inadequate Human Factors and Usability Validation: Under 21 CFR 820.30, design validation requires testing under actual or simulated use conditions. Investigators routinely issue observations when sponsors conduct benchtop mechanical testing but fail to perform rigorous human factors validation with representative patient cohorts (such as geriatric or rheumatoid arthritis patients testing autoinjector grip ergonomics).
- Failure to Meet the 5-Day Information Sharing Mandate: In co-development or licensing partnerships, contract manufacturing agreements often specify monthly or quarterly safety data reconciliation. This violates 21 CFR 4.103, which strictly mandates sharing death, serious injury, or malfunction information within 5 calendar days.
- Neglecting Container-Closure Integrity in Stability Programs: Testing only the chemical potency of a drug while neglecting physical container-closure integrity (CCI) testing under microbial ingress or helium leak detection protocols violates 21 CFR 211.166 when applied to combination prefilled systems.
Worked scenario: compliance framework for a prefilled autoinjector
To illustrate how 21 CFR Part 4 functions in practice, consider an illustrative composite scenario of a biopharmaceutical manufacturer developing a monoclonal antibody prefilled autoinjector for subcutaneous rheumatoid arthritis therapy:
- Jurisdiction and Application Pathway: Under 21 CFR Part 3, the Primary Mode of Action (PMOA) is achieved by the biological monoclonal antibody binding to target inflammatory cytokines. Therefore, CBER or CDER is designated as the lead center, and the premarket approval dossier is submitted as a Biologics License Application (BLA) under PHS Act Section 351.
- Quality Operating System Selection (§ 4.4(b)(1)): Because the commercial sponsor operates a licensed biologics manufacturing facility, it selects a streamlined CGMP operating system anchored in drug CGMPs (21 CFR Parts 210 and 211) and biologics standards (21 CFR Parts 600 through 680). It supplements this baseline by integrating the six mandatory device provisions of 21 CFR 820.20, 820.30, 820.50, 820.100, 820.170, and 820.200.
- Executing Design Controls (§ 820.30): The engineering and quality teams establish a Design History File (DHF). User needs specify that pediatric or arthritic patients must successfully self-administer a 1.0 mL dose in under 10 seconds without visual needle exposure. Design outputs define tight tolerances for spring force, glass barrel silicone coating thickness, and needle gauge. Verification protocols confirm injection kinetics across 500 autoinjectors at 2°C to 8°C and at ambient room temperature.
- Supplier and Purchasing Quality Controls (§ 820.50): The autoinjector device housing is molded by a specialized contract manufacturer. The sponsor audits the vendor, classifies the molded components as critical combination assemblies, and implements statistical incoming inspection plans measuring dimensional tolerances and spring constant metrics.
- Analytical Release and Stability Integration (§§ 211.165 and 211.166): Finished batch release testing incorporates high-performance liquid chromatography (HPLC) for monomer purity, cell-based bioassays for potency, and chromogenic endotoxin testing, combined with automated optical inspection for glass syringe microfractures and mechanical actuation testing. Stability chambers store samples in horizontal and vertical orientations to evaluate stopper lubricity over 24 months.
- Postmarketing Safety Architecture (Subpart B): The pharmacovigilance team configures its safety database to submit adverse event reports to FAERS. The system is programmed with automated business rules: any report of needle retraction failure or autoinjector jamming is classified as a device malfunction report, triggering a 30-day submission to the FDA and notification to the device molding partner within 5 calendar days.
The data table below compares the regulatory burden, quality base systems, and postmarketing reporting obligations across combination product architectures under 21 CFR Part 4.
Frequently asked questions
What is the difference between a single-entity and co-packaged combination product?
Under 21 CFR 3.2(e), a single-entity combination product physically or chemically integrates its constituent parts into an indivisible unit, such as a prefilled syringe or drug-eluting stent. A co-packaged combination product packages two or more distinct constituent parts together in a single retail package, such as a reconstitution kit containing a lyophilized drug vial and an empty specialized delivery syringe. Both architectures are eligible for streamlined CGMP operating systems under 21 CFR 4.4(b).
Can a manufacturer use a device quality system as the base for a drug-device combination product?
Yes. Under 21 CFR 4.4(b)(2), a manufacturer may operate under a medical device quality management system (21 CFR Part 820 / ISO 13485) as its baseline, provided it formally integrates eight specific drug CGMP provisions from 21 CFR Part 211: component testing (§ 211.84), yield calculation (§ 211.103), tamper-evident packaging (§ 211.132), expiration dating (§ 211.137), batch release testing (§ 211.165), stability testing (§ 211.166), special testing for sterility/pyrogens (§ 211.167), and reserve samples (§ 211.170).
What are the six device provisions required under a drug-based streamlined CGMP system?
Under 21 CFR 4.4(b)(1), a facility operating under a drug CGMP baseline (21 CFR Parts 210/211) must incorporate six specific device quality system requirements from 21 CFR Part 820: Management Responsibility (§ 820.20), Design Controls (§ 820.30), Purchasing Controls (§ 820.50), Corrective and Preventive Action (§ 820.100), Installation (§ 820.170), and Servicing (§ 820.200).
How does the 2026 FDA QMSR rule affect 21 CFR Part 4 compliance?
The FDA Quality Management System Regulation (QMSR) Final Rule (89 FR 7496), effective February 2, 2026, amends 21 CFR Part 820 by incorporating ISO 13485:2016 by reference. Conforming amendments to 21 CFR Part 4 harmonize device citations and terminology, meaning that streamlined operating systems incorporating device controls under § 4.4(b)(1) now align directly with ISO 13485 clauses for management, design development, purchasing, and CAPA.
What is the timeline for sharing adverse event information between constituent part applicants?
Under 21 CFR 4.103, when constituent parts are held by separate marketing applicants, an applicant that receives information regarding an event involving death, serious injury, or a recurring device malfunction must provide that information to the other constituent part applicant within 5 calendar days of receipt.
Are cross-labeled combination products eligible for streamlined CGMP systems under § 4.4(b)?
No. The streamlined CGMP operating system provisions under 21 CFR 4.4(b) apply strictly to single-entity and co-packaged combination products. Because cross-labeled constituent parts are manufactured, packaged, and distributed independently, each facility must comply fully with the standalone regulations applicable to its specific constituent part.
Sources
- U.S. Food and Drug Administration. 21 CFR Part 4: Regulation of Combination Products. Electronic Code of Federal Regulations (eCFR). https://www.ecfr.gov/current/title-21/part-4. Accessed 2026-11-14.
- U.S. Food and Drug Administration. Current Good Manufacturing Practice Requirements for Combination Products: Final Rule. Federal Register (78 FR 4321). https://www.federalregister.gov/documents/2013/01/22/2013-01068/current-good-manufacturing-practice-requirements-for-combination-products. Accessed 2026-11-14.
- U.S. Food and Drug Administration. Guidance for Industry and FDA Staff: Current Good Manufacturing Practice Requirements for Combination Products. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/current-good-manufacturing-practice-requirements-combination-products. Accessed 2026-11-14.
- U.S. Food and Drug Administration. Guidance for Industry and FDA Staff: Postmarketing Safety Reporting for Combination Products. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/postmarketing-safety-reporting-combination-products. Accessed 2026-11-14.
- U.S. Food and Drug Administration. Medical Devices; Quality System Regulation Amendments (QMSR Final Rule). Federal Register (89 FR 7496). https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments. Accessed 2026-11-14.
- U.S. Food and Drug Administration. Guidance for Industry: Principles of Premarket Pathways for Combination Products. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/principles-premarket-pathways-combination-products. Accessed 2026-11-14.
Data: 21 CFR Part 4 Regulatory Requirements by Combination Product Architecture
Comparative analysis of Current Good Manufacturing Practice (CGMP) base operating systems, mandatory bridging regulations under 21 CFR 4.4(b), lead center jurisdiction under 21 CFR Part 3, and postmarketing safety reporting (PMSR) pathways under 21 CFR Part 4 Subpart B, compiled directly from 21 CFR Part 4, FDA combination product guidance documents, and the 2024 QMSR Final Rule.
Period covered: 2013/2026. Coverage: United States.
| Combination Architecture | Base CGMP System | Mandatory Bridging Regulations | FDA Lead Center | Primary PMSR Pathway | Cross-Applicant Sharing Window (days) |
|---|---|---|---|---|---|
| Drug-Device Single-Entity (Pharma-Base) | 21 CFR Part 210/211 (Drug CGMP) | 21 CFR 820.20, 820.30, 820.50, 820.100, 820.170, 820.200 | CDER or CBER | FAERS (15-day alert / 30-day ICSR) plus 5-day / 30-day MDR | 5 |
| Drug-Device Single-Entity (Device-Base) | 21 CFR Part 820 / ISO 13485 (QMSR) | 21 CFR 211.84, 211.103, 211.132, 211.137, 211.165, 211.166, 211.167, 211.170 | CDRH | eMDR (30-day malfunction / 5-day report) plus 15-day / 30-day ICSR | 5 |
| Biologic-Device Co-Packaged Product | 21 CFR Parts 600-680 & 210/211 | 21 CFR 820.20, 820.30, 820.50, 820.100, 820.170, 820.200 plus BPDR under 21 CFR 600.14 | CBER | FAERS or eMDR plus BPDR within 45 days | 5 |
| HCT/P Combination Product | 21 CFR Part 1271 (GTP) & Base System | 21 CFR Part 1271 Subparts C/D plus applicable 21 CFR 4.4(b) provisions | CBER or CDRH | FAERS or eMDR plus 21 CFR 1271.350 adverse reaction reporting | 5 |
| Cross-Labeled Combination Product | Separate Independent Systems | Not eligible for 4.4(b) streamlined system; full compliance for each part | Assigned by PMOA under 21 CFR Part 3 | Individual application gateways (FAERS for drugs, eMDR for devices) | 5 |
Sources
- eCFR Title 21 Part 4: Regulation of Combination Products (accessed 2026-11-14)
- FDA Final Guidance: Current Good Manufacturing Practice Requirements for Combination Products (accessed 2026-11-14)
- FDA Final Guidance: Postmarketing Safety Reporting for Combination Products (accessed 2026-11-14)
- FDA Quality Management System Regulation (QMSR) Final Rule (89 FR 7496) (accessed 2026-11-14)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.