Magnetic Materials: Types, Specs and Lab Testing

A practitioner guide to magnetic materials: IEC 60404-1 classification, soft vs hard specs, VSM calibration via NIST SRM 772a, and lab safety protocols.

Written and maintained by CASRAI Editorial Board

Last updated

Short answer

Magnetic materials exhibit macroscopic magnetization under an applied magnetic field. Governed by IEC 60404-1 and ASTM A340, they divide into soft magnetic materials (coercivity Hc ≤ 1,000 A/m, like electrical steel and Permalloy) and hard permanent materials (Hc > 1,000 A/m, like NdFeB and SmCo) for research instrumentation, shielding, and beamline assemblies.

Definition and scope

In materials physics, magnetic materials are defined by their response to an applied magnetic field strength (H). Governed under ASTM A340 by ASTM International Committee A06, the resulting magnetic flux density (B) within a medium follows the constitutive relation:

B = μ₀(H + M) = μH

where μ₀ is the permeability of free space (4π × 10⁻⁷ H/m), M is magnetization, μ is magnetic permeability, and relative permeability is μᵣ = μ / μ₀. Volumetric magnetic susceptibility (χ = M / H) dictates the magnitude and sign of this response.

In materials science research, magnetic materials span solid-state physics, spintronics, and electron microscopy shielding. Fundamental research into novel magnetic compositions receives federal funding through programs like the NSF Division of Materials Research.

How it works or how it is used in practice

Macroscopic magnetism arises from atomic electron spin and orbital angular momentum, yielding dipole moments measured in Bohr magnetons (μB ≈ 9.274 × 10⁻²⁴ J/T). In condensed matter, quantum exchange coupling aligns spins parallel below the Curie temperature (T_C), creating spontaneous magnetization.

To minimize magnetostatic energy, bulk ferromagnetic materials partition into magnetic domains (Weiss domains). Each domain is magnetized to saturation (M_s), but domains point in different crystallographic directions separated by domain walls, leaving the bulk material demagnetized.

Applying an external field H drives technical magnetization through three sequential mechanisms:

  • Reversible Wall Displacement: Low fields flex domain walls elastically without hysteresis loss.
  • Irreversible Depinning: Intermediate fields force walls past grain boundaries and precipitates, producing discrete Barkhausen jumps.
  • Domain Rotation: High fields rotate moments into alignment with H against magnetocrystalline anisotropy.

Plotting B versus H traces a hysteresis loop. Key ASTM A340 parameters include saturation induction (B_s), remanence (B_r at H = 0), and normal coercivity (H_c, reverse field to reduce B to zero). Intrinsic coercivity (H_cJ) measures the reverse field to drive M to zero. Loop area defines hysteresis core loss (W_h = ∮ H dB).

Types or related concepts

IEC 60404-1 and magnetic physics categorize materials by spin ordering and coercivity thresholds.

Class Spin Ordering Susceptibility (χ) Coercivity (H_c) Bound Representative Materials Key Research & Lab Uses
Diamagnetic Paired electrons; orbital precession opposes H Negative (χ ≈ -10⁻⁵ to -10⁻⁹) None (zero remanence) Cu, Bi, Au, pyrolytic graphite Sample mounts, levitation balances
Paramagnetic Unpaired spins; randomized by thermal agitation Small, positive (χ ≈ 10⁻³ to 10⁻⁵) None (zero remanence) Al, Pt, Ti, liquid O₂ EPR markers, cryogenic cooling salts
Ferromagnetic Parallel exchange coupling below T_C Large, positive (χ >> 1) Variable by alloy structure Fe, Co, Ni, Permendur Electromagnet poles, flux yokes
Ferrimagnetic Antiparallel sublattices with unequal moments Moderate to high, positive Low to moderate coercivity Mn-Zn ferrites, Ni-Zn ferrites, YIG High-frequency RF inductors, isolators
Antiferromagnetic Antiparallel sublattices with equal opposing moments Small, positive; peaks at T_N Zero net remanence Cr, MnO, NiO, FeMn Exchange bias pinning in MTJ devices
Soft Magnetic High domain wall mobility; low pinning High permeability (μᵣ up to > 100,000) H_c ≤ 1,000 A/m (IEC 60404-1) Silicon steel, Permalloy, Mu-metal Transformer laminations, magnetic shielding
Hard Magnetic High magnetocrystalline anisotropy Low permeability (μᵣ ≈ 1.05 to 1.2) H_c > 1,000 A/m (IEC 60404-1) NdFeB, SmCo, Alnico, Sr-ferrite Permanent assemblies, magnetic stir bars, ion pumps

Where it matters for labs, clinics or supply decisions

Working with magnetic materials requires attention to shielding performance, metrological calibration, and occupational safety.

Electromagnetic Shielding in Analytical Facilities

Facilities housing electron microscopes (TEM/SEM) and SQUID magnetometers require attenuation of ambient magnetic fields. Mu-metal (an 80% nickel-iron alloy) provides relative permeability exceeding 50,000 to 100,000 at low fields, routing flux lines through shield walls around the instrument chamber. Because mechanical shock or machining pins domain walls and drops permeability by up to 90%, Mu-metal components must receive a final dry-hydrogen anneal at 1100 °C to 1150 °C.

Magnetometer Calibration and Reference Standards

Under ISO/IEC 17025 calibration frameworks, laboratory magnetometers require traceable physical standards. NIST maintains Standard Reference Material SRM 772a (Magnetic Moment Standard – Nickel Sphere). SRM 772a is an annealed 99.999% pure nickel sphere (diameter 2.383 mm, mass 63.16 mg) certified to a magnetic moment of 3.47 mA·m² ± 0.01 mA·m² (3.47 emu) at 298 K in an applied field of 398 kA/m (5,000 Oe). Researchers must handle SRM 772a exclusively with non-magnetic PTFE or ceramic tweezers to prevent scratching and iron contamination.

High-Field Safety and the 5 Gauss Perimeter

Superconducting NMR and MRI facilities establish a 5 Gauss (0.5 mT) safety perimeter per ACGIH and OSHA guidelines. Within this perimeter:

  • Individuals with cardiac pacemakers or ICDs are strictly excluded due to magnetic switch malfunction risks.
  • Ferromagnetic tools and cylinders are prohibited to prevent violent projectile attraction (F_mag ∝ B ∇B).
  • Cryostat sample mounts must use certified non-magnetic materials (such as brass or titanium grade 5) rather than standard stainless steels.

Common misunderstandings

  • Misunderstanding: All austenitic stainless steels are non-magnetic. Annealed 304 and 316 steels are nominally paramagnetic, but cold drawing or machining triggers transformation into strain-induced martensite. This ferromagnetic phase produces background signal distortion in sensitive magnetometer sample rods.
  • Misunderstanding: The Curie temperature is a magnet’s maximum operating temperature. The Curie temperature (T_C) is where ferromagnetic order vanishes (~312 °C for NdFeB). However, maximum operating temperature is dictated by the temperature coefficient of coercivity, restricting standard NdFeB to 80 °C before irreversible demagnetization occurs.
  • Misunderstanding: Shielding materials absorb magnetic fields. Static magnetic shielding does not absorb flux lines. High-permeability alloys provide a path of low magnetic reluctance that diverts flux lines around the shielded cavity.
  • Misunderstanding: Remanence and coercivity measure the same property. Remanence (B_r) reflects retained flux density at zero field, whereas coercivity (H_c) quantifies resistance to reverse demagnetizing fields. High remanence alone does not ensure a stable permanent magnet without sufficient coercivity.

Frequently asked questions

What is the quantitative boundary between soft and hard magnetic materials?

Under IEC 60404-1, materials with coercivity Hc ≤ 1,000 A/m (~12.5 Oe) are classified as soft magnetic materials (low core loss, high permeability). Materials with Hc > 1,000 A/m are classified as hard or permanent magnetic materials (high demagnetization resistance).

How is NIST SRM 772a used in laboratory calibration?

NIST SRM 772a is a certified 99.999% pure nickel sphere (mass 63.16 mg) delivering a magnetic moment of 3.47 mA·m² at 298 K in a 398 kA/m field. It is mounted in vibrating sample magnetometers (VSM) and SQUID systems to calibrate coil sensitivity and ensure traceable moment measurements.

Why are ceramic or PTFE tweezers mandatory for magnetic reference standards?

Non-metallic, non-magnetic tweezers prevent scratching, mechanical strain, and ferromagnetic iron transfer. Ferromagnetic contamination or plastic deformation alters the sample’s mass and demagnetizing factor, invalidating the calibration certificate.

What does the 5 Gauss boundary signify in research facilities?

The 5 Gauss (0.5 mT) line is the safety threshold defined by ACGIH and OSHA around high-field superconducting magnets. Inside this boundary, active cardiac implants risk malfunction and loose ferromagnetic tools become dangerous projectile hazards.

How does mechanical stress affect Mu-metal shielding?

Mu-metal relies on an unconstrained crystal lattice for high permeability (μᵣ > 50,000). Bending or cutting creates dislocation tangles that pin domain walls, reducing shielding effectiveness by over 80% until re-annealed in dry hydrogen at 1100 °C.

Sources

  • IEC — IEC 60404-1:2016, Magnetic materials — Part 1: Classification. https://webstore.iec.ch/publication/24434. Checked 2026-11-08.
  • ASTM International — ASTM A340-23, Standard Terminology of Symbols and Definitions Relating to Magnetic Testing. https://www.astm.org/a0340-23.html. Checked 2026-11-08.
  • NIST — Standard Reference Material 772a: Magnetic Moment Standard – Nickel Sphere. https://www.nist.gov/srm. Checked 2026-11-08.
  • NIST Materials Measurement Laboratory — Magnetic Materials Characterization and Metrology Protocols. https://www.nist.gov/mml/materials-science-and-engineering-division/magnetic-materials. Checked 2026-11-08.
  • ACGIH — Threshold Limit Values for Static Magnetic Fields. Checked 2026-11-08.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Magnetic Materials: Types, Specs and Lab Testing

Ask your first 2 questions free below. Subscribers get 150 a day for $29 a month.

An AI assistant specialized in research administration. It cites the sources behind every answer, labels web answers and says when it can't answer.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays free to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →