PacBio Revio: Specifications, SPRQ-Nx Chemistry, Output and Facility Needs

PacBio Revio explained: four parallel SMRT Cells, SPRQ-Nx chemistry, yield by library length, BAM output and facility requirements, from PacBio’s documents.

Written and maintained by CASRAI Editorial Board

Published

On this page

The PacBio Revio is a long-read DNA sequencing instrument from Pacific Biosciences that runs up to four Revio SMRT Cells at the same time on four independent stages and produces HiFi reads, with base modifications called on the instrument in every run. According to PacBio, it delivers up to 480 Gb of HiFi data per day with SPRQ-Nx chemistry, and it is sold for research use only, not for diagnostic procedures.

This page collects what a researcher, core-facility manager or lab planner needs to know about the system: what the platform does, how library length determines run time and yield, how much data and floor space a site must plan for, and how Revio differs from the earlier Sequel II and IIe systems. Every figure below comes from PacBio’s own documents, read on 2026-11-14; where a number is a vendor claim it is described as one. For what long reads are and how HiFi reads are aligned, see the companion guide on PacBio DNA sequencing.

Short answer: what the Revio does

Revio is PacBio’s production-scale HiFi sequencer. It uses nanofabricated Revio SMRT Cells, four independent stages so that several cells can be sequenced in parallel, and onboard computing that performs basecalling with Google DeepConsensus and methylation calling during every run (Revio specification sheet, REV09, 2 June 2026). PacBio’s product page gives the headline figures: up to 4 SMRT Cells simultaneously, up to 480 Gb of HiFi data per day, a read accuracy of 99.95% (Q33) and a median read quality of at least Q30. Typical HiFi reads are 15-20 kb, and library-dependent lengths range from 0.5 to 25 kb.

Definition and scope

Revio is a single instrument model, but what it can do depends on the chemistry loaded. The specification sheet describes the system ‘optimized with SPRQ-Nx chemistry’. SPRQ is the current sequencing chemistry; SPRQ-Nx extends it so that one SMRT Cell can be used for multiple acquisitions, which PacBio’s product page describes as multi-use SMRT Cells. The consumables are listed by name in the specification sheet: the Revio SPRQ polymerase kit (reagents supporting 24 SMRT Cells), the Revio SMRT Cell tray (4 cells) and an Nx tray variant, and the Revio SPRQ sequencing plate (supporting 4 SMRT Cells) alongside an Nx plate that supports 4 acquisitions.

The scope of the platform is DNA and cDNA sequencing with long, accurate reads. PacBio lists whole-genome variant calling with phasing and methylation, assembly of complex genomes, targeted sequencing with PureTarget, hybrid capture or amplicon libraries, and single-cell and full-length transcriptome sequencing with Kinnex kits. The instrument is labelled for research use only.

How the Revio works in practice

Parallel stages and acquisitions

The Revio has four independent stages. Each stage holds one SMRT Cell, so up to four cells run in parallel, and the stages are independent, so different samples and different applications can be sequenced at the same time. PacBio’s launch announcement on 25 October 2022 described a Revio SMRT Cell with three-fold higher density than the earlier SMRT Cell 8M, giving 25 million ZMWs (zero-mode waveguides, the nanoscale wells in which single molecules are read) per cell, or up to 100 million across four cells. With SPRQ-Nx chemistry a single cell can be used for more than one acquisition, which is why the specification sheet counts samples per acquisition rather than per cell.

Library length decides run time and yield

The most useful relationship in the specification sheet is between library length, data-collection time and HiFi yield per SMRT Cell. Shorter libraries run for 12 hours and report yield in reads; most standard long-read libraries run for 24 hours; and the longest libraries, 20-25 kb, run for 30 hours. The data table below the article sets these values side by side, including the Q30+ base fraction PacBio specifies for each library range. Three footnotes from the sheet matter for planning. First, libraries of 0.5-5 kb are supported only with single-acquisition Revio SMRT Cells with SPRQ or SPRQ-Nx chemistry. Second, run time means the data-collection step, which determines the time between processing SMRT Cells. Third, the specified yield assumes high-quality samples prepared following best practices, so a poorly prepared library should not be planned at the table’s numbers.

The sheet also notes that adaptive loading, which increases the consistency of yield per SMRT Cell, adds about four hours to run time and therefore reduces the maximum number of SMRT Cells per year.

Onboard compute, basecalling and methylation

Revio performs the computation on the instrument. PacBio’s product page says all processing happens on the system, with GPUs producing HiFi reads in BAM format. The launch announcement described Revio as the first PacBio system with NVIDIA GPUs and a 20-fold increase in computing power over the Sequel IIe. Methylation is called in every standard run without bisulfite conversion: the specification sheet lists 5mC and 5hmC at CpG sites and 6mA for native DNA, and notes that the 6mA caller is designed for the Fiber-seq chromatin assay. PacBio’s technical overview for instrument control software v13.3 with SPRQ chemistry lists on-instrument 6mA calling among the changes in that release.

Outputs and what happens next

The instrument writes hifi_reads.bam files of about 60 GB each, and the technical overview lists fail_reads.bam alongside them. According to the same overview, on-instrument demultiplexing applies only to symmetrically indexed samples; samples with asymmetric indexes are demultiplexed off the instrument in SMRT Link. Because the native output is an unaligned BAM rather than FASTQ, downstream steps start with conversion or direct alignment. The guides on FASTQ format and SAM, BAM and CRAM explain the formats, and choosing a de novo assembler covers the assembly route that HiFi reads are often used for.

Applications and samples per run

The specification sheet requires 500 ng of DNA per sample and lists a subset of key applications with the number of samples per Revio acquisition, samples per year and expected coverage. The table below reproduces it. Annual throughput is calculated by PacBio on 2,500 SMRT Cells per year for 12-hour runs, 1,250 for 24-hour runs and 1,050 for 30-hour runs, and the sheet states that expected coverages are estimates.

Library Application Samples per acquisition Samples per year Expected coverage
0.5-5 kb Amplicon sequencing more than 1,000 more than 2.5 million 50x
5-10 kb Microbial genome 384 480,000 30x
5-10 kb PureTarget repeat expansion panel 48 60,000 200x
5-10 kb Ampli-Fi human genome 1 1,250 20x
15-20 kb Human genome 2 2,500 20x
15-20 kb Human methylation profiling 8 10,000 5x
15-20 kb Transcriptome, Kinnex full-length RNA kit 6 7,500 10 million reads

Source: PacBio Revio system + SPRQ-Nx chemistry specification sheet, REV09, 2 June 2026.

Revio compared with Sequel II and IIe

Revio is the successor to the Sequel II and Sequel IIe systems, and PacBio’s own documents quantify the difference in two places. The launch announcement says the Revio SMRT Cell has three-fold the density of the SMRT Cell 8M, uses 50 percent fewer consumables than the Sequel IIe, and provides a 20-fold increase in computing power; PacBio’s headline claim at launch was 15 times more HiFi data, which is a vendor claim and should be read as such. The technical overview adds a per-SMRT-Cell comparison for Kinnex applications across Sequel II/IIe, Revio and Revio with SPRQ chemistry.

Kinnex kit Sequel II/IIe Revio Revio SPRQ
Single-cell RNA 30-40M 80-100M 100-120M
Full-length RNA 15-20M 30-40M 50-60M
16S RNA 20-25M 50-60M 60-80M

Figures are reads per SMRT Cell, from the Revio technical overview (ICS v13.3 and SPRQ chemistry training), Kinnex application updates slide. The same slide shows how that extra throughput is meant to be used: PacBio lists 3-plex for the Kinnex full-length RNA kit on Sequel II/IIe, 4-plex on Revio and 6-plex on Revio SPRQ, and notes that with SPRQ chemistry Kinnex library preparation can typically provide enough library material to run four or more Revio SMRT Cells.

The same training deck states that v13.3 with SPRQ brings a four-fold reduction in DNA input, on-instrument 6mA calling, SMRT Link Cloud support and 480 Gb of HiFi data per run on four cells. Any quoted Revio figure should be checked against the chemistry it assumes; the specification sheet’s 480 Gb per day belongs to current SPRQ-Nx chemistry.

Where it matters for labs, core facilities and procurement

Planning a Revio is as much a facilities exercise as a sequencing one. The specification sheet gives the physical and environmental requirements, the main points of which follow. The instrument is 92.7 x 91.4 x 174.5 cm, weighs 465 kg (1,025 lb) uncrated, needs 243.8 x 138.5 cm of floor space, and arrives in a crate weighing 635 kg. It operates at 19-25 C and 20-80% relative humidity, non-condensing, below 2,250 m altitude, with ventilation of 13,000 BTU (3,800 W) and noise below 75 dB at 3.3 m. Power is 200-240 VAC at 50-60 Hz, with a maximum of 5 kVA, 4.8 kW and 24 A. The network connection is 1 or 10 GbE copper, and the instrument runs Ubuntu 22.

Data handling is the other planning item. At about 60 GB per hifi_reads.bam file and up to 78 TB per year of output, storage and transfer capacity are sizing decisions, not afterthoughts. PacBio’s SMRT Link Cloud support in v13.3 is one option for off-instrument analysis; whether it suits an institution is a data-governance question for that institution. Facilities that do not own an instrument can buy sequencing as a service; the guide to the next-generation sequencing service buying guide covers the questions to ask a provider.

Common misunderstandings

Revio is not a clinical diagnostic system

PacBio’s specification sheet and training documents carry the statement ‘Research use only. Not for use in diagnostic procedures.’ Results from Revio runs are research data.

Q30 and Q33 are not the same figure

PacBio’s product page gives a read accuracy of 99.95%, expressed as Q33, and a median read quality of at least Q30. The specification sheet’s Q30+ column is something else again: the fraction of bases at Q30 or above, 95%, 90% or 85% depending on library length. Treat them as three different statistics, and quote the one a source actually states.

Yield per day is not yield per run, and not guaranteed

The 480 Gb per day figure is a throughput claim across four SMRT Cells under SPRQ-Nx chemistry. Per-cell yield depends on library length and, per the sheet’s footnote, on sample quality and preparation. A 24-hour run of a 15-20 kb library is specified at 100-120 Gb per SMRT Cell, not at the day’s maximum for every sample.

Methylation does not need a separate library

PacBio states that methylation is detected in every standard run without bisulfite conversion, for 5mC and 5hmC at CpG sites and for 6mA in native DNA. The 6mA caller is designed for Fiber-seq, not as a general 6mA survey tool.

Frequently asked questions

What is the PacBio Revio?

It is PacBio’s long-read sequencing system for research use. It runs up to four SMRT Cells in parallel on independent stages and outputs HiFi reads with on-instrument basecalling and methylation calling.

How much data does the Revio produce?

PacBio specifies up to 480 Gb of HiFi data per day with SPRQ-Nx chemistry. Per SMRT Cell, yield ranges from 35-70 Gb for 5-10 kb libraries to 100-120 Gb for 15-20 kb and 20-25 kb libraries, assuming high-quality samples.

How much DNA does a Revio sample need?

The specification sheet states an input of 500 ng of DNA per sample. PacBio’s v13.3 training material describes a four-fold reduction in DNA input with SPRQ chemistry compared with the earlier chemistry.

What file does the Revio produce?

hifi_reads.bam, about 60 GB each, plus fail_reads.bam according to the technical overview. The data is unaligned BAM, not FASTQ.

What does SPRQ-Nx mean?

SPRQ-Nx is PacBio’s chemistry that extends SPRQ by enabling multiple acquisitions per SMRT Cell. The specification sheet pairs it with Nx SMRT Cell trays and Nx sequencing plates.

Is the Revio approved for diagnostics?

No. PacBio labels it research use only, not for use in diagnostic procedures.

Sources

Data: PacBio Revio library length, run time, Q30+ bases and HiFi yield per SMRT Cell

Specified performance of the PacBio Revio system with SPRQ-Nx chemistry by library length range, compiled from the vendor specification sheet REV09 of 2 June 2026. Yield assumes high-quality samples; 0.5-5 kb yield is reported in reads.

Period covered: 2026.

PacBio Revio library length, run time, Q30+ bases and HiFi yield per SMRT Cell
Library length (kb)Data-collection run time (hours)Q30+ bases (percent)HiFi yield per SMRT Cell
0.5-512956-8 million reads
5-10249035-70 Gb
10-15249070-100 Gb
15-202490100-120 Gb
20-253085100-120 Gb

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about PacBio Revio: Specifications, SPRQ-Nx Chemistry, Output and Facility Needs

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →