Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General product and regulatory information only; this page is not medical advice and does not describe any surgical procedure.
Short answer
An acetabular reamer is an orthopedic cutting instrument whose name points to the acetabulum, the hip socket. In FDA’s database, a device named “Reamer” carries product code HTO and a device named “bit, drill” carries HTW. Both sit under 21 CFR 888.4540, are Class I and are exempt from 510(k) premarket notification (FDA product classification database, checked 2026-10-30).
Definition and scope
“Acetabular reamer” is a descriptive trade name, not an FDA device type of its own. The regulation that covers reamers, 21 CFR 888.4540, is titled “Orthopedic manual surgical instrument.” It defines that device as “a nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery,” and its list of included generic types names “reamer” (eCFR, version dated 2026-10-01, checked 2026-10-30).
The same list also names the drill brace, countersink, burr, trephine and file, which is why a reamer is best understood as one member of a boring and shaping group within the broader orthopedic surgical instruments family. The text as written describes nonpowered devices. A powered reamer or drill is not covered by that wording, so a powered product has to be checked against its own FDA listing rather than assumed to share this classification.
How the term is used in classification and catalogues
The regulation does not use the shop-floor names people search for. Catalogues commonly label items with the terms below, but FDA’s database does not treat any of them as a separate device type.
Hip reamer, bone reamer and acetabular reamer
These are descriptive names for a reamer used on bone. “Reamer” is the generic type named in 21 CFR 888.4540; “hip” and “acetabular” describe the anatomical site the product is named for, and “bone” the tissue.
Flexible reamer and cannulated reamer
“Flexible” and “cannulated” (hollow-centred) describe a design feature of a product. Neither word appears in the regulation text or in the FDA product code names for HTO or HTW, so a feature name alone does not tell you the FDA classification of an item.
Orthopedic drill, orthopedic drill bit and cannulated drill bit
FDA’s database lists a “bit, drill” under HTW with the same regulation, 888.4540. The regulation text itself does not use the words “drill bit”; it names the drill brace and countersink among its examples. The product code is therefore the reliable identifier for a drill bit, not the regulation’s example list.
Comparison: reamer (HTO) and drill bit (HTW) in FDA’s database
| Field | Reamer | Bit, drill |
|---|---|---|
| Product code | HTO | HTW |
| Regulation | 21 CFR 888.4540 | 21 CFR 888.4540 |
| Regulation description | Orthopedic manual surgical instrument | Orthopedic manual surgical instrument |
| Device class | Class I | Class I |
| 510(k) | Exempt | Exempt |
Sources: FDA Product Classification database entries for HTO and HTW, checked 2026-10-30.
Where it matters for labs, clinics and supply decisions
For a staff member checking a catalogue entry, the points below come directly from the FDA records and the regulation. They are a records check, not a purchasing recommendation.
- Match the product code, not the shop name. HTO identifies a reamer and HTW a drill bit; the descriptive words “acetabular”, “hip”, “flexible” or “cannulated” do not change the FDA type.
- Read the regulation wording. 888.4540 describes a nonpowered hand-held device. A powered item needs its own listing check.
- Exemption has limits. The exemption from 510(k) is “subject to the limitations in 21 CFR 888.9”, and the FDA database page advises confirming exemption limits under 21 CFR 862 through 892.
- Exempt does not mean unregistered. FDA’s database page notes that exempt Class I manufacturers must still register their establishments.
- Class I means general controls. Per 888.4540(b), the classification is Class I (general controls).
For neighbouring instruments in the same regulation group, see the CASRAI pages on the trephine, osteotome and bone file. Related background is on the CASRAI page for the Medical Device Amendments of 1976.
Common misunderstandings
- “Acetabular reamer is its own FDA device type.” It is not; the FDA generic type is “reamer” under 888.4540.
- “A drill bit and a reamer are different regulations.” In FDA’s database both codes map to the same regulation and class.
- “Class I and 510(k) exempt means no FDA obligations.” General controls still apply and the regulation’s exemption is bounded by 21 CFR 888.9.
- “All reamers and drills fall under 888.4540.” The regulation text describes nonpowered hand-held devices.
FAQ
Is a reamer a Class I device?
Yes. FDA lists the device “Reamer” (HTO) as Class I under 21 CFR 888.4540 and exempt from 510(k) notification, subject to the limitations in 21 CFR 888.9 (checked 2026-10-30).
Is a drill bit the same FDA device type as a reamer?
No, they have different product codes (HTW for “bit, drill”, HTO for “Reamer”), but both are listed under 21 CFR 888.4540 as Class I and 510(k) exempt.
Is a hip reamer the same as an acetabular reamer?
The two names are descriptive, not FDA categories. Neither appears in the regulation; both describe a reamer, the generic type that does.
Does the regulation cover powered reamers?
The text of 888.4540 describes a “nonpowered hand-held device”. A powered product should be checked against its own FDA listing.
Sources
- FDA, Product Classification database, product code HTO (Reamer), accessdata.fda.gov, checked 2026-10-30.
- FDA, Product Classification database, product code HTW (bit, drill), accessdata.fda.gov, checked 2026-10-30.
- 21 CFR 888.4540, Orthopedic manual surgical instrument, eCFR, version dated 2026-10-01, checked 2026-10-30.