SpectraMax is the name Molecular Devices uses for its family of microplate readers, which measure light absorbance, fluorescence, luminescence and related signals from samples held in multi-well plates. The family runs from the SpectraMax Plus 384, an absorbance-only microplate spectrophotometer with a built-in cuvette port, to the SpectraMax iD5e, a five-mode reader that combines monochromator and filter optics. BMG Labtech’s CLARIOstar Plus is a competing multi-mode reader built around a different optical design. All of these are described by their manufacturers as laboratory research instruments, and nothing here is clinical or diagnostic guidance.
What SpectraMax refers to
“SpectraMax” is a product line, not one instrument. Searching the bare word also returns unrelated trademarks and software, so this page covers only the Molecular Devices microplate readers used in research laboratories and core facilities. The line is sold under several model names; the manufacturer documents read for this page cover the Plus 384, the iD5e and the older Paradigm, and the rest of the family follows the same pattern of an absorbance-only reader at one end and a multi-mode reader at the other. If you are new to the instrument class, start with the general explainer on what a plate reader is, which covers the detection modes in plain terms.
SpectraMax Plus 384: absorbance with a cuvette port
Molecular Devices describes the Plus 384 as a microplate spectrophotometer with a built-in cuvette port. It reads one sample or up to 384 samples, and accepts any standard cuvette, a 12 x 75 mm tube, or a 96-well or 384-well plate. The datasheet lists a wavelength range of 190 to 1000 nm selected by monochromator in 1.0 nm increments, a 2 nm bandwidth, a wavelength accuracy of plus or minus 1.0 nm, a photometric range of 0 to 4.000 OD, and a xenon flash lamp. Read time for a full plate is given as 9 seconds in normal mode and 5 seconds in speed-read mode for 96 wells, and 29 seconds and 16 seconds for 384 wells. Operating temperature is ambient plus 4 to 45 degrees C, and the instrument weighs 13.6 kg.
The datasheet names DNA quantitation, colorimetric assays, ELISA and EIA, immunoassays, dissolution profiles and enzyme kinetics as applications. The cuvette port is the practical point of difference from a plate-only reader: the same instrument can read a cuvette sample, which is why the Plus 384 overlaps with the territory of a conventional UV-Vis spectrophotometer. For nucleic acid work specifically, see the CASRAI guide on using a spectrophotometer for DNA concentration.
How PathCheck pathlength correction works
Absorbance in a cuvette is read over a fixed 1 cm path. In a well, the path depends on the liquid volume, so the same solution gives different readings at different fill volumes. The Plus 384 datasheet describes a PathCheck sensor, which Molecular Devices calls the only patented technology for measuring the optical pathlength of a sample in a microplate. Used with SoftMax Pro software, it normalizes readings to the equivalent of a 1 cm cuvette. The datasheet gives a working range of 50 to 300 microlitres in a 96-well plate and 20 to 100 microlitres in a 384-well plate.
The datasheet’s own worked example makes the effect concrete: a 100 microlitre sample, with a pathlength of roughly 0.3 cm, reads 2.8 OD, and the corrected value is about 9.2 OD, the value that a 1 cm cuvette would give. Two consequences follow, both stated by the manufacturer. First, where the extinction coefficient of the analyte is known, the correction can remove the need for a standard curve. Second, because the sensor measures path, it can be used to check pipettor or dispenser volumes. Note that these are vendor statements about the vendor’s own feature; a laboratory relying on pathlength-corrected values should confirm the correction against its own assay in the same way it would validate any other method.
SpectraMax iD5e: five detection modes in one reader
The iD5e brochure describes a multi-mode reader with five modes: absorbance, fluorescence, luminescence, time-resolved fluorescence (TRF) and tunable fluorescence polarization. It uses hybrid optics, a monochromator together with filters, and an ultra-cooled photomultiplier tube (PMT). Other documented features include a touchscreen, a temperature range from ambient to 66 degrees C, more than 200 preconfigured protocols, and near-field communication that identifies the installed filter set.
The brochure lists these options: bottom-read luminescence, TR-FRET including HTRF, a western blot module, dual injectors (SmartInject, with a 10 microlitre dead volume), a CO2 and O2 gas mixer with advanced shaking, the SpectraDrop micro-volume plate for 2 to 4 microlitre samples, and a SpectraCuvette adapter plate. For near-infrared work the brochure claims its filters are 10 to 40 times more sensitive than conventional monochromator-based detection; that is a manufacturer claim and has not been independently checked here. The iD5e measures 40.1 x 53.2 x 59.8 cm and weighs 40 kg.
SpectraMax Paradigm: the user-upgradeable platform
An older Molecular Devices datasheet (copyright 2013) describes the SpectraMax Paradigm as a user-upgradeable multi-mode detection platform. It supports plate formats from 6 to 1536 wells, uses an LED or xenon flash lamp, takes detection cartridges, offers injectors as an option, operates at ambient plus 4 to 45 degrees C and weighs 45 kg. Because the datasheet is from 2013, check the current Molecular Devices catalogue before assuming the model is still sold in this form.
How SpectraMax compares with CLARIOstar
BMG Labtech’s CLARIOstar Plus page describes a different optical approach from the hybrid design above: dual linear variable filter (LVF) monochromators combined with a linear variable dichroic. Its documented modes are absorbance, fluorescence intensity (including FRET), luminescence (including BRET), fluorescence polarization, AlphaScreen, nephelometry, time-resolved fluorescence and TR-FRET. The absorbance spectrometer covers 220 to 1000 nm with precision of 0.5 nm or better; fluorescence intensity with the LVF monochromators covers 320 to 740 nm, extended to 840 nm with the red PMT. It supports up to two injectors with a volume range of 3 to 500 microlitres (optionally to 2 mL), and an atmospheric control unit for O2 and CO2 from 0.1 to 20 percent.
Set against the iD5e documentation, the main contrasts the two vendors’ own pages support are these. The iD5e pairs a monochromator with filters; CLARIOstar Plus uses LVF monochromators throughout. CLARIOstar lists AlphaScreen and nephelometry as modes, which the iD5e brochure does not list among its five modes. The data table below lays these figures out side by side. Treat it as a map of what each manufacturer states, not a performance ranking: sensitivity, dynamic range and speed depend on the assay, and head-to-head numbers would need to come from your own tests or a published comparison. For the general question of when a multi-mode reader is worth having over a single-mode one, see the CASRAI comparison of multimode and single-mode plate readers.
Validation, compliance and maintenance
Both documented SpectraMax lines tie into the manufacturer’s validation tools. The Plus 384 datasheet lists SpectraTest validation with NIST-traceable standards, IQ/OQ/PQ tools and 21 CFR Part 11 tools, and states that the instrument is for research use only. The iD5e brochure lists SoftMax Pro for research use, SoftMax Pro GxP for 21 CFR Part 11 and EudraLex Annex 11, and SpectraTest validation plates for absorbance, fluorescence and luminescence. In practice this means that software choice, not just the instrument, determines whether a reader can sit in a regulated workflow: a laboratory operating under GxP needs the GxP software edition, whereas an academic core facility may run the research edition. Routine performance checks and servicing apply regardless of the edition; the CASRAI guide on microplate reader calibration and maintenance covers that side.
Common misunderstandings
“SpectraMax is one instrument”
It is a family. The Plus 384 is absorbance only, with a cuvette port; the iD5e is a five-mode reader; the Paradigm is a cartridge-based platform. A protocol written for one model’s modes or wavelength range will not necessarily run on another.
“Pathlength correction makes plate readings identical to a cuvette”
The manufacturer states that PathCheck normalizes to a 1 cm equivalent within a stated volume range (50 to 300 microlitres in 96-well plates, 20 to 100 microlitres in 384-well plates). Outside that range, or for assays where the meniscus or sample properties matter, the claim is not made.
“Multi-mode means better”
A multi-mode reader offers more detection modes, not necessarily better absorbance performance. For absorbance-only work such as ELISA readout or DNA quantitation, the Plus 384 datasheet lists these applications directly.
“A research-use reader can be dropped into a regulated lab”
The Plus 384 datasheet states research use only, and the vendor separates research and GxP software editions. Whether a given setup meets a regulated requirement is a decision for the laboratory’s quality system, not something a datasheet settles.
Sources
- Molecular Devices, SpectraMax Plus 384 Microplate Reader datasheet (read 2026-11-16)
- Molecular Devices, SpectraMax iD5e Multi-Mode Microplate Reader brochure (read 2026-11-16)
- Molecular Devices, SpectraMax Paradigm Multi-Mode Microplate Reader datasheet (2013) (read 2026-11-16)
- BMG Labtech, CLARIOstar Plus product page (read 2026-11-16)
Data: Manufacturer-stated specifications of SpectraMax and CLARIOstar Plus microplate readers
Specifications compiled from each manufacturer's own datasheet, brochure or product page for three Molecular Devices SpectraMax models and the BMG Labtech CLARIOstar Plus. A dash means the source read does not state the value. Values are vendor claims, not independent measurements.
Period covered: 2013/2026.
| Instrument | Vendor | Detection modes stated | Absorbance wavelength range (nm) | Operating temperature | Weight (kg) |
|---|---|---|---|---|---|
| SpectraMax Plus 384 | Molecular Devices | Absorbance (with cuvette port) | 190-1000 | Ambient +4 to 45 C | 13.6 |
| SpectraMax iD5e | Molecular Devices | Absorbance, fluorescence, luminescence, TRF, tunable FP | — | Ambient to 66 C | 40 |
| SpectraMax Paradigm (2013 datasheet) | Molecular Devices | User-upgradeable multi-mode (cartridges) | — | Ambient +4 to 45 C | 45 |
| CLARIOstar Plus | BMG Labtech | Absorbance, FI, luminescence, FP, AlphaScreen, nephelometry, TRF, TR-FRET | 220-1000 | — | — |
Sources
- Molecular Devices, SpectraMax Plus 384 Microplate Reader datasheet (accessed 2026-11-16)