Written and maintained by CASRAI Editorial Board
Published
On this page
General product information only, drawn from the manufacturer’s instructions for use and the FDA record. It is not medical advice and does not replace the prescriber’s or manufacturer’s directions.
Short answer
The PARI Trek S is a small, portable, DC-powered air compressor that supplies compressed air to jet nebulizers. PARI’s instructions for use (IFU) describe it as nonsterile, prescription-use only, for adult and pediatric patients, in hospital, clinic or home settings. FDA cleared it through 510(k) K060357 on April 5, 2006, under product code CAF and 21 CFR 868.5630.
Definition and scope
The Trek S is the compressor, not the nebulizer. It produces the airflow; a separate jet nebulizer cup turns liquid medication into an inhalable aerosol. The IFU states that only nebulizers and tubing authorized by PARI should be used with it, and that the nebulizer is for a single patient. For the general device category, see our guides to what a nebulizer is and the nebulizer compressor.
Federal law restricts the device to sale by or on the order of a physician, according to the IFU. The IFU for the combination kit is document 047D0205 Rev A (11/15).
How it is used in practice
Per the IFU, the compressor connects to the nebulizer by tubing and runs from an AC adapter, a 12 V DC adapter, or an optional rechargeable lithium-ion battery. The IFU also lists these practical points for staff who handle the equipment:
- The compressor has auto-resettable thermal protection.
- It should be kept away from screens and monitors because of possible interference.
- It should not be used while operating a vehicle.
- The battery should be charged fully (more than two hours) before first use. Indicator lights show more than 75%, more than 50%, more than 25% and under 25% remaining. A battery in storage should be kept fully charged in a cool, dry place and recharged every six months.
- The nebulizer must be disinfected or sterilized between treatments, not only cleaned. The IFU gives the exact methods; follow the document rather than a summary.
Treatment and cleaning procedures are intentionally not reproduced here. Use the current IFU from PARI (PARI.com/TrekS) and the prescriber’s directions. Filter questions are covered in nebulizer filters.
What is in the kit, and what the IFU says to replace
The package checklist in the IFU for the combination kit lists the items below. This is a useful reference when receiving or auditing a kit.
| Item | What the IFU says |
|---|---|
| Compressor | Portable, DC powered; Class II equipment |
| Air filter | Listed in the package checklist |
| Nebulizer outlet and Wing Tip tubing | Listed in the package checklist |
| PARI LC Sprint reusable nebulizer | Two included; the IFU advises keeping a second reusable nebulizer on hand; replace the nebulizer cup every six months |
| Mouthpiece with expiratory valve | Listed in the package checklist |
| AC adapter and 12 V DC adapter | Listed in the package checklist |
| Carrying case | Listed in the package checklist |
| Rechargeable Li-ion battery | Optional; recharge every six months if unused |
Specifications and the compatibility check
Most pages about this device list specifications without explaining how they relate. The IFU lets you check the compressor against the nebulizer it is paired with.
| Parameter | Value in the IFU |
|---|---|
| Maximum pressure / maximum flow | 35 psi (2.4 bar) / 8.5 LPM |
| Operating point (measured with LC Sprint or LC Plus) | 14.5 psi (1.0 bar) at 4.0 LPM |
| LC Sprint requirement for the specified droplet characteristics | 3.5 to 8 LPM at a minimum of 20 psi |
| Weight | 0.8 lb compressor only; 1.2 lb with battery |
| Dimensions | 4.5 x 3.5 x 1.75 in |
| Electrical input | 12 VDC, 1.67 A, 18 W |
| Ambient operation | 50 to 104 degrees F |
| Transport and storage | 13 to 158 degrees F; 10 to 95% humidity; 500 to 1060 hPa |
| Certification | IEC/EN 60601-1 and IEC/EN 60601-1-2 |
How to read it: the compressor’s maximum (35 psi, 8.5 LPM) covers the range the LC Sprint needs (3.5 to 8 LPM at a minimum of 20 psi). The 14.5 psi, 4.0 LPM figure is the operating point the IFU measures with the LC Sprint or LC Plus, not a separate limit. These numbers are one more reason the IFU limits use to PARI-authorized nebulizers and tubing: another cup has its own requirements.
Where it matters for clinics and supply staff
- Regulatory identity. FDA’s 510(k) database lists K060357, “TREK S NEBULIZER COMPRESSOR”, applicant Pari Innovative Manufacturers, Inc. (Midlothian, VA), product code CAF (Nebulizer, Direct Patient Interface), regulation 868.5630, decision Substantially Equivalent on April 5, 2006. Use these identifiers when matching an item to a record.
- Prescription status. The IFU says prescription use only; supply records should reflect that.
- Consumables. The six-month nebulizer cup interval and the second-nebulizer recommendation are the manufacturer’s stated replacement points.
- Coding. Billing and coding questions for compressors and nebulizers are covered in HCPCS codes for nebulizers; broader equipment context is in respiratory therapy equipment.
Common misunderstandings
- “Trek” is the nebulizer. The Trek S is the compressor; the LC Sprint is the nebulizer cup in the kit.
- Cleaning is enough. The IFU warns that the nebulizer must be disinfected or sterilized between treatments, not just cleaned.
- It can be shared. The nebulizer is for single-patient use only, per the IFU.
- It is sold sterile. The IFU describes the device as nonsterile.
FAQ
What is the PARI Trek S?
A portable, DC-powered air compressor for jet nebulizers, cleared by FDA in 2006 (K060357), prescription use only.
Who can use it?
The IFU states adult and pediatric patients, in hospital, clinic or home environments, on a prescriber’s order.
Does it include a nebulizer?
The combination kit IFU lists two PARI LC Sprint reusable nebulizers, plus tubing, mouthpiece, adapters and a carrying case. Check the kit’s own checklist for what you received.
How much does it weigh?
0.8 lb for the compressor alone and 1.2 lb with the battery, per the IFU.
How often should the nebulizer cup be replaced?
Every six months, according to the IFU.
Sources
- PARI, Trek S combination kit instructions for use, 047D0205 Rev A 11/15, PDF, read 2026-11-08.
- FDA, 510(k) Premarket Notification K060357, FDA 510(k) database record, read 2026-11-08.
- 21 CFR 868.5630, Nebulizer, Legal Information Institute, section title checked 2026-11-08.