Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
Short answer: most autoclave faults come from trapped air, a leaking door gasket, low steam supply, an overloaded chamber or wet steam. Work out which phase failed (heat-up, hold or drying), check the load and gasket first, then the steam supply and sensors, and treat any wet pack or positive indicator as a failed load.
An autoclave that won’t run right doesn’t just waste a cycle — it can mean a load goes back on the bench unsterile, a research protocol slips a day, or an EHS record shows a failed cycle that has to be explained. Most autoclave problems fall into a small set of recurring failure modes, and most of them are diagnosable from the symptoms alone, without opening the unit up. This guide walks through the common problems — failure to reach temperature or pressure, wet loads, door and seal issues, aborted cycles, error codes, and failed indicators — and how to fix or triage each one. For the underlying physical hazards of steam and pressure and how to avoid burns, see Autoclave Operating Safety: Avoiding Steam Burns and Pressure-Related Injuries; this guide assumes you already know how to operate the unit safely and focuses specifically on what to do when it isn’t behaving as expected.
Start With the Basics: What “Not Working” Usually Means
Before troubleshooting a specific symptom, it helps to know what a normal cycle looks like: the chamber pressurizes to roughly 15 psi (103 kPa) above atmospheric, holds at a minimum of 121°C (250°F) for the programmed sterilization time (commonly 15–30 minutes, longer for large or dense loads), then exhausts and, for liquid loads, cools before the door unlocks. Almost every troubleshooting scenario below is a deviation from one of those three phases: the unit can’t reach pressure/temperature, it reaches temperature but the load still isn’t sterile or dry, or the cycle stops before it should. Identifying which phase failed is the fastest way to narrow down the cause.
Keep a simple running log of failures (date, cycle type, load, symptom, error code if any). A single failed cycle is often operator error or a one-off; the same symptom recurring across different operators and loads points to a mechanical or maintenance issue that needs a service call.
Problem: Autoclave Won’t Reach Pressure or Temperature
This is the most common complaint and almost always traces back to air trapped in the chamber, since air doesn’t transmit heat the way saturated steam does — a chamber full of air and steam can show pressure on the gauge while the actual temperature inside a load is well below sterilizing temperature.
- Overloading or improper packing. A tightly packed or overloaded chamber blocks steam circulation and traps air pockets, especially in the center of the load. Load loosely, leave space between items, and don’t exceed the manufacturer’s rated capacity.
- Wrong cycle for the load. A gravity-displacement cycle relies on steam pushing air out the bottom of the chamber by gravity alone, which struggles with dense or porous loads (wrapped instrument packs, textiles). A pre-vacuum (dynamic-air-removal) cycle actively pulls a vacuum before injecting steam and clears trapped air far more reliably for those loads. If gravity cycles are consistently under-performing on packed goods, the fix is often to switch cycle type, not to service the unit.
- Door seal or gasket leak. A worn, cracked, or debris-fouled door gasket lets steam escape as fast as it’s generated, so the chamber never builds full pressure. Inspect the gasket for flat spots, cracks, or embedded debris; clean the gasket channel and door face with a damp cloth (no abrasives or solvents that degrade the rubber/silicone). A gasket that won’t hold a seal needs replacement — this is routine, inexpensive maintenance, not a sign of a failing unit.
- Steam supply or generator problem. On self-generating (electrically heated) units, scale buildup on the heating element from hard or untreated water reduces heating efficiency and can trip a low-water safety cutoff. On house-steam-fed units, low building steam pressure will produce the same symptom. Check the water reservoir/reservoir fill line and, if scale is visible, descale per the manufacturer’s schedule.
- Faulty pressure or temperature sensor. If the chamber is physically hot and pressurized but the controller reads low or errors out, the sensor or its wiring — not the physical process — may be the fault. This needs a qualified service technician; don’t attempt to bypass a safety interlock or sensor reading to force a cycle to run.
Pressure Reads Correctly but Temperature Is Low
When the gauge shows the target pressure but the temperature reading is low, Tuttnauer’s troubleshooting guidance identifies this as the signature of high non-condensable gas (NCG) content in the chamber or steam: air and other gases add to the pressure without carrying heat the way saturated steam does. The same guidance lists the other causes to rule out: low incoming steam supply pressure, a slow air leak at the door seal or chamber, a drifted temperature sensor, and a failed heating element or steam control valve.
- Check incoming steam pressure. On house-steam units, measure supply pressure at the autoclave inlet and compare it with the manufacturer’s specified minimum. Below that minimum the chamber cannot reach its set point, and this is easy to overlook because nothing is wrong with the autoclave itself.
- Look for air leaks. A slow leak at the gasket or a chamber fitting lets air in during heat-up and exhaust.
- Verify the sensor. Compare the controller reading against an independent reference probe; a drifted sensor can make a good cycle look short, or a short cycle look good.
- Escalate. A failed heating element or steam control valve, or a steam supply with high NCG, needs a qualified technician and a steam-quality test (see below).
Problem: Wet Loads or “Wet Packs” After the Cycle
The load reached temperature and the cycle completed normally, but items come out visibly wet, or wrapped packs show moisture soaked through the wrap. This is a drying-phase problem, not a sterilization failure, but a wet pack is still typically treated as non-sterile because moisture can wick contamination through wrapping material.
- Drying time too short. Wrapped and porous loads need an active drying phase (vacuum-assisted drying on pre-vacuum units, or an extended dry time on gravity units) after the sterilization hold. Increase dry time before assuming a mechanical fault, especially for dense or heavily wrapped loads.
- Overloading. The same packing problems that block steam penetration also block drying — steam condensate has nowhere to go in an overpacked chamber.
- Liquids packed too close to solid/wrapped goods. Condensate and residual moisture from liquid containers can drip onto adjacent wrapped items. Separate liquid loads from wrapped instrument or textile loads where possible, and always run liquids on a dedicated slow-exhaust liquid cycle.
- Steam quality (wet steam). Steam carrying excess entrained water (common with scaled or poorly maintained steam generators) deposits more condensate into the load than dry saturated steam does. Persistent wet-pack problems across many loads and operators, even with adequate dry time, point to a steam-quality or generator issue that needs a technician.
Wet-pack causes and a first-check order
Tuttnauer lists these causes of wet packs: an overloaded chamber, heavy items stacked on lighter ones, wet steam (dryness fraction below 0.97), inadequate vacuum pulses, insufficient drying time, cold instruments placed in a hot chamber, wrapping that steam cannot penetrate, and water hammer. It also cites AAMI ST79 for the rule that a wet pack is non-sterile and must be reprocessed.
- Check load weight and arrangement: heavy metal on the bottom shelf, lighter textile and plastics above, nothing packed tight.
- Confirm the chamber drain is clear.
- Pre-warm or lighten metal trays and dense instrument sets so they do not condense excess steam.
- Extend the dry time within the range validated for your unit and load (the CDC drying times in the table below are a reference range, not a replacement for your validated cycle).
- If wet packs continue from correctly built loads across several consecutive cycles, test steam quality.
Steam Quality: the Limits Behind Wet Packs and Failed Indicators
On house-steam and generator-fed units, poor steam quality produces wet packs, slow heat-up and failed air-removal tests that no change in loading will cure. Tuttnauer’s guidance gives three limits, measured at the supply and following the EN 285 test methodology: a dryness fraction of at least 0.97 (below 0.97 is a failure), non-condensable gas of no more than 3.5 percent by volume, and superheat below 25°C above the saturation temperature at the measured pressure. Feed-water quality for self-generating units is covered by AAMI ST108. These are manufacturer-published limits; confirm the figures required for your unit in its manual and your validation protocol. The data table below this guide lists them next to the other monitoring checks.
Problem: Door Won’t Open, Won’t Close, or Won’t Latch
- Won’t open after a cycle. Most autoclaves interlock the door so it physically cannot unlock until chamber pressure has returned to atmospheric (and, on liquid cycles, until the load has cooled below a safe threshold). This is a safety feature, not a fault — wait out the full vent/cool-down phase. If the door stays locked well past a completed, vented cycle, the pressure or door-interlock sensor may be malfunctioning and needs service; never force or defeat a door interlock.
- Won’t seal/latch closed. Usually a gasket problem (see above), a misaligned door, or debris on the door face or chamber rim. Clean both surfaces and re-check alignment before assuming a mechanical failure.
Problem: Cycle Aborts Partway Through or Throws an Error Code
Manufacturer error codes vary, but they generally cluster into the same handful of underlying faults: low water level, door not fully sealed/locked, chamber over- or under-pressure relative to the programmed setpoint, sensor fault, or a timeout waiting for the chamber to reach temperature. Always check the specific error code against your unit’s manual before troubleshooting blind — codes are the fastest path to the actual cause and save time versus working through every possibility above.
- Note the exact code and the point in the cycle it occurred (during heat-up, during the sterilization hold, during exhaust) — this is the single most useful piece of information for both self-diagnosis and for a service technician if you need to call one.
- A cycle that aborts on the same step every time (e.g., always failing to reach setpoint within the time limit) points to a reproducible mechanical cause — gasket, sensor, generator — rather than a one-off operator error.
- Never override a safety abort to force a cycle to complete. An abort exists because a condition needed for either safe operation or effective sterilization wasn’t met; overriding it can produce a load that looks sterilized but isn’t, or create a real injury risk.
Problem: Failed Biological or Chemical Indicator Results
Even when a cycle completes without an error, indicator results tell you whether it actually achieved sterilization — and a failure here is a process failure, not a cosmetic one.
- Chemical indicators (autoclave tape, integrators) didn’t change color/pattern correctly. Confirm the indicator was placed correctly (inside the load, not just taped to the outside) and rated for the cycle type and temperature used. A correctly placed indicator that still fails points to an actual exposure problem — time, temperature, or steam penetration — not a bad indicator.
- Biological indicator (spore test, e.g. Geobacillus stearothermophilus ampoules or strips) shows growth after incubation. Treat this as a real sterilization failure: quarantine anything processed since the last passing biological indicator, do not release those loads as sterile, and take the unit out of service until the cause is found and a repeat test passes. Common causes are the same mechanical issues above — incomplete air removal, overloading, a leaking gasket, or a miscalibrated sensor.
- Bowie-Dick test failure (pre-vacuum units only). A failed daily Bowie-Dick test specifically indicates the vacuum system isn’t removing air effectively from the chamber, which is the pre-vacuum cycle’s core function. This should be resolved, with a passing retest, before the unit is used for any load requiring full sterility assurance.
Problem: Unusual Noise, Odor, or Visible Leaks
- Hissing or whistling during the hold phase. A small, continuous hiss usually indicates a gasket or fitting leak; the chamber may still reach setpoint pressure but will take longer and use more water/energy than normal, and the leak will typically worsen over time.
- Burning or plastic odor. Stop the cycle and inspect the load for melted, non-autoclave-rated plastic or scorched material contacting a heating element or chamber wall. This is a load-compatibility issue, not a unit fault, but repeated occurrences can damage chamber surfaces or seals.
- Visible water leaking from the base or door. Usually a failed gasket, a cracked water reservoir, or a fitting/hose failure on self-generating units. Take the unit out of service and have it inspected rather than continuing to run cycles around a known leak.
Troubleshoot Yourself vs. Call for Service
A reasonable rule of thumb for lab staff without biomedical-equipment training:
- Safe to troubleshoot yourself: load packing/spacing, choosing the correct cycle type, dry-time adjustments, gasket cleaning and visual inspection, and confirming indicator placement.
- Call for qualified service: anything involving the pressure vessel, door interlock, heating element, steam generator, control electronics, or a repeated safety-interlock abort. Autoclaves are pressure vessels and, in many jurisdictions, subject to periodic third-party pressure-vessel inspection separate from routine EHS/biosafety oversight — don’t attempt electrical or pressure-system repairs in-house unless your facilities team is specifically qualified and authorized to do so.
Preventive Maintenance That Prevents Most of This List
Most of the failures above are far less common on units with a basic preventive-maintenance routine:
- Daily: visual gasket/door check, water reservoir level (self-generating units), and a Bowie-Dick test on pre-vacuum units before the first processing cycle of the day.
- Per load: correct packing density, correct cycle selection, correct indicator placement.
- Scheduled (per manufacturer, typically quarterly to annually): gasket replacement, descaling, sensor/gauge calibration verification, and a full service inspection — the same cadence and inspection points covered in-depth for chamber testing methods in a dedicated validation guide.
- Ongoing: a log of every cycle’s parameters and every indicator result, which is what actually turns a one-off failed cycle into an identifiable pattern instead of a mystery.
Bowie-Dick Test Failures in Detail
The Bowie-Dick test checks that a pre-vacuum (dynamic-air-removal) sterilizer removes air and that steam penetrates a test pack evenly. STERIS’s guide and the CDC describe it as a daily test run in an empty chamber before the first load; the CDC cites an exposure of 134°C for 3.5 minutes, and STERIS describes a programmed exposure of 3.5 to 4 minutes.
Avoid a false failure
- Run a warm-up cycle first. A skipped warm-up is the most common cause of a failed test on a healthy machine.
- Run the chamber empty, with the test pack placed over the drain, with the cart or rack present as in normal use.
Real causes of a failed test
If the test fails after a proper warm-up, the likely causes are a degraded vacuum pump, a gasket leak, a steam trap fault, non-condensable gas in the steam supply, or a vacuum sensor fault. STERIS notes that a unit failing the test can still run gravity or similar non-pre-vacuum cycles; it must not run pre-vacuum cycles until it passes. After installation, relocation or major repair, qualification requires three consecutive passing cycles; routine daily use needs one pass.
What to Do After a Positive Biological Indicator
Biological indicators use Geobacillus stearothermophilus spores, the organism the CDC specifies for steam sterilizers. The CDC’s sterilizing-practices guidance calls for monitoring at least weekly, daily for units used for several loads a day, and with every load containing an implant. Rapid-readout indicators are read at 1 hour for 132°C gravity cycles and at 3 hours for 121°C gravity or 132°C vacuum cycles. When a result is positive, Tuttnauer’s decision flow runs as follows:
- Quarantine every load processed since the last confirmed negative indicator.
- Check the unprocessed control indicator from the same lot. If the control is negative the lot is defective and the result is invalid.
- Review the cycle printout for parameter excursions.
- Confirm the indicator sat at the cold spot documented in the performance-qualification protocol.
- Repeat with new biological indicators and a process challenge device (PCD) at the cold spot.
- If the repeat is positive, take the autoclave out of service and arrange inspection.
The CDC adds a nuance for non-implant loads: where mechanical and chemical indicators show a correct cycle, a single positive spore test does not by itself call for recalling every load, but the unit should be retested immediately and taken out of service if the repeat is also positive. Implant loads are held to the stricter standard.
Chemical indicator control technique
To tell whether a chemical indicator or the cycle failed, process a control indicator in open air, outside the chamber, and compare it with the indicator from the load. Common non-cycle causes of a failed chemical indicator are an indicator type that does not match the cycle, and an expired indicator.
Reference Numbers: Cycle and Drying Times
The CDC’s 2008 Guideline for Disinfection and Sterilization (modified from AAMI) gives these minimum exposure and drying times for wrapped items. They are reference values for diagnosing a cycle that looks too short; use the times validated for your own unit and load.
| Sterilizer type | Item | Exposure at 121°C | Exposure at 132°C | Drying time |
|---|---|---|---|---|
| Gravity displacement | Wrapped instruments | 30 min | 15 min | 15–30 min |
| Gravity displacement | Textile packs | 30 min | 25 min | 15 min |
| Gravity displacement | Wrapped utensils | 30 min | 15 min | 15–30 min |
| Dynamic-air-removal (prevacuum) | Wrapped instruments | Not applicable | 4 min | 20–30 min |
| Dynamic-air-removal (prevacuum) | Textile packs | Not applicable | 4 min | 5–20 min |
| Dynamic-air-removal (prevacuum) | Wrapped utensils | Not applicable | 4 min | 20 min |
For laboratory microbiological waste the CDC cites 45 minutes at 121°C in a gravity sterilizer for a 10 lb load, longer than for clean instruments because trapped air slows steam penetration into bags and containers.
When to Take the Autoclave Out of Service, and What to Record
Tuttnauer lists these triggers for removing a unit from service: a confirmed positive biological indicator; a Bowie-Dick failure with no identified root cause; persistent wet packs from correctly configured loads over several consecutive cycles; any safety-system fault (pressure relief, door interlock, temperature cutoff, vacuum); a cycle parameter excursion outside the validated range; and a steam-quality failure (NCG above 3.5 percent, superheat above 25°C, or dryness below 0.97).
For each failure, the record described by Tuttnauer with reference to AAMI ST79 should hold: autoclave ID and sequential cycle number; date and time; failure type (result, alarm code or symptom); load traceability back to the last confirmed passing result; corrective action and who performed it; the verification result (retest or service confirmation); disposition of affected loads (reprocessed, quarantined or recalled); and operator name with supervisor notification.
Frequently Asked Questions
Why does my autoclave show it reached pressure but the load isn’t sterile?
Pressure and temperature aren’t the same thing when air is trapped in the chamber — a chamber can register pressure from a mix of air and steam while the actual temperature at the load is below the sterilizing threshold. This is why biological and chemical indicators exist: they confirm actual sterilizing conditions were achieved at the load, not just that the gauge hit a number.
Is it normal for a load to come out wet?
No — a properly run cycle with adequate dry time should produce a dry load. Occasional light condensation on non-porous items (glassware, metal) can be normal, but a soaked wrapped pack is not, and should be treated as compromised sterility, not just a cosmetic issue.
Why won’t the door open after the cycle finishes?
The door interlock is designed to stay locked until the chamber has vented back to atmospheric pressure (and cooled, for liquid loads) as a safety measure against a sudden steam or superheated-liquid release. If it stays locked well beyond that point, the pressure sensor or interlock mechanism likely needs inspection.
What should I do if a biological indicator comes back positive?
Treat it as a genuine sterilization failure: quarantine everything processed since the last passing biological indicator, take the unit out of service, investigate the likely mechanical cause, and don’t return the unit to use until a repeat biological indicator test passes.
How often should an autoclave be serviced?
Daily and per-load checks (gasket, water level, indicator placement, Bowie-Dick test on pre-vacuum units) plus a scheduled technician service and calibration check, typically quarterly to annually depending on usage and manufacturer recommendation, catches most of the failure modes in this guide before they cause a failed load.
Why is my autoclave not reaching temperature?
Check, in order: load density and cycle type, the door gasket, incoming steam supply pressure against the manufacturer’s minimum, and sensor accuracy against a reference probe. If pressure is right but temperature is low, suspect non-condensable gas in the steam.
Are wet packs sterile?
No. Tuttnauer, citing AAMI ST79, states that wet packs are non-sterile and must be reprocessed.
How often should biological indicators and Bowie-Dick tests be run?
Per the CDC and STERIS: a Bowie-Dick test daily before the first load on pre-vacuum units; biological indicators at least weekly, daily where several loads run each day, and with every implant load.
What do I do after a failed Bowie-Dick test?
Confirm a warm-up cycle was run and the pack was placed over the drain, repeat the test, and if it fails again do not use pre-vacuum cycles until the cause (vacuum pump, gasket, steam trap, NCG, vacuum sensor) is fixed and the unit passes.
How long should an autoclave run for waste?
The CDC cites 45 minutes at 121°C in a gravity sterilizer for a 10 lb load of microbiological waste; follow your validated waste cycle.
What is a good steam dryness fraction?
At least 0.97, per Tuttnauer.
Sources
- CDC, Steam Sterilization (Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008): https://www.cdc.gov/infection-control/hcp/disinfection-sterilization/steam-sterilization.html (2026-11-21)
- CDC, Steam Sterilization Cycle Times: https://www.cdc.gov/infection-control/hcp/disinfection-and-sterilization/steam-sterilization-cycle-times.html (2026-11-21)
- CDC, Sterilizing Practices: https://cdc.gov/infection-control/hcp/disinfection-sterilization/sterilizing-practices.html (2026-11-21)
- STERIS, Bowie-Dick tests and troubleshooting guide: https://www.steris.com/healthcare/knowledge-center/sterile-processing/bowie-dick-tests-and-troubleshooting-guide (2026-11-21)
- Tuttnauer, Autoclave troubleshooting: sterilization failures (manufacturer guidance; defer to your unit’s manual and AAMI ST79): https://tuttnauer.com/knowledge-center/sterile-processing/autoclave/autoclave-troubleshooting-sterilization-failures (2026-11-21)
Data: Autoclave monitoring checks and steam-quality limits used to diagnose sterilization failures
Compiled from CDC's 2008 Guideline for Disinfection and Sterilization (steam sterilization and sterilizing-practices pages), STERIS Bowie-Dick guidance and Tuttnauer troubleshooting guidance: how often each check is run, what it applies to and the limit or pass condition quoted by the sources.
Period covered: 2008/2026. Coverage: United States.
| Check or measure | Applies to | Minimum frequency or trigger | Limit or pass condition | Source |
|---|---|---|---|---|
| Bowie-Dick air-removal test | Pre-vacuum (dynamic-air-removal) sterilizers | Daily, before the first load, empty chamber | Passing result; programmed exposure 3.5 to 4 minutes; warm-up cycle first; CDC cites 134 C for 3.5 minutes | CDC; STERIS |
| Bowie-Dick qualification | Pre-vacuum sterilizers after installation, relocation or major repair | Once per event | Three consecutive passing cycles | STERIS |
| Biological indicator (Geobacillus stearothermophilus) | Steam sterilizers | At least weekly; daily where several loads per day; every implant load | No growth; positive result: repeat test, stop use if repeats stay positive | CDC; Tuttnauer |
| Rapid-readout biological indicator, 132 C gravity | Gravity sterilizers | Per BI schedule | Read at 1 hour | CDC |
| Rapid-readout biological indicator, 121 C gravity or 132 C vacuum | Gravity and vacuum sterilizers | Per BI schedule | Read at 3 hours | CDC |
| Helix test | Hollow instruments | Daily where hollow instruments are processed | Passing result | Tuttnauer |
| Leak-rate test | Pre-vacuum chamber | Weekly or after any chamber service | Passing result | Tuttnauer |
| Steam dryness fraction | Steam supply | When wet packs persist | At least 0.97 | Tuttnauer |
| Non-condensable gas in steam | Steam supply | When air removal or heating is poor | 3.5 percent by volume or less | Tuttnauer |
| Superheat | Steam supply | When Bowie-Dick or indicators fail | Below 25 C above saturation temperature | Tuttnauer |
Sources
- CDC Steam Sterilization (Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008) (accessed 2026-11-21)
- CDC Steam Sterilization Cycle Times (accessed 2026-11-21)
- CDC Sterilizing Practices (accessed 2026-11-21)
- STERIS Bowie-Dick tests and troubleshooting guide (accessed 2026-11-21)
- Tuttnauer autoclave troubleshooting: sterilization failures (accessed 2026-11-21)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.







