Pirani Gauge: Principle, Range, Gas Correction and Zero Adjustment

How a Pirani gauge works, its accuracy by pressure band, gas correction factors and the manufacturer zero-adjustment steps, from a Lesker manual.

Written and maintained by CASRAI Editorial Board

Published

On this page

A Pirani gauge measures low gas pressure from the heat a gas carries away from a hot wire. This page is for lab staff and vacuum-system users who need to read one correctly: what it can and cannot measure, how accurate it is in each pressure band, how the reading changes with the gas, and how the zero is checked. Figures come from the Kurt J. Lesker KJLC PIR operating manual (tkna44e1, 2017-03), read 2026-11-07.

What a Pirani gauge is used for

A Pirani gauge is a thermal-conductivity vacuum gauge used to monitor pressure in vacuum systems. The Museum of Mass Spectrometry describes a gauge head of this type being used to monitor the roughing pressure in front of a diffusion pump. Typical users are core-facility staff, vacuum-system operators and anyone running pumps, freeze-dryers or coating tools such as those described in the CASRAI guides on lyophilization and sputtering. It is not a general-purpose sensor: the manufacturer states that it must not be used for flammable or combustible gases in mixtures containing oxidants within the explosion range.

What is the working principle of a vacuum gauge of this type?

The sensing element is a thin filament heated by an electric current, to roughly 100 to 150 °C according to the Museum of Mass Spectrometry. Gas molecules that collide with the filament carry heat away, so at higher pressure more heat is lost. The filament forms one arm of a Wheatstone bridge, and modern gauges adjust the bridge voltage to keep the filament at constant resistance and temperature; that voltage is then a measure of pressure. Lesker’s manual names the measurement principle as thermal conductance according to Pirani.

Pressure range and accuracy

The Lesker KJLC PIR is specified for 5×10⁻⁴ to 1000 mbar (for air, O₂, CO and N₂). The useful point for operators is that accuracy is not uniform across that range:

Pressure band (N₂) Stated accuracy
1×10⁻³ to 100 mbar ±15% of reading
5×10⁻⁴ to 1×10⁻³ mbar ±50% of reading
100 to 1000 mbar ±50% of reading

Source: Kurt J. Lesker KJLC PIR manual, Technical Data, read 2026-11-07. Other models differ, so check the specification of the gauge you own.

Gas dependence and correction factors

The reading is gas dependent. The manual states the pressure indication applies to dry air, O₂, CO and N₂; for any other gas it must be corrected, and a controller may let you set a calibration factor. Below 1 mbar the manual gives the rule peff = C × pressure reading:

Gas C Gas C
He 0.8 H₂ 0.5
Ne 1.4 air, O₂, CO, N₂ 1.0
Ar 1.7 CO₂ 0.9
Kr 2.4 water vapor 0.5
Xe 3.0 freon 12 0.7

Source: Lesker manual, calibration factors for the pressure range below 1 mbar. Worked example: a display reading of 2×10⁻² mbar in argon corresponds to an effective pressure of 1.7 × 2×10⁻² = 3.4×10⁻² mbar. The manual gives this factor rule only for the range below 1 mbar.

How can I calibrate a Pirani vacuum gauge?

The Lesker gauge is factory calibrated. The manual notes that long operation or contamination can cause zero drift and says to check the zero periodically and adjust if necessary, under the same ambient conditions and in the same mounting orientation as normal use. Its two-point adjustment works as follows:

  1. ATM adjustment: check that the seal, centering ring and filter are clean, operate the gauge at atmospheric pressure for at least 10 minutes, then press the adjustment button with a pin (maximum 1.1 mm diameter). The default is 1000 mbar (10 V).
  2. HV adjustment: evacuate to well below 10⁻⁴ mbar (recommended) or to the 10⁻⁴ to 10⁻² mbar range, wait at least 2 minutes, then press the button. The default is 1.2×10⁻⁴ mbar (1.1 V).

The manual allows adjusting to other values only when the exact pressure is known from a reference measurement. It also advises letting the gauge stabilize for at least 10 minutes after power-up and operating it continuously. Other makes use different procedures, so follow the manual for your model.

Types of vacuum gauges: where the Pirani sits

A Pirani gauge is the thermal-conductivity type. Its sourced range, down to about 10⁻⁴ mbar, covers rough and medium vacuum monitoring; pressures far below that need a different measuring principle, and this page does not rank or compare other gauge families because the sources used here describe only the Pirani.

Is 0 barg a vacuum?

That depends on the pressure reference, which the sources used here do not define, so this page makes no claim. Check whether your instrument or controller reports absolute or gauge pressure before comparing readings with the mbar figures above, which are the manufacturer’s gauge specifications.

Common mistakes

  • Reading a non-air gas without applying the correction factor.
  • Trusting the display above 100 mbar or near 5×10⁻⁴ mbar, where the stated accuracy is ±50% of reading.
  • Skipping the zero check after contamination or long operation.
  • Adjusting the zero with a different orientation or ambient conditions than normal use.
  • Using the gauge on flammable gas and oxidant mixtures in the explosion range.

Contamination failures and consumable parts such as the filament are not covered by the Lesker warranty.

FAQ

What is the Pirani gauge used for?

Measuring low gas pressure in vacuum systems from heat loss of a heated filament, for example roughing-line monitoring.

What is the working principle of a vacuum gauge?

For a Pirani gauge, thermal conductivity: gas cools a heated filament in a bridge circuit, and the bridge signal maps to pressure.

How can I calibrate a Pirani vacuum gauge?

Follow the manufacturer’s zero adjustment: at atmospheric pressure, then at high vacuum, as set out above.

Sources

  • Kurt J. Lesker Company, KJLC PIR Standard Pirani Gauge Operating Manual, tkna44e1 (2017-03): lesker.com, read 2026-11-07.
  • Museum of Mass Spectrometry, Pirani vacuum gauge: ms-museum.org, read 2026-11-07.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Pirani Gauge: Principle, Range, Gas Correction and Zero Adjustment

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →