Written and maintained by CASRAI Editorial Board
Published
On this page
The LUKE arm is the nickname of a powered, multi-joint upper-limb prosthesis that the US Food and Drug Administration lists under the device name DEKA ARM SYSTEM. FDA granted it through the De Novo pathway (record DEN120016, decision date 9 May 2014) and classifies it as a Class II prescription device under 21 CFR 890.3450. LUKE stands for Life Under Kinetic Evolution, and the arm was developed by DEKA with DARPA funding.
This page is general reference information from regulators, government agencies and peer-reviewed studies. It is not medical advice and does not describe fitting or programming.
What the LUKE arm is
DARPA describes the LUKE arm as developed by inventor Dean Kamen and colleagues at DEKA Research and Development Corp. as part of DARPA’s Revolutionizing Prosthetics program, with additional funding from the US Army Medical Research and Materiel Command. The program launched in 2006 with the goal of an advanced electromechanical prosthetic upper limb with near-natural control. DEKA worked with DARPA and the Department of Veterans Affairs (VA) Rehabilitation Research and Development Service under a federal interagency agreement, and engaged directly with amputees in a number of studies, including VA studies. Mobius Bionics was launched in July 2016 to bring the arm to market.
The VA’s 30 June 2017 press release, issued when the VA Secretary unveiled the arm in New York, spells out the name (Life Under Kinetic Evolution), calls the arm a collaboration between VA, DARPA and industry, and says it was the product of nearly eight years of testing and research. It says the entire arm can move as one unit, which reduces the labor-intensive process of controlling one joint at a time, and that it features the first commercially available powered shoulder, with up to 10 powered degrees of freedom.
The research literature does not call it the LUKE arm. The VA-funded studies refer to it as the DEKA Arm, and FDA’s database uses DEKA ARM SYSTEM. Searching for any of the three names is searching for the same device family.
The FDA record: DEN120016
openFDA, the agency’s public data service, holds a premarket record for the device. It is useful because the nickname, the press coverage and the regulatory identity of the arm are easy to confuse, and the record fixes the dates and the category.
| Field | Value in the FDA record |
|---|---|
| Device name | DEKA ARM SYSTEM |
| Applicant | Deka Integrated Solutions Corporation, Manchester, New Hampshire |
| Record number | DEN120016 (a De Novo request, not a 510(k) number) |
| Date received | 15 June 2012 |
| Decision date | 9 May 2014 (decision code DENG, granted) |
| Product code | PAE |
| Review committee | Physical Medicine |
| Classification | Class II, 21 CFR 890.3450 |
The two dates matter because they are often blurred. FDA received the request in June 2012 and granted it in May 2014, so the interval between submission and decision was nearly two years. The record number begins with DEN because this was a De Novo classification request, the route for a new type of low-to-moderate-risk device with no legally marketed predecessor; see the CASRAI entry on FDA De Novo classification. Granting a De Novo request lets FDA create a new device type with its own classification regulation, and that is what happened here.
How FDA classifies this type of arm
The product code PAE is titled “Upper Extremity Prosthesis With Multiple Simultaneous Degrees Of Freedom And Controlled Via Cutaneous Electromyography” in FDA’s classification database, with the definition: an upper extremity prosthesis intended to replace partially or fully amputated or congenitally absent upper extremities and to provide multiple and simultaneous degrees of freedom and functionality. The database marks it as Class II, not implanted, and not exempt from good manufacturing practice requirements.
The governing rule is 21 CFR 890.3450, added to the Code of Federal Regulations on 18 October 2016 (81 FR 71612). It identifies the device as a prescription upper extremity prosthesis that includes a simultaneously powered elbow and/or shoulder with more than two simultaneous powered degrees of freedom, controlled by non-implanted electrical components. Simple, manually controlled electrical components such as switches do not count as the electronic inputs the rule has in mind. Where arm components are meant to be used as a system, the rule requires that the system include all degrees of freedom of the total upper extremity prosthesis.
Special controls that make Class II possible
Class II (special controls) means a manufacturer must meet the conditions the regulation lists in addition to FDA’s general controls. For this device type the regulation requires:
- Analysis or testing that validates electronic compatibility, electrical safety, thermal safety, mechanical safety, battery performance and safety, and wireless performance where applicable.
- Software verification, validation and hazard analysis.
- Non-clinical performance data: mechanical bench data including durability testing, simulated-use testing of arm commands and safeguards under worst-case conditions, verification and validation of force sensors and the hand release button where applicable, and checks on flame-retardant materials, liquid and particle ingress, sensor and actuator performance, and motor and brake performance.
- Non-clinical and clinical performance testing demonstrating the accuracy of device features and safeguards.
- Biocompatibility of elements that may contact the patient.
- Documented clinical experience and human factors testing demonstrating safe and effective use and capturing adverse events observed during clinical use.
- Labeling for the prosthetist and a user guide.
The labeling requirement is unusually detailed. The prosthetist and user guide must include warnings, cautions and limitations, including a warning against activities that may put the user at greater risk, with driving given as an example. It must also cover the clinical training needed for safe use: instructions on assembling the device in all available configurations, fitting the patient, available programs and how to program the device, all controls, inputs and outputs, all modes or states, all safety features, and maintenance. It must state the patient population for which the device has been demonstrated effective and include a detailed summary of the non-clinical and clinical testing. These labeling duties are written for the prosthetist as well as the user. Reporting of device-related adverse events follows the general rules summarised under medical device reporting (MDR).
What the clinical evidence says
Most peer-reviewed evidence on the arm comes from a VA-linked research group at the Providence VA Medical Center and collaborating sites. Four PubMed-indexed papers, summarised in the data table below, show what the studies did and what they found. All findings come from the abstracts and are reported as the authors state them.
Adaptation after home use
Resnik and colleagues followed 22 participants (18 completers) in a quasi-experimental time series after prosthetic training. Quality of life scores did not change between the end of laboratory training and the end of home use, while scores improved for one activity measure, two measures of self-reported function and three dexterity measures (p below 0.05). For participants who withdrew early, dexterity and activity scores worsened, perceived disability increased and prosthesis satisfaction decreased after four weeks of home use. The authors conclude that completers adapted to the arm, and suggest a need for continued therapy after intensive prosthetic training ends.
Control methods: inertial sensors compared with EMG pattern recognition
The arm has historically been operated mainly through inertial measurement units (IMUs). A multi-site study of 36 people with transradial or transhumeral amputation compared IMU control with two prototype electromyography pattern recognition (EMG-PR) controls. For the transradial group, the authors concluded that IMUs were the more effective control method than the EMG-PR prototypes in that study, and said further research is needed to refine EMG-PR for multi-degree-of-freedom devices. For the transhumeral group, no differences were observed by control type. A companion usability paper found that EMG-PR users, 11 of the 36, rated skill, comfort and weight worse than the 25 IMU users, and that usability and satisfaction declined after home use for EMG-PR users.
This is useful context for the FDA product code. The code title refers to control via cutaneous electromyography, while the published trials compared that control approach with IMUs on the same arm; the arm itself is a platform that can be operated by more than one input type.
Who adopts it, and who withdraws
A study in the American Journal of Physical Medicine and Rehabilitation examined cognition, adoption and withdrawal. Participants withdrawn by study staff performed significantly worse than self-withdrawn participants on tests of processing speed, set-shifting and memory encoding. The arm configuration (transradial, transhumeral or shoulder-based) was a stronger predictor of skill retention than neuropsychological test performance. The authors suggest this may be because non-cognitive user demands such as device weight statistically mask the influence of cognition at higher configuration levels, and that neuropsychological assessment warrants consideration in device candidacy evaluations.
Where it matters for labs, clinics and supply decisions
For clinics and rehabilitation research units, the regulatory and trial record above shows what evidence supports the device type: a De Novo grant, a regulation that lists its required testing and labeling, and a small number of VA-site studies with 22 to 36 participants. Those are the facts available in the sources; they do not show how the arm compares with other prosthetic hands or arms. Readers who want to place it among other upper-limb devices can start with CASRAI’s guide to the prosthetic hand.
For research staff reading the trial literature, two design details are worth noting. The studies are quasi-experimental and non-randomized, with small samples, and the control-type comparison was between non-randomized groups. Withdrawal was substantial: 4 of 22 participants did not complete one study, and the cognitive study is entirely about who stays and who leaves. Outcomes in these papers should therefore be read as results for people who completed training, as the authors themselves frame them.
Common misunderstandings
- LUKE is not the legal device name. The FDA record says DEKA ARM SYSTEM; LUKE is the name used by the VA and DARPA, spelled out by the VA as Life Under Kinetic Evolution.
- FDA grant does not mean general consumer availability. The regulation classifies the arm as a prescription device with labeling directed at a prosthetist and a trained user.
- The 2014 date is a decision date, not a launch date. The FDA decision was 9 May 2014; DARPA records that Mobius Bionics was launched in July 2016 and the VA unveiled the arm on 30 June 2017.
- Ten degrees of freedom is a maximum. The VA says up to 10 powered degrees of freedom. The regulation only requires more than two simultaneous powered degrees of freedom and a powered elbow and/or shoulder.
- Class II is not the same as a clinical-benefit claim. Classification states the controls required to market the device type; it does not rank the arm against other prostheses.
Related concepts
The FDA is the regulator whose records are quoted above; see FDA for how the agency fits into research-support work. The device classes and special-controls vocabulary on this page come from the framework created by the Medical Device Amendments of 1976.
Sources
- openFDA device 510(k) and De Novo record DEN120016, DEKA ARM SYSTEM, read 2026-11-17.
- openFDA device classification, regulation number 890.3450, product code PAE, read 2026-11-17.
- 21 CFR 890.3450, Upper extremity prosthesis (Legal Information Institute), read 2026-11-17.
- VA news release: VA Secretary Shulkin unveils world’s most advanced commercial prosthesis, 30 June 2017, read 2026-11-17.
- DARPA: LUKE arm, Revolutionizing Prosthetics, read 2026-11-17.
- Resnik L et al. Prosthet Orthot Int 2018;42(6):571-582 (PMID 29779455), read 2026-11-17.
- Resnik LJ et al. PLoS One 2018;13(10):e0204854 (PMID 30335781), read 2026-11-17.
- Resnik L et al. IEEE J Transl Eng Health Med 2019;7:2100113 (PMID 30680253), read 2026-11-17.
- Am J Phys Med Rehabil 2019, cognition, adoption and withdrawal (PMID 31135465), read 2026-11-17.
Data: Peer-reviewed studies of the DEKA Arm (LUKE arm): design, participants and control method
Four PubMed-indexed studies of the DEKA Arm compiled from their abstracts: publication year, study design, participant numbers and the question each study addressed, with the PubMed identifier for each record.
Period covered: 2018/2019. Coverage: United States.
| Study | PMID | Year (year) | Design | Participants | Question addressed |
|---|---|---|---|---|---|
| Resnik et al., Prosthet Orthot Int | 29779455 | 2018 | Quasi-experimental time series | 22 (18 completers) | Change in function, quality of life and community integration after training and home use |
| Resnik et al., PLoS One | 30335781 | 2018 | Quasi-experimental, multi-site, repeated measures, non-randomized groups | 36 (transradial and transhumeral) | EMG pattern recognition versus IMU control compared with personal prosthesis |
| Resnik et al., IEEE J Transl Eng Health Med | 30680253 | 2019 | Usability and satisfaction comparison by control type | 36 (11 EMG-PR, 25 IMU) | User ratings of usability and satisfaction by control method and amputation level |
| Am J Phys Med Rehabil | 31135465 | 2019 | T tests, Wilcoxon rank-sum tests and multivariable regression | Not stated in the abstract | Role of cognition in adoption, skill retention and study withdrawal |
Sources
- PubMed 29779455 (accessed 2026-11-17)
- PubMed 30335781 (accessed 2026-11-17)
- PubMed 30680253 (accessed 2026-11-17)
- PubMed 31135465 (accessed 2026-11-17)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.