General labeling information, not medical advice. For how a specific device must be handled or used, follow that device’s own instructions for use.
Short answer
The fragile symbol on a medical device label, listed in ISO 15223-1 as reference 5.3.1 (“Fragile, handle with care”), indicates a medical device that can be broken or damaged if not handled carefully. It is a handling mark for the device; it says nothing about MRI conditions, the maker or the legal representative, which have their own symbols.
Definition and scope
Medical device labels use graphic symbols alongside or instead of words. The standard that lists the general-purpose ones is ISO 15223-1; MRI-related markings come from a separate ASTM standard, F2503. CooperSurgical’s published symbols glossary lists the fragile mark with the wording “Indicates a medical device that can be broken or damaged if not handled carefully” and cites ISO 15223-1 reference 5.3.1 (checked 1 November 2026). That wording is the manufacturer’s own explanation of the symbol.
For context on the broader family of marks, see the guide to medical device labeling symbols, which covers sterile and single-use marks, and the guides to the date of manufacture symbol and the WEEE symbol.
How the symbol is used in practice
A symbol on a US medical device label cannot stand alone by default. Under 21 CFR 801.15(c)(1)(i), a symbol shown without adjacent explanatory text must either be contained in a standard that FDA recognizes under section 514(c) of the Federal Food, Drug, and Cosmetic Act, or be established in a standard developed by a standards development organization (SDO) that the manufacturer determines is likely to be read and understood by the ordinary individual. In both cases it must be used according to that standard and be explained in a symbols glossary. The labeling must include that glossary on paper or electronically, and the package label must carry a prominent statement saying where to find it. Under 21 CFR 801.15(c)(1)(ii), using a symbol that does not meet these conditions renders the device misbranded under section 502(c) (21 CFR 801.15, checked 1 November 2026).
So a fragile symbol that appears without words is meant to be paired with a glossary entry. That is why manufacturers such as CooperSurgical publish one.
Related symbols people look up with it
Searches for the fragile symbol often sit beside other label marks. The table gives the CooperSurgical glossary’s definition of each, with the standard reference the glossary cites.
| Symbol | Standard reference given | Glossary definition |
|---|---|---|
| Fragile, handle with care | ISO 15223-1, 5.3.1 | Indicates a medical device that can be broken or damaged if not handled carefully. |
| Manufacturer | ISO 15223-1, 5.1.1 | Indicates the medical device manufacturer. |
| Authorized representative | ISO 15223-1, 5.1.2 | Indicates the authorized representative in the identified country or jurisdiction. |
| MR Conditional | ASTM F2503, Table 2; 7.4.6.1 | An item with demonstrated safety in the MR environment within defined conditions. |
| MR Unsafe | ASTM F2503, Table 2; 7.4.9.1 | An item which poses unacceptable risks to the patient, medical staff or other persons within the MR environment. |
Manufacturer symbol
Reference 5.1.1 indicates the medical device manufacturer. It identifies who makes the device, which is a different question from how to handle it.
EU rep and authorized representative symbol
Reference 5.1.2 indicates the authorized representative in the identified country or jurisdiction. In the glossary the symbol carries a placeholder for a country code, so the mark is read together with the code printed on the label.
MR Conditional symbol and MRI safety labels
MRI safety labels describe an item’s relationship to the magnetic resonance environment rather than its physical fragility. MR Conditional means demonstrated safety in that environment within defined conditions; MR Unsafe means the item poses unacceptable risks there. “Conditional” signals that the conditions are stated by the labeling for that specific item, so the symbol alone does not tell a reader what the conditions are. FDA publishes the guidance “Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment” (issued October 2023), which applies to all medical devices that might be used in the MR environment (FDA guidance page, checked 1 November 2026).
Where it matters for labs, clinics and supply decisions
A short reading checklist built from the sources above:
- Find the symbol’s reference number in the manufacturer’s glossary (for example 5.3.1 for fragile).
- Find the statement on the package label saying where the symbols glossary is located, as 21 CFR 801.15 requires for symbols without text.
- Treat the fragile symbol as a handling note about the device itself, and read MR symbols as statements about the MR environment that depend on the labeled conditions.
The hemoclip guide shows how a specific device type is described in its own labeling: hemoclip types and sizes.
Common misunderstandings
- Fragile symbol and shipping marks are not the same thing. The fragile mark in ISO 15223-1 (5.3.1) is a medical device label symbol, described as indicating a device that can be broken or damaged if not handled carefully.
- MR Conditional does not mean MR safe. It means safety was demonstrated only within defined conditions.
- A symbol without a glossary is not enough in the US. 21 CFR 801.15 ties unaccompanied symbols to a recognized or SDO standard and a labeling glossary.
- The manufacturer and authorized representative marks are different. 5.1.1 and 5.1.2 identify different parties.
FAQ
What does the fragile symbol mean on a medical device label?
CooperSurgical’s glossary gives ISO 15223-1 reference 5.3.1, “Fragile, handle with care”, and explains it as indicating a medical device that can be broken or damaged if not handled carefully.
Is the fragile symbol the same as the manufacturer or authorized representative symbol?
No. In the same glossary, 5.1.1 indicates the medical device manufacturer and 5.1.2 indicates the authorized representative in the identified country or jurisdiction.
Is MR Conditional the same as MR safe?
No. The glossary describes MR Conditional (ASTM F2503) as an item with demonstrated safety in the MR environment within defined conditions, and MR Unsafe as an item that poses unacceptable risks there.
Can a US device label show a symbol with no words?
Under 21 CFR 801.15(c)(1)(i), yes, if it comes from an FDA-recognized standard or a qualifying SDO standard, is used as that standard specifies, and is explained in a symbols glossary in the labeling, with a prominent statement on the package label saying where the glossary is.
Sources
- CooperSurgical, Symbols Glossary, https://www.coopersurgical.com/symbols_glossary/, checked 1 November 2026.
- 21 CFR 801.15, Medical devices; prominence of required label statements, https://www.law.cornell.edu/cfr/text/21/801.15, checked 1 November 2026.
- FDA, Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, https://www.fda.gov/regulatory-information/search-fda-guidance-documents/testing-and-labeling-medical-devices-safety-magnetic-resonance-mr-environment, checked 1 November 2026.
- CASRAI guide to ISO 15223-1.