Pulp Tester: What It Is, How FDA Classifies It and What the IFU Says to Check

A pulp tester sends high frequency current to the dental pulp to help judge vitality. FDA class, product code EAT and IFU points to check.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

General product and regulatory information only; this is not medical or dental advice and does not describe how to test a patient.

Short answer

A pulp tester is a dental device used to help evaluate whether the pulp of a tooth is vital. The FDA regulation defines it as an AC or battery powered device that sends high frequency current through an electrode to stimulate nerve tissue in the dental pulp. It is a Class II device (product code EAT) that is exempt from 510(k) premarket notification, subject to the limitations in 21 CFR 872.9.

Definition and scope

The regulation, 21 CFR 872.1720, reads: “A pulp tester is an AC or battery powered device intended to evaluate the pulpal vitality of teeth by employing high frequency current transmitted by an electrode to stimulate the nerve tissue in the dental pulp.” The FDA lists the device name as “Tester, pulp” in its dental panel. The rule was first published at 52 FR 30097 (12 August 1987) and amended at 84 FR 71812 (30 December 2019).

Electric pulp tester

“Electric pulp tester” is the everyday name for the device the regulation describes: the definition itself is electrical, since it turns on high frequency current delivered by an electrode. Related tests that use cold or heat are a different kind of stimulus and are not what 21 CFR 872.1720 defines. For a cold-based approach, see CASRAI’s guide to Endo Ice and the dental cold spray.

How it is used in practice

Manufacturer instructions for use (IFU) set the intended use. The Parkell Digitest 3 IFU, for example, describes the unit as a diagnostic instrument intended to assist in determining the vitality of the dental pulp. The wording matters: the device assists a clinician’s determination and is not described as producing a diagnosis by itself. The same IFU states that the unit is contraindicated for use on a patient, or by an operator, wearing a cardiac pacemaker, and says not to modify the device. Step-by-step clinical technique belongs to the manufacturer’s IFU and professional training, and is not covered here.

Regulatory identity at a glance

Item What the source states Source
Regulation 21 CFR 872.1720 (pulp tester) eCFR
FDA product code EAT, device name “Tester, pulp” FDA Product Classification database
Device class Class II (special controls) 21 CFR 872.1720(b)
Premarket notification Exempt from 510(k), subject to the limitations in 872.9 21 CFR 872.1720(b)
Panel and review office Dental; Division of Dental and ENT Devices FDA Product Classification database
GMP exemption Not exempt FDA Product Classification database
Implanted / life-sustaining No / No FDA Product Classification database
Recognized standard listed ISO 7405:2018 (biocompatibility evaluation of dental devices) FDA Product Classification database

Because the class is II and the device is 510(k) exempt, a pulp tester does not need an individual FDA clearance letter to be marketed, though other requirements still apply. For how clearance works when a device is not exempt, see the guide to 510(k) and De Novo submissions.

Where it matters for clinics and supply decisions

The following checklist is drawn only from the regulation and from one manufacturer’s IFU and product listing (Parkell Digitest 3). It shows which specification points to look for in any model’s documentation; it does not rank products.

Point to find in the documentation What the Digitest 3 sources say
Intended use statement Diagnostic instrument to assist in determining pulp vitality (IFU)
Power source 9V battery supplied with the unit (retailer product listing of the manufacturer’s kit)
Probe tips Four autoclavable probe tips in the kit (retailer product listing)
Main unit and sterilization IFU states the unit must not be autoclaved; reprocessing instructions are in the IFU
Ground connection Ground wire with clip supplied (retailer product listing)
Stimulus settings Three stimulus rates (retailer product listing)
Contraindications Pacemaker: patient or operator (IFU)
Modification Do not modify the device (IFU)

The practical point is that parts of one kit can differ in how they are sterilized: probe tips and the main unit are not treated the same way. Reprocessing steps should always come from the IFU of the specific model. For general background on steam sterilization equipment, see what an autoclave is and types of autoclaves.

Common misunderstandings

  • “It diagnoses the tooth.” The IFU wording is that it assists in determining vitality; interpretation sits with the clinician.
  • “Class II means it was individually cleared.” This device type is exempt from 510(k) notification under 21 CFR 872.1720(b), subject to 872.9.
  • “Every pulp test is an electric pulp test.” The regulatory definition is specific to high frequency current stimulating nerve tissue; cold-based methods are separate.
  • “The whole kit can go in the autoclave.” For the model cited here, the IFU says the unit must not be autoclaved.

Frequently asked questions

What is a pulp tester?

Under 21 CFR 872.1720 it is an AC or battery powered device that uses high frequency current delivered by an electrode to stimulate nerve tissue in the dental pulp, intended to evaluate pulpal vitality of teeth.

Is an electric pulp tester the same thing?

Yes in everyday use: the regulatory definition is itself electrical, so the two names describe the same device type.

What FDA class is a pulp tester?

Class II (special controls), product code EAT, exempt from 510(k) premarket notification subject to the limitations in 21 CFR 872.9.

Can the whole kit be autoclaved?

Not necessarily. For the Digitest 3 cited here, the manual states the unit itself must not be autoclaved, while the retailer listing describes the probe tips as autoclavable. Always follow the IFU of the specific model.

Sources

All sources checked on 2026-10-31.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Pulp Tester: What It Is, How FDA Classifies It and What the IFU Says to Check

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →