Ct Values in qPCR: What the Cycle Threshold Is and What Moves It

What a Ct (Cq) value in qPCR is, why a higher Ct means less template, and the threshold, baseline and reporting factors that change it.

Written and maintained by CASRAI Editorial Board

Last updated

Short answer

A Ct (cycle threshold) value is the cycle number at which a reaction’s fluorescence curve crosses a threshold line. It is a relative measure of how much target was in the reaction: less starting template means a higher Ct. Cq is the current name for the same quantity, and a Ct is only comparable when threshold, baseline and chemistry are held consistent.

Definition and scope: Ct, Cq and cycle threshold

Thermo Fisher’s real-time PCR learning centre defines Ct as the intersection between an amplification curve and a threshold line, and describes it as a relative measure of the concentration of target in the PCR reaction. The MIQE 2.0 guidelines (Clinical Chemistry 2025;71(6):634) use Cq, the quantification cycle, “formerly known as Ct or Cp”. So “ct values in qpcr”, “cq values qpcr” and “cycle threshold qpcr” all point at one number; Cq is the term the guideline asks authors to report.

This page covers what the number means and what moves it. For converting Ct values into fold changes, see the delta-delta Ct data analysis guide; for quantifying against standards, see building and validating a qPCR standard curve.

How a Ct is read in practice

Each cycle the instrument records fluorescence. The baseline is subtracted to give a baseline-corrected signal (Thermo Fisher’s delta Rn is Rn minus baseline), and a threshold line is drawn across the amplification plots. Thermo Fisher states that the threshold must be set in the linear phase of the amplification plot. The cycle where each curve crosses that line is its Ct.

Because the signal needs more cycles to reach the threshold when there is less template, the Ct value increases with a decreasing amount of template. A sample with a Ct of 22 therefore started with more target than one at 28 run under the same conditions, but the Ct itself is a cycle number, not a concentration.

For background on the chemistry that produces the fluorescence, see the qPCR and RT-qPCR guide and the TaqMan qPCR guide.

What moves a Ct: a checklist

MIQE 2.0 states that Cq depends on several factors, including the sample processing method, fluorescence acquisition and data analysis. The analysis choices it singles out are the ones most often left unreported. Use this list to check whether two Ct values can be put side by side.

Factor Why it changes the Ct Source
Threshold placement Must sit in the linear phase; a different level gives a different crossing cycle. Thermo Fisher uses the same threshold level when comparing reactions. Thermo Fisher; MIQE 2.0
Fixed or per-run threshold MIQE 2.0 asks whether thresholds are consistent across runs or determined individually for each run. MIQE 2.0
Baseline correction The baseline must be subtracted from each reaction individually. Sloped baselines can significantly affect the shape and position of amplification curves, and manual adjustment of the early cycles defining the baseline may be required. MIQE 2.0
Reagents and run conditions Ct values from PCR reactions run under different conditions or with different reagents cannot be compared directly. Thermo Fisher
Sample processing and acquisition Named by MIQE 2.0 as factors on which Cq depends. MIQE 2.0

Reaction efficiency and the standard curve

Thermo Fisher notes that a standard-curve slope of -3.3 ± 10% reflects an efficiency of 100% ± 10%. The slope links Ct to template amount across a dilution series, which is why efficiency should be checked before Ct differences are turned into quantities. The melt curve analysis guide covers a separate check, whether the signal that produced the Ct came from the intended product.

Reporting Ct and Cq values: a MIQE 2.0 checklist

  • State the method of baseline correction and Cq determination.
  • Report the frequency and Cq of negative controls.
  • Collect raw fluorescence data and make it available on request.
  • Call an unknown positive only if its Cq precedes that of the negative control by at least 5 cycles.
  • Avoid further statistical tests on Cq values themselves.

The wider reporting framework is laid out in the MIQE qPCR reporting checklist.

Common misunderstandings

Ct and Cq are different measurements

They are not. MIQE 2.0 introduced Cq as the replacement name for Ct (and Cp); the value is read the same way.

A Ct is a concentration

It is a cycle number and a relative measure of target. Converting it to an amount needs a standard curve or a comparative method with checked efficiency.

A late Ct means a negative result

For validated qualitative assays, MIQE 2.0 says a positive reaction should be reported irrespective of the Cq value, and that the LLOQ or LOD should not be used as cut-offs for reporting negative findings. It also notes that the LOD cannot be below 3 target copies, that variability increases below 10 copies per reaction, and that at 90% efficiency the LLOQ is reached at 35 cycles.

FAQ

What is the difference between Ct and Cq?

Nothing in the measurement. Cq (quantification cycle) is the term used in MIQE 2.0, formerly known as Ct or Cp.

What is the threshold in qPCR?

A fluorescence level set in the linear phase of the amplification plot. The cycle at which a curve crosses it is the Ct.

Why does Ct change between runs?

Reagents, run conditions, sample processing, fluorescence acquisition, baseline correction and threshold choice can all shift it. Thermo Fisher says Ct values from reactions run under different conditions or reagents cannot be compared directly.

Does a high Ct value mean the sample is negative?

Not by itself. In qualitative assays MIQE 2.0 says positives should be reported regardless of Cq, and LOD or LLOQ should not be used as cut-offs for negatives.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Ct Values in qPCR: What the Cycle Threshold Is and What Moves It

Ask your first 2 questions free below. Subscribers get 150 a day for $29 a month.

An AI assistant specialized in research administration. It cites the sources behind every answer, labels web answers and says when it can't answer.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays free to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →