Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
An Ambu bag is a self-inflating bag-valve-mask resuscitator, and its parts are the ventilation bag, an oxygen inlet, an intake valve, a patient valve and a patient connector, plus an oxygen reservoir and, on some models, a pressure relief (pop-off) valve and a PEEP valve. Buyers and clinic staff who need to read a product listing or inspect a unit can use this page to name each part, say what it does and see which parts vary by model.
This page describes products and general use; it is not medical advice.
What an Ambu bag is made of
“Ambu bag” is the common name for a bag-valve-mask (BVM) resuscitator, as described by the manufacturer and in a peer-reviewed review. For what the device is for and how it differs from other oxygen equipment, see what is a bag-valve-mask. Here the focus is the components, the way a peer-reviewed review in BJA Education lists them (Lee and Yip, 2023). Who uses it: clinic staff, buyers and students who need to name and check the components.
Parts of an Ambu bag, part by part
| Part | What it does | Fixed or model-dependent |
|---|---|---|
| Ventilation bag | Squeezed by the user to deliver breaths; recoils on its own during expiration and refills from an oxygen source, room air or both. | Present on every unit; size varies by model |
| Oxygen inlet | Connects the unit to an oxygen source. | Present |
| Intake valve | Made of three soft flap valves (see below) that control which way gas moves at the back of the bag. | Present |
| Oxygen reservoir bag | Fills from the oxygen supply so the unit can deliver the highest oxygen concentration available. | Listed as a component; confirm it is included with the exact model |
| Patient valve | Keeps gas moving one way only and directs expired gas out through the expiratory port, which prevents rebreathing. Has an inspiratory port, an expiratory port and a patient connector. | Present; design varies |
| Patient (mask) connector | Joins the patient valve to a face mask or an airway device. | Present |
| Pressure relief (pop-off) valve | Fitted at the inspiratory port of the patient valve to limit excessive pressure. | Usually found on paediatric models |
| PEEP valve | Attaches to the expiratory port of the patient valve. | Optional; only on some models |
Source: Lee CP, Yip YY, “Understanding your bag-valve-mask resuscitator”, BJA Education 2023;23(6):208-211 (see Sources below).
Valves explained: intake, patient and pop-off
The three flap valves in the intake valve
- Intake disc membrane: makes sure gas goes forward to the patient during manual ventilation and not back into the reservoir bag.
- Inlet reservoir flap valve: opens to let room air in when the reservoir bag is empty. It is what lets the bag self-inflate.
- Outlet reservoir flap valve: lets excess gas escape to the atmosphere.
Patient valve designs
The review names two designs: the Ambu single shutter valve and the duckbill valve. It notes that a duckbill design may or may not include an additional expiratory valve. That is a reason to read the model’s description rather than assume all patient valves are built alike.
Pop-off valve
The pop-off valve sits at the inspiratory port of the patient valve and is usually on paediatric models. Pressure settings differ by model, so check the model’s own documentation. See the Ambu bag pop-off valve guide for that part in detail.
Sizes: one manufacturer’s example
The BJA Education review gives bag volumes for one product, the Laerdal Silicone Resuscitator. They show why the two bags on a unit are different parts and why adult and paediatric models are not interchangeable in the listing. They are an example for that model, not a standard.
| Part | Adult model | Paediatric model |
|---|---|---|
| Ventilation bag | 1600 ml | 500 ml |
| Oxygen reservoir bag | 2600 ml | 600 ml |
Source: Lee and Yip 2023, as above. For sizes by manufacturer for infants, see the infant bag-valve-mask guide.
What to verify on the exact model
These checks are drawn from the cited sources and are not a universal rule.
| Check | Why it matters | Source |
|---|---|---|
| Which parts come with the exact model | Reservoir, pop-off valve and PEEP valve are described as model-dependent or optional, and the manufacturer notes units are available with a variety of pre-attached accessories. | Lee and Yip 2023; Ambu SPUR II product page |
| Patient valve design | Single shutter and duckbill designs differ, and some duckbill designs add an expiratory valve. | Lee and Yip 2023 |
| Materials statement | The manufacturer of the SPUR II states its bag body is made from SEBS rather than PVC, with no PVC, DEHP or phthalates. This is a manufacturer claim for that product. | Ambu SPUR II product page |
| MR status | The same page lists the SPUR II as MR Conditional, so MR status is stated per model. | Ambu SPUR II product page |
| Instructions for use and datasheets | The product page lists an IFU and datasheets as downloads; check the model’s own IFU for the standard it declares. | Ambu SPUR II product page |
Source: Ambu SPUR II product page (see Sources below). For equipment that sits alongside the bag, the emergency oxygen kit checklist lists what else is stocked. Suppliers of related respiratory consumables are grouped under anesthesia circuits.
Safe use and handling basics
Follow the model’s instructions for use, which is the document that governs that product. Training in manual ventilation is separate from reading a parts list, and nothing here replaces it. The reservoir bag and the oxygen inlet only help if an oxygen source is connected, and the outlet reservoir flap valve lets excess gas escape to the atmosphere. Performance figures differ by model, so take them from the model’s own documentation.
Common mistakes
- Confusing the two bags. The ventilation bag is squeezed; the oxygen reservoir bag only stores oxygen.
- Assuming every model has a pop-off valve. It is usually a paediatric feature, so check the listing.
- Assuming a PEEP valve is included. It is optional and fitted only on some models.
- Treating the brand name as a specification. “Ambu bag” is shorthand; configuration is in the model’s description and IFU.
FAQ
What are the parts of an Ambu bag?
The ventilation bag, oxygen inlet, intake valve, oxygen reservoir bag, patient valve, patient connector, and on some models a pressure relief (pop-off) valve and a PEEP valve (Lee and Yip, 2023).
Which parts vary by model?
The pop-off valve is usually found on paediatric models, the PEEP valve is fitted on some models only, and the patient valve design differs (single shutter or duckbill). Bag volumes also differ by model.
What does the patient valve do?
It keeps gas moving one way and directs expired gas out through the expiratory port, which prevents rebreathing.
Is the reservoir bag the same as the ventilation bag?
No. The ventilation bag is the one that is squeezed to deliver breaths; the reservoir bag fills from the oxygen supply.
Sources
- Lee CP, Yip YY. “Understanding your bag-valve-mask resuscitator.” BJA Education 2023;23(6):208-211. https://pmc.ncbi.nlm.nih.gov/articles/PMC10201401. Checked 3 November 2026.
- Ambu SPUR II product page. https://www.ambuusa.com/products/airway-management/resuscitators/product/ambu-spur-ii. Checked 3 November 2026.