Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
Candy cane stirrups are leg supports that mount to the side rail of a surgical or operating table. Two manufacturer listings, checked on 2026-10-31, describe the same product class with very different numbers, so rated capacity, height range and mounting socket have to be read per model rather than assumed. This page compares those listings side by side and explains the regulatory frame that applies to the table they attach to.
This page describes products and general use; it is not medical advice.
Who uses them: the SchureMed spec sheet describes its stirrups for gynecologic, urologic or laparoscopic procedures, so the people who specify and stock them are operating-room and procedure-room staff and the buyers who supply those rooms.
What are candy cane stirrups?
Candy cane stirrups are adjustable surgical-table stirrups: a support that holds a patient’s leg and attaches to the table through a socket. The SchureMed listing describes them as helping to position patients for a wide range of surgical procedures, with adjustable heights and angles. Each stirrup is a separate piece, and the USA Medical & Surgical listing sells one stirrup per package.
They are an accessory to a table, not a table. For the table itself, see the CASRAI guide on the exam table, which covers table types, FDA class and accessible heights.
How are they built?
The two listings describe construction in different terms, and the points below are the ones the makers state:
- Material: the USA Medical & Surgical listing states 304 stainless steel construction.
- Clearance: that listing describes a large 8-inch arc for additional clearance; the SchureMed sheet states an 8-inch (20 cm) leg clearance.
- Height adjustment: the SchureMed sheet says the height is locked by twisting the ergonomic handle; the USA Medical & Surgical listing describes an integral clamp that adjusts over its height range.
- Strap: the USA Medical & Surgical listing includes a 1-inch cotton strap to make 9-inch loops.
- Mounting: both attach to a table-side socket that is sold separately (see the table below).
How do two listings compare?
The table shows only values stated on each manufacturer’s own page. Values were checked on 2026-10-31.
| Specification | SchureMed Candy Cane Stirrups 800-0012 (Soma Tech) | Candy Cane Stirrup (USA Medical & Surgical listing) |
|---|---|---|
| Weight capacity | 350 pounds | 60 lbs. |
| Height range | 28 in to 44 in (71 cm to 112 cm), vertical adjustment | 23 in to 41 in, integral clamp adjustment |
| Leg clearance | 8 in (20 cm) | 8 in arc |
| Table connection | Table side rail, via Schure Socket XL (sold separately) | Standard Clark socket (sold separately) |
| Stated material | Not stated on the spec sheet | 304 stainless steel |
| Supplied per package | Not stated on the spec sheet | 1 stirrup, with 1-inch cotton strap |
Sources: Soma Tech spec sheet 800-0012 (revised 4/21/2020) and the USA Medical & Surgical product page, both linked under Sources.
Which specifications matter when you read a spec sheet?
The comparison above shows why three fields decide whether a model fits a given room:
- Rated capacity. The two listings state 350 lb and 60 lb. They are different products with different ratings, and neither number can be treated as a norm for the class.
- Height range. The ranges are 28 to 44 in and 23 to 41 in. Compare the range against the table and the procedures the room is used for, using the maker’s figures.
- Socket. One listing names the Schure Socket XL and the other a standard Clark socket, and both state the socket is sold separately. Neither listing says it works with the other’s socket, so confirm the socket on the table before relying on a stirrup model.
For material, padding, cleaning and reprocessing, use the manufacturer’s instructions for use for the specific model.
How are stirrups regulated?
21 CFR 884.4900 defines an obstetric table as “a device with adjustable sections designed to support a patient in the various positions required during obstetric and gynecologic procedures”, and says this generic type “may include the following accessories: patient equipment, support attachments, and cabinets for warming instruments and disposing of wastes.” It classifies the type as Class II (special controls), exempt from premarket notification subject to the limitations in 21 CFR 884.9. The openFDA classification database lists product code HDD for “Table, Obstetrical, Ac-Powered (And Accessories)” (checked 2026-10-31).
That regulation does not name stirrups, and neither manufacturer listing states a product code or class for its stirrup. Do not assume a specific stirrup model carries the table’s classification; ask the manufacturer. Related room-stocking context is in the CASRAI guides on the urology clinic supply checklist and on obstetrics and gynecology.
What are the safe handling basics?
Makers state a rated capacity, a height-locking mechanism (the SchureMed sheet describes a twist handle) and a named socket for each model. Follow the manufacturer’s instructions for use and the facility’s protocols for everything else.
What are common mistakes when specifying them?
- Treating one listing’s capacity as representative: the two listings differ by almost six times (350 lb versus 60 lb).
- Ordering a stirrup without the socket: both listings state the socket is sold separately.
- Assuming sockets are interchangeable: the listings name different sockets and do not state cross-compatibility.
- Assuming the regulation for obstetric tables settles a stirrup’s class: it names “support attachments” as accessories but does not name stirrups.
FAQ
Do all candy cane stirrups have the same weight capacity?
No. The two listings checked state 350 lb (SchureMed 800-0012) and 60 lb (USA Medical & Surgical), so read the rating for each model.
Is the mounting socket included?
Both listings state the socket is sold separately: the Schure Socket XL for the SchureMed model and a standard Clark socket for the USA Medical & Surgical model.
Are candy cane stirrups regulated like an obstetric table?
21 CFR 884.4900 covers obstetric tables and lists support attachments among their accessories, but it does not name stirrups. Ask the manufacturer which classification applies to a specific model.
Sources
All sources checked on 2026-10-31.
- Soma Tech International, SchureMed Candy Cane Stirrups 800-0012 spec sheet, revised 4/21/2020: https://www.somatechnology.com/pdfFiles/Candy-Cane-Stirrups-SchureMed-800-0012.pdf and product page https://www.somatechnology.com/surgical-table-stirrups/schuremed-candy-cane-stirrups
- USA Medical & Surgical, Surgery Table Candy Cane Stirrup: https://www.usamedicalsurgical.com/surgery-table-candy-cane-stirrup/
- 21 CFR 884.4900, Obstetric table and accessories (Cornell LII): https://www.law.cornell.edu/cfr/text/21/884.4900
- openFDA device classification, product code HDD: https://api.fda.gov/device/classification.json?search=product_code:HDD