Implant Stability: Primary vs Secondary, ISQ and Torque Explained

Primary vs secondary implant stability, how insertion torque, ISQ and Periotest measure it, and how stability differs from survival figures.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

Implant stability is how firmly a dental implant is held in the surrounding bone. Primary stability is the mechanical hold present at placement; secondary stability is the biological hold that builds as bone remodels around the implant (osseointegration). Clinicians estimate it with insertion torque, resonance frequency analysis (ISQ) or Periotest values.

This page describes general concepts reported in the research literature; it is not medical advice and does not describe a clinical procedure. Whether a specific implant is stable is judged by the treating dentist.

What implant stability means: primary vs secondary

A systematic review of cortical bone thickness and implant stability defines primary stability as the absence of mobility of the implant after insertion. It depends on how well the fixture mechanically engages the surrounding bone, which in turn depends on the surgical method (drill size relative to fixture size) and on implant morphology such as shape and surface roughness.

Secondary stability, sometimes called biologic stability, is the stability that comes from remodeling and functional regeneration of bone around the implant, that is, osseointegration. The same review notes that during healing, insufficient primary stability can allow excessive micromotion at the bone-implant interface (reported as above roughly 50-100 μm), which can interfere with osseointegration and lead to fibrous scar tissue instead of bone contact.

Osseointegration itself is defined in the historical literature as the formation of a direct structural and functional connection between ordered, living bone and the surface of a load-carrying implant. For related biomaterial and implant context, see the guide to orthopedic implants.

How implant stability is measured

No single number captures stability across the whole healing period. The measures differ in what they record and when they can be read. The comparison below is compiled from the systematic review on cortical bone thickness, a study of resonance frequency analysis across bone types, and a literature review of ISQ measurements.

Measure What it records When it can be read Limits stated in the sources
Insertion torque (IT), in Ncm Resistance felt as the implant is seated, a proxy for primary stability Only at placement Secondary stability cannot be assessed with a torque ratchet or micromotor; no consensus on the minimum torque needed for osseointegration; very high torque can compress bone and may hinder microcirculation
Resonance frequency analysis (RFA), reported as ISQ on a 1-100 scale Stiffness of the implant-bone system; higher ISQ means higher stability At placement, during healing and after the prosthesis is loaded Described as noninvasive; the true ISQ cutoff separating success from early failure has not been determined
Periotest value (PTV) Stiffness of the implant-bone continuum from a percussion-type reading Any time after placement Strongly dependent on the direction and position of excitation, which is why RFA is usually preferred
Insertion energy A dynamic parameter proposed as an alternative to peak torque At placement Described as preliminary

Insertion torque

The systematic review reports that oral surgeons usually refer to 20-40 Ncm as the “ideal” range, but it also states that consensus on the minimum insertion torque needed for osseointegration is lacking. These are reported ranges from the literature, not settings for anyone to apply. Because torque can only be obtained at the moment of placement, it says nothing about how stability changes afterwards.

Resonance frequency analysis and the ISQ scale

RFA uses a transducer attached to the implant to estimate stiffness and converts the result to the implant stability quotient (ISQ), from 1 to 100. The review names the Osstell device with a SmartPeg transducer as the system it discusses. It reports that ISQ values of 57-82 denote successfully osseointegrated implants and that ISQ below 50 predicts a high risk of failure. A later literature review of ISQ measurements concluded, however, that the true cutoff separating success from early failure is still not determined, so these bands are best read as reported associations rather than thresholds.

Periotest

The Periotest assesses stiffness of the implant-bone continuum, but its reading depends heavily on where and from which direction the instrument is applied. The review adds that even minimal clinical mobility is considered a symptom of implant failure.

How stability changes over the first weeks

Primary stability does not simply persist. As the bone immediately around the implant remodels, the mechanical hold can fall while the biological hold is still building, so total stability can dip before it recovers. The literature review of ISQ measurements describes the transition from primary-dominated to secondary-dominated stability as taking roughly 5-8 weeks.

One small clinical study illustrates the pattern. Ten patients received 15 implants in the mandible, and ISQ was measured with an Osstell Mentor at surgery and at weeks 1, 2, 3, 4, 5, 7 and 11. Mean ISQ was 77.2 at placement, fell to a lowest mean of 72.13 at week 4, and had risen to 75.6 by week 11, still below the baseline value. The same study treated a primary ISQ of 47 or less as questionable stability. With only 15 implants in 10 patients, this is an illustration of the dip-and-recovery shape, not a norm to compare an individual implant against.

Factors that influence implant stability

  • Bone density and quality. Studies often classify bone with the four-type Lekholm-Zarb scale, based on radiographic evaluation and the resistance felt during drilling. A study of 65 implants in 59 patients measured ISQ immediately, at one month and at three months, and noted that the relationship between bone density and primary or secondary stability remains under debate.
  • Cortical bone thickness. The systematic review found that bone density alone is assessed more often than cortical thickness, that cortical thickness correlates with higher primary stability, and that little is known about its influence on secondary stability and marginal bone loss. The review itself flags scant literature and methodological variability.
  • Implant design. Shape and surface roughness affect primary stability.
  • Surgical technique. The relationship between drill size and fixture size is one of the main determinants cited for primary stability.

Imaging methods used in research settings to examine bone around implants are covered in the guides to micro-CT and to the dental implant drill. For the biological evaluation of implant materials in bone, see ISO 10993-6.

How many years does an implant last? Stability is not survival

There is no single lifespan for a dental implant. The literature reports survival: the percentage of implants still in place at a stated time point, in a stated group of studies. Two meta-analyses show how much the number depends on its conditions.

A meta-analysis of 18 prospective studies of contemporary implant systems reported 10-year implant-level survival of 96.4% (95% CI 95.2-97.5%), with a prediction interval of 91.5-99.4%. A sensitivity analysis that accounted for loss to follow-up gave 93.2% (90.1-95.8%), with a wider prediction interval of 76.6-100%, and the authors noted a possible doubling of the risk of implant loss in older age groups.

A 2024 meta-analysis covering up to 20 years for screw-shaped, rough-surface titanium implants reported 92% (95% CI 82-97%) from three prospective studies (237 implants) and 88% (78-94%) from five retrospective studies (1,440 implants). After imputation for missing data, the prospective-study figure fell to 78%. The authors described failure causes as multifactorial and concluded that follow-up should not end after insertion or after 10 years. They also note that few studies exist at this length. The data table below sets these figures side by side.

Who is the father of osseointegration?

Per-Ingvar Brånemark is widely credited as the founder of modern osseointegration and coined the term. According to a historical review of his career, he found by chance in 1952 at Lund that titanium bonded to bone, while using a titanium chamber in a rabbit leg to observe blood flow. He placed the first titanium implants in a patient with complete edentulism in 1965 and died on 20 December 2014 in Gothenburg at the age of 85. “Father of osseointegration” is a common honorific rather than a formal title.

Signs of a failing implant: what clinicians record

Whether an implant is failing is a clinical judgement made by a dentist, not something to assess at home. The clinical examination described in a 10-year private-practice study covers probing depth, bleeding on probing, suppuration, mobility and radiographs. That study used operational definitions from the literature: mucositis is bleeding on probing or suppuration without detectable bone loss, peri-implantitis is bone loss above 0.5 mm with or without bleeding or suppuration, and moderate to severe is bone loss above 2.0 mm. The stability literature adds that even minimal clinical mobility is considered a symptom of implant failure. Anyone with symptoms around an implant should contact their dentist.

The “3/2 rule” and implant spacing

People searching for the “3/2 rule” are usually looking for implant spacing rules of thumb. The sources read for this page do not name an official “3/2 rule”; they describe spacing guidance. A 2025 case report and review states that recommendations currently exist for a minimum of 1.5 mm between an implant and an adjacent tooth and at least 3 mm between two adjacent implants, a guidance that originated with Tarnow and colleagues, and that a 1 mm distance between implants is poorly documented. A clinical implant dentistry paper advocates at least 1.5-2 mm from adjacent teeth, especially in the anterior region, and 3-4 mm between implants. This is spacing guidance, not a stability measure.

Common misunderstandings

  • Stability is not survival, and survival is not success. Stability describes the implant-bone hold at a moment; survival counts implants still present; neither captures bone loss or soft-tissue health on its own.
  • A high ISQ at placement does not guarantee the outcome. The ISQ cutoff separating success from early failure has not been determined, and values fall and rise during healing.
  • There is no agreed “correct” torque. The 20-40 Ncm range is how surgeons commonly describe the ideal, and consensus on a minimum is lacking.
  • More torque is not always better. Very high torque can compress bone and hinder microcirculation.
  • Bone density does not settle the question. Its relationship with stability is still debated, and cortical thickness appears relevant too.

For the wider research context, see what dental research is.

Frequently asked questions

How many years does an implant last?

No single lifespan is established. Published meta-analyses report survival of about 96% at 10 years in prospective studies, falling to about 93% when loss to follow-up is accounted for, and 88-92% in a 20-year analysis of rough-surface titanium implants, with wide intervals.

Who is the father of osseointegration?

Per-Ingvar Brånemark, who coined the term after observing in 1952 that titanium bonds to bone.

What is the 3/2 rule for dental implants?

It is not a named rule in the sources read. It usually refers to spacing guidance of about 3 mm between adjacent implants and 1.5-2 mm from an adjacent tooth.

What are the signs of a failing implant?

Clinicians record bleeding on probing, suppuration, increased probing depth, bone loss on radiographs and mobility. Any mobility is considered a symptom of failure. Diagnosis belongs to a dentist.

Sources

Data: Reported dental implant survival in meta-analyses, by analysis and time frame

Implant-level survival percentages with 95% confidence intervals reported in two meta-analyses of dental implants: a 10-year analysis of 18 prospective studies (primary and sensitivity analyses) and a 20-year analysis of screw-shaped rough-surface titanium implants (prospective, imputed and retrospective). Values were transcribed from the published abstracts and kept with their stated conditions.

Reported dental implant survival in meta-analyses, by analysis and time frame
AnalysisTime frameSurvival (%)95% CI lower (%)95% CI upper (%)Study basis
Howe 2019 primary analysis10 years96.495.297.518 prospective studies
Howe 2019 sensitivity analysis (loss to follow-up)10 years93.290.195.818 prospective studies
Kupka 2024 prospective studiesup to 20 years9282973 prospective studies, 237 patients
Kupka 2024 prospective studies after imputationup to 20 years783 prospective studies, 237 patients
Kupka 2024 retrospective studiesup to 20 years8878945 retrospective studies, 1,440 patients

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Implant Stability: Primary vs Secondary, ISQ and Torque Explained

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →