Anesthesia Machine Diagram: Labeled Parts and How Gas Flows Through Them

An anesthesia machine diagram read left to right: supply, processing, delivery and disposal, which labeled part does what, and which checkout items test it.

Written and maintained by CASRAI Editorial Board

Published · Last updated

This page describes products and general use; it is not medical advice.

On this page

An anesthesia machine diagram is easiest to read as a path: gas enters from a supply, is controlled and mixed, is delivered through a breathing circuit, and is finally taken away. Every labeled box on a schematic belongs to one of those four stages. Once you can place a label in its stage, an unfamiliar diagram stops being a list of names. For the regulatory definition and FDA class table, see what an anesthesia machine is; this page is the reading guide for the picture itself. Who uses it: clinic and biomedical staff, buyers comparing equipment documentation, and students.

Anesthesia machine diagram at a glance: four stages

The University of Detroit Mercy course “The Anesthesia Gas Machine” (Dosch and Tharp, revised February 2026) teaches the machine as a Supply, Processing, Delivery and Disposal model, and its chapters follow that order: gas sources and electricity, then fail-safe devices, flowmeters, hypoxic guard and vaporizers, then breathing circuits, carbon dioxide absorption and ventilators, then scavenging of waste anesthetic gases. A diagram of an anesthesia machine drawn left to right usually follows the same logic, even when the artist does not label the stages.

Parts on the diagram, stage by stage

Stage Label you will see What it is on the diagram
Supply Pipeline inlets and cylinders, pressure gauges Where oxygen and other gases enter, from wall pipelines or cylinders; gauges show supply pressure. A spare oxygen cylinder is also drawn as a source.
Supply Power, battery Mains power and a battery backup for the electrical and electronic parts of the machine.
Processing Flowmeters, flow control valves Where flow of each gas is set; the oxygen and nitrous oxide flowmeters read separately.
Processing Fail-safe, hypoxic guard Safety devices that the Detroit Mercy course groups under Processing with the flowmeters; Goneppanavar and Prabhu describe an oxygen pressure fail-safe mechanism.
Processing Vaporizer Holds liquid volatile anesthetic agent and has a filler cap; it is a separate chapter in the Detroit Mercy course.
Processing Oxygen flush An oxygen flush control, drawn as a button or knob on the front of the machine.
Delivery Breathing circuit, reservoir bag The tubing, bag and connections carrying gas to and from the patient (a circle system in many machines).
Delivery CO2 absorber In a circle system, the canister of absorbent that removes carbon dioxide from exhaled gas.
Delivery Ventilator The mechanical ventilator, which many diagrams show in place of the bag. See what a ventilator is.
Disposal Scavenging system Carries waste anesthetic gas away from the room; the Detroit Mercy course covers scavenging and waste anesthetic gases as its own section.

Sources (checked 2026-11-03): Detroit Mercy course structure; Goneppanavar and Prabhu, Indian J Anaesth 2013 (PMC3821271); 21 CFR 868.5160. Layouts differ between manufacturers.

Diagram of an anesthesia machine versus the FDA definition

Two diagrams of “an anesthesia machine” can legitimately differ. Under 21 CFR 868.5160(a)(1), a gas machine for anesthesia is a device used to administer a general inhalation anesthetic to a patient, continuously or intermittently, and to maintain the patient’s ventilation, and it “may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply.” The word is “may”: a schematic of a simple machine can omit a ventilator, and a modern workstation adds electronics. The regulation classifies the device as Class II (performance standards). The openFDA classification record for product code BSZ, “Gas-Machine, Anesthesia”, lists the same regulation number, 868.5160, and class 2 (checked 2026-11-03). Part (b) covers analgesia machines, which deliver an analgesic such as a nitrous oxide-oxygen mixture with a maximum of 70 percent nitrous oxide; their diagram is shorter still.

Which checkout items test which part of the diagram

Goneppanavar and Prabhu (2013), a review of the anesthesia machine, also describe pre-use checking. They state that no single checklist can satisfactorily test every anesthesia machine because of their complexity and design differences between manufacturers, so the table below maps general check areas to diagram blocks. The mapping is editorial; it is not a checkout procedure, and a facility follows its own checklist and the manufacturer’s instructions for use.

Check area named by Goneppanavar and Prabhu Diagram block When
Cylinder contents and pipeline connections, pressure gauge readings Supply Start of use
Mains connection, battery backup, monitors Supply (electrical) Start of use
Oxygen and nitrous oxide flowmeters, fail-safe behavior, oxygen flush Processing Start of use
Vaporizer agent level and closed filler cap Processing Start of use
Fresh CO2 absorbent, circuit connections, leak test of the breathing system Delivery Start of use; absorbent also between cases
New equipment for the next patient, such as suction tubing and breathing circuitry Delivery Between cases
Scavenging system function Disposal Start of use

Goneppanavar and Prabhu also state that between two anesthetics the availability of fresh absorbent and volatile agent should be verified, and that new equipment intended for the next patient should be tested. A schematic helps here because a missing stage on the diagram is a check area that is easy to forget.

Disposal and delivery on the diagram

The Disposal stage is often drawn as a small box at the edge of the picture, but it is a stage of its own in the Detroit Mercy teaching model. The Delivery stage is where single-use accessories connect; see anesthesia circuits for the product category, and what an endotracheal tube is and what capnography is for the airway and monitoring parts that sit beyond the machine’s own diagram.

Common mistakes when reading an anesthesia machine diagram

  • Treating one schematic as every machine. The regulation says a machine “may include” certain parts, and Goneppanavar and Prabhu note design differences between manufacturers.
  • Reading a labeled picture as a checklist. A checkout covers what to verify and is specific to the machine and facility.
  • Forgetting the Disposal stage because it is drawn small.
  • Confusing an anesthesia gas machine with an analgesia (nitrous oxide-oxygen) machine, which 21 CFR 868.5160 treats as a separate device in part (b).

FAQ

What are the main parts of an anesthesia machine diagram?

Gas supply (pipelines, cylinders, gauges, power), flowmeters with fail-safe devices, a vaporizer, a breathing circuit with bag or ventilator and CO2 absorber, and a scavenging system. The regulation lists the flowmeter, vaporizer, ventilator, breathing circuit with bag and emergency air supply as parts a machine may include.

Is there one standard diagram of an anesthesia machine?

No. The FDA definition is written as “may include”, and published checkout guidance notes that no single checklist suits every machine, so layouts vary by manufacturer and model.

How does gas flow through an anesthesia machine?

From supply, through flow control and the vaporizer, to the breathing circuit and patient, and then to disposal through scavenging.

Is a diagram of an anesthesia machine the same as a checkout checklist?

No. A diagram shows where each part sits in the gas path. A checkout checklist is specific to the machine and the facility, and follows the manufacturer’s instructions for use.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Anesthesia Machine Diagram: Labeled Parts and How Gas Flows Through Them

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →