Pyrolysis chromatography is an analytical hyphenated technique that uses controlled thermal energy in an oxygen-free, inert atmosphere to cleave non-volatile, high-molecular-weight polymers and insoluble organic solids into smaller, volatile fragments for gas chromatographic separation and mass spectrometric detection. By transforming recalcitrant synthetic resins, cross-linked rubbers, and complex macromolecules into reproducible chemical fingerprints called pyrograms, the method enables structural elucidation, copolymer profiling, and quantitative polymer mass determination without requiring hazardous solvent dissolution or complex chemical extraction workflows.

Definition and scope of analytical pyrolysis

Analytical pyrolysis is defined by the International Union of Pure and Applied Chemistry (IUPAC) as the characterization, in an inert atmosphere, of a material or a chemical process by chemical degradation reactions induced by thermal energy. Unlike bulk industrial pyrolysis, which processes metric tons of biomass, tires, or municipal waste to generate synthetic fuels, bio-oil, and carbon black, analytical pyrolysis operates on microgram quantities (typically 20 μg to 300 μg). The primary objective is the reproducible generation of volatile degradation products, termed the pyrolysate, which directly reflect the primary structure, tacticity, monomer sequence, and additive formulation of the precursor material.

Analytical pyrolysis is fundamentally distinct from thermal combustion. While combustion takes place in the presence of oxygen and leads to complete oxidative breakdown into carbon dioxide, water, and mineral ash, analytical pyrolysis is conducted in an ultra-pure inert carrier gas stream, such as helium, nitrogen, or hydrogen. In this non-oxidative environment, thermal energy cleaves specific covalent bonds based on their bond dissociation energies and steric arrangements, preserving structural fidelity across consecutive analyses.

The technique extends standard gas chromatography carrier gases and capillary columns beyond their conventional volatility limitations. Standard gas chromatography requires analytes to possess sufficient vapor pressure at temperatures below 350 °C without undergoing uncontrolled decomposition inside the injector port. Consequently, macromolecular substances such as polyolefins, polyurethanes, cured epoxy resins, vulcanized elastomers, and synthetic paints cannot be analyzed directly by liquid injection. While classical liquid separation techniques such as thin-layer chromatography plate methods or specialized HPLC column selection and stationary phases can separate soluble fractions, they cannot handle insoluble three-dimensional polymer networks. Pyrolysis chromatography bridges this analytical gap by converting solid, insoluble samples directly into chromatographic vapor pulses inside the instrument front-end.

How pyrolysis chromatography works in practice

In a standard pyrolysis-gas chromatography (Py-GC) or pyrolysis-gas chromatography-mass spectrometry (Py-GC/MS) workflow, a solid or viscous sample is weighed into an inert container—such as a deactivated stainless steel cup, a quartz sample tube, or a ferromagnetic foil. The sample is introduced into a pyrolyzer directly coupled to the injection port of a gas chromatograph. Upon rapid heating to temperatures between 450 °C and 800 °C, macromolecular chains undergo reproducible thermal bond cleavage.

The resulting volatile degradation fragments are swept instantaneously by the inert carrier gas into a heated transfer interface and through a split/splitless inlet. Inside the inlet, a split ratio ranging from 1:20 to 1:100 is typically applied to prevent capillary column overload, as the entire pyrolysate is generated in a concentrated vapor pulse lasting milliseconds to a few seconds. The fragments then enter a fused-silica capillary column coated with a bonded stationary phase (commonly a 5% diphenyl / 95% dimethylpolysiloxane phase such as DB-5ms or Ultra ALLOY-5), where they separate according to their boiling points and stationary-phase interactions before reaching a detector.

Primary thermal degradation mechanisms

The composition of the pyrolysate depends on the chemical structure of the polymer backbone and the thermal fragmentation mechanism it undergoes. In analytical polymer chemistry, three primary pyrolytic cleavage pathways predominate:

  • Random Chain Scission: Occurs when covalent bonds along the polymer backbone have similar dissociation energies and lack substituent groups that facilitate unzipping. Homolytic bond cleavage occurs at random positions along the carbon skeleton, generating radical fragments that undergo intramolecular hydrogen transfer and radical disproportionation (β-cleavage). In linear polyethylene (PE), random scission generates regular triplets of α,ω-alkadiene, 1-alkene, and n-alkane at each carbon number from C4 up to C35. Polypropylene (PP) similarly degrades through random scission with extensive branched β-cleavage, producing distinct multi-branched alkene patterns such as 2,4-dimethyl-1-heptene.
  • Depolymerization (Chain Unzipping): Involves initiation by thermal cleavage at chain ends or weak backbone links, followed by rapid, consecutive depropagation (unzipping) of monomeric units. Polymers with quaternary carbon backbones or sterically hindered substituents favor this pathway because the energetic barrier to depropagation is lower than that of transfer reactions. Poly(methyl methacrylate) (PMMA) undergoes nearly quantitative unzipping, regenerating greater than 90% pure methyl methacrylate monomer in its pyrogram. Polystyrene (PS) exhibits a dominant unzipping pathway that yields monomeric styrene (often exceeding 70% of total peak area), along with characteristic styrene dimer (1,3-diphenylpropane) and styrene trimer (2,4,6-triphenyl-1-hexene).
  • Side-Group Elimination (Stripping): Characteristic of polymers with substituent groups attached by bonds that are thermally more labile than the carbon-carbon backbone. At lower temperatures (200 °C to 350 °C), substituent groups cleave, releasing small volatile inorganic or organic molecules and leaving behind a conjugated, unsaturated polyene backbone. In poly(vinyl chloride) (PVC), dehydrochlorination readily occurs, releasing gaseous hydrochloric acid (HCl). The residual polyene backbone subsequently undergoes cyclization and secondary scission at higher temperatures, generating aromatic hydrocarbons including benzene, toluene, indene, and naphthalene. Similarly, poly(vinyl acetate) (PVAc) eliminates acetic acid, yielding an identical aromatic pyrogram fingerprint.

Reactive pyrolysis and in-situ derivatization

Certain condensation polymers, such as polyesters, polycarbonates, polyamides, and natural biomacromolecules like lignin and fatty acids, produce highly polar, non-volatile carboxylic acids and phenolic fragments upon conventional pyrolysis. These polar fragments cause severe peak tailing, irreversible adsorption, and memory effects on standard GC stationary phases. To overcome this limitation, analysts employ reactive pyrolysis (RxPY), most notably thermally assisted hydrolysis and methylation (THM-GC/MS).

In THM-GC/MS, a strong organic base, such as tetramethylammonium hydroxide (TMAH) or trimethylsulfonium hydroxide (TMSH), is co-injected directly into the sample cup with the solid analyte. Under rapid heating (typically 300 °C to 500 °C), TMAH induces simultaneous ester/amide bond saponification and quantitative methylation of the resulting carboxyl and hydroxyl groups. Polyethylene terephthalate (PET), for instance, is cleanly converted into dimethyl terephthalate (DMT), yielding sharp, symmetrical chromatographic peaks that allow accurate stoichiometric quantification.

Pyrolyzer instrument types and design comparison

Commercial analytical pyrolyzers are categorized into three distinct technical architectures: vertical micro-furnace systems, resistively heated platinum filament systems, and Curie-point induction systems. Each configuration exhibits specific thermal dynamics, sample handling characteristics, and operational advantages across laboratory environments.

Instrumentation Type Heating Mechanism & Rate Operating Temperature Range Sample Introduction Format Multi-Step & EGA Capability Primary Analytical Strengths
Vertical Micro-Furnace (e.g., Frontier EGA/PY-3030D) Continuous ceramic furnace with low thermal mass; programmable ramps up to 600 °C/min or isothermal drops Ambient +10 °C to 1050 °C (±0.1 °C stability) Deactivated stainless steel cups dropped vertically by gravity Full evolved gas analysis (EGA), double-shot, and heart-cut GC/MS Exceptional thermal reproducibility; handles solids, liquids, and waxes; fully automated autosamplers available
Resistively Heated Filament (e.g., CDS Pyroprobe 6150/6200) Platinum ribbon or coil heated resistively; pulse rates up to 20,000 °C/s Ambient to 1300 °C (dynamic micro-sensor control) Quartz tubes with quartz wool plugs or drop-in sample chambers (DISC) Multi-step thermal desorption (up to 10 steps); optional cryotrap concentration Ultra-fast thermal pulse prevents secondary reactions; flexible pyrolysis heating ramps; ideal for micro-solids
Curie-Point Induction (e.g., JAI JCI-77, GSG Pyromat) High-frequency RF induction coil heating a ferromagnetic alloy wire or foil Fixed Curie points: 160 °C to 1040 °C (e.g., 358 °C, 485 °C, 590 °C, 770 °C) Wrapped directly inside specialized ferromagnetic pyrofoils Limited to stepped foils; no continuous linear thermal programming Absolute temperature reproducibility governed by metallurgy; zero thermocouple drift; simple modular bench installation

Vertical micro-furnace pyrolyzers are widely deployed in industrial research because they support four operational modes on a single sample: Evolved Gas Analysis (EGA-MS), single-shot flash pyrolysis, double-shot thermal desorption followed by pyrolysis, and heart-cut two-dimensional GC/MS. In EGA-MS, a deactivated transfer tube connects the furnace directly to the mass spectrometer while the furnace temperature ramps continuously from 40 °C to 700 °C. The resulting thermogram maps out the desorption temperatures of volatile additives, plasticizers, and the decomposition window of the polymer matrix, providing the exact temperature parameters required for subsequent double-shot chromatographic separation.

Filament pyrolyzers excel in applications requiring extreme thermal ramp rates up to 20,000 °C per second. Rapid pulse heating ensures that the sample reaches the target temperature almost instantaneously, minimizing the duration spent at intermediate temperatures where unwanted secondary condensation reactions can occur. Modern Curie-point systems utilize precision-engineered ferromagnetic alloy foils; when subjected to an RF electromagnetic field, induction heats the alloy until it reaches its Curie point (Tc), at which point ferromagnetism ceases and heating halts abruptly within 0.2 seconds. This metallurgical physics mechanism eliminates thermocouple calibration errors entirely across multi-laboratory round-robin studies.

Where pyrolysis chromatography matters in laboratory workflows

Pyrolysis chromatography is a core analytical methodology across materials science, forensic science, environmental monitoring, and industrial quality control. By generating reproducible fragmentation patterns from insoluble samples, it delivers structural insights that cannot be acquired through conventional spectroscopy or wet chemistry alone. Integrating Py-GC/MS data into a formal Certificate of Analysis (CoA) verification protocol enables polymer manufacturers and converters to confirm batch-to-batch compositional fidelity and detect unapproved resin substitutions.

Rubber and elastomer identification under consensus standards

In the tire and elastomeric goods manufacturing sector, identifying the exact polymer backbone within vulcanized, carbon-black-filled rubber compounds is difficult because cured elastomers cannot be dissolved in organic solvents without breaking chemical cross-links. Two international consensus standards define Py-GC methodology for this purpose:

  • ASTM D3452 (Standard Practice for Rubber—Identification by Pyrolysis-Gas Chromatography): Provides standardized procedures for identifying raw, cured, and uncured rubbers. Thermal decomposition at 550 °C cleaves elastomer networks into diagnostic monomeric and dimeric markers. Natural rubber (polyisoprene, NR) is identified by high yields of isoprene monomer (m/z 68) and dipentene/dl-limonene (m/z 136). Styrene-butadiene rubber (SBR) is identified by styrene (m/z 104) and 4-phenylcyclohexene (m/z 158), while polybutadiene rubber (BR) yields 4-vinylcyclohexene (m/z 54, 79).
  • ISO 17257:2020 (Rubber — Identification of polymers — Pyrolytic gas-chromatographic method using mass-spectrometric detection): Expands upon standard flame ionization detection by incorporating mass spectrometric library matching, allowing unambiguous identification of complex multi-component elastomer blends, such as NR/SBR/BR blends used in modern tire treads, even in the presence of sulfur cross-linkers, stearic acid, and process oils.

Forensic trace examination under ASTM E3296

Forensic laboratories frequently analyze microscopic traces of synthetic polymers recovered from crime scenes, including automotive paint chips, architectural coatings, synthetic fibers, adhesive tapes, and electrical insulation. Under ASTM E3296 (Standard Guide for Using Pyrolysis Gas Chromatography and Pyrolysis Gas Chromatography-Mass Spectrometry in Forensic Polymer Examinations), Py-GC/MS serves as an essential confirmatory technique following preliminary optical microscopy and Fourier transform infrared (FTIR) spectroscopy.

Because multi-layered automotive coatings share similar bulk binder compositions (such as acrylic-melamine or alkyd formulations), FTIR spectra of two distinct paints can appear virtually indistinguishable. Py-GC/MS provides sufficient resolution to separate minor cross-linking agents, monomer ratios (e.g., butyl acrylate vs methyl methacrylate ratios), and proprietary additive chemistry, providing rigorous comparative evidence suitable for judicial presentation.

Environmental microplastics mass quantification under ISO 24187

The detection and quantification of microplastics in water supplies, marine sediments, wastewater effluent, and biological tissues represents one of the fastest-growing applications of Py-GC/MS. The overarching international framework, ISO 24187:2023 (Plastics — Determination of microplastics in environmental matrices), formalizes the role of thermal degradation techniques alongside vibrational spectroscopy.

While micro-FTIR and Raman microscopy provide optical particle counts, particle size distributions, and physical dimensions, they encounter significant limitations when particle sizes drop below 10 μm or when particles are heavily pigmented with carbon black. Furthermore, optical particle counts cannot be readily converted into total environmental mass concentrations. Py-GC/MS functions as a destructive, mass-based quantitative technique: environmental sample filters are pyrolyzed directly, and specific marker pyrolyzates are quantified against matrix-matched calibration curves. Polyethylene (PE) is quantified via 1-alkene peaks, polypropylene (PP) via 2,4-dimethyl-1-heptene, polystyrene (PS) via styrene monomer, and poly(vinyl chloride) (PVC) via benzene, directly reporting environmental polymer mass loading in milligrams per kilogram or micrograms per liter.

As explored across the foundational principles of analytical chemistry, selecting between mass-based thermal quantification and particle-based optical enumeration depends entirely on whether the regulatory inquiry concerns toxicological mass dosage or particulate transport dynamics.

Worked example: Polyethylene vs polystyrene differentiation and degradation pathways

To illustrate the practical interpretation of pyrograms, consider a comparative laboratory experiment evaluating two widely utilized commodity thermoplastics: low-density polyethylene (LDPE) and atactic polystyrene (PS). Both samples were analyzed as 100 μg solid micro-pellets using a vertical micro-furnace pyrolyzer coupled to a GC-MS system under identical operational parameters.

The analytical conditions were configured as follows: flash pyrolysis temperature of 600 °C (held for 12 seconds); pyrolyzer-GC interface temperature of 300 °C; split/splitless inlet at 300 °C with a 1:50 split ratio; ultra-high purity helium carrier gas at a constant flow of 1.0 mL/min; separation on a 30 m × 0.25 mm ID × 0.25 μm df 5% diphenyl / 95% dimethylpolysiloxane capillary column; GC oven programmed from 40 °C (held for 2 minutes) ramped at 10 °C/min to 320 °C (held for 5 minutes); mass spectrometer operating in Electron Ionization (EI, 70 eV) full scan mode from m/z 29 to 550.

Sample 1: Low-Density Polyethylene (LDPE) degradation

Upon exposure to the 600 °C furnace zone, LDPE undergoes extensive thermal degradation via random main-chain scission. Because the linear aliphatic backbone consists entirely of repeating methylene units (-CH2-CH2-) with identical C-C bond energies (approximately 348 kJ/mol), homolytic cleavage occurs stochastically along the polymer chain, generating primary alkyl radical fragments. These radicals undergo 1,5-intramolecular radical transfer (back-biting) and β-scission, yielding a highly structured, repeating chromatographic series of hydrocarbon triplets from C6 up to C30.

At each carbon number n, three distinct peaks elute in precise chromatographic order:

  1. α,ω-Alkadiene: The earliest eluting peak of each triplet, containing terminal double bonds at both ends (e.g., 1,13-tetradecadiene for n=14, base ion m/z 55, 69).
  2. 1-Alkene (α-Olefin): The dominant peak of the triplet, formed by β-cleavage following hydrogen transfer (e.g., 1-tetradecene for n=14, base ion m/z 43, 57). This compound serves as the primary quantifier marker for polyethylene.
  3. n-Alkane: The latest eluting peak of each triplet, representing the fully saturated straight-chain alkane (e.g., n-tetradecane for n=14, base ion m/z 57, 71, molecular ion m/z 198).

This repeating triplet profile forms a distinctive ‘picket-fence’ pyrogram that uniquely identifies polyethylene and allows analysts to assess branching density by inspecting minor iso-alkene peaks between the primary triplets.

Sample 2: Atactic Polystyrene (PS) degradation

In stark contrast to polyethylene, polystyrene undergoes thermal degradation primarily through end-chain radical initiation followed by rapid, continuous unzipping (depolymerization). The presence of bulky, electron-rich phenyl ring substituents substantially stabilizes the radical intermediates formed upon homolytic cleavage of the main chain. Depropagation proceeds with a low activation energy, systematically cleaving monomer units off the radical chain ends.

The resulting pyrogram of polystyrene exhibits an exceptionally simple chromatographic fingerprint dominated by three peaks:

  • Styrene Monomer: Eluting at a retention time of approximately 7.2 minutes under the specified oven ramp, styrene accounts for approximately 75% to 85% of the total integrated peak area. The mass spectrum displays an intense molecular ion at m/z 104 and a tropylium-derived fragment ion at m/z 78.
  • Styrene Dimer (1,3-Diphenylpropane): Formed via 1,3-intramolecular radical transfer (back-biting) followed by chain cleavage, eluting as a minor peak with characteristic mass fragments at m/z 91 and 104.
  • Styrene Trimer (2,4,6-Triphenyl-1-hexene): Generated via 1,5-back-biting radical transfer, eluting late in the chromatographic run at approximately 24.5 minutes, exhibiting diagnostic fragments at m/z 91, 117, and a molecular ion at m/z 312.

Comparing these two worked profiles demonstrates the power of analytical pyrolysis: LDPE yields over sixty individual hydrocarbon peaks across a wide boiling-point spectrum through random scission, whereas PS concentrates its pyrolyzate mass almost entirely into three distinct aromatic structures through radical unzipping. The data table below compares the thermal degradation mechanisms, primary pyrolyzates, characteristic quantifier ions, and optimal furnace temperatures for common synthetic polymers and elastomers.

Common misunderstandings and analytical pitfalls

Despite its analytical power and minimal sample preparation requirements, pyrolysis chromatography requires strict experimental control. Misinterpreting the fundamental physical chemistry of pyrolysis or mismanaging instrumental parameters introduces severe analytical artifacts.

Conflating pyrolysis with oxidative combustion

A frequent misconception is that pyrolysis is synonymous with burning or incinerating a sample. Combustion is an exothermic reaction between organic matter and oxygen that oxidizes carbon and hydrogen into CO2, H2O, and inorganic ash, entirely destroying original molecular structure. Analytical pyrolysis is an endothermic thermal fragmentation conducted in an inert, oxygen-free carrier gas. The presence of even trace atmospheric oxygen—introduced through carrier gas contamination, compromised gas purifiers, or leaking inlet seals—alters the degradation mechanism, producing aldehydes, ketones, and carboxylic acids that obscure native polymer fingerprints and rapidly degrade capillary column stationary phases.

Assuming pyrolysis produces non-reproducible chaos

Another persistent myth is that shattering macromolecules at 600 °C produces a random, uncontrollable mixture of fragments. In reality, thermal cleavage is governed strictly by chemical kinetics and thermodynamics. Covalent bonds cleave at precise energy thresholds, and radical intermediates rearrange along predictable, reproducible reaction pathways. When sample mass, heating rate, peak temperature, and carrier gas velocity are held constant, Py-GC pyrograms achieve retention-time relative standard deviations (RSD) below 0.5% and peak area relative standard deviations below 3.0%, comparable to standard liquid injection.

Sample mass overloading and secondary reactions

The most common operator error in analytical pyrolysis is introducing excessive sample mass into the pyrolyzer. While analysts accustomed to gravimetric tools might consider 1 mg to 5 mg a ‘micro’ sample, in analytical pyrolysis an optimal sample mass is between 20 μg and 200 μg. Introducing several milligrams of polymer introduces two critical defects:

  • Internal Temperature Gradients: Polymers possess low thermal conductivities. A large sample mass heats unevenly from the outside inward, resulting in slow, non-isothermal degradation that generates broad, distorted pyrogram peaks.
  • Secondary Bimolecular Reactions: In a dense sample bed, primary volatile fragments cannot escape instantaneously into the carrier gas stream. Trapped primary radicals collide with surrounding matrix molecules, undergoing unwanted secondary radical combinations, condensation, and aromatization to form refractory carbonaceous char and polycyclic aromatic hydrocarbons (PAHs) that were never present in the original sample.

Transfer interface cold spots and memory effects

Pyrolysis products span a wide molecular weight range, often including high-boiling dimers, trimers, and oligomers. The heated transfer line connecting the pyrolyzer furnace to the GC inlet must be maintained at an elevated temperature (typically 280 °C to 320 °C) with complete thermal uniformity. Any unheated zone, cold spot, or unpurged dead volume causes heavy pyrolyzates to condense along the flow path. This results in severe chromatographic peak tailing, irreversible loss of high-mass diagnostic markers, and persistent carryover or ‘ghost peaks’ in subsequent analytical runs.

Confusing microplastic mass concentration with particle counts

In environmental microplastics analysis, laboratories must distinguish between particle counting and mass determination. Because Py-GC/MS thermal degradation completely destroys the physical integrity of microplastic particles, it cannot provide information on particle morphology, count, aspect ratio, color, or size distribution. Attempting to report microplastic particle concentrations from Py-GC/MS data without complementary optical microscopy constitutes a fundamental methodological error. Py-GC/MS specifically determines the total mass fraction of polymer types, providing invaluable bulk mass loading data that complements optical particle enumeration.

Sources

  • ASTM International. ASTM D3452 – Standard Practice for Rubber—Identification by Pyrolysis-Gas Chromatography. https://www.astm.org/d3452-06r21.html (accessed 2026-11-15).
  • ASTM International. ASTM E3296 – Standard Guide for Using Pyrolysis Gas Chromatography and Pyrolysis Gas Chromatography-Mass Spectrometry in Forensic Polymer Examinations. https://www.astm.org/e3296-22.html (accessed 2026-11-15).
  • International Organization for Standardization. ISO 17257:2020 – Rubber — Identification of polymers — Pyrolytic gas-chromatographic method using mass-spectrometric detection. https://www.iso.org/standard/71616.html (accessed 2026-11-15).
  • International Organization for Standardization. ISO 24187:2023 – Principles for the determination of microplastics in environmental matrices using pyrolysis-gas chromatography-mass spectrometry. https://www.iso.org/standard/78114.html (accessed 2026-11-15).
  • Frontier Laboratories Ltd. Multi-Shot Pyrolyzer EGA/PY-3030D Operation and Technical Manual. https://www.frontier-lab.com (accessed 2026-11-15).
  • CDS Analytical LLC. Pyroprobe 6000 Series (Models 6150 and 6200) Pyrolysis Systems Manual. https://www.cdsanalytical.com (accessed 2026-11-15).
  • Japan Analytical Industry Co., Ltd. (JAI). Curie-Point Pyrolyzer JCI-77 and Pyrofoil Operating Specifications. https://www.jai.co.jp (accessed 2026-11-15).
  • International Union of Pure and Applied Chemistry (IUPAC). Nomenclature and Terminology for Analytical Pyrolysis (IUPAC Recommendations 1993). Pure and Applied Chemistry, Vol. 65, No. 11, pp. 2405–2412. https://doi.org/10.1351/pac199365112405 (accessed 2026-11-15).

Data: Analytical Pyrolysis Parameters and Characteristic Pyrolyzate Markers for Common Polymers

Comprehensive comparison of thermal degradation mechanisms, optimal flash pyrolysis temperatures, primary characteristic pyrolyzate markers, and quantifier ions for key commercial polymers and elastomers analyzed via Py-GC/MS, compiled from ASTM D3452, ASTM E3296, ISO 17257, and instrument manufacturer application libraries.

Analytical Pyrolysis Parameters and Characteristic Pyrolyzate Markers for Common Polymers
Polymer or ElastomerPrimary Degradation MechanismOptimal Pyrolysis Temperature (°C)Primary Pyrolyzate MarkerQuantifier Target Ion (m/z)Relevant Standard Method
Low-Density Polyethylene (LDPE)Random chain scission600C10–C20 Alkadiene / 1-Alkene / n-Alkane triplets57ASTM E3296 / ISO 24187
Polypropylene (PP)Random chain scission with β-cleavage6002,4-Dimethyl-1-heptene and branched oligomers70ASTM E3296 / ISO 24187
Polystyrene (PS)Depolymerization / unzipping550Styrene monomer, dimer (1,3-diphenylpropane), trimer104ASTM E3296 / ISO 24187
Poly(methyl methacrylate) (PMMA)Depolymerization / unzipping500Methyl methacrylate monomer100ASTM E3296
Poly(vinyl chloride) (PVC)Side-group elimination (dehydrochlorination) & cyclization600Benzene, toluene, naphthalene (post-HCl stripping)78ASTM E3296 / ISO 24187
Polyethylene terephthalate (PET)Ester cleavage & decarboxylation600Divinyl terephthalate, benzoic acid, vinyl benzoate122ISO 24187
Natural Rubber (Polyisoprene / NR)Depolymerization & cyclization550Isoprene, dipentene (dl-limonene)68ASTM D3452 / ISO 17257
Styrene-Butadiene Rubber (SBR)Simultaneous scission & cyclization550Styrene, 4-phenylcyclohexene, 4-vinylcyclohexene104ASTM D3452 / ISO 17257
Polybutadiene Rubber (BR)Main chain scission & cyclization5504-Vinylcyclohexene, 1,3-butadiene54ASTM D3452 / ISO 17257
Polychloroprene (Neoprene / CR)Side-group dehydrochlorination & scission550Chloroprene monomer, 1-chlorocyclohexene88ASTM D3452 / ISO 17257

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Pyrolysis Chromatography: Principles, Instrumentation & Lab Guide

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →