Short answer: ISO 14644-2:2015 is the part of the ISO 14644 cleanroom series that sets minimum requirements for a monitoring plan showing a cleanroom or clean zone keeps its air cleanliness, measured as airborne particle concentration. It does not define the cleanliness classes themselves; that is Part 1. ISO’s public catalogue entry for Part 2 states no requalification interval or frequency.
This guide is for lab, clinic and manufacturing staff who need to know what Part 2 is for, what it leaves out, and which records to ask a cleanroom supplier or certifier about. It is general information, not a substitute for the standard itself.
Definition and scope
ISO’s catalogue lists the standard as ISO 14644-2:2015 Cleanrooms and associated controlled environments, Part 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration. It is Edition 2, published December 2015, 14 pages. Its abstract says it “specifies minimum requirements for a monitoring plan for cleanroom or clean zone performance related to air cleanliness by particle concentration, based upon parameters that measure or affect airborne particle concentration.”
The same page states what is out of scope: it “does not address condition monitoring of aspects such as vibration or general maintenance of the engineering systems,” and it does not provide for monitoring particle populations outside the specified lower threshold particle-size range of 0,1 µm to 5 µm. Concentrations of ultrafine particles (smaller than 0,1 µm) “will be addressed in a separate standard.” On 1 November 2026 the ISO page showed the stage “International Standard under systematic review” (90.20); that is a routine review stage and does not mean the standard has been replaced.
How Part 2 fits with Part 1 and regulator practice
The three documents below do different jobs. The last column names the source for each row.
| Document | What it does | Source |
|---|---|---|
| ISO 14644-1:2015 | Classifies air cleanliness by airborne particle concentration; only particle sizes from 0,1 µm to 5 µm are considered for classification, using light-scattering airborne particle counters. | ISO catalogue, standard 53394 |
| ISO 14644-2:2015 | Sets minimum requirements for a monitoring plan covering the same particle-concentration performance. | ISO catalogue, standard 53393 |
| FDA aseptic processing guidance (2004, nonbinding) | Recommends regular monitoring during each production shift, suggests a remote counting system, and says a result outside the established classification at a location should be investigated, with trending evaluated. It does not cite Part 2 by name. | FDA, Sterile Drug Products Produced by Aseptic Processing |
In short: Part 1 answers “which class is this room?”, Part 2 answers “what plan shows it stays that way?”, and a regulator’s guidance describes how monitoring is applied and how excursions are handled in a specific setting. For the class limits themselves see cleanroom classifications and ISO Class 5. As an example of a class mapping, the FDA guidance’s Table 1 lists ISO 5 as 3,520 particles per m³ at 0.5 µm or larger, and notes that ISO 5 equals the older Class 100.
2015 edition versus the withdrawn 2000 edition
ISO’s catalogue shows a first edition, ISO 14644-2:2000 (7 pages), with status Withdrawn and revised by the 2015 edition. The title changed from a testing-and-monitoring formulation to the current “Monitoring to provide evidence of cleanroom performance” wording, which places the monitoring plan at the centre. If a supplier document cites the 2000 edition, it cites a withdrawn text.
Where it matters for labs, clinics and supply decisions
For pharmaceutical settings, pharmaceutical environmental monitoring shows how monitoring programmes are applied; for general lab rooms see laboratory environmental monitoring and clean room labs. The FDA guidance supports one practical point: routine particle monitoring is useful for rapidly detecting significant deviations from qualified norms, and the investigation should match the severity of the excursion.
Records to ask a supplier or certifier about
This checklist is built only from the statements above; it is a set of questions, not a compliance test.
- Which edition of ISO 14644-2 is cited (2015 is the current edition shown by ISO; 2000 is withdrawn)?
- Is there a written monitoring plan for particle concentration, and which parameters does it use?
- Which ISO 14644-1 class is stated, and does the record say the conditions of measurement?
- Does the plan say whether counting is remote or portable, and why? The FDA guidance recommends remote systems for aseptic processing; see portable versus remote particle counters.
- What happens when a result exceeds the class limit: is there an investigation record and trend review?
- Does the plan claim anything outside the standard’s scope, such as vibration or ultrafine particles?
Common misunderstandings
- “Part 2 is the class table.” No. Classification is Part 1.
- “Part 2 fixes a monitoring interval.” The public catalogue entry states no interval. Take any interval from the standard itself or from your quality system, not from a summary.
- “It covers all cleanroom monitoring.” It covers particle concentration in the 0,1 to 5 µm range, not vibration, maintenance, or ultrafine particles.
- “The FDA guidance is Part 2.” The FDA document is separate, nonbinding guidance and does not cite Part 2.
FAQ
What is ISO 14644-2?
The ISO standard setting minimum requirements for a monitoring plan for cleanroom air cleanliness by particle concentration, in its 2015 edition.
Is ISO 14644 2 still current?
The 2015 edition is listed as Published; the 2000 edition is Withdrawn. ISO’s page showed the 2015 edition under systematic review on 1 November 2026.
Does ISO 14644-2 define cleanroom classes?
No. Classes come from ISO 14644-1.
Sources
- ISO, ISO 14644-2:2015: https://www.iso.org/standard/53393.html, checked 1 November 2026.
- ISO, ISO 14644-2:2000 (withdrawn): https://www.iso.org/standard/25008.html, checked 1 November 2026.
- ISO, ISO 14644-1:2015: https://www.iso.org/standard/53394.html, checked 1 November 2026.
- FDA, Guidance for Industry: Sterile Drug Products Produced by Aseptic Processing, September 2004, https://www.fda.gov/media/71026/download, checked 1 November 2026.