Skip to main content
v2026.11,986 entries · CC-BY 4.0

Cryo-EM Workflow: From Purified Sample to Deposited Map

The single particle cryo-EM workflow stage by stage, with the check at each gate and a benchmark of ice thickness, collection hours and resolution.

Written and maintained by CASRAI Editorial Board

Published

On this page

A cryo-EM workflow for single particle analysis runs from a purified, stable sample through vitrification on a grid, screening, data collection on a cryo-electron microscope, image processing to a 3D density map, and finally deposition of the map in a public archive such as EMDB. The stages are not strictly one-way: the quality of the sample and the ice decides how good the map can be, so a poor screening result sends the project back to purification or vitrification rather than forward to collection.

Definition and scope

This page describes the single particle analysis (SPA) route through cryo-electron microscopy, in which many copies of a purified macromolecule are frozen in a thin layer of vitreous ice, imaged, and computationally averaged into a three-dimensional reconstruction. In the review by Vilas, Carazo and Sorzano (Chemical Reviews, 2022), SPA rests on an assumption that the particles are identical copies in the same conformational state, seen from different directions. That assumption shapes the whole workflow: the bench stages try to deliver a homogeneous, intact sample, and the computational stages sort particles by view and, where needed, by conformation.

The workflow is described here at the level of what each stage is for, what it produces and what is checked before the next stage. It is not a protocol: the settings reported in published benchmarks depend on the specimen, grid type, microscope and software, and are shown below as one documented example rather than as recommended values. For the instrument itself see cryo-EM microscope; for the supports the sample is frozen on see electron microscopy grids.

The cryo-EM workflow stage by stage

Thermo Fisher Scientific’s single particle analysis workflow page and the Chemical Reviews article together give a consistent sequence. The table below pairs each stage with what it produces and the check that decides whether the project moves on. The check column is where the workflow loops back.

Stage What happens Output Check before the next stage
1. Expression and purification The target is expressed and purified, using affinity purification and, where a tag is present, tag removal by protease Purified protein or complex Quality assessment of the sample
2. Sample quality assessment Methods listed by Thermo Fisher include differential scanning fluorimetry (DSF), SEC-MALS, mass photometry, negative stain and native mass spectrometry Evidence that the sample is well defined and stable Is the sample homogeneous and stable enough to vitrify?
3. Vitrification Sample on a grid is cooled rapidly so that water forms vitreous rather than crystalline ice, often with a semi-automated plunge freezer Vitrified grids Grid screening
4. Grid screening Grids are inspected for particle distribution, concentration, stability and ice thickness A decision on which grids go to collection If distribution or ice is poor, return to vitrification or the sample
5. Data collection Movies are recorded on the microscope Raw movies Micrograph quality, for example ice and defocus
6. Image processing Movie alignment, CTF estimation, particle picking, 2D classification, initial volume, refinement and 3D classification 3D density map Map quality and resolution
7. Validation and deposition The map is assessed and deposited through the wwPDB deposition portal Public EMDB entry; raw data may go to EMPIAR End of workflow

Sample preparation: expression, purification and quality assessment

Thermo Fisher’s workflow page states that well-defined and stable samples are critical for reliable vitrification and structural reconstruction. Its sample stage covers expression, purification by affinity methods with protease removal of tags, and a set of quality assessment techniques: differential scanning fluorimetry, SEC-MALS, mass photometry, negative stain electron microscopy and native mass spectrometry. The practical consequence is that cryo-EM time is only spent on samples that have already shown they are homogeneous.

The same page describes two approaches for difficult samples. GraFix and AgarFix cross-linking are named for fragile assemblies, and scaffolding is described for small proteins, typically below about 50 kDa, that are hard to see against the ice background. It also names apoferritin as a standard specimen used for optimization and quality control of the microscope and workflow.

Vitrification and grid screening

Vitrification means rapid cooling of the sample into vitreous ice. Thermo Fisher describes semi-automated plunge freezing for this step. Screening then asks four questions of each grid: how particles are distributed, whether the concentration is suitable, whether the sample is stable on the grid, and how thick the ice is. Screening is a gate, not a formality. A project that reaches the long collection session with the wrong ice or a poorly distributed sample has spent its most expensive resource on data that cannot be rescued computationally.

Ice thickness is the clearest place to see why. In the benchmarking study by Kim and colleagues (Frontiers in Molecular Biosciences, 2018), all six of the sub-3 Å maps came from datasets with ice reported at 10 to 20 nm, while the dataset with ice of 100 to 250 nm reached 3.0 Å with all its micrographs and 4.6 Å with its first 700. The authors state that ice thickness is a primary driver of data quality. The data table at the end of this page reproduces the nine reconstructions they report.

Data collection

Collection turns the screened grid into movies. In the Kim 2018 benchmark the specimen was rabbit muscle aldolase at 1.5 mg/ml, a homotetrameric enzyme of about 150 kDa, on gold R1.2/1.3 300-mesh grids. Data were recorded on a Titan Krios at 300 keV with a Gatan K2 Summit detector, a pixel size of 0.855 Å at the detector, a nominal defocus of -1.0 to -2.0 μm and a total dose of 60 to 70 electrons per Ų. These are the reported conditions of one benchmark with an aldolase test specimen, not settings to copy for another sample.

The same paper shows what collection costs in time. The three full datasets contained 699, 1,635 and 1,614 micrographs, collected over 18, 52 and 40 hours and reconstructed at 2.5, 3.0 and 2.4 Å. Collection hours and micrograph counts grow together, but resolution does not follow them: the largest and longest dataset was the one with thick ice and had the lowest resolution of the three. More data cannot replace better ice.

Image processing: from movies to a map

Vilas and colleagues describe the computational pipeline as a series of steps, each with a named purpose:

  • Movie alignment and gain correction compensates for motion during exposure and for detector response.
  • CTF estimation fits the contrast transfer function to the power spectrum of each micrograph; CTFFIND4 and gCTF are the programs named in the review.
  • Particle picking locates the particles. The review contrasts analytical approaches with deep-learning pickers such as Topaz and crYOLO.
  • 2D classification groups particles by view and removes picks that are not usable particles.
  • Initial volume, refinement and 3D classification build and improve the 3D map and separate conformational states.
  • Map quality assessment and sharpening evaluate and improve the final density.

The review notes that the signal-to-noise ratio of individual micrographs is low, roughly 0.1 to 0.01, which is why averaging many particles is necessary at all. In the Kim 2018 benchmark the pipeline used MotionCor2 for motion correction, CTFFind4 for CTF estimation, and cryoSPARC and RELION 2.0 for classification and refinement, with resolution reported using the 0.143 FSC criterion. For a worked look at one of those programs see the CryoSPARC guide.

Validation and deposition

The end of the workflow is a map that others can inspect. The Electron Microscopy Data Bank (EMDB) describes itself as a public repository for cryogenic-sample electron microscopy volumes, with deposition through the wwPDB deposition portal; its home page showed 64,595 entries on 2026-11-18. Kim and colleagues deposited their nine maps in EMDB (EMD-7528 to EMD-7617) and the raw datasets in EMPIAR as EMPIAR-10182, -10183 and -10184, which lets anyone re-run the processing. The raw-data side is covered in EMPIAR: the Electron Microscopy Public Image Archive.

Where the workflow loops back

Because the stages depend on each other, the practical question at every gate is whether to go on or return. The sources support three loops. Sample quality assessment can send the project back to purification or to stabilization with cross-linking or scaffolding. Grid screening can send it back to vitrification when the distribution, concentration or ice is unsuitable. Map quality can send it back to classification, picking or, if the particles themselves are heterogeneous, to the sample. The benchmark above shows why the earlier loops are cheaper: the cost of a poor grid shows up as extra collection hours or a lower-resolution map, not as an error message.

Common misunderstandings

Cryo-EM is only about the microscope

The instrument is one stage of seven. Thermo Fisher’s workflow page places expression, purification and quality assessment first and treats them as critical, and the benchmark data show that ice thickness has an effect on resolution that a longer collection did not offset.

Software can fix any dataset

Processing programs classify and average particles, but they work on what was recorded. SPA assumes identical copies in the same conformational state seen from different directions; heterogeneous or damaged samples undermine that premise before any program runs.

More micrographs always give a better map

In Kim 2018 the 1,635-micrograph, 52-hour dataset reached 3.0 Å, while the 699-micrograph, 18-hour dataset reached 2.5 Å. Ice thickness differed between them (100 to 250 nm against 10 to 20 nm).

The workflow ends at the map

Deposition is part of the workflow. EMDB holds the maps and EMPIAR the raw images, and deposited data are what make a reported structure checkable by other groups.

Cryo-EM sits next to other electron microscopy methods that share its sample-preparation vocabulary. Room-temperature preparation for TEM is covered in transmission electron microscopy sample preparation, and the combination of fluorescence and electron imaging is covered in correlative light and electron microscopy workflows. Cryo-SEM is a separate technique and is not the single particle workflow described here.

Sources

Data: Aldolase cryo-EM benchmark: ice thickness, collection time and resolution (Kim et al. 2018)

Nine reconstructions from three aldolase datasets reported in the Frontiers in Molecular Biosciences 2018 benchmarking paper, compiled from its Tables 1 and 2 and results text: micrographs used, data collection hours, reported ice thickness, global resolution and EMDB accession.

Period covered: 2017/2018.

Aldolase cryo-EM benchmark: ice thickness, collection time and resolution (Kim et al. 2018)
ReconstructionMicrographs usedCollection time (hours)Ice thickness (nm)Global resolution (angstrom)EMDB accession
Dataset 1, all micrographs6991810-202.5EMD-7616
Dataset 1, first 500 micrographs50013.510-202.8EMD-7614
Dataset 1, ice under 25 nm only5351810-202.5EMD-7617
Dataset 2, all micrographs163552100-2503EMD-7551
Dataset 2, first 700 micrographs70024100-2504.6EMD-7615
Dataset 2, ice under 25 nm only6352100-2503.5EMD-7562
Dataset 3, all micrographs16144010-202.4EMD-7541
Dataset 3, first 382 micrographs3821010-202.8EMD-7528
Dataset 3, ice under 25 nm only11084010-202.4EMD-7550

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Cryo-EM Workflow: From Purified Sample to Deposited Map

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →