Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General product and regulatory information only. This page is not medical advice and does not describe how to place, use or care for any catheter; clinical decisions belong to the treating team and the manufacturer’s instructions for use.
Short answer
A peritoneal drain is a catheter that leaves the abdominal (peritoneal) cavity so fluid can be drained out of it. In the FDA’s classification, the long-term version used for refractory ascites is a tunneled, indwelling catheter filed under product code PNG, Class II, and it is not the same device family as a peritoneal dialysis catheter even though both sit under one regulation number.
Definition and scope
“Peritoneal drain” is a descriptive phrase, not a regulatory term. FDA’s product classification database lists code PNG as “Peritoneal, Drainage Catheter For Refractory Ascites, Long-Term Indwelling”, defined as a device for drainage of refractory ascites with long-term occurrence from the peritoneal cavity. The record shows Class 2, a 510(k) submission route, the Gastroenterology/Urology panel and regulation number 876.5630 (checked 2026-10-30).
That regulation, 21 CFR 876.5630, is titled “Peritoneal dialysis system and accessories”. Its identification describes a peritoneal access device (a flexible tube passing through the abdominal wall into the peritoneal cavity, for short-term use under 30 days or long-term use), an administration set and a dialysate source. Code PNG reuses this regulation number for a drainage purpose, so the number alone does not tell you whether a product is for dialysis or for ascites drainage. The product code and the labeled indication do.
Ascites is the accumulation of fluid in the peritoneal cavity. This page covers the device and its classification only, not diagnosis or treatment of ascites.
How it is used in practice
The cleared labeling for the PleurX Peritoneal Catheter System (510(k) K201155, decision date 2020-10-21, product code PNG) states that it is indicated for intermittent, long-term drainage of symptomatic, recurrent, malignant and non-malignant ascites in adults unresponsive to medical management. The summary describes a percutaneously tunneled indwelling catheter with internal fenestrations, radiopaque markings, a tissue cuff and an external valve.
The later PeritX Peritoneal Catheter System (510(k) K241946, Bard Peripheral Vascular, decision date 2024-12-06) is described as a family of sterile, single-use indwelling peritoneal catheters plus accessories, including valve and access kits, drainage lines, and 1 L and 2 L vacuum-initiated drainage kits.
One peer-reviewed observational study of 51 patients in Goettingen (Petzold et al., Cancers 2021;13(12):2926) describes tunnelled peritoneal PleurX catheters as a permanent solution to drain ascites and gives the device as a 15.5 Fr, 71 cm fenestrated silicone catheter with a one-way valve and a polyester cuff. That is one centre’s experience, in a population where 27 of 51 patients died within 30 days, and its rates should not be read as general figures for the device.
Types and related terms
The same object is described with different words depending on who is writing. The table lists only wording found in the sources above.
| Term | Where it appears | What it points to |
|---|---|---|
| Peritoneal drainage catheter | FDA code PNG device name | The regulatory category: drainage of refractory ascites, long-term indwelling, Class 2 |
| Indwelling peritoneal catheter | PeritX 510(k) K241946 description | A sterile, single-use catheter left in place, sold with accessory kits |
| Tunneled peritoneal catheter | K201155 technological characteristics; Petzold et al. 2021 | A catheter passed through a subcutaneous tunnel with a tissue cuff (the study reports a 5-8 cm tunnel) |
| Peritoneal access device | 21 CFR 876.5630 | The dialysis-system component, short- or long-term; a different intended use from code PNG |
Peritoneal drainage catheter
This is the formal name for the PNG category. The openFDA 510(k) database lists PNG clearances from several manufacturers, including PleurX, PeritX, ASEPT, Aspira and Passio systems.
Tunneled and indwelling peritoneal catheters
Both phrases describe the same long-term design: the catheter stays in place between drainage sessions, the tunnel and cuff anchor it, and a valve at the outer end closes it. Related CASRAI pages cover other drains and catheter vocabulary, including the external ventricular drain and catheter lumen.
Where it matters for labs, clinics and supply decisions
Three distinctions come straight from the sources and prevent most mix-ups when a request names a “peritoneal catheter”:
- Purpose. Check the product code and indication. Code PNG is for refractory ascites drainage; items under the dialysis system description in 21 CFR 876.5630 are for peritoneal dialysis. A dialysis supply list, such as the outpatient dialysis center supply checklist, does not cover ascites drainage kits.
- Catheter versus kit. The PeritX family lists the catheter, valve kit, access kit, lockable drainage lines and 1 L and 2 L vacuum drainage kits as separate items under one clearance. The catheter and the drainage components are not interchangeable terms.
- Labeled population. The K201155 indication is adults, with further limits stated in the labeling. Any use outside the cleared labeling is a question for the clinical team and the manufacturer, not for a procurement list.
For background on dialysis equipment as a separate topic, see what is a dialysis machine.
Common misunderstandings
“It is a dialysis catheter.”
Not by intended use. It shares regulation number 876.5630 with dialysis access devices, but its PNG classification defines it as a drainage catheter for refractory ascites.
“It is the same as a paracentesis needle.”
The K201155 labeling describes the indwelling catheter as a device for intermittent, long-term drainage, a different product category from a needle used for a single-session paracentesis.
“The study numbers show how safe it is.”
The Goettingen series is a single-centre observational study of 51 patients with advanced disease. It reported complications requiring explant in four patients (cellulitis in two, peritonitis in one, dislocation in one). Those are that cohort’s figures, not general rates.
FAQ
What is a peritoneal drain?
A catheter that drains fluid from the peritoneal cavity. The long-term indwelling version for refractory ascites is FDA product code PNG, Class 2.
What is the difference between a tunneled and an indwelling peritoneal catheter?
The sources use the words for the same long-term device: “indwelling” says it stays in place, “tunneled” says it runs through a subcutaneous tunnel with a cuff.
Is a peritoneal drainage catheter the same as a peritoneal dialysis catheter?
No. Both fall under 21 CFR 876.5630, but code PNG is defined for ascites drainage, while the regulation’s identification describes dialysis systems.
Is it single use?
The PeritX 510(k) describes its catheters and accessories as sterile and single-use; check the instructions for use of any specific product.
Sources
- FDA, Product Classification, code PNG: accessdata.fda.gov, and openFDA device classification and 510(k) records, checked 2026-10-30.
- 21 CFR 876.5630, Peritoneal dialysis system and accessories: ecfr.gov, checked 2026-10-30.
- 510(k) K201155, PleurX Peritoneal Catheter System, decision 2020-10-21, and K241946, PeritX Peritoneal Catheter System, decision 2024-12-06, summaries checked 2026-10-30 (openFDA 510(k) records, product code PNG).