Written and maintained by CASRAI Editorial Board
Published · Last updated
This page describes products and general use; it is not medical advice.
On this page
An ETCO2 detector is a small, single-patient device that attaches to an airway and shows whether exhaled carbon dioxide (CO2) is present. Clinic staff, buyers and students usually look for the same things: what type it is, which size matches which patient weight, and how it differs from a capnometer or capnography monitor. This page answers those, using a manufacturer’s published specifications and the American Society of Anesthesiologists (ASA) monitoring standard.
Who uses it: clinicians who manage an endotracheal tube (ETT) or other airway, and the staff who stock and maintain airway supplies.
What an ETCO2 detector is
“ETCO2” stands for end-tidal carbon dioxide, the CO2 level at the end of an exhaled breath. In supply catalogues, “ETCO2 detector” most often means a disposable colorimetric CO2 detector: a device that sits between the endotracheal tube and the breathing circuit and changes color when it senses exhaled CO2. One widely listed example, the Medtronic Nellcor Easy Cap II adult and PediCap pediatric detectors, describes a color change from purple to yellow in response to exhaled CO2, and states its purpose as helping clinicians verify proper ETT placement. For the broader physiology and the monitor-based method, see our guide to what capnography is.
How it is built
A colorimetric detector is a short in-line housing with standard airway connectors at each end. In the Medtronic example both sizes have a 15 mm inside-diameter connector on the patient end and a 15 mm outside-diameter connector on the circuit end, and the product is latex-free. The detector responds breath by breath and gives continuous visual feedback for up to two hours. It has no screen, number or alarm: the reading is the color the user sees.
Colorimetric detector vs capnometer vs capnography
These are often confused because all of them involve end-tidal CO2. The ASA Standards for Basic Anesthetic Monitoring (last amended October 15, 2025; checked 2026-10-30) separate two jobs: confirming that an airway is placed correctly, and monitoring ventilation continually afterward.
| Device | What it shows | How the ASA standard treats it |
|---|---|---|
| Colorimetric ETCO2 detector | Color change in response to exhaled CO2 (Medtronic example: purple to yellow); no numeric value | Not named among the quantitative methods listed in the standard |
| Capnometer | A numeric end-tidal CO2 value | Named as a quantitative method for continual end-tidal CO2 analysis |
| Capnography monitor | Numeric value plus a waveform | Named as a quantitative method for continual end-tidal CO2 analysis |
The standard says correct positioning of an endotracheal tube or supraglottic airway “must be verified by clinical assessment and by identification of carbon dioxide in the expired gas.” It then requires continual end-tidal CO2 analysis from placement until extubation, removal or transfer, “using a quantitative method such as capnography, capnometry or mass spectroscopy.” The practical reading for buyers and students is that a color-change detector and a quantitative monitor are different tools covered by different parts of that wording. Which one applies in a given setting is a decision for the responsible clinicians and the facility’s own policy.
Sidestream vs mainstream capnography
Quantitative monitors are described by where the CO2 is measured. In a mainstream monitor the infrared sensor sits in the breathing circuit close to the patient, between the tracheal tube and the Y-piece, so gas is measured where it is exhaled. In a sidestream monitor a small gas sample is drawn through a sampling line and analyzed away from the patient (PubMed Central, comparison of mainstream and sidestream techniques, checked 2026-10-30). Both are monitor-based methods, not disposable colorimetric detectors.
Adult and pediatric sizes
The table below uses the manufacturer’s published specifications for one product family (Medtronic Nellcor Easy Cap II and PediCap, checked 2026-10-30). Other brands have their own values, so check the instructions for use (IFU) of the product you stock.
| Specification | Adult (Easy Cap II) | Pediatric (PediCap) |
|---|---|---|
| Recommended patient weight | Over 15 kg | 1 kg to 15 kg |
| Internal volume | 22 ± 3 mL | 3 mL |
| Resistance to flow | 4.4 ± 1.0 cm H2O at 60 L/min | 2.5 ± 0.5 cm H2O at 10 L/min |
| Patient-end connector | 15 mm ID | 15 mm ID |
| Circuit-end connector | 15 mm OD | 15 mm OD |
| Use period | Up to 2 hours | Up to 2 hours |
| Order codes | EASYCAPII (24 per box), EASYCAPII6 (6 per box) | PEDICAP (24 per box), PEDICAP6 (6 per box) |
How to choose
The manufacturer states its sizes by patient weight band (over 15 kg for adult, 1 kg to 15 kg for pediatric) and lists 15 mm connectors for both. The ASA standard separates verifying airway placement from continual quantitative end-tidal CO2 analysis. Product choice is a decision for trained clinicians under facility protocol. For the airway devices that these detectors connect to, see our guides on laryngeal mask airway sizes and the pop-off valve on a manual resuscitator. Related airway and monitoring supplies are listed in the diagnostic and monitoring equipment category.
Safe use and handling basics
- The manufacturer states that the Easy Cap II and PediCap are latex-free.
- Follow the manufacturer’s instructions for use (IFU) for the specific model; this page does not replace it.
- Product choice and interpretation of any reading are decisions for trained clinicians under facility protocol.
Common mistakes
- Treating a colorimetric detector as continual quantitative monitoring. The ASA standard lists capnography, capnometry or mass spectroscopy for that purpose.
- Confusing the monitor types: sidestream and mainstream describe where a capnography monitor measures CO2, not types of disposable detector.
Frequently asked questions
What is an ETCO2 detector?
It is an in-line device that indicates exhaled CO2. The common disposable type changes color (purple to yellow in the Medtronic example) and is used to help verify endotracheal tube placement.
Is an ETCO2 detector the same as capnography?
No. Capnography and capnometry are quantitative methods that the ASA standard lists for continual end-tidal CO2 analysis. A colorimetric detector gives a color indication rather than a numeric value or waveform.
What is the difference between sidestream and mainstream capnography?
Mainstream places the sensor at the airway in the circuit. Sidestream draws a gas sample through a sample line to a sensor in the monitor.
How long can a colorimetric detector be used?
The Medtronic Easy Cap II and PediCap are listed for use up to two hours. Other products differ, so check the IFU.
Sources
- Medtronic, Nellcor adult and pediatric colorimetric CO2 detector product page (checked 2026-10-30): https://www.medtronic.com/covidien/en-us/products/intubation/nellcor-adult-pediatric-colorimetric-co2-detector.html
- American Society of Anesthesiologists, Standards for Basic Anesthetic Monitoring, last amended October 15, 2025 (checked 2026-10-30): https://www.asahq.org/standards-and-practice-parameters/standards-for-basic-anesthetic-monitoring
- PubMed Central, Capnogram slope and ventilation dead space parameters: comparison of mainstream and sidestream techniques (checked 2026-10-30): https://pmc.ncbi.nlm.nih.gov/articles/PMC4913399/