SpeedVac: How a Vacuum Concentrator Works, Traps and Pumps

What a SpeedVac (centrifugal vacuum concentrator) does, how its cold trap and pump work, and how it differs from freeze drying and rotary evaporation.

Written and maintained by CASRAI Editorial Board

Published

On this page

A SpeedVac is a centrifugal vacuum concentrator: an instrument that removes solvent from small samples in tubes, plates or vials by spinning them in a vacuum chamber with gentle heat, leaving the dissolved material concentrated or dry at the bottom of the container. SpeedVac is a Thermo Scientific Savant trademark, and the generic names are vacuum concentrator and centrifugal evaporator. This guide explains how the three forces work together, what the solvent trap and pump do, and where vacuum concentration sits next to freeze drying and rotary evaporation.

Definition and scope: SpeedVac, vacuum concentrator and centrifugal evaporator

Thermo Fisher’s SpeedVac white book defines centrifugal vacuum concentration as a technique that removes solvents in order to concentrate or dry biological and non-biological samples, and states that it combines three things: centrifugation, vacuum and heat. It is used on small-volume samples held in tubes, microplates and vials, which is the practical dividing line from bulk solvent removal in a flask.

The word SpeedVac names a product line. The same white book dates the technology to 1968, when Savant created it (with a collaboration at Rockefeller University shown in the company timeline), and records that Thermo Electron later bought Savant. Other manufacturers sell instruments that work on the same principle under different names, so a methods section or purchase specification is clearer when it says “vacuum concentrator” or “centrifugal evaporator” and names the model separately.

How a SpeedVac works

Each of the three forces solves a specific problem of evaporating a small liquid sample.

Centrifugation

According to the white book, the sample is spun at roughly 200 to 500 x g. That force keeps the liquid pressed to the bottom of the container, which prevents bumping, boiling over and sample loss when pressure drops, and it makes the dissolved solute settle at the bottom of the tube as the solvent leaves. To translate a rotor speed into a force for your own rotor, see RCF versus RPM in centrifugation.

Vacuum

Lowering the pressure lets the solvent evaporate while the sample stays liquid at a temperature below ambient-pressure boiling. The white book adds that evaporating under vacuum also prevents oxidation of the sample.

Heat

Evaporation is strongly cooling. The white book notes that without added heat an aqueous solution could freeze during the run, so concentrators provide chamber heaters, and some models use radiant lamps. The SPD120 operation manual describes a pre-heat function that brings the chamber to 45 C before a run, and a manual run mode that locks the cover.

The vapour path: cold trap, pump and chemical trap

The concentrator itself is only part of a system. The SPD120 manual describes the standalone unit as one component of a system that also includes a refrigerated vapour trap and a vacuum pump, with optional chemical trap cartridges for acid-neutralising and volatile-radioactivity applications. The manual’s caution is that the cold trap must sit in line between the vacuum source and the concentrator, so that vapour is condensed before it reaches the pump.

Cold trap, chemical trap and pump in practice

The white book gives specific working guidance on the vapour path. It describes the cold trap as the primary trap, condensing 85 to 95 percent of solvent vapours, and the chemical trap as a secondary line of protection. Its listed practices are:

  • Start every run with a clean, empty glass condensation flask in the cold trap.
  • Check trap efficiency by thawing the trap and measuring the collected solvent against what was removed from the samples; the white book puts the expected recovery at 85 to 95 percent.
  • If recovery is poor, use a colder trap or place two traps in series.
  • Use the vacuum gauge, and be ready to stop the run when the samples are dry, because dry samples can warm up.
  • Pre-warm the chamber (45 C in the SPD120 manual) before loading.
  • For acidic samples, fit an activated-alumina filter to the pump oil.

In its expert question-and-answer section the white book is direct that you should not run an oil pump, or an oil-free pump, without a cold trap.

Which pump: oil-free diaphragm or rotary vane oil pump

The white book contrasts the two. Oil pumps are faster, but it notes they can cause bumping and cross-contamination and need frequent oil changes. Oil-free diaphragm pumps are slower but need no oil changes. Its general recommendation is an oil-free pump, with the exceptions of freeze drying inside the concentrator and high-boiling solvents such as DMSO, both of which need a deeper vacuum from a rotary vane oil pump. For how that pump type works, see What is a rotary vane pump.

Solvent classes and what they ask of the system

The white book groups solvents by how demanding they are on the instrument, from least to most:

Solvent class (white book grouping) Examples named in the source
Aqueous solutions Least demanding class
Non-aggressive organics Ethanol, methanol, acetonitrile
Aggressive organics Methylene chloride, toluene
Strong acids and bases Most demanding class; see the chemical trap and pump oil filter options above
High-boiling solvents DMSO, DMF (the white book names DMSO as an exception to its oil-free pump recommendation)

The grouping runs from least to most demanding on the instrument, with high-boiling solvents handled as a separate case.

Vacuum concentration, freeze drying and rotary evaporation compared

Choosing between concentration techniques is mostly a question of sample type and scale. The white book states that for non-biological samples, such as products of organic synthesis, metabolites and pesticides, vacuum evaporation is the technique of choice. For biological samples it generally recommends freeze drying when the aim is long-term storage, quick reconstitution and retained activity, while DNA and RNA in aqueous or low-organic solutions are mostly dried by vacuum concentration. It also states that a rotary evaporator uses the same ingredients of vacuum, heat and condensation, but works on bulk liquid, whereas a SpeedVac works in tubes, plates and vials.

CASRAI covers the neighbouring techniques separately: What is a rotary evaporator for bulk solvent removal, and the lyophilization and freeze-drying guide for sublimation-based drying. The data table below this guide sets out the protective components of the vapour path as the manufacturer documents them.

Where it matters for labs and supply decisions

Safety limits from the manual

The SPD120 manual warns against using the instrument in a radioactive, highly reactive or explosive atmosphere, and against processing explosive, radioactive or highly reactive substances. It also cautions that only the recommended rotors should be used and that rotor loads must always be balanced, loaded symmetrically (a tube is not needed in every holder), because an imbalance causes vibration that damages the bearings. The general principles are the same as for any centrifuge; see centrifuge rotor balancing and safety.

What to specify as a system

Because the concentrator, cold trap and pump are described as a system, a requisition or SOP that names only the centrifuge unit leaves out the parts that protect the pump and capture solvent. The white book also distinguishes integrated systems from modular ones, where the trap and pump are separate units, and lists chemical trap cartridges as accessories: activated carbon to adsorb organic vapours and volatile radioactivity, and soda lime to neutralise acid.

Common misunderstandings

  • SpeedVac is not a generic term. It is a brand; vacuum concentrator is the generic name.
  • It is not a small rotary evaporator. The ingredients overlap, but the format differs: tubes, plates and vials rather than a rotating flask.
  • The vacuum alone does not dry the sample. Without heat the sample can cool toward freezing; without spinning it can bump.
  • A dry sample is not a finished run. The white book advises being ready to stop when the samples are dry because they can heat.
  • The pump does not replace the trap. Both the white book and the manual put a cold trap between the samples and the pump.
  • It is not for everything. Explosive, radioactive and highly reactive materials are excluded in the manual.

Frequently asked questions

What is a SpeedVac?

A SpeedVac is a Thermo Scientific Savant centrifugal vacuum concentrator: it removes solvent from samples in tubes, plates or vials using centrifugation, vacuum and heat together.

Is a vacuum concentrator the same as a SpeedVac?

Vacuum concentrator, or centrifugal evaporator, is the generic name for the technique and instrument class. SpeedVac is one manufacturer’s trademark within it.

Why does the sample need heat under vacuum?

Evaporation cools the sample strongly. According to the manufacturer’s white book, aqueous solutions could freeze without heat, so chamber heaters (and on some models lamps) offset the cooling.

What does the cold trap do?

It condenses solvent vapour before it reaches the pump. The white book describes it as the primary trap, capturing 85 to 95 percent of solvent vapours, with a chemical trap as the secondary line.

Oil pump or oil-free pump?

The white book generally recommends an oil-free pump, except for freeze drying in the concentrator and high-boiling solvents such as DMSO, which need the deeper vacuum of a rotary vane oil pump.

Sources

Data: Vapour-path protection components of a centrifugal vacuum concentrator, as documented by the manufacturer

Compares the protective components between a centrifugal vacuum concentrator and its vacuum pump (cold trap, chemical trap cartridges, pump oil filter, pump type), compiled from the SpeedVac white book and the SPD120 operation manual.

Vapour-path protection components of a centrifugal vacuum concentrator, as documented by the manufacturer
ComponentRole stated by the manufacturerSpecific figure or option statedSource document
Cold trapPrimary trap that condenses solvent vapours; must sit in line between the vacuum source and the concentratorCondenses 85 to 95 percent of solvent vapours; use a colder trap or two traps in series if recovery is lowWhite book; SPD120 manual
Chemical trap, activated carbonSecondary trap that adsorbs organic vapours and volatile radioactivityAccessory cartridgeWhite book; SPD120 manual
Chemical trap, soda limeSecondary trap that neutralises acid vapourAccessory cartridgeWhite book; SPD120 manual
Pump oil filter, activated aluminaProtects a rotary vane pump when acid samples are runFitted to the pump oilWhite book
Oil-free diaphragm pumpGenerally recommended vacuum source; slower, no oil changesFreeze drying in the concentrator and high-boiling solvents such as DMSO are the stated exceptionsWhite book
Rotary vane oil pumpDeeper vacuum; faster, but can cause bumping and cross-contamination and needs frequent oil changesNeeded for freeze drying in the concentrator and for high-boiling solvents such as DMSOWhite book

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about SpeedVac: How a Vacuum Concentrator Works, Traps and Pumps

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →