ECG Baseline Wander: Causes, Filter Cutoffs and What to Check

What causes ECG baseline wander, why filtering it distorts the ST segment, the AHA cut-offs, and what to check on the machine and electrodes.

Written and maintained by CASRAI Editorial Board

Published · Last updated

This page describes products and general use; it is not medical advice.

On this page

ECG baseline wander is slow, low-frequency drift of the ECG trace away from its flat reference line. Clinic staff and buyers who see it usually want to know what causes it, whether the machine’s filter setting is part of the answer, and what to check on the device and electrodes. Who uses this page: clinic staff running ECG machines, technicians, and people comparing or specifying ECG equipment. For the wider device, see what an ECG machine is.

What ECG baseline wander is

Baseline wander (BW) is described by Fasano and Villani (2013) as “a kind of noise affecting almost all bioelectrical signals”, with the electrocardiogram the worst affected. Li and Boulanger (2021) describe it as low-frequency noise caused by respiration and body motion, and note that baseline wander and motion artifacts can lead to false diagnoses. In signal terms it is additive low-frequency noise, modelled by Fasano and Villani in the range 0 to 0.8 Hz, extending to 1 Hz or more during stress tests. For the general idea of separating signal from noise, see what signal processing is.

What causes it

The two papers name the same main causes. The right-hand column is a plain reading of where each one sits, not a diagnostic rule.

Cause (as named in the sources) Where it comes from Source
Respiration Breathing moves the chest and the signal baseline together Fasano and Villani 2013; Li and Boulanger 2021
Patient movement / body motion Movement of the body during recording Fasano and Villani 2013; Li and Boulanger 2021
Fluctuations of impedance between electrodes and skin The electrode-skin contact, which is the equipment-side item: the electrode and how it is attached Fasano and Villani 2013

Electrode micromovement produces related artifacts in other biosignals; see EMG electrodes for that context.

Why filtering baseline wander is a trade-off

The short answer: drift and the heartbeat signal overlap in frequency, so removing drift with a simple filter can distort the ECG. Fasano and Villani explain that baseline wander and the ECG share the low-frequency band, and that distortion there “negatively affects the shape of the ST segment”. The simplest fix is a high-pass filter with a cut-off around 0.8 Hz, but separating the two purely by frequency “may induce unacceptable artifactual distortions in the ST segment”.

Fasano and Villani report that the American Heart Association (AHA) recommends a cut-off of 0.05 Hz for routine filters, or 0.67 Hz or below for linear digital filters with zero phase distortion, and that some residual baseline wander may still remain after filtering. Li and Boulanger add that fixed digital filters can introduce nonlinear phase distortion and displace key points of the waveform, and that noise-free ECG spans roughly 0.1 to 100 Hz.

Filter approaches compared

Approach What the sources say Source
High-pass filter, cut-off about 0.8 Hz Simplest approach; frequency-domain separation can distort the ST segment Fasano and Villani 2013
AHA routine filter, 0.05 Hz Recommended cut-off for routine filters; residual wander may remain AHA recommendation as summarised by Fasano and Villani 2013
AHA linear digital filter with zero phase distortion, 0.67 Hz or below Recommended cut-off for this filter type AHA recommendation as summarised by Fasano and Villani 2013
Adaptive, median and wavelet-based filtering Other common approaches; adaptive filters work well when noise is limited to a fixed frequency range Fasano and Villani 2013; Li and Boulanger 2021
Fixed digital filters (general) Can introduce nonlinear phase distortion and displace key points; perform poorly on noise near ECG frequencies Li and Boulanger 2021

What to check on the device and electrodes

These are questions to ask, drawn from the causes and trade-offs above, not rules from a standard.

  • Electrode-skin contact. Impedance fluctuation at the electrode-skin interface is a named cause, so the electrode and its attachment are the equipment-side items to ask about. The electrode and device instructions for use govern preparation and placement.
  • Movement and breathing. The sources list respiration and body motion as causes of drift, alongside electrode-skin impedance.
  • The device’s stated low-frequency cut-off. Ask whether the manufacturer’s documentation states the cut-off and what type of filter is used (for example linear digital with zero phase distortion), so it can be compared with the AHA figures above.
  • Whether filtering is described at all. Documentation that explains the filter helps staff understand why a trace looks the way it does.

Equipment of this kind is listed under diagnostic, monitoring and imaging supplies.

Common mistakes

  • Treating all drift as a machine fault, when the sources name respiration, movement and electrode-skin impedance as the main causes.
  • Assuming a stronger high-pass filter is always better; the sources describe ST-segment and phase distortion as the cost.
  • Assuming filtering removes drift completely; residual baseline wander may remain.

FAQ

What is baseline wander on an ECG?

Slow, low-frequency drift of the trace, mainly caused by respiration, patient movement and changes in electrode-skin impedance (Fasano and Villani, 2013).

Which frequencies does it occupy?

Fasano and Villani model it as 0 to 0.8 Hz, extending to 1 Hz or more during stress tests.

What cut-off does the AHA recommend?

As summarised by Fasano and Villani: 0.05 Hz for routine filters, or 0.67 Hz or below for linear digital filters with zero phase distortion.

Why not just filter it harder?

Because the drift overlaps the low-frequency ECG band, and aggressive filtering can distort the ST segment.

Sources

  • Fasano A, Villani V. Baseline Wander Removal in ECG and AHA Recommendations. Computing in Cardiology 2013;40:1171-1174. pprd.cinc.org (checked 2026-11-03).
  • Li H, Boulanger P. An Automatic Method to Reduce Baseline Wander and Motion Artifacts on Ambulatory Electrocardiogram Signals. Sensors 2021;21(24):8169. PMC8708403 (checked 2026-11-03).
  • Origin of the AHA recommendation: Kligfield P et al., The Electrocardiogram and Its Technology, Circulation 2007;115:1306-1324, doi 10.1161/CIRCULATIONAHA.106.180200. Figures on this page are cited as summarised by Fasano and Villani.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about ECG Baseline Wander: Causes, Filter Cutoffs and What to Check

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →