General product and standards information for lab and clinic staff, not medical advice.
Short answer
ASTM F1671 is a laboratory test method that checks whether a protective clothing material, or a seam in it, lets a virus-sized challenge through under continuous liquid contact. It uses the Phi-X174 bacteriophage as a stand-in for bloodborne viruses such as hepatitis B, hepatitis C and HIV, and a pass means no viral penetration was detected. It tests material performance, not a whole garment’s design.
Definition and scope
The standard is titled “Resistance of Materials Used in Protective Clothing to Penetration by Blood-Borne Pathogens Using Phi-X174 Bacteriophage Penetration as a Test System”. The edition listed as active on the ASTM store (checked 2026-10-30) is F1671/F1671M-26; earlier editions (for example -22, -13 and -07) exist, so check which edition a manufacturer cites.
- What it measures: resistance of protective clothing materials to penetration by blood-borne pathogens, using a surrogate microbe and continuous liquid contact. Pass or fail rests on whether viral penetration is detected.
- What it models: penetration by hepatitis B, hepatitis C and HIV. Other pathogens are considered case by case.
- What it covers: material performance or certain constructions such as seams. It does not address overall garment design.
- What it is used on: specimens cut from finished items such as gloves and gowns, and candidate materials.
How the test is used in practice
ASTM describes the method as built on Test Method F903, the liquid-penetration method for chemical protective clothing. The specimen sits in a penetration cell and is challenged with a suspension of the Phi-X174 phage. The suspension’s surface tension is adjusted to about 0.042 N/m to approximate blood, and the cell is pressurised to 13.8 kPa (2 psig). Because liquid alone can pass unseen, ASTM states that visual detection is not sufficient, so a biological assay for the virus is used to decide pass or fail.
NIOSH (CDC) describes the outcome the same way: the bacteriophage serves as a surrogate for human pathogenic viruses, and materials pass when no liquid is observed to penetrate the specimen and no bacteriophage is detected.
How F1671 relates to F1670 and F903
Three ASTM methods are often named together. Based on the ASTM F1671 scope page and the NIOSH standards page:
| Method | What it assesses | Role relative to F1671 |
|---|---|---|
| ASTM F903 | Liquid chemical penetration through protective clothing materials | Source of the test conditions F1671 is based on; its optional techniques (alternate challenge fluids, fluorescent dye) are useful for screening and failure analysis |
| ASTM F1670 | Resistance of a material to synthetic blood penetration | A screening test. ASTM says passing it does not replace the need to run F1671; materials that pass F1670 should then be tested with F1671 |
| ASTM F1671 | Viral penetration, using Phi-X174 bacteriophage as the surrogate | The confirmation step, with pass or fail based on detecting viral penetration |
What a pass does and does not show
A claim that a material “meets F1671” is narrower than it sounds. Read against the standard’s own scope notes, a pass tells you less than a garment-level claim would:
- Material or seam, not garment design. The method covers material performance and certain constructions such as seams, not overall design.
- Test pressure. The cell is pressurised to 13.8 kPa. ASTM notes that mechanical pressures above 345 kPa (50 psig) can occur in actual clinical use and that the test does not simulate every real-use stress.
- Absorbent inner liners. ASTM cautions the method is not always effective for materials with thick inner liners that absorb the assay fluid.
- Pathogens beyond HBV, HCV and HIV. The surrogate models those three viruses; other pathogens are handled case by case.
- Edition. The claim is only as specific as the edition cited.
Where it matters for labs, clinics and supply decisions
Gown and drape documentation often cites F1670 and F1671 together. The site’s isolation and surgical gown selection guide describes Level 4 gowns as relying on F1670/F1671, while Levels 1 to 3 rely on AATCC 42 and 127. NIOSH notes that ANSI/AAMI PB70 classifies protective apparel by liquid barrier performance and was accepted by the FDA in 2004.
For staff reading a specification sheet, the limits above translate into questions the document itself should answer:
- Which edition of F1671 is cited, and was it run on the material, the seams, or both?
- Is F1670 data being presented as if it were F1671 data?
- Does the sheet name Phi-X174 as the surrogate and state the result as no viral penetration detected?
For the wider exposure-control context in research settings, see the OSHA bloodborne pathogens standard in research laboratories and the overview of PPE types and how OSHA and CDC say to choose.
Common misunderstandings
Is passing F1670 the same as passing F1671?
No. F1670 uses synthetic blood and is a screening step; ASTM says a pass does not replace F1671.
Can a visible leak check stand in for the virus test?
No. ASTM states that visual detection alone is insufficient, which is why a biological assay is used.
Does F1671 use MS2 phage?
No. The standard names Phi-X174 as the surrogate, so a document naming another phage is not describing this method.
Does a pass certify the whole gown?
No. The method addresses materials and certain constructions such as seams, not overall garment design.
FAQ
What does ASTM F1671 test?
It tests whether a protective clothing material resists penetration by a surrogate virus (Phi-X174 bacteriophage) under continuous liquid contact, with pass or fail based on detecting viral penetration.
What is the difference between ASTM F1670 and F1671?
F1670 evaluates resistance to synthetic blood penetration and serves as a screen. F1671 uses a bacteriophage to test for viral penetration and is not replaced by an F1670 pass.
What is ASTM F903?
F903 assesses liquid chemical penetration through protective clothing materials. F1671 is based on it and takes its test conditions from it.
Sources
- ASTM International, F1671/F1671M-26 (active): https://store.astm.org/f1671_f1671m-26.html, and F1671-07 scope page: https://store.astm.org/f1671-07.html. Checked 2026-10-30. Summarised, not reproduced.
- NIOSH / CDC, Standards and Specifications for protective clothing: https://www.cdc.gov/niosh/healthcare/protective-clothing/standards.html. Page updated 2026-10-28; checked 2026-10-30.