Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General product information only, not medical advice. Do not change oxygen settings or equipment without your clinician or supplier. Figures were checked against the sources listed below on 2026-11-01.
A stationary (home or facility) oxygen concentrator typically draws about 280 to 600 watts while running, depending on the model and flow setting; portable units draw less. One measured example, the Drive DeVilbiss Compact 525, is rated at approximately 290 W at 2 l/min and approximately 312 W at 5 l/min, which works out to about 7 kWh per day if it ran around the clock. This guide shows how to find the figure for a specific machine, convert it to kWh, and check the outlet and backup power.
Short answer: how much electricity does an oxygen concentrator use?
- Stationary units that deliver 5 l/min: about 280 to 350 W per concentrator, per the Sustainable Energy for All (SEforALL) oxygen concentrator factsheet. Units delivering more than 5 l/min: about 400 to 600 W.
- Range quoted for health-facility units: 300 to 600 W, per the PATH and Clinton Health Access Initiative brief on oxygen concentrators.
- Efficiency benchmark: PATH states that concentrators should have a power consumption of less than 70 W per l/min. At 5 l/min, that benchmark is a ceiling of 350 W.
- Portable concentrators consume less power than stationary counterparts (SEforALL). For their battery runtime, see how long an oxygen concentrator battery lasts.
How much power does an oxygen machine use?
“Oxygen machine” usually means the same device: an electrically powered oxygen concentrator. The answer is the same as above, a few hundred watts for a stationary unit, and the exact figure is printed on the machine’s rating label or in its specification sheet. For what the device does and why it needs mains power, see what an oxygen concentrator is. Unlike a cylinder, it cannot run without power; the trade-offs are covered in oxygen concentrator vs oxygen tank.
What you need to estimate your own use
- The model name and the rating label or specification sheet (or instructions for use).
- The flow setting the machine is normally run at, as prescribed. This guide does not suggest settings.
- Roughly how many hours per day the machine runs.
Steps: estimating watts and kWh
- Find the rated power. Look for watts (W) or amps (A) and volts (V) on the label or specification sheet.
- Note what the number means. Some sheets give a figure at a stated flow, others an average or a maximum. The Drive DeVilbiss Compact 525 sheet gives two: approximately 290 W at 2 l/min and approximately 312 W at 5 l/min.
- Convert to energy. kWh = watts x hours / 1000.
- Compare with the benchmark. Divide watts by the flow in l/min. PATH’s benchmark is under 70 W per l/min.
- Check the supply. Confirm voltage and frequency match your outlets (see below), and plan for outages.
How to read the wattage on the label or spec sheet
| Label wording | What it tells you | Example from a sourced sheet |
|---|---|---|
| Power consumption at a stated flow | Draw while delivering that flow | Drive DeVilbiss Compact 525: approx. 290 W at 2 l/min; approx. 312 W at 5 l/min |
| Voltage, frequency, current | Supply the unit is built for | Same sheet: 230 V, 50 Hz, 3.2 A |
| Power management feature | Draw may fall at lower flow | Same sheet: “Turn-Down Technology” provides less power consumption at lower flow rates |
The same sheet lists a flow range of 0.5 to 5 l/min, oxygen concentration of 93% +/-3% and a weight of 16.3 kg. It is one example unit, not a recommendation, and other models differ.
Worked example: watts to kWh per day
This is an illustration of the arithmetic using the two figures above, not a statement about any particular household. It involves no tariff or money.
| Rated draw | Hours/day | kWh per day | kWh per 30 days |
|---|---|---|---|
| 290 W (at 2 l/min) | 24 | 6.96 | 208.8 |
| 312 W (at 5 l/min) | 24 | 7.488 | 224.64 |
| 350 W (70 W per l/min x 5 l/min, the PATH ceiling) | 24 | 8.4 | 252 |
For fewer hours, scale the result: 312 W for 8 hours is 312 x 8 / 1000 = 2.496 kWh. Multiply your kWh by your supplier’s rate from your own bill if you want a cost; this guide does not state rates.
What affects power draw
- Flow setting. The sheet example draws less at 2 l/min than at 5 l/min.
- Model class. SEforALL separates 5 l/min units (280 to 350 W) from units above 5 l/min (400 to 600 W).
- Portable vs stationary. Portable units consume less than stationary ones (SEforALL), though they run on batteries.
- Continuous operation. SEforALL notes concentrators are designed for continuous operation, so daily hours can be high.
Power supply and backup checks
- Voltage and frequency. PATH names 120 V / 60 Hz and 220 V / 50 Hz as the two most common configurations, and SEforALL notes separate models for 110-120 VAC (typically 60 Hz) and 220-240 VAC (typically 50 Hz). Use the model built for your supply.
- Continuous, reliable power. SEforALL calls a source of continuous and reliable power critical.
- Backup. PATH says back-up sources such as generators or uninterruptible power supplies can help provide oxygen during outages, and that voltage stabilizers and surge suppressors can help protect the electronics and extend unit life.
- Alarms. The sheet example has a power failure alarm; PATH also lists an audio alarm for power supply failure.
Home safety topics such as open flames and tubing are covered in home oxygen safety.
What can go wrong
- Using a model built for a different voltage or frequency than the outlet supplies.
- Assuming one flat wattage when the sheet gives figures at different flows.
- No plan for a power outage when the machine is the oxygen source.
When to get professional help
Ask the equipment supplier or your clinician about the right model, settings and backup oxygen plan for your situation. Ask a qualified electrician if you are unsure about circuits or outlets. For outages and equipment alarms, follow your supplier’s instructions.
FAQ
How much electricity does an oxygen concentrator use?
Stationary units typically draw about 280 to 600 W (SEforALL, PATH). One 5 l/min example is rated approximately 312 W at 5 l/min, about 7.5 kWh over 24 hours.
How much power does an oxygen machine use?
The same as a stationary concentrator: a few hundred watts. Check the label or specification sheet for the figure for your model.
Does the flow setting change the power used?
It can. The Drive DeVilbiss Compact 525 sheet lists approximately 290 W at 2 l/min and approximately 312 W at 5 l/min and describes power management that reduces consumption at lower flow.
How do I convert watts to kWh?
Multiply watts by hours of use and divide by 1000.
Sources
- Drive DeVilbiss Compact 525 5 L oxygen concentrator specification sheet (hosted by Air Liquide Healthcare Australia, 2026-03), https://au.healthcare.airliquide.com/sites/alh_au/files/2026-03/drive_oxygen_concentrator_5l_525ks_spec_sheet.pdf. Checked 2026-11-01.
- PATH and Clinton Health Access Initiative, “O2 generation and storage: oxygen concentrators”, v2, July 2021, https://media.path.org/documents/O2_generation_and_storage_oxygen_concentrators_v2.pdf. Checked 2026-11-01.
- Sustainable Energy for All, “Efficiency: Oxygen Concentrator” factsheet, https://www.seforall.org/system/files/2020-05/Efficiency-Oxygen-Concentrator.pdf. Checked 2026-11-01.