Written and maintained by CASRAI Editorial Board
Published · Last updated
This page describes products and general use; it is not medical advice.
On this page
“Autoclavable gloves” is used for three different products: gloves worn to take hot items out of an autoclave, reusable gloves that can themselves go through a steam cycle, and single-use exam gloves that one laboratory study autoclaved as a stand-in for sterile surgical gloves. They are not interchangeable, and the first thing to settle is which one you mean. Heat-resistant autoclave gloves are used by laboratory and clinic staff who unload sterilizers; the exam-glove evidence below concerns laboratory rodent surgery only.
What “autoclavable gloves” means
An autoclave sterilizes with pressurized steam. Gloves come into the picture in two ways: the person unloading the machine needs hand protection from heat, and some workflows want a glove that has itself been sterilized. A glove that protects against heat is not necessarily one that survives steam sterilization, and the reverse is also true. Product pages for heat-resistant gloves describe what the glove protects the wearer from, not whether the glove itself is sterilized.
Types and variants
| Type | Typical use | What the source states |
|---|---|---|
| Heat-resistant cotton terrycloth gloves (example: Bel-Art Clavies biohazard autoclave gloves) | Removing hot objects from an autoclave or oven | Uncoated, latex-free cotton terrycloth; protects from temperatures up to 450 °F (232 °C); not waterproof; not for handling dry ice. The product page does not say the gloves are themselves autoclaved. (Thermo Fisher product page, checked 2026-10-30) |
| Single-use nitrile exam gloves, autoclaved | Substitute for sterile surgical gloves in laboratory rodent surgery (one study) | 270 °F for 4 minutes with a 20-minute dry time; 98% of autoclaved nitrile exam gloves met performance criteria and none failed to don. (LeMoine et al., JAALAS 2015) |
| Single-use latex exam gloves, autoclaved | Same study | 76% met performance criteria; 23% could not be donned after autoclaving. (LeMoine et al., JAALAS 2015) |
What the autoclaved exam-glove study found
The peer-reviewed evidence on putting exam gloves through an autoclave cited here is a 2015 performance analysis of exam gloves used for aseptic rodent surgery. The authors compared autoclaved and disinfected exam gloves with sterile surgical gloves. The exam gloves were thinner: nitrile 0.09 mm and latex 0.14 mm, against 0.15 mm and 0.22 mm for sterile nitrile and latex surgical gloves.
- Autoclaved nitrile exam gloves: 98% passed visual inspection, stretch tests and simulated suturing; sterile nitrile surgical gloves: 88%.
- Autoclaved latex exam gloves: 76% passed; sterile latex surgical gloves: 100%.
- In a pressure test, autoclaved nitrile exam gloves peaked at 20.50 cm H2O versus 30.10 cm H2O for sterile nitrile surgical gloves.
- Surface disinfection with HP-PA gave a 94% pass rate, 70% isopropyl alcohol 47%.
The authors concluded the results support HP-PA-disinfected latex and nitrile exam gloves, or autoclaved nitrile exam gloves, as viable cost-effective alternatives to sterile surgical gloves for rodent surgeries. In vivo outcomes were not evaluated, and the authors say facilities should monitor postoperative complication rates. The finding is specific to laboratory rodent surgery and should not be read as covering human clinical use or exam gloves in general.
Sizes and specifications that matter
| Specification | Value (Bel-Art Clavies, Thermo Fisher listing) |
|---|---|
| Maximum temperature | 450 °F (232 °C) |
| Style | Full-hand mitt with gauntlet cuff |
| Overall length | 18.5 in (47 cm) |
| Cuff length | 11 in (28 cm) |
| Size | Universal |
| Care | Machine wash or air dry |
Nitrile exam gloves: what the term covers
Nitrile exam gloves are disposable gloves made from synthetic nitrile rubber and sold as patient examination gloves. FDA classifies a patient examination glove as a disposable device worn on the examiner’s hand or finger to prevent contamination between patient and examiner, in Class I (general controls) under 21 CFR 880.6250, and regulates medical gloves as Class I reserved devices that need a 510(k) premarket notification. FDA’s glove guidance points manufacturers to ASTM D6319, the standard specification for nitrile examination gloves, or an equivalent consensus standard, and to ASTM D5151 for detecting holes. FDA also requires manufacturers to identify the materials on the package labeling, and its consumer page names nitrile, polyvinyl chloride (PVC) and polyurethane as synthetic alternatives to natural rubber latex. (FDA, Medical Gloves; FDA Medical Glove Guidance Manual; 21 CFR 880.6250, checked 2026-10-31.) See the guides to types of gloves, gloves for latex allergy and whether nitrile exam gloves are food safe.
FDA’s consumer page on medical gloves also says to never reuse, wash or disinfect medical gloves. That sits alongside, not inside, the laboratory rodent-surgery study above: the study tested autoclaving as a cost-saving substitute in an animal-research setting and did not evaluate in vivo outcomes, so it is not evidence that exam gloves are cleared for steam sterilization.
Sterile surgical gloves: how they differ from exam gloves
FDA defines a surgeon’s glove as a device worn on the hands of operating room personnel to protect a surgical wound from contamination, Class I under 21 CFR 878.4460. Surgeon’s gloves are expected to be sterile when offered to hospitals, clinics and surgeons, and FDA’s guidance states that it has not cleared any 510(k) for non-sterile surgeon’s gloves. FDA recommends sterilizing medical gloves to a sterility assurance level of 10-6, meaning a 1 in 1,000,000 probability that a glove is non-sterile, with a validated sterilization process, and notes that some glove materials are adversely affected by certain sterilization methods. The physical-properties standard FDA names for surgical gloves is ASTM D3577 (rubber surgical gloves), as opposed to ASTM D3578 (rubber examination gloves) and ASTM D6319 (nitrile examination gloves). (FDA Medical Glove Guidance Manual; 21 CFR 878.4460, checked 2026-10-31.)
| Feature | Patient examination glove | Surgeon’s glove |
|---|---|---|
| Regulation | 21 CFR 880.6250, Class I | 21 CFR 878.4460, Class I |
| Stated purpose | Prevent contamination between patient and examiner | Protect a surgical wound from contamination |
| Sterility | Supplied sterile or non-sterile | Expected to be sterile when offered to end users |
| ASTM specification named by FDA | D3578 (rubber), D6319 (nitrile) | D3577 (rubber surgical) |
For sizing and donning of sterile gloves see surgical gloves sizes and how to put on sterile gloves. This is general product information, not a statement of what suits a particular procedure.
How to choose
The selection rule that applies is regulatory. Under OSHA 29 CFR 1910.138, employers must select and require hand protection when hands are exposed to hazards that include thermal burns and harmful temperature extremes, and must base the selection on an evaluation of the hand protection’s performance characteristics relative to the task, the conditions present, the duration of use, and the hazards identified. In practice, that means matching a glove’s stated rating to what comes out of your sterilizer, and checking the manufacturer’s own statement of what the glove is rated for. See the guide to autoclave operating safety and steam burns for the hazards at unloading.
Safe use and handling basics
- Use a glove rated for the temperature involved; the example above lists 450 °F (232 °C) as its limit.
- A gauntlet cuff covers more of the forearm; the example’s cuff is 11 in long.
- The example glove is not waterproof, so it is not a substitute for protection against wet or liquid hazards.
- Do not use the example glove to handle dry ice.
Cycle parameters for your own load belong in your autoclave’s instructions and your facility’s protocol; see how long unwrapped items should be sterilized and the overview of a tabletop autoclave for a research lab.
Common mistakes
- Assuming a heat-resistant glove can itself be autoclaved. The manufacturer listing describes heat protection and washing, not sterilization of the glove.
- Assuming any exam glove survives an autoclave. In the study, nearly a quarter of autoclaved latex exam gloves could not be donned.
- Generalizing the rodent-surgery study to other settings.
- Mixing up disposal and reuse questions: for what happens to used nitrile gloves, see the nitrile glove recycling program guide.
- Treating a non-waterproof glove as protection against wet loads.
FAQ
What is the difference between heat-resistant autoclave gloves and gloves you can autoclave?
Heat-resistant autoclave gloves protect the wearer’s hands when removing hot items from an autoclave or oven. Gloves you can autoclave are gloves that are themselves put through a steam cycle. The Bel-Art Clavies listing describes the first kind and does not state that the gloves are autoclaved.
Can you autoclave nitrile exam gloves?
One laboratory study autoclaved nitrile exam gloves at 270 °F for 4 minutes with a 20-minute dry time, and 98% met its performance criteria for rodent surgery. It did not evaluate in vivo outcomes and does not cover other uses.
Can you autoclave latex exam gloves?
In the same study only 76% of autoclaved latex exam gloves met the criteria and 23% could not be donned, against 0% of nitrile.
What temperature do heat-resistant autoclave gloves handle?
The example listing states protection from temperatures up to 450 °F (232 °C). Check the rating on the specific product you use.
What does OSHA require for hand protection?
29 CFR 1910.138 requires employers to select hand protection based on the task, conditions, duration of use and hazards identified, including thermal burns and temperature extremes.
Are nitrile exam gloves sterile?
Not by definition. FDA describes patient examination gloves as disposable devices worn to prevent contamination between patient and examiner, and they are offered sterile or non-sterile; check the package labeling for the sterility statement and the materials.
What is the difference between exam gloves and sterile surgical gloves?
Surgeon’s gloves (21 CFR 878.4460) are intended to protect a surgical wound and are expected to be sterile when sold to end users; patient examination gloves (21 CFR 880.6250) are intended to prevent contamination between patient and examiner. FDA names different ASTM specifications for each (D3577 for surgical, D3578 and D6319 for examination).
Can nitrile exam gloves replace sterile surgical gloves?
Only one laboratory rodent-surgery study supports autoclaved nitrile exam gloves as an alternative, and its authors did not evaluate in vivo outcomes. FDA treats the two glove types as separate devices, and its consumer guidance says never to reuse, wash or disinfect medical gloves.
Sources
- Thermo Fisher Scientific, Bel-Art Clavies heat-resistant biohazard autoclave gloves, product 11394298, product page, checked 2026-10-30.
- LeMoine DM, Bergdall VK, Freed C. Performance Analysis of Exam Gloves Used for Aseptic Rodent Surgery. J Am Assoc Lab Anim Sci 2015;54(3):311-316. PMC4460945, checked 2026-10-30.
- OSHA, 29 CFR 1910.138 Hand protection, Cornell LII text, checked 2026-10-30.
- FDA, Medical Gloves, https://www.fda.gov/medical-devices/personal-protective-equipment-infection-control/medical-gloves, checked 2026-10-31.
- FDA, Medical Glove Guidance Manual (January 2008), https://www.fda.gov/regulatory-information/search-fda-guidance-documents/medical-glove-guidance-manual, checked 2026-10-31.
- 21 CFR 880.6250 Non-powdered patient examination glove, https://www.law.cornell.edu/cfr/text/21/880.6250, checked 2026-10-31.
- 21 CFR 878.4460 Non-powdered surgeon’s glove, https://www.law.cornell.edu/cfr/text/21/878.4460, checked 2026-10-31.