Tensile Test: Machine, Specimens and ASTM/ISO Standards

How a tensile test works, the specimen and machine, which ASTM or ISO standard fits metals, plastics and rubber, and how to read the curve.

Written and maintained by CASRAI Editorial Board

Last updated

A tensile test pulls a specimen of known shape apart at a controlled rate while recording force and elongation, so that strength and ductility can be calculated from the result. This guide is for materials-science and quality-control lab staff who need to pick the right standard, understand the machine and specimen, and read the resulting stress-strain curve. The right standard depends on the material: ASTM E8/E8M for metals, ASTM D638 for plastics, ASTM D412 for rubber.

What is a tensile test?

A tensile test (also called a tension test or tensile strength test) is a standardized, quasistatic, destructive method. The specimen is clamped between two crossheads, one of which moves at constant velocity until the specimen fails. Force and elongation are recorded throughout, and stress and strain are computed from the specimen’s original dimensions (ZEISS, read 2026-11-07).

ASTM E8/E8M states that tension tests provide information on the strength and ductility of materials under uniaxial tensile stress, and that the results are useful for comparison of materials, alloy development, quality control and design (ASTM E8/E8M-25, section 4.1). The standard lists the properties it determines for metals: yield strength, yield point elongation, tensile strength, elongation and reduction of area.

How a tensile test machine works

A tensile tester (universal testing machine in tension mode) has a fixed crosshead and a movable crosshead, driven hydraulically or electrically, with grips that hold the specimen. Table-top machines start at about 3 kN and the range extends up to 2,000 kN (ZEISS). Strain is measured with an extensometer, either a contact clip-on type or a non-contact optical type.

The force measurement of the machine is itself checked against a standard: ISO 7500-1 covers verification of the force measuring system of tension/compression testing machines, and ISO 5893 covers testing machines for plastics and rubber (both named by ZEISS).

Tensile test specimens, including the dog bone shape

ZEISS names two basic specimen forms: round specimens for solid material and flat specimens for sheet metal. The enlarged ends are gripped; the narrower central section is where deformation and failure are intended to happen. Specimens with wide gripped ends and a narrow middle are informally called dog bone specimens. Exact dimensions come from the governing standard, not from this page.

For round metal specimens, ASTM E8/E8M states: “The gauge lengths for most round specimens are required to be 4D for E8 and 5D for E8M” (D is the specimen diameter; powder-metallurgy materials are treated separately in the standard). Reading E8 and E8M as the same gauge length is a common mix-up, so confirm which of the two you are working to.

ASTM and ISO tensile test standards by material

Standard Material it covers What it governs
ASTM E8/E8M-25 (active, updated 2025-08-26) Metallic materials, any form, room temperature Tension testing; yield strength, yield point elongation, tensile strength, elongation, reduction of area
ISO 6892-1 (DIN EN ISO 6892-1) Metals Tensile test method for metals (named by ZEISS)
ISO 7500-1 Testing machine Verification of machine force measurement (named by ZEISS)
ISO 5893 Testing machine for plastics and rubber Machine requirements (named by ZEISS)
ASTM D638 Plastics See ASTM D638
ASTM D412 Vulcanized rubber and thermoplastic elastomers See ASTM D412

To locate and cite a standard, see the guides to ASTM Compass and citing ASTM standards.

Reading the stress-strain curve

ZEISS describes these features of the curve, and each corresponds to a reported property:

Curve feature Reported property
Slope of the initial straight (Hooke) line Young’s modulus
Upper and lower points where the material yields Upper and lower yield strength
Maximum stress reached Ultimate tensile strength; necking begins after the maximum force
Strain at maximum force Uniform strain
Strain at fracture Failure strain

Common mistakes

  • Treating E8 and E8M as interchangeable when the round-specimen gauge length differs (4D versus 5D).
  • Using a metals standard for plastic or rubber specimens, or the reverse; pick the standard by material first.
  • Reading dimensions from a summary page rather than the current edition of the standard; ASTM E8/E8M-25 is the active edition as checked on 2026-11-07.
  • Skipping verification of the machine’s force system (ISO 7500-1 for the metals side).

Frequently asked questions

What is a tensile test?

A destructive test in which a specimen is stretched at a controlled rate until it fails, while force and elongation are recorded to calculate stress and strain (ZEISS).

What does a tensile test measure?

For metals under ASTM E8/E8M: yield strength, yield point elongation, tensile strength, elongation and reduction of area.

Which standard applies to a tensile test?

It depends on the material. See the table above: ASTM E8/E8M and ISO 6892-1 for metals, ASTM D638 for plastics, ASTM D412 for rubber.

What is the difference between a tensile tester and a tensile test machine?

The terms are used interchangeably for the same instrument: a frame with fixed and movable crossheads, grips, a force measuring system and an extensometer.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Tensile Test: Machine, Specimens and ASTM/ISO Standards

Ask your first 2 questions free below. Subscribers get 150 a day for $29 a month.

An AI assistant specialized in research administration. It cites the sources behind every answer, labels web answers and says when it can't answer.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays free to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →