Short answer: Form FDA 3500 is the MedWatch form for voluntary reports of adverse events, product use errors, product quality problems and therapeutic failures. FDA designed it for healthcare professionals; Form FDA 3500B is the consumer and patient version, and Form FDA 3500A is the separate mandatory-reporting form. This page is general information, not medical or legal advice.
What Form FDA 3500 is
FDA’s MedWatch forms page describes Form FDA 3500 as designed for use by healthcare professionals, who can submit it through MedWatch Online or download it, complete it and submit it to MedWatch (FDA page checked 2026-11-01; the page states content current as of 10/22/2025). FDA’s instructions say health professionals or consumers may use the form for voluntary reporting for prescription and over-the-counter medicines, biologics, medical devices, combination products, cosmetics and cannabinoid hemp products such as those containing CBD.
The reporter does not need to be certain of a cause. FDA’s instructions state that it is not necessary to be certain of a cause and effect relationship, that suspicion of an association is sufficient reason to report, and that submitting a report is not an admission that medical personnel or the product caused or contributed to the event. For adverse events, FDA asks reporters to limit submissions to serious events: death, life-threatening events, hospitalization (initial or prolonged), disability or permanent damage, congenital anomaly or birth defect, required intervention to prevent permanent impairment or damage (devices), and other serious important medical events.
Form 3500, 3500B or 3500A: which one?
The three MedWatch forms differ by reporter role and by whether the report is voluntary or mandatory. The table below uses only FDA’s own descriptions. For the mandatory side in depth, see the section-by-section guide to Form 3500A and the overview in FDA MedWatch.
| Form | FDA’s description | Who it is for |
|---|---|---|
| Form FDA 3500 | Voluntary reporting; designed for use by healthcare professionals | Healthcare professionals (consumers may also use it) |
| Form FDA 3500B | A consumer/patient-friendly version of the 3500 reporting form | Consumers and patients |
| Form FDA 3500A | Mandatory reports of adverse events and device malfunctions, used only when electronic submission is not required | IND researchers, manufacturers, distributors, importers and user facility personnel |
Form FDA 3500B and ‘medwatch 3500’
FDA says its 3500B is customized to make completion easier for non-health professionals, and that a consumer may continue to use the 3500. People searching for medwatch 3500 or medwatch form 3500 are looking for this same form: MedWatch is the FDA safety reporting program, and 3500 is the form number. The 3500B can be completed online or downloaded and submitted to MedWatch.
Reports that do not go on Form 3500
FDA’s instructions send some reports elsewhere:
| Event or product | Where FDA says it goes |
|---|---|
| Vaccine adverse events | Vaccine Adverse Event Reporting System (VAERS) |
| Tobacco product problems, including e-cigarettes | Safety Reporting Portal |
| Animal drug, device, pet food and livestock feed problems | FDA veterinary product reporting (fda.gov/vetproductreporting) |
| Events with investigational (IND) study drugs | As required in the study protocol; generally not submitted to MedWatch as voluntary reports |
The seven sections, A to G
FDA’s instructions organize Form FDA 3500 into these sections (source: FDA instructions page, checked 2026-11-01):
- A, Patient Information: identifier, age or date of birth, sex, weight, ethnicity and race. The identifier must not be the patient’s name or social security number; initials or a patient number that lets the reporter find the case again are what FDA describes.
- B, Adverse Event or Product Problem: type of report, outcome, event date, report date, a description of the event, relevant tests and relevant history.
- C, Product Availability: whether the product is available for evaluation and whether a picture exists. FDA says not to send the product to FDA.
- D, Suspect Products: medicines, biologics and cosmetics; up to two suspect products per form.
- E, Suspect Medical Device: brand name, common device name, product code, manufacturer, identification numbers, operator, implant and explant dates, reprocessing and third-party servicing.
- F, Other (Concomitant) Medical Products: other products in use that the reporter does not think were involved.
- G, Reporter: contact details, whether the reporter is a health professional, occupation, other places the report was sent, and the option to withhold identity from the manufacturer.
What to gather before filing: a device and lab-supply checklist
For a suspected device or supply problem, Section E collects device identifiers. FDA’s instructions ask for these identifiers exactly as they appear on the device or its labeling, including spaces and hyphens:
- Brand name and common device name (FDA asks reporters not to use broad terms such as ‘catheter’, ‘valve’ or ‘screw’).
- Manufacturer name, city and state; if a single-use device was reprocessed and reused, the reprocessor is identified instead.
- Model number, lot number, catalog number, expiration date and serial number, as printed on the device, label or packaging.
- The Unique Device Identifier (UDI), recorded with all numbers, letters, parentheses and symbols; it is found below the barcode.
- Operator of the device, implant and explant dates if applicable, and whether a third party ever serviced the device.
- Whether the product is still available for evaluation, whether it was returned to the manufacturer (and when), and whether a picture exists (Section C).
FDA also states that a separate complete form goes in for each patient (with an exception for one device that affected several patients), and that if more than one suspect device is involved, a form with only Section E is attached for each additional device. FDA does not define a minimum data set; its instruction is to complete all sections that apply.
Confidentiality and institutional reporting
FDA states that the patient’s identity is held in strict confidence and protected to the fullest extent of the law, and that it will not disclose the reporter’s identity to the public in response to a Freedom of Information Act request. In serious cases the reporter’s identity may be shared with the manufacturer unless block G5 is marked.
A separate duty can apply to institutions. FDA’s instructions note that if a serious adverse event with a device occurred in a device user facility (other than a doctor’s office), that facility may be legally required to report to FDA and/or the manufacturer, and ask the reporter to notify the person who handles such reporting. That mandatory route uses Form 3500A or its electronic equivalent; see Safe Medical Devices Act reporting duties, the FDA definition of serious injury and MAUDE reports.
Common misunderstandings
- ‘Voluntary’ does not mean unimportant. It means the reporter is not under a legal obligation to file; facilities and manufacturers may have one through a different form.
- Certainty is not required. FDA’s instructions say suspicion of an association is enough.
- The product is not mailed to FDA. Section C asks whether it is available, and FDA says not to send it.
- Do not put names in the narrative. FDA’s instructions say not to identify any patient, physician or institution by name in block B5; the reporter’s identity belongs in Section G.
- 3500B is not a different program. It is the consumer-friendly version of the same voluntary form.
FAQ
What is MedWatch form 3500?
It is FDA’s voluntary adverse event and product problem reporting form, Form FDA 3500, designed for healthcare professionals and submitted through MedWatch Online or as a downloaded form.
What is the difference between Form FDA 3500 and 3500B?
Form FDA 3500B is FDA’s consumer and patient-friendly version of the same voluntary report. Form 3500 is designed for healthcare professionals.
What is the difference between Form FDA 3500 and 3500A?
3500A is for mandatory reports from IND researchers, manufacturers, distributors, importers and user facility personnel, and FDA says it is used only when electronic submission is not required. 3500 is voluntary.
Can a consumer use Form FDA 3500?
Yes. FDA’s instructions say consumers may continue to use the 3500, and point them to the 3500B as an easier option.
Sources
- U.S. Food and Drug Administration, Instructions for Completing Form FDA 3500, checked 2026-11-01.
- U.S. Food and Drug Administration, MedWatch Forms for FDA Safety Reporting, checked 2026-11-01 (content current as of 10/22/2025).