Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General product information for lab, clinic and supply staff. It is not medical advice and does not describe how to perform any procedure.
Short answer
A contrast injector, also called a power injector, is a motor-driven device that pushes contrast material from a syringe into a patient’s circulation at a set flow rate, volume and pressure limit during imaging. In FDA’s vocabulary the CT-type device is filed under product code IZQ, “Injector, Contrast Medium, Automatic”, and the cardiac-catheterization version is regulated as an angiographic injector and syringe. The injector sets its own limits; the catheter or IV set attached to it carries separate labelled limits.
Definition and scope
“Contrast injector”, “power injector”, “CT contrast injector” and “contrast media injector” are different names for one device family. The FDA sources below describe it from two angles.
- Angiographic injector and syringe. 21 CFR 870.1650 defines it as a device that consists of a syringe and a high-pressure injector, used to inject contrast material into the heart, great vessels and coronary arteries so they can be studied by x-ray photography. It is Class II (special controls).
- CT injector. The FDA/NLM AccessGUDID database lists CT injector systems under product code IZQ. Its device-category definitions describe AC-powered electromechanical injectors, with syringes capable of delivering media with the pressure, flow range and volume required for CT studies. One category is stationary (fixed to a room rack) and one is mobile (on a wheeled pedestal), and the system may synchronise media delivery with the x-ray generator during scanning.
The injector is not the scanner. For the scanner itself, see what a CT scanner is; the injector is an add-on system that works alongside it. For how a device is defined and classified more broadly, see what a medical device is.
How a contrast injector is used in practice
At the level of general product information, an injector system has a syringe that holds the contrast material, a drive that moves the syringe plunger, and a connection to the patient’s vascular access. The settings an operator works with are the flow rate, the volume and a pressure limit. The CT category definition above names exactly those three quantities (pressure, flow range and volume) as what the system must be capable of delivering.
Because the syringe and tubing sit between the injector and the patient, a department stocks the disposable interface that matches its injector model. The radiology and imaging center supply checklist makes the point that the injector interface has its own tubing and connector specification set by the injector manufacturer, so the connector set should match the specific injector model rather than assuming generic IV tubing is compatible. Setup and operating steps belong in the manufacturer’s instructions for use and the facility’s own protocols, not here.
Types: how the same family appears in FDA records
| Where you meet the term | What the record says | Source |
|---|---|---|
| Angiographic (cath lab) injector | “Angiographic injector and syringe”: a syringe plus a high-pressure injector for contrast into the heart, great vessels and coronary arteries; Class II (special controls) | 21 CFR 870.1650 |
| CT injector, stationary | AC-powered electromechanical injector system with syringes for the pressure, flow range and volume required for CT studies; fixed to a room rack | AccessGUDID, product code IZQ |
| CT injector, mobile | Same capability description; on a wheeled pedestal | AccessGUDID, product code IZQ |
The practical use of this table is search vocabulary. Searching device databases for “contrast injector” alone can miss records; the product name “Injector, Contrast Medium, Automatic” (IZQ) and the regulation title “angiographic injector and syringe” are the terms FDA’s own records use.
Where it matters for labs, clinics and supply decisions
A checklist: who owns which limit
The injector and everything between it and the patient each have their own labelled limits. Checking only one of them leaves a gap.
- Injector limits (flow range, volume, pressure). Owned by the injector manufacturer and stated in its instructions for use. The FDA category definition says only that the system must be capable of the pressure, flow range and volume required for CT studies; it gives no numbers.
- Catheter limits. Owned by the catheter manufacturer and stated in that catheter’s own instructions for use. “Power-injectable” is a catheter label, not an injector specification. The double lumen PICC guide shows the range: one manufacturer’s instructions describe power-injection use with stated limits, while another’s say not to use that catheter for high-pressure injection.
- Connector and tubing sets. Matched to the injector model, as above.
Two catheters that look alike can therefore differ in whether they may be connected to a power injector at all. Read each catheter’s own instructions for use; see also what a PICC line is for the device background.
Servicing
Maintenance intervals and performance checks are set by each manufacturer. The FDA records above state no maintenance intervals.
Common misunderstandings
- “21 CFR 870.1650 is the CT injector regulation.” It is the angiographic injector and syringe regulation. CT injectors appear in AccessGUDID under product code IZQ; the AccessGUDID record names no separate CFR section for them.
- “A power-injectable catheter makes the injector settings irrelevant.” The catheter label and the injector settings are separate limits; both apply.
- “Any IV tubing fits any injector.” The injector manufacturer specifies the connector and tubing interface.
- “All injector records carry the same labelling.” They do not. The AccessGUDID record for a CT injector unit states that its labelling does not contain MRI safety information; that is a statement about that one record, not about injectors in general.
FAQ
What is a contrast injector?
A motor-driven syringe system that delivers contrast material into a patient’s circulation at a controlled flow rate, volume and pressure limit during imaging. FDA files the CT type under product code IZQ and regulates the angiographic type under 21 CFR 870.1650.
Is a power injector the same as a CT contrast injector?
“Power injector” is the general name for the device family; “CT contrast injector” names the version used for CT studies. The angiographic version is regulated separately as an angiographic injector and syringe.
Does the injector or the catheter set the pressure limit?
Both have limits. The injector manufacturer states the injector’s capabilities; the catheter’s instructions for use state what that catheter is labelled to withstand. Check each document.
What FDA class is an angiographic injector?
Class II (special controls) under 21 CFR 870.1650, checked 2026-11-01. A non-patient-contacting balloon inflation syringe variant is exempt from premarket notification, subject to the limitations in the regulation.
Sources
- U.S. FDA, 21 CFR 870.1650 Angiographic injector and syringe, eCFR (ecfr.gov), checked 2026-11-01.
- FDA / National Library of Medicine, AccessGUDID record for a CT injector unit, product code IZQ “Injector, Contrast Medium, Automatic” (accessgudid.nlm.nih.gov), checked 2026-11-01.
- primePICC and Argon First PICC instructions for use, as summarised in CASRAI’s double lumen PICC guide, checked 2026-11-01.