Luminometer: How It Works, Types, and Read Modes

Luminometer explained: PMT detection, injectors, tube vs microplate vs imaging types, and flash, glow and photon-counting read modes, from named sources.

Written and maintained by CASRAI Editorial Board

Published

On this page

A luminometer is an instrument that measures the light emitted by a sample. Because the light comes from a chemical or biological reaction, the instrument needs no excitation lamp: a photomultiplier tube (PMT) in a light-tight chamber counts the photons, and the signal is read as a measure of how much of the reacting substance (such as ATP or a luciferase reporter) is present. Tube, microplate and camera-based imaging formats trade sensitivity against throughput.

Definition and scope

Berthold’s glossary defines a luminometer as an instrument that measures emissions of light emitted by a sample. The term covers dedicated luminescence readers and, in practice, the luminescence mode of multimode instruments. It is distinct from a fluorometer: fluorescence needs an excitation light source and emission filters, while luminescence does not, because the sample produces its own light. For the broader instrument family, see what a plate reader is and multimode versus single-mode plate readers.

Luminescence is attractive for sensitive assays. Berthold states that luminescence is typically 10 to 1,000 times more sensitive than fluorescence and 100 to 10,000 times more sensitive than absorbance, and that luminescence signals are typically linear over 6 to 8 logs. Danaher Life Sciences similarly notes that luminescence detection is often in the femtomole or even attomole range.

How a luminometer works

Photomultiplier tube and dark chamber

The core detector is normally a PMT. Photons striking the photocathode release electrons, which are multiplied along a chain of dynodes into a measurable pulse or current. The sample sits in a dark chamber so that stray light does not reach the detector. Promega’s GloMax 20/20 product page, for example, describes a low-noise PMT placed directly below the sample well.

Injectors

Some luminometers include reagent injectors. They matter for fast reactions: Berthold notes that flash luminescence, which decays in about 1 to 2 seconds, often needs an injector so that the reaction starts inside the chamber while the detector is already recording. The GloMax 20/20 page lists an injector system among its features.

Filters

Filters and monochromators are usually unnecessary because all emitted wavelengths are pooled into one signal. Berthold names bioluminescence resonance energy transfer (BRET) as the exception, where two emission wavelengths must be separated.

Types of luminometer

Berthold distinguishes tube luminometers, microplate luminometers and in vivo imaging systems. The data table below compares them using the glossary’s descriptions.

Tube luminometers

These read one sample at a time and are loaded by hand. They are available as benchtop, portable and hand-held units, and benchtop models are the most sensitive. The Promega GloMax 20/20 accepted 1.5 ml microcentrifuge tubes and 35 mm dishes, with 12 x 50 mm tubes as an option, and Promega listed read types of glow, flash, kinetic and repeat. Promega’s page states that the instrument was retired on 1 September 2023, with accessories and service available to 1 September 2028; it is cited here as an example of the format, not as a current product.

Microplate luminometers

These read 96-well and 384-well plates, giving higher throughput. Berthold describes them as more complex and expensive than tube units and generally less sensitive than benchtop tube instruments. Danaher adds a plate-level point: white, opaque plates reflect light, which increases the measured signal, and reduce crosstalk between wells.

Imaging systems

In vivo imaging systems use a camera rather than a PMT. They are less sensitive, but they can localise the emission within living animals or plants, which a tube or plate reader cannot do.

Read modes: flash, glow and photon counting

The read mode should match the kinetics of the assay chemistry. The table below combines Berthold’s and Danaher’s descriptions.

Mode or setting What the source says Implication for the instrument
Flash Fast kinetics, about 1 to 2 seconds (Berthold); a brief intense burst needing careful timing, suited to rapid high-throughput work (Danaher) Injectors are often needed (Berthold)
Glow A prolonged, stable signal suitable for longitudinal quantification (Danaher) Timing is less critical because the signal is stable
Photon counting Counts individual pulses; gives the best sensitivity and is best for luminescence (Berthold) Preferred for weak signals
Analog (current) mode Measures current; suited to high intensities and more common in fluorescence (Berthold) Used when the signal would be too intense to count individually

Assays that use a luminometer

Berthold lists reporter-gene assays, cell viability, proliferation and cytotoxicity (ATP-based) assays, hygiene monitoring, immunoassays, caspase and kinase assays, reactive oxygen species, SPARCL, BRET and water toxicity testing. Danaher groups luminescence assays into bioluminescence (luciferase, ATP, reporter), chemiluminescence and BRET. CASRAI guides on individual methods show what each readout assumes: the ATP assay and CellTiter-Glo guides cover ATP-dependent luminescent viability readouts, the luciferase reporter assay guide covers reporter expression, and luciferin explains the substrate behind many bioluminescent signals.

Settings and workflow factors that affect the reading

Danaher describes microplate readers and luminometers as designed to detect low-intensity light from samples in multiwell plates, and states that settings such as integration time, sensitivity and well geometry can significantly affect signal detection. Increasing integration time in the detector can help resolve weak signals. Detection limit is the minimum amount of analyte producing a signal distinguishable from background noise, and dynamic range is the concentration range over which the signal is linear and distinguishable from background.

The same source describes a typical luminescence protocol as three steps: reagent addition, incubation and light-emission measurement. In add-and-read formats no washing or separation steps are needed. Luminescent reagents such as luciferin and enzymes are added to the wells; reagents should be freshly prepared and equilibrated to room temperature, substrates are often stored at -20 C and protected from light, and stabilizing reagents can raise signal duration and intensity. Cross-well contamination is a listed risk, mitigated by clean tips, dedicated reservoirs and avoiding carry-over between wells. Emission time ranges from microseconds to hours depending on the assay type.

Matching an assay to a format (editorial decision aid)

The following reading of the sources above is an interpretation, not a rule from any one manufacturer.

  • Few samples and a need for the highest sensitivity: Berthold identifies benchtop tube luminometers as the most sensitive.
  • Many samples in 96- or 384-well plates: a microplate luminometer, accepting somewhat lower sensitivity (Berthold), with white opaque plates to raise signal and reduce crosstalk (Danaher).
  • Fast flash chemistry: choose an instrument with injectors (Berthold), and keep read timing consistent (Danaher).
  • Stable glow chemistry measured over time: no injector is required by the sources; longitudinal quantification suits the stable signal (Danaher).
  • Location of signal in a living animal or plant: an imaging system with a camera (Berthold).

Instrument facts to record in a core facility or lab

This checklist is drawn only from the specifications the sources above treat as defining, and is intended as a record-keeping aid rather than a purchasing rule.

  • Detector: PMT or camera (Berthold).
  • Format: tube, 96-well or 384-well plate, or imaging stage (Berthold); for tube units, the tube and dish sizes accepted (Promega GloMax 20/20 page).
  • Injectors: present or absent, and whether the intended assay is a flash assay (Berthold).
  • Read modes offered: glow, flash, kinetic, repeat (Promega); photon counting or analog (Berthold).
  • Linear dynamic range: as stated by the manufacturer for that model; Promega stated over 8 logs for the GloMax 20/20.
  • Plate type: white opaque plates for signal and crosstalk (Danaher).

Common misunderstandings

  • A luminometer is not a fluorometer. Luminescence needs no excitation source because the reaction emits the light (Berthold, Danaher).
  • Filters are not normally needed. The exception is BRET (Berthold).
  • Higher throughput is not higher sensitivity. Microplate units are generally less sensitive than benchtop tube units, and camera systems are less sensitive again (Berthold).
  • Flash and glow need different timing. Danaher notes that flash signals decay rapidly, so consistent read timing matters most for them, while glow gives consistent, long-lasting signals.

FAQ

What does a luminometer measure?

It measures light emitted by a sample from a chemical or biological reaction, expressed as photon counts or current from a PMT. No excitation light is used.

What is the difference between a luminometer and a plate reader?

A luminometer is dedicated to light emission. A plate reader reads multiwell plates and may include luminescence alongside absorbance and fluorescence; see the plate reader guide.

Do I need an injector?

For flash assays with signals lasting a second or two, Berthold says injectors are often needed. For stable glow signals, the Danaher description indicates measurement can be made over a longer window.

Sources

  • Berthold Technologies, Luminometer glossary entry, berthold.com, read 2026-11-12.
  • Promega, GloMax 20/20 Luminometer product page, promega.com, read 2026-11-12.
  • Danaher Life Sciences, Luminescence assays, lifesciences.danaher.com, read 2026-11-12.

Data: Luminometer formats compared: detector, sample handling, throughput and relative sensitivity

Compiled from the Berthold glossary description of tube luminometers, microplate luminometers and in vivo imaging systems; wording is condensed and no values are estimated.

Luminometer formats compared: detector, sample handling, throughput and relative sensitivity
FormatDetectorSample handlingThroughputRelative sensitivity
Tube luminometer (benchtop, portable, hand-held)PMTOne sample at a time, manual loadingSingle sampleBenchtop units most sensitive
Microplate luminometerPMT96-well or 384-well platesHigher throughputGenerally less sensitive than benchtop tube units
In vivo imaging systemCameraLiving animals or plantsImaging of whole subjectLess sensitive; can localise emission

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Luminometer: How It Works, Types, and Read Modes

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →