RNeasy is QIAGEN’s family of silica-membrane spin-column kits for purifying total RNA: the sample is lysed in a guanidine-isothiocyanate buffer, ethanol is added, the lysate is bound to the column membrane, contaminants are washed away and the RNA is eluted in water. The kits differ mainly in the sample they are built for (cells, tissue, fatty tissue, FFPE sections), the scale (Micro to Maxi, or 96-well) and whether genomic DNA is removed by a column (Plus kits) or by an on-column DNase step.
What RNeasy is
RNeasy is a brand family rather than a single product. According to QIAGEN’s RNeasy kits page (read 2026-11-19), the core chemistry is the same across the family: lysis in Buffer RLT, homogenization, addition of ethanol, binding to a silica membrane in a spin column, wash steps, an optional on-column DNase digestion, and elution in RNase-free water. What changes from kit to kit is the sample type, the amount of starting material the column is sized for and the lysis or clean-up add-ons.
For a generic explanation of what is inside any RNA extraction kit, and the kit-size table for Mini, Midi and Maxi, see What is an RNA extraction kit. This page covers what that guide does not: why each RNeasy buffer is there, which variant fits which sample, and how long the kits and their buffers keep.
How does RNeasy work?
Each step of the workflow has a stated purpose. The table below summarizes what QIAGEN’s kit page says about each buffer and step.
| Step or buffer | What it does according to QIAGEN |
|---|---|
| Buffer RLT (lysis) | Guanidine-isothiocyanate lysis buffer. Beta-mercaptoethanol is added to it before use. |
| Homogenization | Sample is homogenized in the lysis buffer before binding. |
| Ethanol addition | Added to the lysate before it is loaded on the column. |
| Binding | Lysate is loaded on the spin column; RNA binds to the membrane. |
| Buffer RW1 (stringent wash) | Removes carbohydrates, proteins and fatty acids. RNAs longer than about 200 nucleotides stay bound. |
| Buffer RPE (mild wash) | Mild wash. Ethanol must be added to the buffer before first use. |
| On-column DNase (optional) | Buffer RDD with the RNase-Free DNase Set digests genomic DNA bound to the membrane. |
| Elution | RNA is released from the membrane in RNase-free water. |
What the 200-nucleotide cutoff means for small RNA
The statement that RNA larger than about 200 nucleotides remains bound after the stringent wash is the single most useful fact for deciding whether an RNeasy kit suits an experiment. Total RNA purified this way is mainly long RNA: mRNA and ribosomal RNA. If the work depends on short RNAs such as microRNAs, QIAGEN’s description of the standard purification is in terms of the larger fraction, so check the kit’s own documentation for small-RNA recovery before relying on it.
The Plus kits push the same logic further. QIAGEN’s RNeasy Plus page (read 2026-11-19) says RNAs shorter than 200 nucleotides, such as rRNA fragments and tRNA, are largely excluded, which enriches for mRNA.
Why the chemistry binds RNA and not the rest
RNeasy is one implementation of a general spin-column principle. Thermo Fisher’s spin column purification page (read 2026-11-19) explains that nucleic acids bind silica in the presence of chaotropic agents, and gives guanidine isothiocyanate as a common example in RNA lysis buffers. Under those conditions nucleic acids bind tightly to the silica membrane while proteins and other contaminants do not, so centrifuging the column washes unbound contaminants away. Further centrifugation with an elution buffer, typically RNase-free water or a low-salt buffer for RNA, releases the purified RNA into a collection tube. The same page describes DNase treatment as the step that removes contaminating DNA to help ensure RNA purity, which is the role the RDD and DNase set play in an RNeasy workflow.
Seen this way, the buffers in the table are an ordered sequence: lyse and denature, bind under chaotropic and ethanol conditions, wash twice at different stringencies, optionally digest DNA, then elute. Because each step builds on the one before, a preparation error such as ethanol not added to Buffer RPE before first use affects the run from that step onward, which is why QIAGEN’s kit instructions call out those preparation steps.
The FFPE variant as a worked case of chemistry changing with the sample
The RNeasy FFPE kit page (read 2026-11-19) shows how the family adapts to a harder sample. It is specified for sections from one 5 µm section up to four 10 µm sections of FFPE tissue, with an elution volume of 14-30 µl. Special lysis and incubation conditions reverse formaldehyde modification of the RNA: samples are lysed with proteinase K for about 15 minutes and then incubated at 80 °C for 15 minutes. DNase treatment uses RNase-Free DNase I with DNase Booster Buffer, purification uses silica technology on RNeasy MinElute spin columns, and deparaffinization solution is listed as a separate purchase rather than part of the kit. For the MinElute columns, QIAGEN’s FAQ on that page says they should be stored at 2-8 °C on arrival, and that up to 4 weeks at room temperature (15-25 °C) does not affect performance.
The FFPE case separates what is common to the family (silica column, wash, elute) from what is sample-specific (the lysis and crosslink reversal). Reading the FFPE page alongside the Mini page is a quick way to see which parts of the family are shared and which are sample-specific.
Which RNeasy kit fits which sample?
The data table below this guide compares the main variants by sample type, capacity and elution volume. The notes here explain when each applies, as stated on QIAGEN’s pages.
Scale: Micro, Mini, Midi, Maxi and 96
RNeasy Micro is for very small samples: less than 5 mg of tissue or fewer than 5 x 10^5 cells, with a 45 µg binding capacity and a small elution volume of 10-14 µl on a MinElute column. RNeasy Mini handles 0.5-30 mg of tissue or 10 to 1 x 10^7 cells, with a 100 µg capacity. Midi and Maxi are the larger-scale columns, for 20-250 mg and 150 mg-1 g of tissue respectively. The RNeasy 96 format moves the same chemistry to a plate for 10 to 5 x 10^5 cultured cells per well, using vacuum or centrifugation.
The practical rule that follows from these stated ranges is to choose the column by the amount of sample you actually have, because each size is specified for its own input range and binding capacity.
Sample type: lipid tissue, FFPE and microarray tissue
QIAGEN lists dedicated variants where the default lysis is not enough. RNeasy Lipid Tissue Mini and Midi are for fatty tissue, up to 100 mg and 500 mg respectively. The RNeasy FFPE kit is for microdissected formalin-fixed, paraffin-embedded tissue. RNeasy Microarray Tissue Mini uses QIAzol lysis and then a spin column. QIAwave versions of the Mini kit use concentrated buffers and eco-oriented packaging with the same chemistry.
Genomic DNA removal: on-column DNase or Plus kits
There are two routes to DNA-free RNA in the family. The standard kits offer an optional on-column DNase digestion (Buffer RDD plus the RNase-Free DNase Set). The RNeasy Plus kits instead pass the lysate, in Buffer RLT Plus, through a gDNA Eliminator column, where high salt binds genomic DNA, and only then add ethanol and load the RNeasy column. Plus Micro is rated for up to 5 x 10^5 cells or 5 mg tissue and up to 45 µg RNA; Plus Mini for up to 10^7 cells or 30 mg tissue and up to 100 µg; Plus 96 is the plate format; and Plus Universal is for fatty or otherwise difficult tissue, using QIAzol and a gDNA eliminator solution without a column.
Which route to choose depends on the downstream application and the sample, not on a universal ranking. The Plus approach enriches long RNA and avoids an enzyme step; the DNase route keeps the standard column chemistry. If the later assay is sensitive to residual genomic DNA, such as RT-PCR, no-RT controls remain the way to check whichever route you used.
How long do RNeasy kits last?
QIAGEN states that the Mini, Midi and Maxi kits are stored dry at room temperature, 15-25 °C, and are stable for at least 9 months under those conditions. Buffer RLT is stable for at least 9 months at room temperature, but once beta-mercaptoethanol is added it is stable for 1 month. In practice this makes the beta-mercaptoethanol addition a per-batch decision: add it to the volume you expect to use, not to the whole bottle.
The MinElute columns that come with the RNeasy Micro kit are stored differently: at 2-8 °C, with up to 4 weeks at room temperature acceptable and no storage above 25 °C. For the Plus kits, the product page does not state a shelf life, so use the expiry and storage information printed on the kit and in its handbook.
How much does an RNeasy kit cost?
This page does not list prices. Price depends on the variant, the number of preparations in the kit, the supplier and any institutional contract, so a quote from the supplier or a purchasing portal is the reliable source. What can be compared from published documentation is the number and size of preparations each kit is built for, and the sample amount per column, as in the data table below.
Which RNA isolation kit is the best?
No kit is best in general, and this page makes no such ranking. The defensible way to choose within the RNeasy family is to match the kit to the sample: cultured cells or small tissue amounts to Mini or Micro, fatty tissue to a Lipid Tissue or Plus Universal variant, fixed archival sections to FFPE, high-throughput plates to the 96 format. Then decide on genomic DNA removal and check the result.
Common misunderstandings
- “RNeasy” means one kit. It is a family; the Plus, Lipid, FFPE and Micro variants differ in sample, scale and chemistry.
- The kit removes all DNA automatically. Only the optional on-column DNase step or the Plus gDNA Eliminator column addresses genomic DNA.
- Total RNA includes everything. The 200-nucleotide behaviour of the wash and the Plus kits means short RNAs are not the target.
- A purified sample is a good sample. Yield and purity do not show integrity; see RNA integrity number (RIN) for how integrity is measured.
- RNase is only a kit issue. Eluted RNA still needs RNase-free handling; see RNase inhibitors.
Preparations can also be run on instruments: the Mini kit is described as usable manually or on the QIAcube Connect, covered in QIAcube. The purified RNA then feeds downstream work such as RNA-seq.
Frequently asked questions
Which RNA isolation kit is the best?
There is no single best kit. Within RNeasy, the right one is set by sample type, amount of starting material and whether genomic DNA must be removed, as laid out in the variant section above.
How does RNeasy work?
Cells or tissue are lysed in guanidine-isothiocyanate Buffer RLT, ethanol is added, and the lysate is loaded on a silica membrane. Washes remove proteins, carbohydrates and fatty acids, an optional DNase step removes genomic DNA, and the RNA is eluted in water.
How much does a Qiagen RNA extraction kit cost?
Price varies by variant, kit size and supplier, and is not stated here. Request a quote from the supplier.
How long do RNeasy kits last?
QIAGEN states Mini, Midi and Maxi kits are stable for at least 9 months stored dry at 15-25 °C. Buffer RLT with beta-mercaptoethanol added is stable for 1 month. MinElute columns are stored at 2-8 °C.
Sources
- QIAGEN, RNeasy kits (total RNA purification), read 2026-11-19.
- QIAGEN, RNeasy Plus kits, read 2026-11-19.
- QIAGEN, RNeasy FFPE Kit, read 2026-11-19.
- Thermo Fisher Scientific, Spin column purification, read 2026-11-19.
Data: RNeasy kit variants: sample type, capacity and elution as stated on QIAGEN product pages
Original compilation of the sample range, RNA binding capacity, elution volume and genomic DNA handling that QIAGEN's RNeasy kits page and RNeasy Plus kits page state for the main variants, compiled for comparison.
Period covered: 2026.
| Kit variant | Sample input | Binding capacity (µg) | Elution volume (µl) | Genomic DNA handling |
|---|---|---|---|---|
| RNeasy Micro | 45 | 10-14 | Optional on-column DNase | |
| RNeasy Mini | 0.5-30 mg tissue or 10 to 1 x 10^7 cells | 100 | 30-100 | Optional on-column DNase |
| RNeasy Midi | 20-250 mg tissue | 1000 | Not stated | Optional on-column DNase |
| RNeasy Maxi | 150 mg-1 g tissue | 6000 | Not stated | Optional on-column DNase |
| RNeasy 96 | 10 to 5 x 10^5 cultured cells | Not stated | 45-140 | Optional on-column DNase |
| RNeasy Plus Micro | Up to 5 x 10^5 cells or 5 mg tissue | 45 | 14 | gDNA Eliminator column |
| RNeasy Plus Mini | Up to 10^7 cells or 30 mg tissue | 100 | 30-50 | gDNA Eliminator column |
Sources
- QIAGEN RNeasy kits (accessed 2026-11-19)
- QIAGEN RNeasy Plus kits (accessed 2026-11-19)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.