Plate Washer: What It Is, How It Works and How to Maintain It

How a microplate washer dispenses and aspirates wash buffer, the types and specs vendors state, cleaning routines and troubleshooting.

Written and maintained by CASRAI Editorial Board

Published

On this page

A plate washer, or microplate washer, is a laboratory instrument that dispenses wash buffer into the wells of a microplate and then removes it by controlled aspiration, so that unbound material is cleared from each well between assay steps. It is most often used for ELISA and other plate-based binding assays, and common models handle 96-well and 384-well plates.

What a plate washer is and what it is for

In a plate-based immunoassay, antigen, antibody or enzyme that has not bound to the solid phase has to be removed before the next reagent is added. Manufacturer documentation describes this as the purpose of the washer: the SkanWasher 400 user guide states that the instrument washes unbound antigens, antibodies and enzymes away from the solid phase in ELISA. A 2015 SelectScience article puts the reason in terms of the result: excess reagent must be removed from the plates to prevent high background, and adding and aspirating wash buffer with a washer gives precision and speed that manual washing does not.

Users are mainly laboratory staff and core-facility users running plate assays. The instrument sits between the pipetting steps and the plate reader in the workflow, and it works on standard microplates, so the wells it washes follow the footprint and well geometry described in 96-well plate dimensions. Do not confuse it with a washer-disinfector, which cleans reusable instruments and is a different device. The Molecular Devices MultiWash+ brochure also carries the note “For research use only. Not for use in diagnostic procedures”, which applies to that model and shows that intended-use statements differ by instrument.

How a plate washer works

Every washer in the documents read for this page repeats the same cycle: dispense wash buffer into the wells, optionally soak or shake, aspirate the liquid, and repeat for the programmed number of cycles. What differs between instruments is how liquid is delivered and removed.

Wash head and probes

The SkanWasher 400 has a wash head with 96 paired probes: the longer probes aspirate and the shorter probes dispense. The head can be removed for cleaning. The MultiWash+ brochure describes crosswise aspiration, which it says reduces the residual volume left in each well. A microplate lift positions the plate under the head on the SkanWasher.

Liquid handling

The two Molecular Devices instruments differ in how they move liquid. The SkanWasher 400 has an internal vacuum pump and an internal pressure source, and its bottles withstand up to 0.6 bar; the user guide says to fill a bottle to about 80% and leave roughly 20% air space, and to use one reservoir at a time. The MultiWash+ is described as vacuum and pressure free, using on-board pumps, with four wash or rinse bottles and one waste bottle. The Thermo Scientific Wellwash guidance describes non-pressurized bottles that the user supplies.

Programmable wash steps

The MultiWash+ brochure lists 20 stored wash protocols with up to 8 cycles each, adjustable dispense speed and volume, adjustable aspiration speed and time, soak times and three shaking modes. The SkanWasher 400 exposes aspirate and dispense probe height and position, inlet flow (gentle or hard), an overflow wash option, a dispense-only option and an adjustable lift speed. The aspirate probe must sit close to the well bottom to leave a low residual volume, and the adjustable distance on that instrument is 6 mm.

Sensors

The SkanWasher 400 has three safety sensors: pressure pump level, plate in lift, and waste reservoir space. The Wellwash range is described as having liquid-level sensors, plate sensors and an aerosol-resistant cover. SelectScience explains the benefit of level sensors on the Wellwash: they warn before wash solution runs out, so a plate does not pass through with unwashed wells, and they help prevent waste-bottle leaks and spills.

Types and variants

The documents read describe the following variations. Each row is stated by the named source for that instrument or article, and none of it should be read as a ranking.

Variant What the source says Source
96-well and 384-well washers MultiWash+ handles both formats; 384-well washing uses a 16-way manifold; 12- and 16-way manifolds are optional accessories. Wellwash and Wellwash Versa are described as 96/384-well. Molecular Devices brochure; Thermo Scientific poster
Pressurized and non-pressurized bottles SkanWasher 400 uses an internal pressure source with bottles rated to 0.6 bar. The Wellwash uses non-pressurized bottles. SkanWasher 400 user guide; Thermo Scientific poster
Cell-washing head The Wellwash Versa offers a special cell-washing head. Thermo Scientific poster
Angled or adjustable dispensing Washers with angled or adjustable dispensing pins can wash wells without disrupting the cell layer; aiming at the well wall gives a gentler wash. SelectScience, 2015
Washer with built-in stacker The Titertek-Berthold Zoom HT is described with a built-in microplate stacker and self-emptying vacuum system for high throughput. SelectScience, 2015
Dispense-only operation SkanWasher 400 offers a dispense-only mode alongside normal wash cycles. SkanWasher 400 user guide

Specifications that matter and how to read them

The data table below the article compares what three manufacturer documents state about plate formats, liquid delivery, aspiration and built-in protection. Two points about reading such specifications are drawn from the same documents.

  • Residual volume is conditional. The MultiWash+ brochure gives a residual volume of less than 1 µL per well, a dispense range of 50 to 2000 µL in 50 µL steps and a precision below 5% at 300 µL across the plate. Those figures belong to that instrument under the manufacturer’s test conditions; the Wellwash poster, in contrast, lists the settings that determine residual volume in practice (aspiration height, speed and mode), so a figure from one document does not transfer to another model.
  • Time depends on volume and cycles. For a 96-well plate at 300 µL, the MultiWash+ brochure gives 1 minute 9 seconds for one wash and 3 minutes 35 seconds for three washes. The same document gives a fits-the-ANSI/SBS-footprint statement, a dead volume of 15.62 mL and a weight of 8 kg.

Criteria the manufacturers and the literature state

No standard defines which washer a laboratory should buy, so the only criteria that can be stated are those found in the sources read. SelectScience names speed and constant flow as the key factor for laboratories with many samples, adjustable dispense and aspiration rates for weakly adherent cells, dispensing angle for adherent cell layers, and liquid-level sensors for protection against unwashed wells and spills; it says only that there are many other factors to consider. The manufacturer documents add the plate formats supported (96, 384), whether the bottles are pressurized, whether cell-washing heads exist, and what maintenance the instrument requires. Check each instrument’s own documentation for these points before relying on any figure.

Safe use, priming and decontamination

The following points are taken from the manufacturers’ own guidance and apply to the models named; they are not a general procedure.

  • Prime first. The Thermo Scientific poster advises priming at first use each day and checking that liquid comes through all tips, then flushing the channels with distilled water after use. Its containers and tubes can be autoclaved up to 10 times.
  • Daily rinse. The SkanWasher 400 user guide specifies a daily rinse program with distilled water (a 5-second rinse and 5-second soak, three times) and says to empty the wash head at the end of the day.
  • Decontaminate after infectious material. The same guide says to decontaminate the instrument after use with infectious material, using laboratory decontaminants according to their directions.
  • Weekly cleaning. The guide describes a weekly clean, or one after contaminated solution, with a 2% bleach or decontaminant solution (never above 10% bleach), a 5-minute probe soak and five rinses with distilled water, with a caution about the chemical resistance of the parts.
  • Bottle pressure. On the pressurized SkanWasher, fill bottles to about 80% and use one reservoir at a time, as above.
  • Maintenance accessories. The MultiWash+ brochure lists a maintenance kit with cleaning needles.

Follow your institution’s biosafety and chemical-waste rules for the contents of the waste bottle; the manufacturer guidance above addresses only the cleaning of the instrument.

Common mistakes and troubleshooting

The Thermo Scientific good washing practice poster lists causes for the problems that most often affect plate washing. The table reproduces them for orientation.

Symptom Causes and actions listed by the manufacturer
High residual volume Adjust aspiration height, aspiration speed or sweep aspiration mode; clean tips with the de-clogging tool; rinse the head; check the head is fitted.
Low dispense volume Clean the inlet filter; check that step parameters were not reset to default after a head or well-type change; clean tips; recalibrate.
High dispense volume Check that step parameters were not reset to default after a head or well-type change; recalibrate.

The same poster describes the conditions for a good wash: the head near the bottom of the well (the default), a high aspiration speed, a well offset slightly off-centre (0.5 to 1.0 mm for flat-bottom wells), tips that are not clogged and the correct dispense volume. The SkanWasher 400 guide gives the same reason for the aspirate probe position: it must be close to the well bottom to leave little liquid behind.

The recurring mistakes in these documents are therefore practical: skipping the daily prime or rinse, letting tips clog, changing the plate type without re-checking the step parameters, and setting the aspiration probe too far from the well bottom.

For the assay itself, including where washes sit in the protocol, see ELISA protocol basics and direct ELISA.

Sources

  • Molecular Devices, MultiWash+ 96- and 384-well microplate washer brochure (2017; product discontinued), https://www.moleculardevices.com/sites/default/files/en/assets/brochures/br/96-and-384-well-multiwash-microplate-washer.pdf, read 2026-11-17.
  • Molecular Devices, SkanWasher 400 user guide, https://moleculardevices.com/sites/default/files/en/assets/user-guide/br/skanwasher-400-user-guide.pdf, read 2026-11-17.
  • Thermo Scientific, Your guide to good washing practice (Wellwash and Wellwash Versa poster), https://assets.thermofisher.cn/TFS-Assets/LCD/brochures/LCD-Thermo-Scientific-Good-Washing-Practice-Poster-EN.pdf, read 2026-11-17.
  • SelectScience, How choosing the right microplate washer can improve assay performance (13 April 2015), https://www.selectscience.net/article/how-choosing-the-right-microplate-washer-can-improve-assay-performance, read 2026-11-17.

Data: What three manufacturer documents state about microplate washer formats, liquid delivery and aspiration

Original compilation of the plate formats, liquid-delivery design, aspiration features and built-in protections that three manufacturer documents (two Molecular Devices, one Thermo Scientific) state for their microplate washers.

Period covered: 2017/2026.

What three manufacturer documents state about microplate washer formats, liquid delivery and aspiration
Washer documentPlate formatsLiquid deliveryAspiration and wash settingsBuilt-in protections
Molecular Devices MultiWash+ (brochure)96- and 384-well (384 via 16-way manifold)On-board pumps, vacuum and pressure free; 4 wash/rinse bottles and 1 waste bottle; dispense 50-2000 uL in 50 uL stepsCrosswise aspiration; 20 protocols of up to 8 cycles; adjustable aspiration speed and time; soak times; three shaking modes; residual under 1 uL per wellProduct labelled for research use only
Molecular Devices SkanWasher 400 (user guide)96-well (96 paired probes)Internal vacuum pump and internal pressure source; bottles withstand up to 0.6 barAdjustable aspirate and dispense probe height and position; gentle or hard inlet flow; overflow wash; dispense-only; adjustable lift speedThree safety sensors: pressure pump level, plate in lift, waste reservoir space
Thermo Scientific Wellwash / Wellwash Versa (poster)96- and 384-wellNon-pressurized bottles (user supplies bottles); auto-prime and auto-rinseAdjustable aspiration height, speed and sweep aspiration mode; well offset 0.5-1.0 mm for flat-bottom wellsLiquid-level sensors, plate sensors, aerosol-resistant cover; special cell-washing head on Versa

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Plate Washer: What It Is, How It Works and How to Maintain It

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →