Written and maintained by CASRAI Editorial Board
Published
On this page
Biacore is a family of surface plasmon resonance (SPR) instruments made by Cytiva for label-free, real-time analysis of molecular interactions: a binding partner is attached to a gold-coated sensor chip, a sample flows over it, and the instrument records binding as it happens, from which affinity and on/off-rate (kinetic) values are derived. The model names researchers meet most often are the Biacore T200 and S200 (four flow cells per chip) and the newer Biacore 1 series (1K, 1K+ and 1S+, six flow cells per chip), with the higher-throughput Biacore 8 series alongside.
Definition and scope
In Cytiva’s own documents a Biacore system is described as a “system for real-time label-free analysis of molecular interactions” (Biacore S200 Operating Instructions). It is a bench-top analytical instrument, usually housed in a core facility or a protein-characterisation lab, and it is distinct from the assay it runs: the instrument supplies the optics, the microfluidics and the control and evaluation software, while the researcher designs the experiment (what is captured on the chip, what is injected, at which concentrations). For the assay logic and the maths of turning a sensorgram into a dissociation constant, see our guide to SPR binding kinetics and KD calculation; this page is about the instrument family itself.
“Biacore” is a brand, not a single model. The same word covers older single-channel systems, the four-flow-cell T200 and S200, the six-flow-cell 1 series and the 8 series. That is why searches for “Biacore” and “Biacore T200” overlap, and why the practical question for a researcher, a core-facility manager or a quality reviewer is usually “which generation, and what can it do that the previous one could not?” The sections below answer that from Cytiva’s own documentation.
How a Biacore system works
The detection principle
The Biacore S200 manual explains the physics. SPR occurs in thin conducting films at an interface between media of different refractive index. Under total internal reflection, light leaks an evanescent field across the interface; at a particular combination of angle of incidence and wavelength it excites plasmons in the gold film, which shows up as a drop in the intensity of the reflected light at the “SPR angle”. Because the evanescent field penetrates the solution next to the gold, that angle depends on the refractive index within roughly 300 nm of the chip surface. When analyte binds to a molecule fixed on the chip, solute concentration at the surface rises, the refractive index changes and the SPR angle shifts. The manual states that the measured response relates to the mass of analyte bound and is largely independent of the nature of the analyte, and that the same principle measures both the amount of molecule attached to the chip and the amount of analyte that later binds.
The same manual notes a practical consequence: the reduced reflected intensity is not conventional absorption by the sample, because the signal-generating light is totally internally reflected inside the optical unit and only the evanescent wave enters the sample. Measurements can therefore be made on turbid or even opaque solutions without interference from conventional light absorption or scattering.
Chip, fluidics and sample handling
In the S200 manual the sensor chip is a gold-coated glass slide mounted on a frame in a cassette. An integrated microfluidic cartridge (IFC) with micro-channels and membrane valves forms the flow cells when the chip is docked; on the S200 there are four, optimised for use as pairs (flow cells 1 and 2, flow cells 3 and 4). Two syringe pumps, one peristaltic pump, a degasser and an autosampler with an injection needle move buffer and sample. One syringe pump keeps a low buffer flow through the IFC; the other aspirates samples and reagents and delivers them to the chip surface.
The S200 specification table gives figures a lab can plan around: a typical limit of detection below 1 RU (approximately 1 pg/mm² for proteins on Sensor Chip CM5), sample volumes of 10 to 425 µL depending on application, 48 hours of unattended operation, and typical analysis times of 2 to 15 minutes per sample. Analysis temperature is 4°C to 45°C but not more than 20°C below ambient, and the manual flags that SPR is temperature sensitive: a run does not start if the surface temperature is unstable. These are S200 values; other models have their own specification tables.
What a run produces
The output is a sensorgram: response in resonance units (RU) against time, with a baseline, an association phase while analyte is injected and a dissociation phase in buffer. Cytiva’s 1 series data file describes two ways of collecting kinetic data. In multi-cycle kinetics each analyte concentration gets its own cycle. In single-cycle kinetics a series of concentrations (normally 3 to 9) is injected one after another in the same cycle, followed by one dissociation time; the data file states this gives slow interactions a better chance to reach saturation and shortens assay time compared with multi-cycle kinetics, with one worked example using a one-hour dissociation after five injections.
The model families
Biacore T200 and S200: four flow cells
The T200 and S200 are the reference point in Cytiva’s later comparisons: four flow cells per chip, used for example as three active flow cells plus one reference for screening, or as two active plus two reference when each active surface gets its own reference. A user searching for “Biacore T200” is typically looking at a core-facility instrument of this generation, and the S200 operating instructions are the most detailed public technical description of how the hardware is built.
Biacore 1 series: 1K, 1K+ and 1S+
Cytiva’s data file (CY29857, October 2022) describes the 1 series as a “one needle platform” with a six-flow-cell, one-channel flow system and a single software platform, in three configurations: Biacore 1K, 1K+ and 1S+. The 1K can be upgraded to the 1K+. The systems use the Series S sensor chip, and the file says methods transfer to other 1 series systems and to the higher-throughput Biacore 8 series. All three are described as built for analysis in a GxP-regulated environment.
The flow cells can be addressed singly (1 to 6) and in pairs (1/2, 3/4, 5/6) on every 1 series system. The 1K+ and 1S+ can in addition address them in series, as quadruplets (1234, 3456) or all six in sequence (123456). That serial addressing is what drives the screening saving described below.
The data file separates the 1S+ on sensitivity and speed: sensorgrams can be separated in the milli-resonance-unit range, data are collected at 40 Hz, off-rates up to 6 s⁻¹ can be determined, and, across the 1 series, very slow off-rates down to 10⁻⁶ s⁻¹ and fragment-scale affinities (KD in the millimolar range) are claimed. Concentration analysis is supported on all three through the Biacore Insight software extension “Concentration and Potency”, with detection reported down to 1 pM; the limit of detection in the file’s example is the average plus three standard deviations of ten buffer replicates. These are vendor-reported performance claims from Cytiva’s data file, not independent benchmarks.
Biacore 8 series
Cytiva’s 1 series data file refers to the Biacore 8 series as the “higher throughput systems” and states that application methods transfer between the two series.
The data table below summarises how the flow-cell counts and addressing differ across generations, using only figures stated in Cytiva’s data file.
Worked comparison: what six flow cells change
The clearest information in the 1 series data file is its arithmetic for the same experiment on four versus six flow cells. Each line below is stated in the file; the percentages follow from the counts.
- Yes/no binding, antibody screening, kinetic screening and epitope binning. With one reference flow cell and serial analysis, the T200/S200 gives three active flow cells and one reference; the 1K+ and 1S+ give five active and one reference. Five ligands instead of three is the +60% in the file (5 ÷ 3 ≈ 1.67).
- Worked example: a 15-antibody yes/no screen. Captured on Sensor Chip Protein A, the file reports three cycles of five antibodies each on the 1K+ or 1S+, against five cycles of three antibodies on the T200. One analyte (antigen) injection per cycle means 3 injections instead of 5, which is the roughly 40% lower analyte consumption quoted (3 ÷ 5 = 0.6).
- Kinetic and affinity determination. With one reference flow cell per flow-cell pair, six flow cells give three active plus three reference, against two plus two on four flow cells: a 50% rise in analytical capacity per chip and, per the file, a one-third reduction in sensor chip cost.
- Concentration analysis. No reference flow cell is needed, so all six flow cells can be active, against four on the T200/S200: six components can be measured in one injection of a complex sample.
- 1K versus the older Biacore X100. Restricted to pairwise addressing, the six-flow-cell 1K has three flow-cell pairs per chip against one on the two-flow-cell X100, which the file states triples the number of interaction analyses per chip and cuts sensor chip cost by two thirds.
The practical reading is that the 1 series changes how much each chip and each milligram of analyte buys, not the underlying SPR principle. The savings only apply where the experiment can use the extra flow cells, which is why Cytiva ties each figure to a named application. The file also describes a “Dual” injection command in Biacore Insight Control Software that injects two solutions in sequence with no wash between them, suggested for epitope binning when the first antibody binds the antigen weakly.
Where it matters for labs, quality units and facility decisions
Regulated use: the GxP package
For laboratories working under GxP, the Biacore T200 GxP Handbook (document 28976881) is the clearest public description of what the instrument vendor provides. The GxP Package adds support for GxP work to the Control and Evaluation Software and offers qualification services in connection with installation and service maintenance, so the lab receives documentation for equipment qualification. It provides support for compliance with FDA’s 21 CFR Part 11 (Electronic Records; Electronic Signatures) with respect to electronic records, and states that electronic signatures are not supported in the current version of the product it describes. A quality reviewer should therefore treat “GxP-capable” as a statement about specific software and services in a specific version, to be checked against the version installed.
The same handbook describes “publication”: methods and wizard templates are locked for GxP use so that settings cannot be changed when the published procedure runs, while sample information can still be entered at run time. Changes made to items in an evaluation are logged in an audit trail; altering a published procedure requires publishing a new version, which inactivates the previous one, and the version history is kept. These are the controls that map onto the data-integrity expectations described in our 21 CFR Part 11 entry, our guide to the ALCOA principles and the broader explainer on what data integrity means.
For the 1 series, Cytiva’s data file states only that all three systems are built for analysis in a GxP-regulated environment; it does not give Part 11 detail, so a lab validating a 1 series instrument should obtain the current software documentation from Cytiva rather than assume the T200 handbook’s wording carries over.
Choosing between generations
Cytiva’s data file frames the choice around sample capacity, sensitivity and throughput: the 1K for six-flow-cell capacity, the 1K+ for in-series addressing, the 1S+ for the highest sensitivity and the fastest off-rates. A facility that already runs a T200 or S200 will find the 1 series documented as method-compatible within its own series and with the 8 series; whether a given T200 method transfers unchanged is not stated in the data file and is worth testing before relying on it.
Common misunderstandings
- “Biacore” is not one instrument. The name spans several generations with different flow-cell counts, addressing options and software.
- SPR does not need a label. The response comes from mass-related refractive-index change at the surface, so neither partner has to be fluorescently or radioactively tagged; one partner must still be immobilised or captured on the chip.
- The signal is not absorbance. The S200 manual states the dip in reflected light is not absorption by the sample, which is why turbid samples can be measured.
- More flow cells do not change the chemistry. The 40% and 50% figures above are design-dependent savings, and apply when reference and active surfaces are arranged as the data file describes.
- Vendor sensitivity figures are conditions-specific. The 1 pM and 10⁻⁶ s⁻¹ values come from named example interactions and a defined limit-of-detection method, so they describe what was achieved in those examples.
Sources
- Cytiva, Biacore 1 series data file CY29857 (October 2022), read 2026-11-16.
- Cytiva, Biacore S200 Operating Instructions 29143106 AD, read 2026-11-16.
- Cytiva, Biacore T200 GxP Handbook 28976881 AG (copy hosted by China Medical University), read 2026-11-16.
- 21 CFR Part 11, Electronic Records; Electronic Signatures, section 11.1 (Legal Information Institute, Cornell Law School), read 2026-11-16.
Data: Biacore model generations compared: flow cells, addressing and experiment layouts
Compares the Biacore X100, T200/S200, 1K, and 1K+/1S+ on flow cells per sensor chip, flow-cell addressing modes and the active/reference layouts Cytiva lists for screening and kinetics. Compiled from Cytiva's Biacore 1 series data file (Tables 1 and 2 and its text) and the Biacore S200 operating instructions.
Period covered: 2022/2026.
| System | Flow cells per chip (count) | Addressing modes | Screening layout (active + reference) | Kinetics/affinity layout (active + reference) |
|---|---|---|---|---|
| Biacore X100 | 2 | Pairwise only | 1 active + 1 reference | 1 active + 1 reference |
| Biacore T200 / S200 | 4 | Singly and in pairs (flow cells optimised as pairs 1+2 and 3+4 in the S200 manual) | 3 active + 1 reference | 2 active + 2 reference |
| Biacore 1K | 6 | Singly and in pairs (1/2, 3/4, 5/6) | Not stated in source | 3 flow-cell pairs per chip (1 active + 1 reference each) |
| Biacore 1K+ / 1S+ | 6 | Singly, in pairs, in quadruplets (1234, 3456) or all six in sequence | 5 active + 1 reference | 3 active + 3 reference |
Sources
- Cytiva Biacore 1 series data file CY29857 (accessed 2026-11-16)
- Cytiva Biacore S200 Operating Instructions 29143106 AD (accessed 2026-11-16)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.