Skip to main content
v2026.11,986 entries · CC-BY 4.0

Internal Quality Control in the Laboratory

Internal quality control (IQC) monitors daily laboratory test precision, detecting analytical errors before patient or experimental results are released.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

Short answer

Internal quality control (IQC) is the continuous, daily operational process by which laboratory personnel test known control materials alongside analytical batches to detect immediate errors and monitor measurement precision and accuracy over time. Required by CLIA and ISO 15189 standards, IQC verifies that an analytical testing system performs reliably before patient or research data are released.

Definition and scope

The World Health Organization (WHO) and Clinical and Laboratory Standards Institute (CLSI) define internal quality control as the operational procedures undertaken by laboratory staff for continuous monitoring of analytical operations. The central goal of IQC is to establish whether analytical test runs are sufficiently reliable and precise to permit releasing examination results.

Within the 12 Quality System Essentials of the WHO/CDC/CLSI Laboratory Quality Management System (LQMS) framework, IQC operates under the Process Control essential. It applies across clinical diagnostics, academic core facilities, and analytical testing laboratories. Governed by a formal standard operating procedure (SOP), IQC addresses the analytical testing phase, positioned within a defined work instruction vs. SOP documentation hierarchy.

How it works or how it is used in practice

Internal quality control operates through the regular analysis of characterized control materials alongside test samples, followed by statistical evaluation against established tolerance limits.

Control materials and assay ranges

Control materials are stable substances with established target values. Unlike calibrators that set instrument response factors, controls are analyzed under identical conditions to unknown samples. Under federal regulations (42 CFR § 493.1256), laboratories conducting quantitative testing must run at least two control materials of different concentrations each day of testing. Qualitative testing requires a positive and negative control each testing day.

Regulations also require control testing upon complete reagent lot changes, major preventive maintenance, or critical component replacement. Overseeing this testing forms a core element of CLIA laboratory director responsibilities and rigorous equipment qualification, validation, and calibration protocols.

Statistical process control and Westgard rules

Laboratories quantify performance using Levey-Jennings charts, plotting daily control measurements against the established mean (μ) and action limits set at ±1s, ±2s, and ±3s. Under CLSI C24-Ed4 guidelines, laboratories evaluate runs using Westgard multi-rules:

  • 12s (Warning): One control measurement exceeds mean ± 2s. This serves strictly as a warning to inspect data; the run is not rejected.
  • 13s (Random error rejection): One control measurement exceeds mean ± 3s. The analytical run is rejected.
  • 22s (Systematic error rejection): Two consecutive control measurements exceed the same +2s or -2s threshold. The run is rejected.
  • R4s (Random error rejection): The difference between two controls within a run exceeds 4s. The run is rejected.
  • 41s (Systematic bias): Four consecutive control measurements exceed the same +1s or -1s limit, signaling systematic shift.
  • 10x (Systematic trend): Ten consecutive control measurements fall on one side of the mean, signaling baseline calibration drift.

Quality management relies on complementary activities that serve separate roles in analytical validation. The table below outlines key operational differences between calibrators, internal quality controls, and external quality assessment, based on ISO 15189 and WHO LQMS standards.

Quality Component Primary Purpose Operator Knows Target? Typical Frequency Regulatory Mandate
Calibrator (Standard) Sets the analytical measurement scale and establishes instrument response curves. Yes; exact assigned values. Periodic (lot change, maintenance, or recalibration interval). Required for method calibration under CLIA and ISO 17025.
Internal Quality Control (IQC) Monitors day-to-day precision, detects immediate errors, and gates batch release. Yes; target mean and ±2s / ±3s ranges known. Daily, per analytical run/batch, or per risk-based IQCP interval. Mandatory for testing release under 42 CFR § 493.1256 and ISO 15189:2022.
External Quality Assessment (EQA / PT) Evaluates assay accuracy and trueness against external peer laboratories. No; blind samples with undisclosed values. Periodic (typically 3 times per year). Mandatory for clinical accreditation under CLIA and ISO 15189.

For institutions evaluating compliance across systems, our guide on ISO 15189 vs. CLIA accreditation frameworks details how accreditation bodies assess internal versus external quality requirements.

Where it matters for labs, clinics or supply decisions

IQC policies directly impact procurement choices, consumable management, and laboratory operating costs.

Third-party vs. in-kit control procurement

ISO 15189:2022 Clause 7.3.7.2 highlights using independent, third-party control materials where feasible. In-kit controls provided by instrument manufacturers often share reagents or calibrator origins with the test kit. While convenient, in-kit controls can mask subtle calibration shifts. Independent third-party controls provide an unbiased assessment of system stability and sensitivity to lot-to-lot reagent variation.

Out-of-control troubleshooting checklist

When an IQC rule violation occurs, laboratory personnel must follow a standardized investigative checklist before resuming testing:

  1. Quarantine testing output: Immediately halt reporting of all patient or research results generated within the affected analytical run.
  2. Inspect physical integrity: Verify control expiration dates, storage temperatures, vial reconstitution times, and instrument error codes.
  3. Review reagent performance: Inspect on-board reagent lot numbers, fluid lines, and calibration logs for signs of drift.
  4. Test fresh control aliquots: Run a freshly prepared control vial to rule out control vial contamination or evaporation.
  5. Perform recalibration or maintenance: If fresh controls confirm the error, recalibrate the assay or perform maintenance.
  6. Conduct patient lookback: Evaluate samples analyzed since the last acceptable QC run to confirm no compromised data were released.
  7. Log root cause and corrective action: Document the incident, identified cause, resolution, and authorization in accordance with quality management guidelines, similar to practices in culture media quality control.

Common misunderstandings

  • Calibrators can serve as daily quality controls: Calibrators adjust instrument response factors. Testing a calibrator as a control creates circular verification that cannot independently detect calibration curve bias.
  • A 12s rule violation requires shutting down the test system: Under CLSI C24 multi-rule systems, the 12s rule is strictly a warning. Because 4.5% of normally distributed data naturally fall between 2s and 3s, treating 12s as an automatic rejection rule leads to unnecessary testing delays and reagent waste.
  • Repeating a failed control until it passes resolves the issue: Repeating a failed run without investigating the root cause masks systematic error. If a control fails due to systematic shift, repeated runs may eventually pass by random chance while the underlying defect remains.
  • IQC eliminates the need for external proficiency testing: IQC measures ongoing precision and system stability within a single laboratory. External Quality Assessment (EQA) remains necessary to evaluate inter-laboratory trueness.

Frequently asked questions

How often must internal quality control be run in a clinical laboratory?

Under CLIA standard 42 CFR § 493.1256, laboratories must run at least two levels of control materials of different concentrations for quantitative tests, or a positive and negative control for qualitative tests, on each day patient specimens are tested. Testing is also required upon reagent lot changes, major maintenance, or critical component replacement.

What is the difference between random error and systematic error in IQC?

Random error refers to unpredictable fluctuations that cause a single measurement to deviate from the mean (such as electrical noise or pipetting variability), typically flagged by a 13s or R4s rule violation. Systematic error represents predictable, directional bias affecting all measurements (such as reagent deterioration or calibration drift), detected by 22s, 41s, or 10x rule violations.

What should a laboratory do when an IQC result violates a Westgard rejection rule?

When a rejection rule is violated, staff must withhold result reporting, quarantine affected samples, investigate the root cause using an out-of-control protocol, execute necessary corrective actions, and verify resolution with fresh controls before releasing subsequent data.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Internal Quality Control in the Laboratory

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →