Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
A plate sealer is a laboratory instrument that closes a microplate with a film or foil so that the samples in its wells are protected from evaporation and contamination during storage, incubation or thermal cycling. The most common design is a thermal (heat) sealer, which presses a heated platen onto the film so it fuses to the rim of every well; the alternative is an adhesive film pressed on by hand, with no machine at all. This guide explains how plate sealers are built, how the manual, semi-automatic and automated roll types differ, which specifications manufacturers publish, and which handling limits they state.
Who uses it: researchers and laboratory technicians who store, archive or amplify samples in 96-well, 384-well, PCR and deepwell plates, and the lab managers who equip those workflows.
What is a plate sealer?
Thermo Scientific describes its ALPS30 as a hand-operated microplate heat sealer that gives a safe and controlled method for sealing plastic sample plates, protecting the samples from evaporation and contamination during storage. Its manual states the intended use more precisely: applying a consistent, tight seal around individual wells, which prevents evaporation loss and cross-contamination between wells. The listed applications are compound storage, sample archiving and PCR.
The word “plate” refers to the microplate, the grid of small wells that most screening, assay and PCR work is run in. The microplate guide covers the plate itself; a sealer is one of the accessory instruments that sit around it. Wikipedia’s microplate article lists plate thermal sealers (for applying heat seals) and de-sealers (for removing them) among the robots and instruments developed to handle microplates, next to liquid handlers, plate movers, stackers, washers and incubators.
Two seal families exist. Heat seals are fused to the plate by the instrument. Adhesive seals are pressed on and rely on a pressure-sensitive coating. Only the first family needs a plate sealer machine, which is why the term “plate sealer” in a purchasing or protocol context normally means a heat sealer. Adhesive and cling films, including the flask and plate wraps used for cell culture, are covered separately in the laboratory sealing film guide.
How a microplate heat sealer is built and works
The manuals for two very different machines show the same core design: a heated plate, a carrier or support that holds the microplate at the right height, and a mechanism that brings the two together for a set time.
On the ALPS30 manual sealer the parts are a seal activation handle, status LEDs, an internal heater plate, an adapter plate carrier, and the power inlet, switch, fuse and voltage selector. The microplate sits in an adapter block on the carrier with the film on top, and the operator pulls the handle down until the heater plate contacts the plate. The sealing temperature is fixed at 340 °F (170 °C) with a tolerance of ±2 °C, and an amber LED flashes when the unit is at temperature. The manual says the sealing time may need to be adjusted empirically for each plate and film type.
The Azenta automated roll heat sealer replaces the handle with an external compressed-air supply and feeds film from a roll. It is supplied with a roll holder, a seal loading tool, plate support adapters A and B, vacuum cups and a spare fuse. The roll is loaded with the loading tool, and a cutter section cuts the film to length. Its sealing temperature and time are programmable over a range rather than fixed (see the specification table below).
A semi-automatic design sits between the two. Porvair Sciences describes its MiniSeal II as an electrically driven, stepper-motor-controlled sealer with 450 watts of heating power, programmable up and down settings for sealing time and temperature, a drawer that presents the plate, and automatic ejection of the sealed plate. The company states that sealing times for most polypropylene plates with its seals are under 3 seconds, and that the unit is supplied with adaptors for standard SBS-format microplates, deepwell microplates and PCR plates, plus a brass plate weight that keeps foils and seals flat during sealing.
Types of plate sealer
The table compares the three machine types described by the manufacturer documents above. Every row is taken from the cited manual or product page.
| Type | Example documented | How the seal is applied | Stated use or positioning |
|---|---|---|---|
| Manual heat sealer | Thermo Scientific ALPS30 | Operator pulls a handle to press the heater plate onto the plate and film; fixed 170 °C | Low-throughput labs; compound storage, sample archiving, PCR |
| Semi-automatic heat sealer | Porvair Sciences MiniSeal II | Electrically driven; programmable time and temperature; drawer presents plate and ejects it sealed | PCR, filter-bottomed, assay and deepwell plates, including distorted or bowed plates |
| Automated roll heat sealer | Azenta automated roll heat sealer | Roll-fed film cut by an internal cutter; driven by compressed air; RS-232 and USB connection | Wide film and plate compatibility, supplied with roll holder and loading tool |
The same documents point to the main axes along which types differ: how much of the process the instrument performs, whether time and temperature are fixed or programmable, whether film comes pre-cut or from a roll, and how much bench space and utility supply the unit needs. The data table after this guide sets the published numbers side by side.
Which plates a sealer has to accommodate
Plate geometry decides whether a sealer fits a given plate. The ALPS30 uses three adapter blocks so that it suits virtually any plate height, and its listed blocks cover 96-well and 384-well microtiter plates, 96-well skirted, semi-skirted and unskirted PCR plates, and deepwell plates. Its manual lists polypropylene, polyethylene and polystyrene as plate materials. The Porvair MiniSeal II page states that it seals plates of any height from 9 to 48 mm, and seals SBS-format microwell plates and tubes.
Wikipedia’s microplate article explains why adapters are needed. The ANSI SLAS standards define a footprint and a bottom-flange geometry that microplate makers generally follow, but deepwell plates have no height defined in the ANSI SLAS 2-2004 standard and typically follow a de facto height of 44 mm. PCR plates have thin walls for thermal conduction and come in full-skirt, semi-skirt and unskirted forms, and the skirted variants other than full-skirt may not comply with the ANSI SLAS footprint. See the 96-well plate dimensions guide for the footprint standard in detail.
Specifications manufacturers publish
Published specifications describe the instrument itself, not the quality of the seal. The units below are the ones documented in the manuals and product pages read for this guide.
- Thermo Scientific ALPS30 manual sealer. Fixed sealing temperature of 340 °F (170 °C), non-adjustable, ±2 °C. Dimensions 8.7 × 9.8 × 9.1 in (220 × 250 × 230 mm), weight 8.8 lb (4 kg), maximum power 350 W. Operating conditions 65 to 85 °F (18 to 30 °C) at 20 to 80% relative humidity, non-condensing. The manual lists no scheduled maintenance, only routine cleaning and inspection and a decontamination procedure.
- Porvair Sciences MiniSeal II. Seal time 0.1 to 9.9 seconds, plate heights 9 to 48 mm, 450 W of heating power, weight 6 kg, footprint smaller than a sheet of A4 paper.
- Azenta automated roll heat sealer. Sealing temperature range 100 to 195 °C, sealing time range 0.1 to 10 s, dimensions 230 × 507 × 276 mm, 27 kg without a roll, 700 W maximum, AC 100 to 240 V, RS-232 and USB connections, working temperature 10 to 30 °C, indoor use only.
The data table after this guide compiles the numeric values from these three sources into one view. The manual unit is fixed at one temperature, while the Azenta unit spans 100 to 195 °C and weighs about seven times as much (27 kg without a roll, against 4 kg).
What the sources say about choosing a sealer
Manufacturer documents do not offer a scoring method, so this section records only the criteria they state or imply, without ranking models.
- Plate formats supported. Check that the unit’s adapters or carrier match the plates in use. The ALPS30 lists its adapter blocks by plate type; the MiniSeal II states a plate height range of 9 to 48 mm.
- Fixed or programmable settings. The ALPS30 has a fixed 170 °C and relies on the user setting time empirically. The MiniSeal II and the Azenta unit allow time to be programmed, and the Azenta unit allows temperature to be programmed as well.
- Throughput class. Thermo designs the ALPS30 for low-throughput labs. The Azenta unit, with roll-fed film, a cutter and a serial interface, is documented as an automated instrument.
- Film compatibility. The ALPS30 manual names pierceable, optically clear and permanent films as the choices. The Azenta manual describes wide compatibility with sealing films and foils. A film has to be specified for the unit and the plate material; the Porvair page, for example, quotes sealing times for polypropylene plates with its own seals.
- Utilities and space. The Azenta unit needs an external compressed-air supply and indoor use; the ALPS30 needs only mains power and a bench footprint of about 220 × 250 mm.
The ALPS30 manual tells the operator to review seal integrity after sealing and to adjust time for the plate and film, so a published specification does not replace a check on the actual plate and film in use.
Safe use and handling basics
These points are drawn from the ALPS30 user manual, because it is the one document read here that gives a full operating and safety sequence. Other models carry their own instructions, and the manufacturer’s manual for the unit in use is the governing document.
- Switch the unit on and wait for the temperature indicator to show the heater plate is at temperature; the ALPS30 flashes an amber LED.
- Load the plate in the correct adapter block on the carrier and place the film on top, right side up.
- Bring the heater plate down onto the plate without applying more force than is needed.
- Release the handle when the sealing time has elapsed. Leaving the handle down too long can damage the plate and degrade the seal.
- Remove the plate with appropriate personal protective equipment and inspect the seal. The plate and film can remain hot for several seconds.
The manual’s stated hazards are specific. The heater plate runs at 340 °F (170 °C) and stays hot for a considerable time after the unit is switched off, which is a severe burn risk. The handle and carrier present a finger-crush hazard. The unit is for plates of biological samples only: it must never be used to seal explosive, volatile or highly reactive substances, and it is not for use in areas with flammable gases. There are no user-serviceable parts inside. A unit that has been stored cold should be allowed to acclimatise before it is powered on.
When the seal is placed matters as much as how: the ALPS30 manual treats the seal as protection against evaporation and cross-contamination, so a poor seal puts exactly those two outcomes at risk. Sealed PCR plates are typically heated next in a thermal cycler, and the PCR tube guide covers the single-tube equivalent of a PCR plate.
Common mistakes the manuals warn against
- Treating sealing time as universal. The ALPS30 manual says the time may need to be adjusted empirically for each plate and film type. A time that works for one combination is not guaranteed for another.
- Holding the handle down too long. The manual states this can damage the plate and degrade the seal.
- Applying extra force. The instructions say not to press harder than needed to bring the heater plate into contact.
- Sealing the wrong contents. Volatile, explosive or highly reactive substances are excluded by the manufacturer.
- Handling the plate straight after sealing. Plate and film stay hot for several seconds, and the heater plate stays hot long after switch-off.
- Using the wrong adapter. The sealer is built around adapter blocks matched to plate type and height, so a plate in the wrong block is not supported by the documented design.
- Using a sealer outside its environment. The ALPS30 specifies 18 to 30 °C and 20 to 80% non-condensing humidity; the Azenta unit specifies 10 to 30 °C and indoor use.
Sources
- Thermo Scientific ALPS30 Manual Heat Sealer User Manual (May 2015), read 2026-11-17.
- Azenta Automated Roll Heat Sealer User Manual, part 347598 Rev. E (4 July 2023), read 2026-11-17.
- MiniSeal II Semi-Automatic Plate Sealer from Porvair Sciences (News-Medical product page), read 2026-11-17.
- Microplate (Wikipedia), describing ANSI SLAS standards, deepwell height and PCR plate skirt types, read 2026-11-17.
Data: Published specifications of three plate sealers: manual, semi-automatic and automated roll
Numeric specifications compiled from the Thermo Scientific ALPS30 user manual, the Porvair Sciences MiniSeal II product page and the Azenta automated roll heat sealer user manual, one row per sealer type.
Period covered: 2015/2023.
| Sealer | Type | Maximum power (W) | Weight (kg) | Sealing temperature | Sealing time |
|---|---|---|---|---|---|
| Thermo Scientific ALPS30 | Manual | 350 | 4 | 170 (fixed, +/-2) | Set by operator, adjusted empirically |
| Porvair Sciences MiniSeal II | Semi-automatic | 450 | 6 | Programmable (range not stated) | 0.1-9.9 |
| Azenta automated roll heat sealer | Automated roll | 700 | 27 | 100-195 | 0.1-10 |
Sources
- Thermo Scientific ALPS30 Manual Heat Sealer User Manual (accessed 2026-11-17)
- MiniSeal II Semi-Automatic Plate Sealer from Porvair Sciences (accessed 2026-11-17)
- Azenta Automated Roll Heat Sealer User Manual 347598 Rev. E (accessed 2026-11-17)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.