Nutator: What It Is and How a Nutating Mixer Works

A nutator tilts and rotates a tray for gentle 3D mixing. Compare speed, tilt and load limits from three manuals, plus loading and cleaning rules.

Written and maintained by CASRAI Editorial Board

Published

On this page

A nutator is a laboratory mixer whose platform tilts and rotates at the same time, producing a gentle three-dimensional rocking and swirling motion that mixes liquid samples in tubes, bottles or flasks without vigorous shaking. Manufacturers also call it a nutating mixer or a 3D platform rotator. Typical bench models turn at a fixed 24 rpm or at a user-set speed between 2 and 80 rpm, and the Benchmark manual describes the motion as suited to tube mixing with no or low foam.

This page describes laboratory equipment and general use as stated in manufacturer instruction manuals. It sets out what a nutator does, how the three manuals we compared differ, the loading and speed rules they share, and the maintenance and troubleshooting points they list.

Short answer: what is a nutator?

A nutator is a benchtop instrument with a flat tray that is tilted from the horizontal and carried round through a rotation, so the tray rocks in two directions in turn. The Conoptics description puts it as a laboratory instrument that creates a gentle three-dimensional rocking motion used to mix biological samples and other materials, combining rocking and swirling movements. The Benchmark Scientific manual describes its nutating mixers as having the angle and speed set for gentle mixing of tubes with no or low foam. The manuals position it for gentle mixing rather than vigorous agitation.

Definition and scope

In the manuals read for this page, the same instrument appears under three names. Fisher Scientific titles its document a nutating mixer and 3D platform rotator manual, VWR calls its product a nutating mixer, and Benchmark Scientific sells a mini 3D (nutating) rocker. In everyday lab speech all three are simply a nutator. The core of the definition is the combination of a tilt angle and a rotation angle: the Fisher manual lists a tilt angle with ±1° accuracy and a rotation angle from 0 to 360°, while Benchmark gives a tilt of ±20° from the horizontal.

What is a nutating mixer?

A nutating mixer is the same instrument as a nutator. The word nutating describes the motion of the tray, and mixer describes the job. Instruction manuals use the two terms interchangeably, and the VWR manual covers fixed-speed and digital variable-speed versions under the one title.

What is a rocking platform, and is it the same thing?

Rocking platform is a related search term, and it is used loosely. Several suppliers label nutating models as rockers, as in the Benchmark mini 3D rocker, so a shopper or a protocol that says rocking platform may mean a nutator. A plain rocker that tilts in a single direction is a different design from the nutating motion the manuals describe, in which tilt and rotation combine. When a protocol names the motion, match the motion rather than the product label. For platforms that move in other ways, see our guides to the orbital shaker, the vortex mixer and the incubator shaker.

How a nutator works in practice

The three manuals describe the same working pattern. Containers are placed on a tray, usually on a dimpled rubber mat, and held with rubber strips. The motor tilts the tray and rotates it, so liquid in each container moves back and forth along the container and swirls round it. The two settings that shape the motion are speed and tilt.

Speed

The Fisher Scientific manual lists a fixed-speed model at 24 rpm (20 rpm on 230 V models) and variable-speed models covering 2 to 60 rpm and 2 to 80 rpm, with speed accuracy of ±1 rpm and a display accuracy of 1 rpm. It adds a timer from 1 minute to 99 hours 59 minutes. The Benchmark mini 3D rocker is a fixed-speed unit at 24 rpm (20 rpm on 230 V) with a gear motor that has no brushes.

Tilt angle

The VWR manual covers a fixed-speed 3D nutating mixer (76595-806) and two digital models named 20 Digital (76595-808) and 7 Digital (76595-810), so tilt is one of the main differences between models: a 20° tilt gives a bigger swing than a 7° tilt. The Benchmark unit tilts ±20° from the horizontal. The Fisher manual states a tilt accuracy of ±1°.

Speed and load interact

The Fisher manual states that the rotation speed is inversely proportional to the load, and it warns that maximum speed and tilt angle may vary with heavy or unbalanced loads. It advises raising the speed from low to high, step by step, to avoid spillage. In practice this means the headline rpm in a specification is a no-load or light-load figure, and a full tray will be run slower.

Types and how the three manuals compare

The data table below compares the models covered by the three manuals. The differences that matter most when staff are deciding which instrument a protocol needs are whether the speed is fixed or adjustable, the tilt angle, and the stated limits on load and environment.

Two patterns stand out. First, the Fisher and VWR documents describe full-size tray units with variable-speed digital versions and a timer, whereas the Benchmark mini rocker is a fixed-speed bench unit with a 15.25 × 20.25 cm (6 × 8 in) platform and a stated load capacity of 1.6 kg (3.5 lb), centred and balanced. Second, the Benchmark unit is rated for an operating environment of 4 to 65 °C and is designed to sit on the floor of a mini incubator, centred and not on shelves, while the Fisher manual specifies 5 to 40 °C, 20 to 85 % humidity and an altitude of up to 2,000 m for indoor use.

Where it matters for labs and supply decisions

Loading checklist from the manuals

  • Place the load at the centre of the tray and distribute it symmetrically around the centre (Fisher). Benchmark says to load the platform from the centre out and warns that overhung loads may cause the rocker to stall or tip.
  • Fasten containers with the rubber strips. The Fisher manual says to make sure to use rubber strips to fasten containers onto the tray.
  • Do not load asymmetrically. Fisher states that asymmetric loading or unloading may affect normal operation and may even cause unpredictable damage.
  • Start slow and increase speed in steps, because speed and load are inversely related.
  • Keep the instrument out of hazardous atmospheres and do not use it with hazardous materials it was not designed for.

Cleaning, spills and maintenance

Both Fisher and Benchmark describe the instrument as needing no routine maintenance beyond keeping it clean. Fisher describes the motor as maintenance free and says to clean surfaces with mild detergent and water and never immerse the unit. Benchmark says to use a damp cloth and to avoid solvents, which can attack the rubber mat and ABS housing. The fixed-speed Fisher model has two fuses, one working and one spare. Disconnect the power before any maintenance.

For spills, the Fisher manual asks for biohazard spills to be cleaned promptly with an approved liquid. A solvent spill is a fire hazard: stop the unit and do not run it again until clean-up is complete and the vapours have dissipated. If liquid gets inside the unit, cut the power, clean the surface, leave the unit in a ventilated dry place for 24 hours before reuse, and contact the manufacturer if it still does not work.

Troubleshooting points

The Fisher troubleshooting table lists a few causes that cover most stoppages. If the tray does not move, check the power connection, the switch and the adaptor, and also check for an overweight or unbalanced load, in which case adjust the weight and position and reduce the speed. If the load moves on the tray, change the dimpled rubber mat or fasten the load more securely. If the tray is loose, tighten the screws. If operation is obstructed, the instrument stops within 2 minutes and sounds an alarm. After a sudden power loss, the unit restarts at its previous parameters, which is worth knowing before leaving a long run unattended.

Shared instruments and equipment records

Where several groups share one nutator, the loading and clean-up rules above are the points a local sign-off sheet should cover, alongside the general approach in our guide to core facility equipment sharing agreements and the overview in biology lab equipment.

Common misunderstandings

A nutator is not a shaker

The manuals position the nutating mixer around gentle, low-foam mixing. Shakers and vortex mixers are chosen for more vigorous agitation, so swapping one for the other changes how the sample is treated.

The rated speed is not the working speed

Rated speeds assume light or balanced loads. Fisher states that maximum speed and tilt angle may vary with heavy or unbalanced loads, and Benchmark gives a capacity limit of 1.6 kg for its mini model. Check both the rpm range and the capacity before assuming a tray of flasks will run at the top of the range.

Higher tilt is not always better

A 20° tilt gives a larger swing than a 7° tilt, but the manuals do not rank one above the other; the right setting depends on the container and the protocol. The manuals do not claim that either tilt suits a given sample type, so this page does not either.

Maintenance free does not mean cleaning free

The motor needs no routine service, but the manuals still require cleaning after spills, avoidance of solvents on the housing, and checks of the mat and straps.

Frequently asked questions

What is a nutator?

A nutator is a laboratory mixer with a tray that tilts and rotates together, giving a gentle three-dimensional rocking and swirling motion for mixing samples. The Conoptics page defines it that way, and the Fisher, VWR and Benchmark manuals describe instruments that work on that principle under the names nutating mixer and 3D platform rotator or rocker.

What is a rocking platform?

It is a loose term for a tray that moves back and forth to mix or agitate samples. Suppliers sometimes use rocker for nutating models, as in the Benchmark mini 3D rocker, so check the motion described rather than relying on the name alone.

Sources

Data: Nutating mixer models: speed, tilt, load and operating environment as stated in three manufacturer manuals

Compiled from the Fisher Scientific nutating mixer manual, the VWR nutating mixer instruction manual (2022) and the Benchmark Scientific H3D1020 mini 3D rocker manual. Each row is one model or model family; a dash means the manual read does not state the value unambiguously for that model.

Nutating mixer models: speed, tilt, load and operating environment as stated in three manufacturer manuals
Model or family (manual)Speed controlSpeed (rpm)Tilt angle (degrees)Stated load or loading ruleOperating temperature (°C)
Fisher Scientific nutating mixer, fixed speedFixed24 (20 on 230 V models)Not stated unambiguouslyCentre and balance the load; fasten with rubber strips5 to 40
Fisher Scientific nutating mixer, variable speedVariable, digital timer 1 min to 99 h 59 min2 to 60 or 2 to 80 depending on modelTilt accuracy ±1Speed is inversely proportional to load; ramp speed up in steps5 to 40
VWR 76595-806 fixed-speed 3D nutating mixerFixed24Not stated unambiguouslySame loading rules as the Fisher manual5 to 40
VWR 76595-808 20 Digital nutating mixerVariable, digital2 to 60 or 2 to 80 per manual20Heavy or unbalanced loads may reduce maximum speed and tilt5 to 40
VWR 76595-810 7 Digital nutating mixerVariable, digital2 to 60 or 2 to 80 per manual7Heavy or unbalanced loads may reduce maximum speed and tilt5 to 40
Benchmark Scientific H3D1020 Mini BioMixer 3DFixed24 (20 on 230 V models)±20 from horizontal1.6 kg (3.5 lb), centred and balanced; platform 15.25 x 20.25 cm4 to 65

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Nutator: What It Is and How a Nutating Mixer Works

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →